K251890 · Zhejiang Chuangxiang Medical Technology Co., Ltd. · FAD · Feb 20, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K251890
Device Name
Disposable Ureteral Stents
Applicant
Zhejiang Chuangxiang Medical Technology Co., Ltd.
Product Code
FAD · Gastroenterology, Urology
Decision Date
Feb 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4620
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Disposable Ureteral Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant and post-traumatic conditions. The stents are placed using endoscopic technique. The indwelling time should not exceed 30 days.
Device Story
Flexible, tubular double-pigtail ureteral stent; radiopaque polycarbonate-based polyurethane with hydrophilic coating; used for temporary internal drainage from ureteropelvic junction to bladder. Placed via endoscopic technique by urologists; indwelling time limited to 30 days. Available in 4.0-8.0 Fr diameters; lengths 14-28 cm. Supplied sterile with pusher tube; optional guide wire, clamp, or silk thread included. Relieves ureteral obstruction; benefits patient by maintaining urine flow. Silk thread removal required after 14 days.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing (flow rate, break strength, pigtail retention, frictional force) and biocompatibility testing (cytotoxicity, sensitization, reactivity, toxicity, pyrogenicity, genotoxicity, implantation).
Technological Characteristics
Radiopaque polycarbonate-based polyurethane with hydrophilic coating. Double pigtail configuration. Dimensions: 4.0-8.0 Fr, 14-28 cm. Sterilized via EtO (ISO 11135:2014). Testing per ASTM F1828-22. Single-use.
Indications for Use
Indicated for temporary internal drainage from the ureteropelvic junction to the bladder in patients (including pediatric) with ureteral obstruction due to benign, malignant, or post-traumatic conditions.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Universa™ Firm Ureteral Stents and Stents Sets (K161236)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 20, 2026
Zhejiang Chuangxiang Medical Technology Co., LTD.
Lucius Long
RA Manager
Room 101-1, 201-1, 301, 401, Building 50
No.650 Hongfeng Road, Donghu Street, Linping District
Hangzhou, 311100
CHINA
Re: K251890
Trade/Device Name: Disposable Ureteral Stents (Types C and D)
Regulation Number: 21 CFR 876.4620
Regulation Name: Ureteral Stent
Regulatory Class: II
Product Code: FAD
Dated: January 20, 2026
Received: January 20, 2026
Dear Lucius Long:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251890 - Lucius Long
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251890 - Lucius Long
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Negeen Haghighi -S
for
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251890 | |
| Device Name Disposable Ureteral Stents (Types C and D) | |
| Indications for Use (Describe) The Disposable Ureteral Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant and post-traumatic conditions. The stents are placed using endoscopic technique. The indwelling time should not exceed 30 days. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K251890
Page: 1 / 5
# 510(k) Summary
## I. Submitter information
510(k) Submitter: Zhejiang Chuangxiang Medical Technology Co., LTD.
Address: Room 101-1, 201-1, 301 & 401, Building 50
No.650 Hongfeng Road, Donghu Street
Linping District, Hangzhou, 311100
Zhejiang Province, P.R. China
Contact Person: Lucius Long
Telephone: 86-571-89167088
Fax: 86-571-89167086
Email: lucius.long@med-nova.com
Preparation Date: February 19, 2026
## II. Subject Device
Common Name: Ureteral Stent
Trade Name of Device: Disposable Ureteral Stents (Types C and D)
Classification Name: Stent, Ureteral
Regulation Class: Class II
Regulation Number: 21 CFR 876.4620
Review Panel: Gastroenterology/Urology
Product Code: FAD
## III. Predicative and Reference Devices
Predicate Device:
510(k) Number: K161236
510(k) Holder: Cook Incorporated
Trade Name of Device: Universa™ Firm Ureteral Stents and Stents Sets
Classification Name: Stent, Ureteral
Regulation Class: Class II
Regulation Number: 21 CFR 876.4620
Review Panel: Gastroenterology/Urology
Product Code: FAD
{5}
科技医学
MEDNOVA
Zhejiang Chuangxiang Medical Technology Co., LTD.
K251890
Page: 2 / 5
Reference Device:
510(k) Number: K243830
510(k) Holder: Shenzhen Trious Medical Technology Co., Ltd
Trade Name of Device: Disposable Ureteral Stent
Classification Name: Stent, Ureteral
Regulation Class: Class II
Regulation Number: 21 CFR 876.4620
Review Panel: Gastroenterology/Urology
Product Code: FAD
## IV. Device Description
The Disposable Ureteral Stents are a set of ureteral stents used for temporary internal drainage from the ureteropelvic junction to the bladder. The subject ureteral stent is a flexible, tubular double pigtail stent composed of radiopaque polycarbonate-based polyurethane with hydrophilic coating. The subject stents include two stent types: Type C and Type D stents. The only difference between the two types of stents is the opening at the stent end (Type C - single opening at one end only vs. Type D - double openings with one at each end).
The Disposable Ureteral Stents are available in 4.0 to 8.0 French (Fr) diameter, with lengths ranging from 14.0 to 28.0 centimeters (cm) (4Fr × 14 cm for pediatric patients and other sizes for adults). The stents are supplied sterile, packaged with a small pusher tube. Depending on configurations, the device package may include an optional guide wire, clamp or silk thread in addition to the stent and pusher tube.
The Disposable Ureteral Stents are not intended for long-term use. The stents are labeled for indwell time not to exceed thirty (30) days and no repeated use. The Silk thread should be removed after 14 days.
## V. Indications for Use
The Disposable Ureteral Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant and post-traumatic conditions. The stents are placed using endoscopic technique. The indwelling time should not exceed 30 days.
{6}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K251890
Page: 3 / 5
VI. Comparison of Technological Characteristics with Predicate Device
| Device Characteristics | K251890 Disposable Ureteral Stents (Types C and D) (Subject Device) | K161236 Universal™ Firm Ureteral Stents and Stent Sets (Predicate Device) |
| --- | --- | --- |
| Manufacturer | Zhejiang Chuangxiang Medical Technology Co., LTD. | Cook Incorporated |
| Regulation No. | 21 CFR 876.4620 | 21 CFR 876.4620 |
| Product Code | FAD | FAD |
| Classification | Class II | Class II |
| Indications For Use | The Disposable Ureteral Stents are used for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant and post-traumatic conditions. The stents are placed using endoscopic technique. The indwelling time should not exceed 30 days. | The Universal® Firm Ureteral Stents and Stent Sets are intended for temporary internal drainage from the ureteropelvic junction to the bladder. Ureteral stents have been used to relieve obstruction in a variety of benign, malignant, and post-traumatic conditions. The stents may be placed using endoscopic, percutaneous, or open surgical techniques. |
| Labeled Indwelling Time | • Not to exceed thirty (30) days for stent
• Thread removed after 14 days | • Not to exceed 12 months for stent
• Tether removed after 14 days |
| Device Components | • Double pigtail ureteral plastic stent (radiopaque marks)
• Pusher tube
• Guidewire (optional)
• Clamp (optional)
• Thread (optional) | • Double pigtail ureteral plastic stent (radiopaque marks) with tethers and pigtail straightener.
• Positioner with radiopaque tip
• Guidewire |
| Stent Material | Polyurethane Hydrophilic coating | Polyurethane Hydrophilic coating |
| Stent Diameter* | 4.0Fr, 4.5Fr, 4.7Fr, 5.0Fr, 6.0Fr, 7.0Fr, 8.0Fr | 5.0Fr, 6.0Fr, 7.0Fr, 8.0Fr |
| Stent Length (cm) | 14, 16, 18, 20, 22, 24, 26, 28 cm | 18, 20, 22, 24, 26, 28, 30, 32 cm |
| Stent Coils | Double pigtails (one at each end) | Double pigtails (one at each end) |
| Pusher Tube/Positioner | • Diameter: 4.0Fr, 4.8Fr, 6.3Fr
• Length: 70, 60, 40cm | • Diameter: 5Fr, 6Fr, 7Fr, 8Fr
• Length: not specified |
| Guide Wire | • Diameter: 0.028” for 4.0Fr and 4.5Fr; 0.032” for 4.7Fr and | • Diameter: 0.035” for 5Fr; 0.038” for 6Fr, 7 Fr and 8 Fr |
{7}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K251890
Page: 4 / 5
| | 5.0Fr; 0.035” for 6.0Fr, 7.0Fr and 8.0Fr
• Length: 150 cm | • Length: not specified |
| --- | --- | --- |
| Sterilization | EtO | EtO |
| Single Use | Yes | Yes |
*The 4.0Fr stent is intended for pediatric patients and the reference device includes this size.
As evidenced by the above table, both the subject and predicate devices have similar intended use, but they have different technological characteristics. Non-clinical testing was conducted on the subject stents, and it was established that the differences in technological characteristics between the subject and predicate devices do not raise different questions of safety or effectiveness.
# VII. Non-Clinical Data
The Disposable Ureteral Stents were tested and demonstrated to be substantially equivalent to the predicate device in safety and performance.
## Biocompatibility Testing:
Biocompatibility of the Disposable Ureteral Stents was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process'" (September 4, 2020). The following biological endpoints were addressed:
- Cytotoxicity
- Sensitization
- Intracutaneous Reactivity
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity
- Subacute Systemic Toxicity
- Muscle Implantation
- Genotoxicity
## Sterility and Shelf Life:
The subject device is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 (ISO 11135:2014) and the device packaging was tested for integrity in maintaining a sterile barrier over a shelf life of 2 years. An accelerated aging study and simulated transportation study were conducted on representative device models to demonstrate that the subject device was able to maintain the device performance over a shelf life of 2 years.
{8}
科技医学
Zhejiang Chuangxiang Medical Technology Co., LTD.
K251890
Page: 5 / 5
# Bench Performance Testing:
A battery of bench testing based on the FDA guidance “Guidance for the Content of Premarket Notification for Ureteral Stents” (1993) and ASTM 1828-22 “Standard specification for ureteral stents” was conducted on the subject and predicate stents using established methods to demonstrate the substantial equivalence to the predicate device.
The performance evaluations included:
- Appearance and dimensions
- Flow rate
- Break strength and elongation
- Pigtail retention strength
- Dynamic frictional force
- Ink adherence
- Guidewire performance and compatibility
- Pusher tube performance and compatibility
- Clamp strength
- Silk thread strength and attachment performance
# VIII. Clinical Testing
No clinical study is included in this submission.
# IX. Conclusions
The proposed device has the similar indications for use and has similar design features and technological characteristic as the predicate device. Non-clinical testing data demonstrates that the proposed device is as safe and effective as the predicate device. Therefore, the subject device is substantially equivalent to the predicate device Universa™ Firm Ureteral Stents and Stents Sets.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.