← Product Code [OHS](/productcode/OHS) · K251825

# Jmoon FlexGlow Light Therapy Device (JD20U SP, JD20U SG) (K251825)

_Shenzhen Ulike Smart Electronics Co., Ltd. · OHS · Oct 1, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251825

## Device Facts

- **Applicant:** Shenzhen Ulike Smart Electronics Co., Ltd.
- **Product Code:** [OHS](/productcode/OHS.md)
- **Decision Date:** Oct 1, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Jmoon FlexGlow Light Therapy Device is an over the counter device that is intended: - the Red light and Infrared light is indicated for use in the treatment of full face wrinkles, - the Blue light and Red light indicated for the treatment of mild to moderate inflammatory acne.

## Device Story

Tabletop LED device; 464 LEDs; emits blue, red, and near-infrared light. User places device on table; positions face/body 1 cm from unit. Three modes: Acne defense (M01), Anti-wrinkle recovery (M02), Dual care (M03). Device delivers specific wavelengths to skin layers to produce photobiological effects. Operated by patient in home environment. Output is light energy; no data processing or clinical decision support. Benefits include reduction of fine lines, crow's feet, and inflammatory acne.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance supported by electrical safety (IEC 60601-1, -1-2, -1-11, -2-57, -2-83), photobiological safety (IEC 62471), software verification/validation, and usability engineering (IEC 60601-1-6).

## Technological Characteristics

Tabletop LED device with 464 LEDs. Wavelengths: Blue (415nm±10nm), Red (630nm±10nm), NIR (830nm±10nm). Power: 24V DC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-2-83, IEC 62471. Connectivity: None. Sterilization: Not applicable.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- BEMER Light Therapy System Applicator: B. Light Clear Evo ([K223919](/device/K223919.md))
- BEMER Light Therapy System Applicator: B.Light Restore Evo ([K210948](/device/K210948.md))
- Omnilux CLEAR™ ([K240089](/device/K240089.md))

## Reference Devices

- Face Patches (Shenzhen Kaiyan Medical Equipment Co., Ltd)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 1, 2025

Shenzhen Ulike Smart Electronics Co., Ltd.
Blue Yang
Registration Director
810, Building 1, Xunmei Science and Technology Plaza
No. 8 Keyuan Rd, Science Park Community, Yuehai Sub-District
Shenzhen,
China

Re: K251825
Trade/Device Name: Jmoon FlexGlow Light Therapy Device (JD20U SP, JD20U SG)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: June 13, 2025
Received: September 2, 2025

Dear Blue Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251825 - Blue Yang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}

K251825 - Blue Yang
Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -
S
Date: 2025.10.01
15:06:03 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251825  |   |
|  Device Name Jmoon FlexGlow Light Therapy Device (JD20U SP, JD20U SG)  |   |
|  Indications for Use (Describe) Jmoon FlexGlow Light Therapy Device is an over the counter device that is intended: - the Red light and Infrared light is indicated for use in the treatment of full face wrinkles, - the Blue light and Red light indicated for the treatment of mild to moderate inflammatory acne.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

510(k) Summary of K251825

I. Submitter

Shenzhen Ulike Smart Electronics Co., Ltd.

Address: 810, Building 1, Xunmei Science and Technology Plaza, No. 8 Keyuan Road, Science Park Community, Yuehai Sub-District, Nanshan District, Shenzhen 518000, Guangdong, P.R. China

Contact person: Blue Yang

Email: blue@ulike.com

The date the summary was prepared: 6/15/2025

II. Device

Name of Device: Jmoon FlexGlow Light Therapy Device

Model(s): JD20U SP, JD20U SG

Common or Usual Name: Light based over the counter wrinkle reduction, Over-The-Counter Powered Light Based Laser For Acne

Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology

Regulatory Class: II

Product Code: OHS, OLP

Regulation Number: 21 CFR 878.4810

III. Legally Marketed Predicate Devices

|  Legally Marketed Predicate Devices | Predicate Device 1 |   | Predicate Device 2 | Predicate Device 3  |
| --- | --- | --- | --- | --- |
|  510(k) number | K223919 |   | K210948 | K240089  |
|  Trade Name | BEMER Light Therapy System
Applicator: B. Light Clear Evo | BEMER Light Therapy System
Applicator: B.Light Restore Evo | Omnilux CLEAR™ | Face Patches  |
|  Manufacturer | BEMER Int. AG |   | Globalmed Technologies | Shenzhen Kaiyan Medical Equipment Co., Ltd  |
|  Device classification | Class II |   | Class II | Class II  |

IV. Device Description

Jmoon FlexGlow Light Therapy Device is an exclusive mirrored tabletop device that contains 464 LEDs. It utilizes LED(light-emitting diode) technology to emit energy in the blue, red and near-infrared spectra for the aim to improve the skin appearance, helping to reduce wrinkles, like fine lines and crow's feet, and treat mild to moderate inflammatory acne. The specific wavelengths

{5}

of light can act on different skin layers to produce different photobiological effects. The device is to be used at 1 cm distance. The flexible and foldable device unit is designed to be placed on a table for easy exposure to the face and body to ensure consistent light treatment. There are three modes: Acne defense Mode(M01), Anti-wrinkle recovery Mode(M02) and Dual care Mode(M03).

2 / 6

# V. Indications for Use

Jmoon FlexGlow Light Therapy Device is an over the counter device that is intended:

- the Red light and Infrared light is indicated for use in the treatment of full face wrinkles,
- the Blue light and Red light is indicated for the treatment of mild to moderate inflammatory acne.

# VI. Comparison of Technological Characteristics with the Predicate Device

Jmoon FlexGlow Light Therapy Device has the same intended use and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, Jmoon FlexGlow Light Therapy Device may be found substantially equivalent to its predicate devices.

Jmoon FlexGlow Light Therapy Device is compared with the following Predicate Devices in

|  Comparis on Items | Subject Device | Predicate Device 1 |   | Predicate Device 2 | Predicate Device 3 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  510(k) number | / | K223919 |   | K210948 | K240089 | /  |
|  Trade Name | Jmoon FlexGlow Light Therapy Device | BEMER Light Therapy System Applicator: B. Light Clear Evo | BEMER Light Therapy System Applicator: B.Light Restore Evo | Omnilux CLEAR™ | Face Patches | /  |
|  Manufactu rer | Shenzhen Ulike Smart Electronics Co., Ltd. | BEMER Int. AG |   | Globalmed Technologies | Shenzhen Kaiyan Medical Equipment Co., Ltd | /  |
|  Regulatio n number | 21 CFR 878.4810 | 21 CFR 878.4810 |   | 21 CFR 878.4810 | 21 CFR 878.4810 | Same  |
|  Product code | OHS, OLP | OLP | OHS | OLP | OHS, OLP, GEX | Same  |
|  Device classifica tion | Class II | Class II |   | Class II | Class II | Same  |
|  Indication s for use/ Intended | Jmoon FlexGlow Light Therapy Device | The device is intended for over-the-counter | Use of light based treatment to reduce wrinkles on | The Omnilux CLEAR acne facemask is | Face Patches (Model: MT-12MA)is an | Note 1  |

{6}

|  use | is an over the counter device that is intended: - the Red light and Infrared light is indicated for use in the treatment of full face wrinkles, - the Blue light and Red light is indicated for the treatment of mild to moderate inflammatory acne. | (OTC) use to treat patients with mild to moderate acne vulgaris on the face. | the face. | an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris on the face. | Over-the Counter(OTC) device intended for treatment of mild to moderate inflammatory acne. Face Patches (Model:MT-12 MC) is an Over-the Counter (OTC) device intended for treatment of mild to moderate inflammatory acne and full-face wrinkles (including periorbital wrinkles) |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Rx or OTC | OTC | OTC |   | OTC | OTC | Same  |
|  Location for use | Face and body | Face |   | Face | Face | Note 2  |
|  Power supply | Main unit:24V D.C.,1.5A Adapter Input:100-240V A.C., 50/60Hz, 1.3A Adapter Output:24V D.C., 1.5A | Control unit : Input voltage 100 -240 V AC / 50 - 60 Hz, 15 VDC / 2A Applicator: 15 V AC |   | unknown | Controller:3.6V, 65mAh lithium battery,0.234Wh | Note 3  |
|  Light source | LED | LED |   | LED | LED | Same  |
|  Wavelength | Red light: 630nm±10nm NIR light: 830nm±10nm Blue light: 415nm±10nm | Red: 645 nm ± 20 nm Blue: 465 nm ± 20 nm | IR: 860 nm ± 20 nm Red: 645 nm ± 20 nm | Red: 630nm +/- 5nm. Blue: 412.5nm +/- 7.5nm | For MT-12MA: 630+415nm For MT-12MC: 630nm(type1)+4 15nm; 830nm+630nm(t | Note 4  |

3 / 6

{7}

|   | M01: 415nm±10nm, 630nm±10nm
M02: 630nm±10nm, 830nm±10nm
M03: 415nm±10nm, 630nm±10nm, 830nm±10nm |  |  |  | ype2); 590nm |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Irradiance | M01: Blue:11.2 mW/cm²
Red: 8.2 mW/cm²
M02: IR: 7 mW/cm²
Red: 8.1 mW/cm²
M03: | Red: 0.76 mW/cm²
Blue: 1.20 mW/cm² | Red: 0.56 mW/cm²
IR: 1.40 mW/cm² | Blue:28 mw/cm²
Red:16 mw/cm²
Total:44mw/cm² | For MT-12MA: 630nm:5
415nm:25
630+415nm:30
For MT-12MC: 630nm(type1):5
415nm:25
Mode1-total:30
830nm:15
630nm(type2):2
0
Mode2-total:35
Mode3-590nm: 35 | Note 5  |
|  Treatment Time | M01(5min)+
MI02 (5min)
M01: 5min
M02: 5min
M03: 10 min
(M01(5min)+
MI02 (5min)) | 8 min per day | 8 min per day | 10 Min | For MT-12MA (630+415nm): 3 minutes per treatment
For MT-12MC (mode1 630+415nm):3 minutes per treatment
For MT-12MC (mode2 630+830nm):9 minutes per treatment
For MT-12MC (mode 3-590nm): 21 minutes per | Note 5  |

4 / 6

{8}

|   |  |  |  | treatment |   |
| --- | --- | --- | --- | --- | --- |
|  Electrical safety | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57
IEC 60601-2-83 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-57
IEC 60601-2-83 |  | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11 | Same  |
|  Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | unknown | Same  |

Note 1:
The acne removal function of Mode 1 of the subject device is consistent with predicate device 1. Mode 3 of the subject device is a continuous bonding of Mode 1 and Mode 2 so it has the functions of acne removal and wrinkle removal.

Note 2:
Though the applicable location is different from the predicate devices, the commonly recognized applicable location for LED device is mostly face or body. The applicable location of the subject device is face and body and usability evaluation has been conducted to verify them can use the device safely and effectively, so this difference will not raise any safety/ effectiveness problems.

Note 3:
Although the power supply of the subject device and predicate devices is not identical, the IEC 60601-1 test demonstrated the safety of the power adapter. The slight difference will not raise safety and effective issue.

Note 4:
The light wavelength is very similar. The wavelength of the subject device is similar to the predicate device 1. Besides, the subject device has passed IEC 62471, IEC60601-2-57 and IEC 60601-2-83 tests, so the slight difference will not raise safety and effective issue.

Note 5:
Although the “Irradiance” and “Treatment Time” are a little different from the predicate devices, the values of each treatment dose for these parameters are within the range of the predicate devices. Dose is equal to Irradiance * Time. Through calculation, it can be concluded that the dose of the subject device are within the range of the predicate devices.

Summary of performance testing
The following performance data were provided in support of the substantial equivalence determination.
1) Electrical Safety and EMC
Electrical safety and EMC testing was performed to, and passed, as per the following standards:
&gt; IEC 60601- 1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
&gt; IEC 60601- 1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard:

5 / 6

{9}

Electromagnetic disturbances - Requirements and tests

&gt; IEC 60601-1-11:2015+A1:2020 Medical Electrical Equipment–Part 1-11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
&gt; IEC 60601-2-57:2011 Medical electrical equipment–Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
&gt; IEC 60601-2-83:2019+2022 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment

2) Light Safety
&gt; IEC 62471:2006 Photobiological safety of lamps and lamp systems

3) Software Verification and Validation

Software documentation consistent with **Basic Documentation Level** was submitted in this 510(k). System testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

4) Usability

The product usability has been evaluated and validated according to the following standard and FDA guidance.

&gt; IEC 60601-1-6:2010+2013+2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
&gt; Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016

Conclusion: Based on the comparison analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate devices.

6 / 6

---

**Source:** [https://fda-staging.innolitics.com/device/K251825](https://fda-staging.innolitics.com/device/K251825)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).