K251788 · Zhuhai DR Medical Instruments Co., Ltd. · DXT · Mar 5, 2026 · Cardiovascular
Device Facts
Record ID
K251788
Device Name
Extension tube
Applicant
Zhuhai DR Medical Instruments Co., Ltd.
Product Code
DXT · Cardiovascular
Decision Date
Mar 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The Extension tube is intended as a connecting line for injection of a contrast and saline during coronary angiography procedures.
Device Story
Extension tube serves as a fluid pathway for contrast media and saline delivery during coronary angiography. Device consists of tubing, male/female luer connectors, Y-type needle access adapters, rotary connectors, and protective caps. Operated by clinicians in clinical settings; device connects mechanical injector syringes to angiographic catheters. Clinician manually prepares device by removing caps, securing luer connections, and purging air bubbles; injector is then programmed to deliver contrast per physician orders. Device facilitates high-pressure fluid transfer (up to 1200 psi) to enable diagnostic imaging. Benefits include reliable, sterile fluid delivery for visualization of coronary vasculature.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing including appearance, leakage, tensile strength, flow rate, check valve performance, particulate contamination, and chemical/extractable analysis per ISO 8536 and USP <788>. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, and hemolysis).
Technological Characteristics
Single-use, sterile (EO) extension tubing. Materials include luer connectors, Y-adapters, and tubing. Pressure rating up to 1200 psi. Conical fittings comply with ISO 80369-7. Lengths 23-404 cm; inner diameters 1.5-2.5 mm. Biocompatibility per ISO 10993. Sterilization validated per ISO 11135.
Indications for Use
Indicated for patients undergoing coronary angiography procedures requiring the delivery of contrast media and saline via a connecting line between an angiographic syringe and catheter.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 5, 2026
Zhuhai DR Medical Instruments Co., Ltd.
Guanglong Zhang
Registration Supervisor
No. 352, Dingwansan Road, Jinwan District
Zhuhai, 519040
China
Re: K251788
Trade/Device Name: Extension tube
Regulation Number: 21 CFR 870.1650
Regulation Name: Angiographic Injector And Syringe
Regulatory Class: Class II
Product Code: DXT
Dated: January 30, 2026
Received: February 2, 2026
Dear Guangdong Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251788 - Guanglong Zhang
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251788 - Guanglong Zhang
Page 3
Sincerely,
# Shruti N. Mistry -S
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251788 | |
| Device Name Extension tube | |
| Indications for Use (Describe) The Extension tube is intended as a connecting line for injection of a contrast and saline during coronary angiography procedures. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Premarket Notification 510(k) Submission
510(k) Summary
# K251788- 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the guidance The 510(k) Program and 21 CFR 880.5860.
510(k) Number: K251788
1. Date of Preparation: March 5, 2026
2. Submitter
Zhuhai DR Medical Instruments Co., Ltd.
No. 352, Dingwansan Road, Jinwan District, 519040 Zhuhai,
PEOPLE’S REPUBLIC OF CHINA
Contact Person: Mr. Guanglong ZHANG
Tel.:+86-13798952608
Email: zhangguanglong@int-medical.com
3. Proposed Device
510(k) submission Number: K251788
Trade Name: Extension tube
Common name: Angiographic injector and syringe
Device Classification Name: Angiographic Injector And Syringe
Review Panel: Cardiovascular
Regulation Number: 21 CFR 870.1650
Regulation Class: Class II
Product Code: DXT
Manufacturer: Zhuhai DR Medical Instruments Co., Ltd.
4. Predicate device
510(k) Number: K244038
Product Name: High Pressure Tubing
Review Panel: Cardiovascular
Regulation Number: 21 CFR 870.1650
Device Classification name: Angiographic Injector And Syringe
Regulation Class: Class II
Product Code: DXT
Manufacturer: Shandong INT Medical Instruments Co., Ltd.
{5}
Premarket Notification 510(k) Submission
510(k) Summary
## 5. Indications for Use
The Extension tube is intended as a connecting line for injection of a contrast and saline during coronary angiography procedures.
## 6. Device description
The Extension tube is made up of connectors, male and/or female luer, caps and tubing, with the maximum injection pressure of 1200psi, as a connecting line for direct injection of contrast media or saline. It is single use and supplied sterile by Ethylene Oxide.
## 7. Comparison to the Predicate Device
Extension tube is substantially equivalent in indication for use, design, sterilization method to the predicate devices, High Pressure Tubing, K244038. The differences in components between the devices do not raise different questions of safety and effectiveness.
{6}
Premarket Notification 510(k) Submission
510(k) Summary
| Item | Proposed Device K251788 | Predicate Device K244038 | Remark |
| --- | --- | --- | --- |
| Manufacturer | Zhuhai DR Medical Instruments Co., Ltd. | Shandong INT Medical Instruments Co., Ltd | / |
| Product code | DXT | DXT | Same |
| Device Classification | Class II, 21 CFR 870.1650
Angiographic Injector And Syringe | Class II, 21 CFR 870.1650
Angiographic Injector And Syringe | Same |
| Indications for Use | The Extension tube is intended as a connecting line for injection of a contrast and saline during coronary angiography procedures. | High Pressure Tubing is indicated for use in PTCA surgery to connect the angiographic catheter and the angiographic syringe to deliver contrast media. | Same |
| Principle of Operation | 1. According to needed function, select suitable type.
2. Take out Extension tube from sterile package, remove screw cover and protective cap at both ends of Extension tube.
3. Through male and female luer connector coordinate so as to assemble High pressure tube to corresponding infusion instruments.
4. Turn on the mains switch. | 1. Using aseptic technique, remove the protective caps from both ends of High Pressure Tubing.
2. Securely attach female luer of High Pressure Tubing to the mechanical injector syringe.
3. Prepare High Pressure Tubing for use by filling with contrast media, carefully removing any air bubbles from the fluid column.
4. Securely attach male luer of High Pressure Tubing to the catheter hub, using a ‘wet to wet’ connection, carefully avoiding air bubbles during the connection, Slowly pull a small negative vacuum on the mechanical injector; gently tap on the connection points and the syringe to draw any small bubbles or particulate into the top of the syringe barrel.
5. Make sure tip of mechanical injector syringe is pointing down toward the table. Program injector and dispense contrast media per physician’s orders. | Same. The operating principles and methods are essentially consistent, with the Predicate Device description being more detailed. The underlying operational principles remain the same. |
{7}
| Models | SKU 72233-01(23cm), SKU 72233-02(150cm), SKU 72233-03(150cm), SKU 72233-04(250cm), SKU 72233-05(150cm), SKU 72233-06(150cm), SKU 72233-07(150cm), SKU 72233-08(120cm), SKU 72233-09(120cm), SKU 72233-10(273cm), SKU 72233-11(275cm), SKU 72233-12(31cm), SKU 72233-09A(180cm), SKU 72233-21(224cm), SKU 72233-22(404cm), All models length range in 23 cm ~ 404 cm. Inner diameter:1.5, 1.8, 2.1, 2.2 and 2.5 mm; Outer diameter:3.6 mm. | 10cm~350cm; Inner diameter:1.8 mm; Outer diameter:3.6 mm. | Different, Comment 1 |
| --- | --- | --- | --- |
| Main Configuration/Components | Male/Female luer connector, Y-type needle access adapter, Rotary connector, tube, Beveled trocar assembly | Tubing, Male/female luer connector, Rotating male luer connector, protective caps. | Different, Comment 2 |
| Use | Single use | Single use | Same |
| Pressure Rating | 65 psi ~1200 psi | 1200 psi | same |
| Conical Fitting | Comply with ISO 80369-7 | Comply with ISO 80369-7 | same |
| Biocompatibility | ISO 10993 series | ISO 10993 series | Same |
| Shelf-life | 5 years | 3 years | Different, Comment 3 |
| Sterilization method | EO sterilization | EO sterilization | Same |
| Sterility assurance level | 10-6 | 10-6 | Same |
{8}
Premarket Notification 510(k) Submission
510(k) Summary
# Comparison Discussion
The following comments was the differences between the subject and predicate device:
## Comment 1 - Models
The length ranges of some models of the subject device are longer than those of the predicate devices, and some of their inner diameters are smaller. The subject device has undergone the flow rate verification in Bench performance testing which was pursuant to ISO 10555-1:2023 and met the requirement. So the difference in its length and inner meter do not introduce new risks and do not affect the device's safety or effectiveness.
## Comment 2 - Main Configuration/Components
The subject device and predicate device have similar components, such as male/female luer connector, tubing and valve. The subject device has different components from the predicate device. The subject device has undergone the examination which was pursuant to ISO 8536-4:2019, ISO 8536-9:2015, ISO 8536-10:2015, ISO 8536-12:2021, ISO 80369-7:2021, AAMI ANSI CN27:2021 and met the requirements. The differences of configurations between the subject device and predicate device do not raise different questions of safety and effectiveness.
## Comment 3 - Shelf-life
The subject device has the longer shelf-life. According to the Package Integrity Test Report which was accessed in accordance with ASTM F1929, ASTM F1886 and ASTM F88 after conducting an accelerated aging test and simulated shipping testing. The results of the accelerated aging study have not shown any significant changes in product characteristics and Packaging integrity.
{9}
Premarket Notification 510(k) Submission
510(k) Summary
# 8. Bench Performance Testing
All necessary bench and non-clinical testing was conducted on Extension tube to support a determination of substantial equivalence to the predicate devices. The non-clinical, bench testing included:
| Items | Standards | Conclusion |
| --- | --- | --- |
| 1. Appearance | ISO 8536-9:2015 | Pass |
| 2. Length | / | Pass |
| 3. Luer connector | ISO 80369-7:2021 | Pass |
| 4. Leakage | ISO 8536-9:2015 | Pass |
| 5. Tensile strength | ISO 8536-9:2015 | Pass |
| 6. Protective caps | ISO 8536-4:2019 | Pass |
| 7. Check valve - Counterflow pressure resistance | ISO 8536-12:2021 | Pass |
| 8. Flow rate | ISO 10555-1:2023 | Pass |
| 9. Check valve - Blocking performance | ISO 8536-12:2021 | Pass |
| 10. Check valve - Opening pressure | ISO 8536-12:2021 | Pass |
| 11. Particulate contamination | USP<788>method 1 | Pass |
| 12. Reducing (oxidizable) matter | ISO 8536-4:2019 | Pass |
| 13. Metal ions | ISO 8536-4:2019 | Pass |
| 14. Titration acidity or alkalinity | ISO 8536-4:2019 | Pass |
| 15. Residue on evaporation | ISO 8536-4:2019 | Pass |
| 16. UV absorption of extract solution | ISO 8536-4:2019 | Pass |
| 17. Residual ethylene oxide | ISO 10993-7:2008 | Pass |
# 9. Biocompatibility Testing
Biocompatibility testing was conducted in compliance with ISO 10993-1, for externally communicating devices with limited exposure (<24 hours) to blood path, indirect, and included below test items.
| Items | Standards | Conclusion |
| --- | --- | --- |
| Cytotoxicity | ISO 10993-5:2009 | Pass |
| Sensitization | ISO 10993-10:2021 | Pass |
| Irritation or intracutaneous reactivity | ISO 10993-23:2021 | Pass |
| Acute Systemic Toxicity | ISO 10993-11:2017 | Pass |
| Material-mediated pyrogenicity | ISO 10993-11:2017 | Pass |
| Hemocompatibility -- Hemolysis testing | ASTM F756-17/ISO10993-4 : 2017 | Pass |
The subject device, Extension tube is subject to biocompatibility test in accordance with ISO 10993-1, the test result demonstrate that Extension tube is safe.
{10}
Premarket Notification 510(k) Submission
510(k) Summary
## 10. Sterilization
The subject device is sterilized by traditional Ethylene Oxide sterilization methods. The Ethylene Oxide sterilization process is validated as per ISO 11135-2014 Sterilization of healthcare products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices, Overkill Method: This method is based on demonstrating that the sterilization of a microbial challenge (biological indicator) exceeds the challenge posed by the bioburden of the product. Confirmatory EO residual testing was conducted on the subject device to confirm that the design differences did not impact residual EO levels of the device. The testing confirmed that EO residuals were within the limits specified in ISO 10993-7.
## 11. Packaging and Shelf Life
The subject device is supplied sterile. The sterile barrier packaging is designed to maintain sterility throughout the labeled shelf life 5 years.
The shelf-life study was planned as per ASTM F1980 recommended conditions. The study was conducted with elevated temperature conditions and with real time conditions. The packaging system was accessed in accordance with ASTM F1929, ASTM F1886 and ASTM F88 after conducting an accelerated aging test and simulated shipping testing. The results of the accelerated aging study have not shown any significant changes in product characteristics and Packaging integrity.
## 12. Clinical Testing
No clinical study is included in this submission.
## 13. Conclusion
The proposed device has the same indications and has similar design features and technological characteristics as the predicate device, High Pressure Tubing, K244038. Non-clinical testing data demonstrates that the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.