Focal One®

K251910 · Edap Technomed, Inc. · PLP · Nov 19, 2025 · Gastroenterology, Urology

Device Facts

Record IDK251910
Device NameFocal One®
ApplicantEdap Technomed, Inc.
Product CodePLP · Gastroenterology, Urology
Decision DateNov 19, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4340
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Focal One® device is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue.

Device Story

Focal One® is a high-intensity focused ultrasound (HIFU) system for prostate tissue ablation. It utilizes an endorectal dynamic focusing probe to deliver ultrasound energy through the rectal wall, focusing on targeted prostate tissue to induce thermal ablation. The system includes a control module, computer, cooling unit, and disposable FocalPak kit. A coupling liquid (Ablasonic®) circulates through a balloon to protect the rectal wall. The device is operated by a physician in a clinical setting. The physician uses integrated imaging (ultrasound and DICOM fusion) to localize the prostate and plan treatment. The system uses pre-set algorithms to guide the stepwise destruction of targeted tissue. The device provides visual feedback via dual touch-screens, mouse, and keyboard. The procedure is minimally invasive and performed on an outpatient basis under anesthesia. The device benefits patients by providing a targeted, non-surgical ablation option for prostate tissue.

Clinical Evidence

Bench testing only. Evidence includes biocompatibility testing, reprocessing and shelf-life validation, electrical safety and EMC testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 60601-2-62), software validation, bench testing of the new imaging ultrasound, and human factors testing.

Technological Characteristics

System includes control module, computer, endorectal dynamic focusing probe (curved array), cooling unit, and disposable accessories. Ablation frequency: 3.0 MHz. Imaging frequency: 6.6–7.7 MHz. Focal distance: 60 mm ± 2 mm. Connectivity: DICOM for image fusion. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 60601-2-62. Software: Pre-set algorithm for protocol management.

Indications for Use

Indicated for transrectal high-intensity focused ultrasound (HIFU) ablation of prostate tissue in patients requiring such intervention.

Regulatory Classification

Identification

A high intensity ultrasound system for prostate tissue ablation is a prescription device that transmits high intensity therapeutic ultrasound energy into the prostate to thermally ablate a defined, targeted volume of tissue, performed under imaging guidance. This classification does not include devices that are intended for the treatment of any specific prostate disease and does not include devices that are intended to ablate non-prostatic tissues/organs.

Special Controls

In combination with the general controls of the FD&C Act, the high intensity ultrasound system for prostate tissue ablation is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Characterization of acoustic pressure and power output at clinically relevant levels; (ii) Measurement of targeting accuracy and reproducibility of high intensity ultrasound output; (iii) Ultrasound-induced heating verification testing at target and non-target tissues; (iv) Electrical safety testing; and (v) Electromagnetic compatibility testing. (2) Software verification, validation, and hazard analysis must be performed. (3) The elements of the device that may contact the patient's mucosal tissue must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the device components that contact the patient's mucosal tissue. (5) Performance data must support shelf life by demonstrating continued sterility of the device or the sterile components, package integrity, and device functionality over the identified shelf life. (6) Performance data must support the instructions for reprocessing all reusable components. (7) *In vivo* testing must demonstrate that the device thermally ablates targeted tissue in a controlled manner without thermal injury to adjacent, non-target tissues.(8) Clinical testing must document the adverse event profile, provide evidence of prostatic ablation, and demonstrate that the device performs as intended under anticipated conditions of use. (9) Training must be provided so that upon completion of the training program, the physician can: (i) Use all safety features of the device; (ii) Accurately target the high intensity ultrasound energy within the desired region of the prostate; and (iii) Perform the ablation procedure in a manner that minimizes damage to non-target tissues. (10) Labeling must include: (i) A section that summarizes the clinical testing results, including the adverse event profile and evidence of prostate ablation achieved; and (ii) An expiration date or shelf life for single use components.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION November 13, 2025 Edap Technomed, Inc. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street Suite 2300 Philadelphia, Pennsylvania 19103 Re: K251910 Trade/Device Name: Focal One® Regulation Number: 21 CFR 876.4340 Regulation Name: High Intensity Ultrasound System For Prostate Tissue Ablation Regulatory Class: II Product Code: PLP Dated: June 20, 2025 Received: June 20, 2025 Dear Janice Hogan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251910 - Janice Hogan Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K251910 - Janice Hogan Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **MARK J. ANTONINO -S** Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Focal One® Page 10 of 58 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251910 | ? | | Please provide the device trade name(s). | | ? | | Focal One® | | | | Please provide your Indications for Use below. | | ? | | The Focal One® is indicated for transrectal high-intensity focused ultrasound (HIFU) ablation of prostate tissue. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K251910 Page 1 of 4 # 510(k) SUMMARY Applicant: EDAP Technomed, Inc. 5321 Industrial Oaks Blvd, Suite 110 AUSTIN, TX 78735 USA Phone: 512 832 7956 Facsimile: 512 684 1313 Contact Person: Hugo EMBERT Date Prepared: June 20, 2025 Proprietary Name: Focal One® Common Name: High intensity ultrasound system for prostate tissue ablation Classification Name: High intensity ultrasound system for prostate tissue ablation Regulatory Class: II Regulation: 21 CFR 876.4340 Product Code: PLP Predicate Device(s): EDAP Technomed, Inc.'s Focal One® (K172721) ## Intended Use: The Focal One® device is indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue. ## Device Description: The Focal One® is an evolution from the previous generation of Focal One® by EDAP as cleared in K172721. The Focal One® consists of the Focal One® module with a software control system, an endorectal dynamic focusing probe, a leg support, and a disposable set (FocalPak) including miscellaneous single use disposables and a coupling liquid pouch. The Focal One® module consists of a motorized and manual endorectal dynamic focusing probe positioning unit, a RF amplifier to power the transducer, a computer to control device operation, a cooling system to cool Ablasonic (coupling liquid) and built-in safety features. It also allows the user to control the treatment while providing different user interfaces: two dual touch-screens, mouse, keyboard and printer. The Focal One® energy is delivered via an endorectal dynamic focusing probe, which includes an imaging system. The high-energy ultrasound waves propagate through the rectal wall and are focused on a portion of the prostate, generating heat and causing the ablation of tissue within the focal area. The process is then repeated in a stepwise fashion to destroy the targeted tissues within the prostate. The endorectal dynamic focusing probe is attached to a support, enabling movements in the longitudinal, transverse, vertical and angular directions, and in rotation. A coupling {5} K251910 Page 2 of 4 liquid pouch (Ablasonic®) included in the FocalPak disposable kit, attached to a stand, maintained at a controlled temperature flows continuously (via a peristaltic pump) into the balloon to preserve the rectal wall from heating associated with the treatment. The module is mounted on four multidirectional wheels allowing an easy positioning of the device in the treatment room. ## Substantial Equivalence The subject Focal One® device is an updated version of the previously cleared Focal One® (K172721) and is substantially equivalent to the Focal One® System cleared in K172721 ("predicate device"). The subject Focal One has the identical indications for use and principles of operation, and similar technological characteristics, as the previously cleared predicate. Minor differences in the technological characteristics between the subject device and the predicate do not raise different questions of safety or efficacy. The ultrasound treatment by the therapeutic transducer remains unchanged compared to the cleared version. The main difference is the inclusion of a new imaging ultrasound to replace the imaging ultrasound in the cleared device. Notably, the modification to the imaging scanner has no impact on the overall device specification. Nor does it affect the user interface or the safety features of the device. Other minor updates have been made, e.g., to replace obsolete components without altering specifications. None of the changes affect the treatment parameters or energy specifications. | | Focal One® – Subject Device | Focal One® – Predicate Device | Comment | | --- | --- | --- | --- | | Manufacturer | EDAP | EDAP | Same | | 510(k) No. | | K172721 | | | Product Code | PLP | PLP | Same | | Intended Use | | | | | Indications for Use | Indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue. | Indicated for transrectal high intensity focused ultrasound (HIFU) ablation of prostate tissue. | Same | | Prescription use | Yes | Yes | Same | | Minimally invasive | Yes | Yes | Same | | Outpatient procedures | Yes | Yes | Same | | Anesthesia required | Yes | Yes | Same | | Physician training required | Yes | Yes | Same | | General Description | | | | {6} K251910 Page 3 of 4 | | Focal One® – Subject Device | Focal One® – Predicate Device | Comment | | --- | --- | --- | --- | | System Components | Control module, Computer and peripherals, Endorectal probe, Cooling unit, Probe holder, Ultrasound scanner, Disposable accessories Probe movement assembly Movement detector, | Control module, Computer and peripherals, Endorectal probe, Cooling unit, Probe holder, Ultrasound scanner, Disposable accessories Probe movement assembly Movement detector, | Same | | Patient position | right lateral decubitus | right lateral decubitus | Same | | Performance Characteristics | | | | | Imaging modality for localization, treatment and control | Ultrasound | Ultrasound | Same | | Fusion of ultrasound with other imaging modalities (DICOM) | Yes | Yes | Same | | Probe type | Curved Array | Curved Array | Same | | Imaging Frequency | Between 6.6 and 7.7 MHz | Between 6.6 and 7.7 MHz | Same | | Longitudinal Imaging frame rate (typical) | NA (Longitudinal image is reconstructed) | NA (Longitudinal image is reconstructed) | Same | | Transverse Imaging frame rate (typical) | 25 FPS | 25 FPS | Same | | Image size | 8x8cm | 8x8cm | Same | | Field of view | 144° | 144° | Same | | Ablation modality | HIFU | HIFU | Same | | Ablation Frequency | 3.0 MHz | 3.0 MHz | Same | {7} K251910 Page 4 of 4 | Performance Characteristics | | | | | --- | --- | --- | --- | | Focal Distance | 60 mm ± 2 mm 32 – 67 mm | 60 mm ± 2 mm 32 – 67 mm | Same | | Probe length (mm) | 60.9 | 60.9 | Same | | Probe diameter (mm) | 38.4 | 38.4 | Same | | Probe neck diameter (mm) | 25 | 25 | Same | | Management of protocols | Pre-set algorithm | Pre-set algorithm | Same | ## Non-Clinical Testing The following non-clinical testing was provided in support of this submission to demonstrate that the new system is equivalent to the cleared system: - Biocompatibility Testing - Reprocessing and Shelf Life Testing - Electrical Safety and Electromagnetic Compatibility (EMC) Testing according to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and IEC 60601-2-62 - Software Documentation and Validation Testing - Bench Testing to assess the new imaging ultrasound of Focal One. - Human Factors Testing ## Conclusion The Focal One and its predicate have the same indications for use and principles of operation, as well as similar technological characteristics. Minor differences in the technological characteristics, including the new ultrasound scanner, do not present different questions of safety or effectiveness as compared to the predicate devices. Performance testing demonstrate that the subject device is as safe and effective as its predicate devices. Thus, the Focal One is substantially equivalent to its predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...