K251862 · Paragon 28, Inc. · KTT · Aug 11, 2025 · Orthopedic
Device Facts
Record ID
K251862
Device Name
External Fixation Bone Distractor
Applicant
Paragon 28, Inc.
Product Code
KTT · Orthopedic
Decision Date
Aug 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The External Fixation Bone Distractor is intended for treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities. The External Fixation Bone Distractor is indicated for adult and pediatric (greater than 2 through 21 years of age) patients.
Device Story
Transverse bone transport system; assists controlled movement of bone segment across defect. Used independently or with Monkey Rings (K232838) or Monkey Bars (K242452) systems to form hybrid frames. Device consists of distractor body with 100mm rod, connection bolts, and half-pins. Operated by orthopedic surgeons in clinical settings. Facilitates gradual distraction of tibial bone segment. Serrated connection bolts prevent rotational movement/loosening. Surgeon adjusts frame to accommodate patient anatomy. Output is mechanical distraction force; aids bone healing/correction of deformities.
Clinical Evidence
No clinical data. Bench testing only; performance evaluated via worst-case analysis of mechanically demanding configurations per ASTM F1541-17.
Indicated for adult and pediatric patients (>2 to 21 years) for treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Monkey Bars Pin to Bar External Fixation System (K242452)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 11, 2025
Paragon 28, Inc.
Jacqueline Sloan
Regulatory Affairs Specialist I
14445 Grasslands Drive
Englewood, Colorado 80112
Re: K251862
Trade/Device Name: External Fixation Bone Distractor
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: KTT
Dated: June 13, 2025
Received: June 17, 2025
Dear Jacqueline Sloan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251862 - Jacqueline Sloan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251862 - Jacqueline Sloan
Page 3
Sincerely,
Lixin Liu-S
Lixin Liu, Ph.D
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
External Fixation Bone Distractor
Page 7 of 28
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251862 | ? |
| Please provide the device trade name(s). | | ? |
| External Fixation Bone Distractor | | |
| Please provide your Indications for Use below. | | ? |
| The External Fixation Bone Distractor is intended for treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities.
The External Fixation Bone Distractor is indicated for adult and pediatric (greater than 2 through 21 years of age) patients. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251862 Page 1 of 5
510(K) SUMMARY
Manufacturer: Paragon 28, Inc.
14445 Grasslands Dr.
Englewood, CO 80112
Contact: Jacqueline Sloan
Regulatory Affairs Specialist I
Paragon 28, Inc.
14445 Grasslands Dr.
Englewood, CO 80112
Phone: 720-617-6843
Jackie.Sloan@zimmerbiomet.com
Date Prepared: August 08, 2025
Device Trade Name: External Fixation Bone Distractor
Device Class and
Common Name: Class II, appliance, fixation, nail/blade/plate combination, multiple component
Classification: 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and accessories.
Product Code: KTT
Indications for Use: The External Fixation Bone Distractor is intended for treatment of non-union or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities.
The External Fixation Bone Distractor is indicated for adult and pediatric (greater than 2 through 21 years of age) patients.
Device Description: The Paragon 28 External fixation Bone Distractor is a transverse bone transport system designed to assist in the controlled movement of a bone segment across a defect. The device can be used independently or in conjunction with the Monkey Rings External Fixation System (K232838) or
{5}
K251862 Page 2 of 5
Monkey Bars Pin to Bar External Fixation System (K242452) to form hybrid frames.
**Primary Predicate:** Truelok Elevate (K242861)
**Reference Device A:** Monkey Rings External Fixation System (K232838)
**Reference Device B:** Monkey Bars Pin to Bar External Fixation System (K242452)
**Technological Comparison:**
The subject device, the External Fixation Bone Distractor, and the predicate device, the TrueLok™ Elevate, share comparable technological characteristics. Both devices are intended for treating extremity conditions by facilitating the gradual formation and distraction of a tibial bone segment over a period of several days. Each incorporates serrated connection bolts to prevent rotational movement and loosening during clinical use. Additionally, both systems support customizable frame configurations to accommodate individual patient anatomy and clinical needs.
The subject device features a distractor with a travel length of 17 mm, compared to 14 mm in the predicate device. This minor difference does not raise different questions of safety or effectiveness, as both devices perform the same function of controlled bone segment transport.
The subject device features a distractor body with a 100mm long, 11mm diameter rod. Reference device A has 11mm threaded pillar lengths 30mm-300mm. The subject device and Reference Device A utilize the same threaded connection for connecting to circular rings.
Regarding fixation, the subject device accommodates half pins with thread diameters of 3 mm, 4 mm, and 5 mm, all with blunt tips, and lengths ranging from 150 mm to 200 mm. The predicate device uses 4 mm diameter half pins, 120 mm in length, with an 18 mm threaded portion.
These dimensional and design differences are minor and do not affect the intended use, performance, or safety profile of the device. The subject device utilizes the same half-pins as Reference device B. The subject device also uses the same pin clamping methods as Reference device B. Therefore, the pins do not raise different questions of safety or effectiveness.
{6}
K251862 Page 3 of 5
Materials used in the subject device include stainless steel (ASTM F138 / AISI 316LVM) for the half pins—matching the predicate—as well as stainless steels (AISI 303, 304, 302, 420, 630), aluminum alloy (EN-AW 6082 T6), and Polyphenylsulfone (PPSU) polymer for frame components. Patient-contacting instruments are also made from stainless steel. All materials are biocompatible and commonly used in orthopedic devices.
{7}
K251862 Page 4 of 5
| Feature | Similarities | Differences | Explanation for Differences |
| --- | --- | --- | --- |
| Intended Use | Both the subject device and predicate (TrueLok Elevate, K242861) are designed to treat extremity conditions by facilitating the formation and gradual distraction of a tibial bone segment. | None | N/A |
| Connection Bolts | Both the subject device and predicate (TrueLok Elevate, K242861) incorporate serrated connection bolts to prevent rotational movement and loosening during clinical use. | None | N/A |
| Frame Configurations | Both the subject device and predicate (TrueLok Elevate, K242861) support customizable frame configurations to accommodate patient anatomy and clinical needs. | None | N/A |
| Distractor Travel Length | Both the subject device and predicate (TrueLok Elevate, K242861) perform controlled bone segment transport via distractors. | Subject device travel length: 17 mm; Predicate device (TrueLok Elevate, K242861) travel length: 14 mm. | The 3 mm longer travel length in the subject device does not impact safety or effectiveness as the function is equivalent. |
| Distractor Rod | Both the subject device and Reference Device A (K232838) have an 11 mm diameter rod. | Subject device rod length: 100 mm; Reference Device A (K232838) threaded pillar lengths range from 30 mm to 300 mm. | The difference in length allows flexibility; both use the same threaded connection ensuring mechanical compatibility. |
| Half Pins | Subject device and Reference Device B (K242452) use the same half pins. | Subject device half pins: 3, 4, and 5 mm diameters; lengths 150–200 mm, blunt tips. Predicate device (TrueLok Elevate, K242861) half pins: 4 mm diameter, 120 mm length, 18 mm threaded portion. | The dimensional differences are minor and do not affect performance, intended use, or safety. The subject device pins align with a reference device known to be safe and effective. |
| Pin Clamping Methods | Both the subject device and Reference Device B (K242452) use identical pin clamping methods. | None | N/A |
| Materials | Materials used in the subject device include stainless steel (ASTM F138 / AISI 316LVM) for the half pins—matching the predicate—as well as stainless steels (AISI 303, 304, 302, 420, 630), aluminum alloy (EN-AW 6082 T6), and PPSU polymer for frame components. | None- The stainless-steel material are the same for the pins in the subject device and the half pins for the predicate. | For the frame components. All materials are biocompatible and commonly used in orthopedic devices. |
{8}
K251862 Page 5 of 5
## Performance Testing:
No additional bench testing was performed for the subject device. Instead, a worst-case analysis was conducted using existing data from other components within the device system. The selected components represent the most mechanically demanding configurations, and their performance was evaluated in accordance with ASTM F1541-17. This approach was conducted under design control procedures, and a rationale for the selection of worst-case components was provided to ensure that the evaluation adequately addresses the mechanical performance of the entire system.
## Conclusions:
The External Fixation Bone Distractor has the same intended use and similar technological characteristics as the predicate device, the TrueLok™ Elevate (K242861). Differences in design do not raise different questions of safety or effectiveness.
No additional bench testing was performed. Instead, a worst-case analysis was conducted using existing data from mechanically demanding components within the system. Testing was performed per ASTM F1541-17 under design control procedures, with rationale provided to support the selected configurations.
The subject device is therefore considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.