The sterile single-use flexible video bronchoscopes are designed to be used with the Himaging endoscopic video processor, monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The Himaging bronchoscope system is applicable to hospital environment or medical office environment. The Himaging bronchoscopes are sterile single-use devices designed for use in adults.
Device Story
System consists of single-use flexible video bronchoscope (DBRS-100HLP) and desk-type endoscopic video processor (HDIS-1000). Device captures airway images via distal LED-illuminated camera; transmits video signal to processor for display on external monitor (not included). Physician operates bronchoscope to navigate airways and pass endotherapy instruments through working channel. Processor provides image management functions: zoom, brightness adjustment, manual white balance, and image saving (BMP/JPG) to internal storage or USB. System intended for hospital/medical office use. Output assists visualization for diagnostic/therapeutic airway procedures. Benefits include sterile single-use design eliminating reprocessing needs and cross-contamination risks.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including optical performance (color reproduction, geometric distortion, resolution), ISO 8600 compliance, photobiological safety (IEC 62471), electrical safety/EMC (IEC 60601 series), biocompatibility (ISO 10993), sterilization validation, and shelf-life/packaging integrity testing.
Technological Characteristics
Single-use flexible video bronchoscope; LED illumination at distal tip. Materials: PP, Pebax, silicone rubber, TPU, PC, SUS 304. Sterilization: Ethylene Oxide (EO). Connectivity: DVI video output. Processor: HDIS-1000, 100-240V AC power. Software: Moderate level of concern. Standards: ISO 8600, ISO 10993, IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, ASTM F1980, ASTM F88, ASTM F1929, ASTM F1886, ASTM D4169.
Indications for Use
Indicated for use in adults for endoscopic procedures and examination within the airways and tracheobronchial tree in hospital or medical office environments.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Ambu® aScope™ 3 Slim 3.8/1.2 and Ambu ® aScope™ 4 Broncho Slim 3.8/1.2; Ambu ® aScope™ 3 Regular 5.0/2.2 and Ambu ® aScope™ 4 Broncho Regular 5.0/2.2; Ambu ® aScope™ 3 Large 5.8/2.8 and Ambu® aScope™ 4 Broncho Large 5.8/2.8; Ambu ® aView Monitor (K173727)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 23, 2025
Himaging Technology (Shanghai) Co., Ltd.
Shumei Xu
Registration Manager
Room 2601 and 2605 (Nominal room 2801 and 2805),
No. 418 Guiping Road, Xuhui District
Shanghai, 200233
China
Re: K251894
Trade/Device Name: Himaging Bronchscope System
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: May 16, 2025
Received: June 20, 2025
Dear Shumei Xu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K251894 - Shumei Xu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251894 - Shumei Xu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Himaging Bronchscope System
Page 10 of 47
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251894 | ? |
| Please provide the device trade name(s). | | ? |
| Himaging Bronchscope System | | |
| Please provide your Indications for Use below. | | ? |
| The sterile single-use flexible video bronchoscopes are designed to be used with the Himaging endoscopic video processor, monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Himaging bronchoscope system is applicable to hospital environment or medical office environment.
The Himaging bronchoscopes are sterile single-use devices designed for use in adults. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251894
510(k) Summary
Date of Preparation: 09/23/2025
1. Sponsor Identification
Himaging Technology (Shanghai) Co., Ltd.
Room 2601 and 2605(Nominal room 2801 and 2805), No. 418 Guiping Road, Xuhui District, 200233, Shanghai, China
Establishment Registration Number: Not assigned yet
Contact Person: Shumei Xu
Position: Registration Manager
Tel: +86-21-5330-8891
Email: shumei.xu@himaging.com
2. Designated Submission Correspondent
Ms. Shumei Xu (Primary Contact Person)
Himaging Technology (Shanghai) Co., Ltd.
Room 2601 and 2605 (Nominal room 2801 and 2805), No. 418 Guiping Road, Xuhui District, 200233, Shanghai, China
Tel: +86-21-5330-8891
Email: shumei.xu@himaging.com
Mr. Xingwang Chen (Alternative Contact Person)
Himaging Technology (Shanghai) Co., Ltd.
Room 2601 and 2605 (Nominal room 2801 and 2805), No. 418 Guiping Road, Xuhui District, 200233, Shanghai, China
Tel: +86-21-5330-8891
Email: xingwang.chen@himaging.com
Page 1 of 13
{5}
Page 2 of 13
3. Identification of Subject Device
Trade Name: Himaging Bronchoscope System
Common Name: Endoscope and accessories
**Regulatory Information**
Classification Name: Bronchoscope (Flexible or Rigid)
Classification: II
Product Code: EOQ
Regulation Number: 21 CFR 874.4680
Review Panel: Ear Nose & Throat;
**Indication for Use:**
The sterile single-use flexible video bronchoscopes are designed to be used with the Himaging endoscopic video processor, monitor, endoscopic equipment and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Himaging bronchoscope system is applicable to hospital environment.
The Himaging bronchoscopes are sterile single-use devices designed for use in adults.
4. Identification of Predicate Device
K Number: K173727
Trade/Device Name: Ambu® aScope™ 3 Slim 3.8/1.2 and Ambu ® aScope™ 4 Broncho Slim 3.8/1.2; Ambu ® aScope™ 3 Regular 5.0/2.2 and Ambu ® aScope™ 4 Broncho Regular 5.0/2.2; Ambu ® aScope™ 3 Large 5.8/2.8 and Ambu® aScope™ 4 Broncho Large 5.8/2.8; Ambu ® aView Monitor
Manufacturer: Ambu A/S
**Indication for Use:**
The aScope 3 and aScope 4 Broncho endoscopes have been designed to be used with the aView monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The aScope 3 system and aScope 4 Broncho system are for use in a hospital environment.
The aScope 3 and aScope 4 Broncho are single-use devices designed for use in adults. They have been evaluated for the following endotracheal tubes (ETT), double lumen tubes (DLT) and endoscopic accessories (EA) sizes:
{6}
Page 3 of 13
| | Minimum ETT inner diameter | Minimum DLT size | EA minimum working channel width |
| --- | --- | --- | --- |
| aScope 3 Slim and aScope 4 Broncho Slim | 5.0 mm | 35 Fr | Up to 1.2 mm |
| aScope 3 Regular aScope 4 Broncho Regular | 6.0 mm | 41 Fr | Up to 2.0 mm |
| aScope 3 Large aScope 4 Broncho Large | 7.0 mm | - | Up to 2.6 mm |
Note: The predicate device has multiple models, the aScope 3 system (aScope 3 Regular Broncho endoscopes and Ambu ® aView Monitor) are selected to compare with the subject device.
5. Identification of Reference Device
K Number: K233671
Trade/Device Name: Ambu® aScope™ 5 Broncho 4.2/2.2, Ambu® aScope™ 5 Broncho 2.7/1.2, Ambu® aView™ 2 Advance
Indication for Use:
aScope™ 5 Broncho is intended for endoscopic procedures and examination within the airways and tracheobronchial tree.
aScope™ 5 Broncho is intended to provide visualization via a compatible Ambu displaying unit, and to allow passing of endotherapy instruments via its working channel.
The aView™ 2 Advance is intended to display live imaging data from compatible Ambu visualization devices.
6. Device Description
The subject device, Himaging Bronchoscope System, is consisting of a single-use flexible video Bronchoscope and an endoscopic video processor. The subject device has been designed to be used for endoscopy within the airways and tracheobronchial tree.
{7}
# 7. Substantially Equivalent (SE) Comparison
Table. 1 General Comparison
| ITEM | Subject Device | Predicate Device K173727 | Reference Device K233671 | Remark |
| --- | --- | --- | --- | --- |
| Product Code | EOQ | EOQ | EOQ | Same |
| Regulation No. | 874.4680 | 874.4680 | 874.4680 | Same |
| Class | II | II | II | Same |
| Manufacturer | Himaging Technology (Shanghai) Co., Ltd. | Ambu A/S | Ambu A/S | Same |
| Product name | Himaging Bronchoscope System | aScope 3 Regular Broncho endoscopes Ambu ® aView Monitor | Ambu® aScope™ 5 Broncho 4.2/2.2 Ambu® aView™ 2 Advance | Same |
| Indication for Use | The sterile single-use flexible video bronchoscopes are designed to be used with the Himaging endoscopic video processor, monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.The Himaging bronchoscope system is applicable to hospital environment.The Himaging bronchoscopes are sterile single-use devices designed for use in adults. | The aScope 3 and aScope 4 Broncho endoscopes have been designed to be used with the aView monitor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The aScope 3 system and aScope 4 Broncho system are for use in a hospital environment. The aScope 3 and aScope 4 Broncho are single-use devices designed for use in adults. | aScope™ 5 Broncho is intended for endoscopic procedures and examination within the airways and tracheobronchial tree. aScope™ 5 Broncho is intended to provide visualization via a compatible Ambu displaying unit, and to allow passing of endotherapy instruments via its working channel. The aView™ 2 Advance is intended to display live imaging data from compatible Ambu visualization devices. | Same |
| Main Configuration | Single-use flexible video bronchoscope (DBRS-100HLP) | Single-use flexible bronchoscope (Ambu® aScope™ 3 | Single-use flexible bronchoscope (Ambu® aScope™ 5 | Different |
{8}
# Different- Main Configuration
The endoscopic video processor of the predicate device which has an extra integrated monitor, is different from the subject device. The endoscopic video processor of the subject system is desk type, and the subject system does not include the external monitor. Based on the results of the comparing testing of image quality (color reproduction, geometric distortion, intensity uniformity, depth of field and resolution), the image quality of the subject endoscopic video processor was equivalent to that of the predicate device. Therefore, the difference will not affect the safety and effectiveness of the subject device.
Table. 1 Specifications Comparison of Endoscopic video processor
| ITEM | Endoscopic Video Processor | Endoscopic Image Processor K173727 | Remark |
| --- | --- | --- | --- |
| Product name | Endoscopic Video Processor | Ambu® aView™ Monitor | |
| Model | HDIS-1000 | aView | / |
| Dimension | 350mm(W)×360mm (D)×92mm (H) | 241mm(W)×175(H)mm×33.5(D) mm | Different |
| Weight | About 5Kg | 1.5Kg | Different |
| Is there an integrated LCD display? | No | Yes | Different |
| Freeze | Create static picture of dynamic image | Create static picture of dynamic image | Same |
{9}
| Zoom | 1x,2x,4x (3 levels) | The size of the live image can be changed between two sizes: 480x480 - Default setting. 560x480 - Expanded view | Different |
| --- | --- | --- | --- |
| Brightness Adjustment | Adjustment levels: 0-9 | Continuous brightness adjustment | Similar |
| Image Saving | Able to save in internal storage or U disk (BMP or JPG format). | Able to save in internal storage or U disk. | Same |
| White balance | Manual | N/A | Different |
| Video Signal Output | DVI signal output port | JACK connector (video-out) and USB port | Different |
| Power Supply | It has only an external power supply, 100-240V AC, 50/60Hz, 60VA. | It has a rechargeable lithium-ion battery. It has two power supply modes, which are internal and external power supply. The external power supply is 100-240V AC, 50/60Hz, 0.6A. | Different |
| Communication with endoscope | Provided | Provided | Same |
# Different-Dimension
The dimension for the subject image processor is different from the predicate device. The dimension is just in physical specification and this difference will not raise any issues on the subject device's safety and effectiveness.
# Different- Weight
The weight for the subject image processor is different from the predicate device. The weight is just in physical specification and this difference will not raise any issues on the subject device's safety and effectiveness.
# Different- Is there an integrated LCD display?
The endoscopic video processor of the subject system is desk type, without an integrated LCD display. In addition, the subject system does not include the external monitor. Based on the results of the comparing testing of image quality (color reproduction, geometric distortion, intensity uniformity, depth of field and resolution), the image quality of the image processor was equivalent to that of the predicate device. Therefore, the difference will not affect the safety and effectiveness of the subject device.
# Different- Zoom
{10}
The ZOOM function of the subject image processor is different from predicate image processor. The ZOOM function is used to adjust the image size without changing the image content; this difference will not raise any issues in safety and effectiveness.
# Similar- Brightness Adjustment
The brightness adjustment function of the subject image processor is different from predicate image processor. The brightness adjustment function is used to adjust the brightness without changing the image content; this difference will not raise any issues in safety and effectiveness.
# Different- White balance
The subject device has the manual white balance function, is different from the predicate device which has not white balance function. The white balance function is used to generate more real color of the image without changing the image content. Based on the results of the comparing testing of image quality (color reproduction, geometric distortion, intensity uniformity, depth of field and resolution), the image quality of subject endoscopic video processor was equivalent to that of the predicate device. Therefore, the difference will not affect the safety and effectiveness of the subject device.
# Different- Video Signal Output
The types of video signal output are different between subject endoscopic video processor and predicate image processor. The subject device has DVI output port, the predicate device has JACK connector (video-out) and USB port. The image quality of the subject device and device have been tested in the quantitative image quality testing (Color Reproduction and Geometric Distortion). The test results demonstrate that the image quality of the subject endoscopic video processor was equivalent to that of the predicate device. Therefore, the difference on video signal output will not affect the safety and effectiveness of the subject device.
# Different- Power Supply
The subject device is only supplied by the external power, is different from the predicate device. There are two power supplies for predicate device, which are internal and external power supply. The external power supply for subject device is same as predicate device. Therefore, the difference on power supply will not affect the safety and effectiveness of the subject device.
Table. 2 Specifications Comparison of Broncho Videoscope
| ITEM | Subject Device | Predicate Device K173727 | Reference Device K233671 | Remark |
| --- | --- | --- | --- | --- |
| Model | DBRS-100HLP | Ambu® aScope™ 3 Regular 5.0/2.2, | Ambu® aScope™ 5 Broncho 4.2/2.2 | / |
| | | aScope™ 5 Broncho 4.2/2.2 | | |
| Model | DBRS-100HLP | Ambu® aScope™ 3 Regular 5.0/2.2, | Ambu® aScope™ 5 Broncho 4.2/2.2 | / |
{11}
| Maximum Outer Diameter of Insertion Section | 5.8mm | 5.5mm | / | Similar |
| --- | --- | --- | --- | --- |
| Minimum Instrument Channel Width | 2.0mm | 2.0mm | / | Same |
| Working Channel | Φ 2.0 mm | Φ 2.0 mm | / | Same |
| Working Length | 600mm | 600mm | / | Same |
| Total Length | 820mm | 820mm | / | Same |
| Bending angle | Up:210°, Down:130° | Up:150°, Down:130° | / | Similar |
| Rotary function | 120° | N/A | 120° | Same |
| Field of View | 100° | 85° | / | Similar |
| Depth of Field | 2~100mm | 8~19mm | / | Different |
| Endoscope Illumination | LED light is placed at the distal end of the endoscope | LED light is placed at the distal end of the endoscope | / | Same |
| Method of Sterilization | EO | EO | / | Same |
Similar- Maximum Outer Diameter of Insertion Section
The maximum outer diameter of insertion part of subject device is similar as the predicate device. Meanwhile, based on ISO 8600 Testing, the test results of maximum insertion portion width demonstrate the subject device meets the product design requirements. Therefore, this item will not affect the safety and effectiveness of the subject device.
Different- Bending angle
The bending angle of subject device is similar as the predicate device. Meanwhile, based on Function Testing of Single-use Flexible Video Bronchoscope, the test results of bending angle demonstrate the subject device meets the product design requirements. Therefore, this item will not affect the safety and effectiveness of the subject device.
Same- Rotary function
The subject device has an extra rotary function than the predicate device aScope 3, but the rotary function of subject device is same as the aScope 5 of Reference Device K233671. Therefore, the difference will not affect the safety and effectiveness of the subject device.
Similar- Field of View
{12}
The field of view of subject device is similar as the predicate device. Meanwhile, based on ISO 8600 Testing, the test results of field of view demonstrate the subject device meets the product design requirements. Therefore, this item will not affect the safety and effectiveness of the subject device.
# Different- Depth of Field
The depth of field of subject device is different from the predicate device. Meanwhile, based on Resolution and Depth of Field Testing, the subject device has good resolution in the intended depth of field. Therefore, this item will not affect the safety and effectiveness of the subject device.
Table. 3 Safety Comparison
| ITEM | Subject Device | Predicate Device K173727 | Remark |
| --- | --- | --- | --- |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Particular requirements | Comply with IEC 60601-2-18 | Comply with IEC 60601-2-18 | Same |
| Patient Contact Material | | | |
| Working channel port | PP | Unknown | Different |
| Suction connector | PP | | |
| Insertion cord (main hose) | Pebax | | |
| Bending section | Silicone rubber Gray color masterbatch | | |
| Working channel exit | TPU | | |
| Camera Lens | Glass | | |
| Transparent cover | PC | | |
| Camera seat | PC with 1% Black color masterbatch | | |
| Two-way stopcock | PC | | |
| Luer taper | PP | | |
| Biopsy cap | Silicone rubber | | |
| Snake bone rubber | Silicone rubber Gray color masterbatch | | |
| Blocking clamp tube | TPU | | |
| Water gas 3-way tube | PP | | |
| Water gas extension tube | TPU | | |
| Device extension tube | Pebax | | |
| Device 3-way | PP | | |
| Multi-lumen tube | Pebax | | |
{13}
| Steel ring | SUS 304 | | |
| --- | --- | --- | --- |
| Working channel tube | TPU | | |
| Biocompatibility | No Cytotoxicity | No Cytotoxicity | Same |
| | No Sensitization | No Sensitization | |
| | No Irritation | No Irritation | |
| Sterilization (Single-use Broncho Videoscope) | | | |
| Method | EO sterilization | EO sterilization | Same |
| SAL | 10^{-6} | 10^{-6} | Same |
## Different - Patient-contact material
The patient-contact material of the predicate device is unknown. However, the biocompatibility tests were conducted on the subject device and the test results show that the patient-contact materials of the subject device do not raise the adverse effect on the patient. The subject device is biocompatible and conforms to ISO 10993 series standards. Therefore, the difference on the patient-contact material will not affect the safety and effectiveness of the subject device.
## 8. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the subject device met all design specifications and was Substantially Equivalent (SE) to the predicate device.
## Performance testing:
The optical performance comparison test was conducted on the aged and un-aged Single-use Flexible Video Bronchoscope and un-aged predicate device. The test results demonstrate that the optical performance of the un-aged endoscope is similar to those of the predicate device, and those of aged endoscope is similar as those of the un-aged endoscope and the optical performance of subject device complies with the product design requirements.
The subject device was tested per the following standard, to evaluate Surface and edges, Maximum insertion portion width, Minimum instrument channel width and FOV.
- ISO 8600-1:2015 Endoscopes - Medical endoscopes and endotherapy devices - Part 1: General requirements
- ISO 8600-3: 2019 Endoscopes - Medical endoscopes and endotherapy devices - Part 3: Determination of field of view and dire
- ISO 8600-4:2014 Endoscopes - Medical endoscopes and endotherapy devices - Part 4: Determination of maximum width of insertion portion
- CIE ISO 11664-2:2022 Colorimetry - Part 2: CIE standard illuminants
- CIE ISO 11664-6:2022 Colorimetry - Part 6: CIEDE2000 colour-difference formula
The test results demonstrate that the subject device complies with the product design requirements.
Page 10 of 13
{14}
The air/water supply test has been conducted on the subject device and predicate device. The test results demonstrate that air/water supply performance of the subject device is similar to those of the predicate device and the air/water supply performance of subject device complies with the product design requirements.
The leakage test of the air/water connector has been conducted on the subject device. The test results demonstrate that the subject device complies with the product design requirements.
The subject device is tested per IEC 62471:2006 Photobiological safety of lamps and lamp systems standard, to evaluate photobiological safety. The test result demonstrated that the subject system complies with the standard requirements.
## Sterilization:
EO/ECH residual testing was conducted on the subject device per ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
## Shelf life:
A maximum shelf life of 2 years has been assigned to the subject endoscope. In order to demonstrate substantial equivalence between the subject and predicate device, the sponsor conducted accelerated aging on the subject device and performed the following tests on the accelerated aged device to provide data for the device shelf life:
- ASTM F1980-2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- Package integrity of aged device;
- Product performance testing
## Packaging integrity:
The package integrity performances of the subject sterility maintenance package include visual inspection, seal strength and dye penetration resistance. The test results demonstrated that the subject device complies with the following standards:
- ASTM F88/F88M-21, Standard Test Method for Seal Strength of Flexible Barrier Materials.
- ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration.
- ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
## Shipping:
A simulated transportation test was conducted on the subject device. The test results demonstrated that the subject device complies with the following standards:
- ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
Page 11 of 13
{15}
Page 12 of 13
# Biocompatibility
Biocompatibility of the subject Himaging Video Bronchoscope System was evaluated in accordance with the FDA guidance “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. The test results demonstrated that the subject device complies with the following standards:
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device Himaging Video Bronchoscope System. The test results demonstrated that the subject device complies with the following standards:
- IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC60601-2-18 :2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-1-2 :2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
# Software Verification and Validation Testing
Software verification and validation testing were conducted on the subject device. The test results demonstrated that the subject device complies with the FDA’s Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” The software for this device was considered as a “moderate” level of concern.
# Cybersecurity testing
Cybersecurity testing was conducted on the subject device. The test results demonstrated that the subject device complies with the FDA’s Guidance entitled “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” Document issued on June 27, 2025.
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
# 9. Clinical Test Conclusion
No clinical study is included in this submission.
{16}
10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.