K251835 · Neocis, Inc. · PLV · Oct 10, 2025 · Dental
Device Facts
Record ID
K251835
Device Name
Yomi Robotic System
Applicant
Neocis, Inc.
Product Code
PLV · Dental
Decision Date
Oct 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4120
Device Class
Class 2
Attributes
AI/ML, Therapeutic
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Automatic segmentation of maxillary sinuses, inferior alveolar nerve, and maxillary and mandibular bone
—
—
—
—
—
—
—
Indications for Use
The Yomi Robotic System (Yomi) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.
Device Story
Yomi Robotic System is a computerized dental stereotaxic instrument for implant surgery and bone reduction. Input: CBCT scan data. Process: YomiPlan software creates 3-D anatomical models; integrates automatic segmentation of maxillary sinuses, inferior alveolar nerve, and bone. Surgeon uses software to plan implant placement; system provides physical, visual, and audible feedback during surgery. Robotic arm holds and guides third-party powered bone cutting instruments. Used in clinical settings by dentists. Benefits: precise navigational guidance for implant site preparation and bone reduction. System includes patient tracking via splints (teeth or bone-mounted) and trackers. Modifications include restorative planning features, dual arch planning, and improved alignment/guidance algorithms. Output: real-time guidance during surgical procedures.
Clinical Evidence
Bench testing only. Software verification and validation conducted per IEC 62304 and FDA guidance. Usability validation performed per human factors guidance. Performance testing confirms substantial equivalence to the predicate.
Technological Characteristics
Dental stereotaxic instrument; robotic arm; PC-based software (Windows 10); connectivity via Ethernet/Wi-Fi/USB. Sensing: patient tracking via splints and trackers. Software: non-adaptive, locked-weight segmentation algorithm. Sterilization: not specified (standard dental instrument accessories).
Indications for Use
Indicated for partially and fully edentulous adult patients qualifying for dental implants; used for pre-operative planning and intra-operative robotic navigational guidance for dental implantation and guided bone reduction (alveoplasty) of the mandible or maxilla.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Neocis, Inc.
Jorge Fernandes
VP Quality and Regulatory
545 NW 26th Street
Miami, Florida 33127
October 10, 2025
Re: K251835
Trade/Device Name: Yomi Robotic System
Regulation Number: 21 CFR 872.4120
Regulation Name: Bone Cutting Instrument And Accessories
Regulatory Class: Class II
Product Code: PLV, QRY
Dated: June 13, 2025
Received: September 19, 2025
Dear Jorge Fernandes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251835 - Jorge Fernandes
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K251835 - Jorge Fernandes
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Yomi Robotic System
Page 11 of 83
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251835 | ? |
| Please provide the device trade name(s). | | ? |
| Yomi Robotic System | | |
| Please provide your Indications for Use below. | | ? |
| The Yomi Robotic System (Yomi) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
yomi. BY NEOCIS
# 510(k) Summary K251835
## I. Submitter
Neocis, Inc.
545 NW 26th Street
Unit 700
Miami, FL 33127
Tel: 1-855-9NEOCIS
Contact Person: Jorge Fernandes, VP, QA-RA
Date Prepared: October 10, 2025
## Device
Trade Name: Yomi Robotic System
Common Name: Dental Stereotaxic Instrument
Classification Name: Bone cutting instrument and accessories (21 CFR 872.4120)
Classification: Class II
Product Code: PLV, QRY
## Predicate Devices
Primary Predicate: Yomi Robotic System (K231018)
Reference Device: Relu Creator (K233925)
## Indications for Use
The Yomi Robotic System (Yomi) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of Yomi Plan is to be used with the Yomi Robotic System.
## Device Description
Yomi Robotic System is a dental stereotaxic instrument and a powered surgical device for bone cutting. Yomi Robotic System is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The Yomi Robotic System is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.
The Yomi Robotic System allows the user to plan the surgery virtually in YomiPlan, cleared for use alone on third-party PCs for preplanning. The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in the planning software. The plan is used for the system to provide physical, visual, and audible feedback to the
Page 1 of 6
{5}
yomi. BY NEOCIS
surgeon during the implant site preparation. The Yomi robotic arm holds and guides a standard FDA-cleared third party powered bone cutting instrument.
The patient tracking portion of Yomi is comprised of linkages from the patient to Yomi, which include the Patient Splint (YomiLink Teeth or YomiLink Bone), Tracker End Effector (TEE), and the Patient Tracker (PT). In cases where YomiLink Teeth is utilized, it is attached to the contralateral side of the patient's mouth over stable teeth using on-label dental materials prior to the presurgical CBCT scan. In cases where YomiLink Bone is utilized, it is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required), or after the scan when using the subject YomiLink Arch device.
The subject of this submission is to: Integrating algorithms that provide automatic segmentation of maxillary sinuses, inferior alveolar nerve, and maxillary and mandibular bone. The integrated software, Relu Creator, was cleared in K233925. The software is not adaptive, it is trained at the manufacturer (Relu), and the weights are locked.
Additionally, since the most recent clearance of Yomi Robotic System (K231018), minor modifications to the Yomi System include the following:
- Planning software improvements
- Restorative planning – Features to support customized crown design
- Dual arch planning – Feature to enable the end user to plan multiple arches in a single case and a single scan
- Patient work volume guidance improvements – Added guidance for the angulation of the patient chair
- Added patient proximity for baseline
- YomiLink Bone (YLB) planning – improved placement of the YLB
- Added proximity threshold lower limit value
- Improved alignment between CT scans and imported .stl objects
- Added ability for user to designate soft tissue thickness to assist in bone reduction planning
- Added max depth information to the implant cursor hover info
- VTK Off-the-Shelf software version update
- Added model details to implant selection
- Added restorative planning case feedback option
- Added additional implant models to the implant library
- Control software and behavior improvements
- Updates to handpiece interaction gestures, and optimization of the response of the control software to guide arm joint limits, singularities and potential wrist / base collisions.
- Hardware improvements Tracker Arm Joint
- Accessory improvements
- Updates to the YomiLink Teeth and intraoral fiducial array
- Minor bug fixes
All other aspects of the Yomi Robotic System remain unchanged from prior clearances.
## Comparison of Technological Characteristics
The following Table 1 provides a summary of the subject Yomi Robotic System features compared to the predicate device, Yomi Robotic System (K231018) and reference device Relu Creator (K233925
Page 2 of 6
{6}
yomi.
BY NEOCIS
Table 1: Comparison of technological characteristics to the predicate (with a supporting reference device)
| Feature | Subject Device: Yomi Robotic System with Automatic Segmentation Algorithm | Primary Predicate: Yomi Robotic System with YomiLink Arch K231018 | Reference Device: Relu Creator (K23392S) | Justification of Differences |
| --- | --- | --- | --- | --- |
| Device Name | Yomi Robotic System | Yomi Robotic System | Relu Creator | Equivalent |
| Device Classification | Class II | Class II | Class II | Equivalent |
| Indications for use | The Yomi Robotic System (Yomi) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third-party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides preoperative planning for dental implantation procedures using the Yomi Robotic System. The output of YomiPlan is to be used with the Yomi Robotic System. | Yomi Robotic System is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi Robotic System is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third-party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides preoperative planning for dental implantation procedures using the Yomi Robotic System. The output of YomiPlan is to be used with the Yomi Robotic System. | Relu Creator is a software program for the management, transfer, and analysis of dental and craniomaxillofacial image information, and can be used to provide design input for dental solutions. It displays and enhances digital images from various sources to support the diagnostic process and treatment planning. It stores and provides these images within the system or across computer systems at different locations. | Equivalent The subject device has the same Indications for Use as the predicate device. The Reference device is a software application that is part of the digital workflow of dental specialists in preoperative planning. |
| Intended User | Dentist | Dentist | Dentist | Equivalent |
| Use environment | Clinical Setting, Doctors Office | Clinical Setting, Doctors Office | N/A | Equivalent |
| Yomi Plan Software Version | v2.7 | v2.4.1 | N/A | Equivalent The only difference is a version change due to the addition of Automatic Segmentation Algorithm functionality and minor LTF changes since the last K231018 clearance. All changes have been successfully verified and, therefore, not considered to affect the overall safety and efficacy profile of Yomi Plan. |
{7}
yomi.
BY NEOCIS
| Feature | Subject Device: Yomi Robotic System with Automatic Segmentation Algorithm | Primary Predicate: Yomi Robotic System with YomiLink Arch K231018 | Reference Device: Relu Creator (K233925) | Justification of Differences |
| --- | --- | --- | --- | --- |
| OS | Windows 10 | Windows 10 | Windows 10 or 11 | Equivalent |
| PC Requirements | PC with 64-bit Windows 10 OS or newer with a minimum of 4 GB of RAM and a 2 GHz dual core processor. Local memory (hard drive) should be a minimum of 100 GB with 7200 RPM or SSD. Connectivity requirements include ethernet, Wi-Fi, USB, or CD drive. | PC with 64-bit Windows 10 OS or newer with a minimum of 4 GB of RAM and a 2 GHz dual core processor. Local memory (hard drive) should be a minimum of 100 GB with 7200 RPM or SSD. Connectivity requirements include ethernet, Wi-Fi, USB, or CD drive. | Windows 10 or 11 with minimum 8 GB of RAM and Intel i5 or Ryzen 5 with Nvidia GPU 4GB VRAM (with CUDA drivers of at least 450.80.02 on Linux and 452.39 on Windows) using Pascal architecture or newer, Ubuntu 20.04 or newer (or an equivalent distribution with glibc 2.28 or newer) Client Hosted Servers: Ubuntu 20.04 or newer (or an equivalent distribution with glibc 2.28 or newer) | Equivalent The software has been successfully verified to perform with the PC specifications of the Yomi Robotic System. Therefore, the difference is considered not to affect the overall safety and efficacy profile of Yomi Plan. |
| Yomi Plan Functions | • Load CT Scanned Image
• Optimize Image
• Plan Procedure (place implant)
• Save Surgical Plan
• Connect to Control software
• Provide Feedback to Surgeon regarding physical location of Drill and Drill components
• Select Surgical Phase
• Set areas for mechanical restriction during surgical operation
• Visualize CT Scanned Image with 2D Slices
• Generate Panoramic reconstruction along arch
• Visualize Panoramic reconstruction with cross sections along panoramic arch
• Map Splint coordinate system to structures in CT Scan
• Define anatomical planes
• Clip CT Scanned Images
• Define Arch for generating panoramic reconstruction
• Provide the user with a means to define a nerve
• Allow the user to plan multiple implants
• Measure distances and angles in the plan
• Integrated Automatic Segmentation Algorithm. | • Load CT Scanned Image
• Optimize Image
• Plan Procedure (place implant)
• Save Surgical Plan
• Connect to Control software
• Provide Feedback to Surgeon regarding physical location of Drill and Drill components
• Select Surgical Phase
• Set areas for mechanical restriction during surgical operation
• Visualize CT Scanned Image with 2D Slices
• Generate Panoramic reconstruction along arch
• Visualize Panoramic reconstruction with cross sections along panoramic arch
• Map Splint coordinate system to structures in CT Scan
• Define anatomical planes
• Clip CT Scanned Images
• Define Arch for generating panoramic reconstruction
• Provide the user with a means to define a nerve
• Allow the user to plan multiple implants
• Measure distances and angles in the plan | The Relu Creator is a software that is part of the digital workflow of dental specialists in preoperative planning. The main purpose is the 3D modeling of the patient anatomy, which is technically called image segmentation and multimodel registration. Based on the model, simulations for preoperative and pretreatment planning can be carried out for dental applications. The 3D modeling (segmentation + registration) is performed on medical images like CBCT, IOS and FS. The preoperative/pretreatment software can be applied in various dental disciplines such as orthodontics, implantology, and maxillofacial surgery. | Equivalent The only change is the addition of Automatic Segmentation Algorithm functionality. Yomi Plan 2.7 with Automatic Segmentation Algorithm functionality was successfully verified and user validated. |
| Software Level of Concern | Enhanced | Enhanced | N/A | Equivalent |
| Installation | Windows Installer .msi file | Windows Installer .msi file | N/A | Equivalent |
Page 4 of 6
{8}
yomi.
BY NEOCIS
| Feature | Subject Device: Yomi Robotic System with Automatic Segmentation Algorithm | Primary Predicate: Yomi Robotic System with YomiLink Arch K231018 | Reference Device: Relu Creator (K233925) | Justification of Differences |
| --- | --- | --- | --- | --- |
| Origin | Proprietary | Proprietary | 3rd Party | Equivalent |
| Software | TeamViewer | TeamViewer | Relu® Engine | Equivalent |
| Yomi Hardware | No changes (activate/use pre-existing network features) | No changes (activate/use pre-existing network features) | N/A | Equivalent |
| Wireless data transmission over LAN | Yes, via integrated hardware, tested according to: • AAMI TIR69: 2017 Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems. • IEEE ANSI C63.27-2017 American National Standard for Evaluation of Wireless Coexistence | Yes, via integrated hardware, tested according to: • AAMI TIR69: 2017 Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems. • IEEE ANSI C63.27-2017 American National Standard for Evaluation of Wireless Coexistence | N/A | Equivalent |
| Interface | Windows based GIU | Windows based GIU | Windows based GIU | Equivalent |
| Wi-Fi | Always active | Always active | N/A | Equivalent |
# Performance Testing
Software and Cybersecurity Verification and Validation Testing
- Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions" published (June 2023). The software for this device was considered an "enhanced" level of concern, since a failure or latent flaw in the software could directly result in result in minor injury to the patient or operator.
- Software Development and Testing was performed per IEC 62304 Edition 1.1 2015 Medical Device Software – Software Lifecycle Processes, FDA Guidance for the Content of Premarket Submissions for Device Software Functions (June 2023), and FDA General Principles of Software Validation; Final Guidance for Industry and FDA Staff (January 2002). The combined testing and analysis of results provides assurance that the device performs as intended.
- Cybersecurity development and testing was performed in accordance with FDA's Guidance for Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (June 2025) and Postmarket Management of Cybersecurity in Medical Devices (December 2016).
# User Validation testing
Usability validation testing has been performed in accordance with the FDA Guidance document "Applying Human Factors and Usability Engineering to Medical Devices" Final Guidance, (February 2016).
Page 5 of 6
{9}
yomi. BY NEOCIS
# Conclusion
The subject of this submission is to modify the software to include automatic segmentation of maxillary sinuses, inferior alveolar nerve, and maxillary and mandibular bone, as well as minor modifications including planning software improvements, control software and behavior improvements, hardware improvements – Tracker Arm Joint, accessory improvements and minor bug fixes. There are no changes to the intended use compared to the predicate device. There are no fundamental changes to the technology. The performance testing demonstrates substantially equivalent performance of the subject device as compared to the predicate.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.