K251777 · Mesi D.O.O. · BZG · Mar 2, 2026 · Anesthesiology
Device Facts
Record ID
K251777
Device Name
MESI mTABLET SPIRO
Applicant
Mesi D.O.O.
Product Code
BZG · Anesthesiology
Decision Date
Mar 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MESI mTABLET SPIRO is indicated for use to conduct assessment of respiratory function by way of measurement of dynamic lung volumes (also known as spirometry) on adult and pediatric patients, ages 5 years and older. The device measures patient respiratory parameters including FVC, FEV1, FEV6, and VC. MESI mTABLET SPIRO is intended to be operated by medical professionals in health care environments including primary care facilities, hospitals and health centers.
Device Story
Automated wireless spirometer for respiratory disease screening/diagnosis; measures dynamic lung volumes (FVC, FEV1, FEV6, VC). Operates on pneumotach principle (Venturi Effect); airflow measured via pressure difference across disposable mouthpiece. System comprises wireless tablet, spirometer module, and charging station. Used in professional clinical environments (primary care, hospitals) by trained healthcare personnel. Input: patient respiratory airflow; Output: numerical/graphical spirometry parameters displayed on tablet. Higher sampling rate (800Hz) than predicate; wireless architecture enables data storage and viewing. Benefits include objective respiratory function assessment to aid clinical decision-making.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including performance testing per ATS/ERS (2019) and ISO 26782:2009, biocompatibility (ISO 10993 series, ISO 18562 series), electrical safety (AAMI ANSI ES 60601-1), EMC (IEC 60601-1-2), wireless coexistence (ANSI/IEEE C63.27), and reprocessing validation (AAMI ST98).
Technological Characteristics
Fleisch pneumotachograph sensing; 800Hz flow sampling. Materials: aluminum housing (reusable), disposable mouthpieces. Connectivity: Bluetooth, Wi-Fi. Power: Rechargeable Li-Polymer battery (8800 mAh tablet, 620 mAh spiro unit). Display: 10.1" color IPS touchscreen (1280x800). Sterilization: Reprocessing validation for reusable housing. Standards: ATS/ERS 2019, ISO 26782, ISO 10993, IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for assessment of respiratory function via dynamic lung volume measurement (spirometry) in adult and pediatric patients aged 5 years and older. Contraindications not specified.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 2, 2026
Mesi D.O.O.
% Elaine Duncan
President
Paladin Medical, Inc.
P.O. Box 560
Stillwater, Minnesota 55082
Re: K251777
Trade/Device Name: MESI mTABLET SPIRO
Regulation Number: 21 CFR 868.1840
Regulation Name: Diagnostic Spirometer
Regulatory Class: Class II
Product Code: BZG
Dated: January 29, 2026
Received: January 29, 2026
Dear Elaine Duncan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K251777 - Elaine Duncan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251777 - Elaine Duncan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
PRAKHYAT SINGH -S
Digitally signed by
PRAKHYAT SINGH -S
Date: 2026.03.02
13:33:40 -05'00'
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251777 | |
| Device Name MESI mTABLET SPIRO | |
| Indications for Use (Describe) The MESI mTABLET SPIRO is indicated for use to conduct assessment of respiratory function by way of measurement of dynamic lung volumes (also known as spirometry) on adult and pediatric patients, ages 5 years and older. The device measures patient respiratory parameters including FVC, FEV1, FEV6, and VC. MESI mTABLET SPIRO is intended to be operated by medical professionals in health care environments including primary care facilities, hospitals and health centers. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(K) Summary
MESI mTABLET SPIRO - K251777
Submitted on behalf of: MESI D.O.O
Date Prepared: March 2, 2026
Address: Lesloskova
cesta 11A,
1000
Ljubijana
Slovenia,
Europe
Telephone: +386 1 620 34 87
+386 8 2015 32 95
Submitted by: Paladin
Medical,
Inc. PO Box
560
Stillwater, MN 55082
Telephone: 715-549-6035
Contact Person: Elaine Duncan, MSME., RAC, FAIMBE, FBSE
President, Paladin Medical, Inc.
Trade name: MESI mTABLET SPIRO
Common name: Automated wireless spirometry system
Classification name: Diagnostic spirometer.
Device classification: Class II
Product classification: 868.1840
Product code: BZG
Classification panel: Anesthesiology; Spirometer, diagnostic
Predicate Device: K212938: Vitalograph Model 6000 Alpha
Common Name: Diagnostic Spirometer
Classification #: 868.1840
Regulatory Class: II
Product Code: BZG
Reference Device: K201046: MESI mTABLET system
Common Name: Plethysmograph, Photoelectric, Pneumatic Or Hydraulic
Classification #: 870.2780
Regulatory Class: II
Product Code: JOM
{5}
510(K) Summary
MESI mTABLET SPIRO - K251777
## Indications for Use
The MESI mTABLET SPIRO is indicated for assessment of respiratory function by way of measurement of dynamic lung volumes (also known as spirometry) on adult and pediatric patients, ages 5 years and older. The device measures patient spirometry parameters including FVC, FEV1, FEV6, and VC. MESI mTABLET SPIRO is intended to be operated by medical professionals in health care environments including primary care facilities, hospitals and health centers.
## Device Description
The MESI mTABLET SPIRO is an automated wireless spirometer device for diagnosing and screening patients with respiratory diseases. The system is intended to perform, view and store spirometry measurements of adult and pediatric patients. It is comprised of a wireless tablet, spirometer module, disposable mouthpieces and charging station module.
The MESI mTABLET SPIRO is intended to be used in a professional clinical environment by trained healthcare personnel who understand the principle of spirometry measurements, can demonstrate the measurement procedure, verify that the system is working as intended, and start the measuring process.
The MESI mTABLET SPIRO works on pneumotach principle that relies on the airflow measurement according to the pressure difference that occurs when a flowing fluid is forced through a narrow section (disposable mouthpiece), resulting in a pressure decrease and a velocity increase, which is also known as the Venturi Effect. Values measured are indirectly calculated from the measured airflow. Spirometry parameters are captured and displayed as a numerical and graphical representation on MESI mTABLET UNIT.
The device is recharged through the AC/DC power supply. The MESI mTABLET SPIRO is not intended to be used while connected to mains electricity.
## Substantial Equivalence Comparison
The MESI mTABLET SPIRO is compared to the predicate device, Vitalograph Model 6000 Alpha (K212938), a Class II diagnostic spirometer classified under 21 CFR 868.1840 (Product Code BZG). The predicate was selected because it has the same intended use, similar indications for use, and utilizes the same Fleisch pneumotachograph measurement principle for assessment of dynamic lung volumes. Both devices are intended for use by healthcare professionals in clinical environments to perform spirometry testing in adult and pediatric populations.
In addition to the primary predicate, the MESI mTABLET ABI (K201046) is referenced as a device-level technology platform to provide technological justification for the shared tablet hardware, wireless communication architecture, display system, and charging hardware. The MESI mTABLET SPIRO incorporates the identical tablet unit, Wi-Fi docking unit, and large charging plate previously cleared under K201046.
The substantial equivalence comparison between the MESI mTABLET SPIRO and the predicate device is summarized in the table below.
{6}
510(K) Summary
MESI mTABLET SPIRO - K251777
| Device | 6000 ALPHA | MESI mTABLET SPIRO | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | K212938
(Predicate Device) | K251777 | |
| Manufacturer | Vitalograph Inc | MESI D.O.O. | |
| Indications for Use / Intended use | The intended use of the Vitalograph Model 6000 Alpha is the simple assessment of respiratory function through the measurement of dynamic lung volumes i.e. spirometry. The device measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC.
The device is designed to be operated by medical professionals trained in respiratory and lung function testing on adults and pediatrics, 5 years and older, in a variety of professional healthcare environments, e.g. primary care, hospitals and occupational health centers. | The MESI mTABLET SPIRO is indicated for use to conduct assessment of respiratory function by way of measurement of dynamic lung volumes (also known as spirometry) on adult and pediatric patients, ages 5 years and older. The device measures patient respiratory parameters including FVC, FEV1, FEV6, and VC.
MESI mTABLET SPIRO is intended to be operated by medical professionals in health care environments including primary care facilities, hospitals and health centers. | Substantially equivalent to the predicate |
| Target population | Adults and pediatrics patients | Adults and pediatrics patients | Same as the predicate |
| Where used | Clinical environment | Clinical environment | Same as the predicate |
| Classification | 21 CFR 868.1840 BZG | 21 CFR 868.1840 BZG | Same as the predicate |
| Measurement technology | Fleisch Pneumotachograph type | Fleisch Pneumotachograph type | Same as the predicate |
{7}
510(K) Summary
MESI mTABLET SPIRO - K251777
| Device | 6000 ALPHA | MESI mTABLET SPIRO | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | K212938
(Predicate Device) | K251777 | |
| Manufacturer | Vitalograph Inc | MESI D.O.O. | |
| Volume calculation | Flow integration sampling 100Hz | Flow integration sampling 800Hz | Same as the predicate but at higher sampling rate |
| Maximum displaced volume | 10L | 14L | MESI mTABLET SPIRO can displace bigger air volume. The difference does not introduce any new risks. |
| Performance Specification | Accuracy: ±2.5%
Not Publicly Available
Not Publicly Available
Not Publicly Available | Accuracy: ±2.5%
Repeatability: ±2.5%
Linearity: ±2.5%
Impedance: within 0.15 kPa/(L/s) | Similar as the predicate, within ATS 2019 guidelines. |
| Parameters measured | FVC, FEV1, FEV6, PEF, MVV and VC | FVC, FEV1, FEV6, and VC | Similar parameters |
| Patient Interface | Flowhead cone | MESI spirometer UNIT and disposable mouthpiece | Substantially equivalent |
| Power Supply | 7.2V, 2.2Ahr NiMH | MESI mTABLET UNIT
AC/DC adaptor: Input: 100-240 V~, 50-60 Hz
Output: 5 Vdc; 5000 mA
Battery type: Rechargeable Lithium-Polymer battery
Capacity: 8800 mAh, Battery operation: more than 8 hours
MESI SPIRO UNIT | Similar battery type, Identical to MESI mTABLET ABI reference device |
{8}
510(K) Summary
MESI mTABLET SPIRO - K251777
| Device | 6000 ALPHA | MESI mTABLET SPIRO | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | K212938
(Predicate Device) | K251777 | |
| Manufacturer | Vitalograph Inc | MESI D.O.O. | |
| | | Battery type: Rechargeable Lithium-Polymer battery, Capacity: 620 mAh Examinations per battery charge: > 150 | |
| Display | Color touchscreen | 10,1" color IPS screen with 1280x800 resolution | Identical to the MESI mTABLET ABI |
| Contact type and duration – Bio-compatibility | Externally Communicating, Limited Duration Tissue and Surface Contact, Skin / Mucosa, | Externally Communicating, Limited Duration Tissue and Surface Contact, Skin / Mucosa, | Same |
| Performance Testing | ATS/ERS (2019)
ISO 23747:2007
ISO 26782:2009 | ATS/ERS2019
ISO 26782:2009 | MESI mTABLET SPIRO is compliant with applicable standards. The difference does not introduce any additional risks. |
| Electrical Safety and EMC | ES 60601-1
IEC 60601-1-2 | ES 60601-1
IEC 60601-1-2:2014 | Same as predicate |
| Working environment | Temperature: 10–40°C | Temperature: 10–40°C
Humidity: 25% to 85% | Substantially Equivalent |
| Transport and storage environment | Not Available | 15° to 50°C (<1 month) / -15° to 40°C (<3 month) / -15° to 25°C (<12 month)
Relative humidity: 25 to 85% (no condensation) | Same as Mesi mTABLET ABI |
{9}
510(K) Summary
MESI mTABLET SPIRO - K251777
| Device | 6000 ALPHA | MESI mTABLET SPIRO | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | K212938
(Predicate Device) | K251777 | |
| Manufacturer | Vitalograph Inc | MESI D.O.O. | |
| | | | |
| Communications | USB, Ethernet, WiFi | Bluetooth, WiFi | Identical to the MESI mTABLET ABI |
| IP degree | IP42 | IP 30 | MESI mTABLET SPIRO has slightly lower IP degree from the predicate and MESI mTABLET ABI.The difference does not introduce any risks. |
## Discussion of Differences
The MESI mTABLET SPIRO has the same intended use and similar indications for use as the Vitalograph Model 6000 Alpha (K212938), and both devices utilize a Fleisch pneumotachograph measurement principle for assessment of dynamic lung volumes. However, several technological differences exist that have been justified. The subject device employs a wireless, tablet-based architecture with Bluetooth and Wi-Fi connectivity, whereas the predicate is a standalone spirometer platform. The MESI mTABLET SPIRO samples flow data at a higher integration rate (800 Hz versus 100 Hz), supports a greater maximum displaced volume (14 L versus 10 L), and measures a more limited set of parameters (FVC, FEV1, FEV6, and VC) compared to the predicate, which also reports PEF and MVV. Additional differences include system configuration, power supply, user interface, and environmental protection ratings. These differences relate to hardware design, data acquisition rate, and feature set rather than to the fundamental measurement principle or intended diagnostic function. Based on performance testing and non-clinical validation, these differences do not raise new or different questions of safety or effectiveness.
## Non-Clinical Testing
The following performance data were provided in support of the substantial equivalence determination.
{10}
510(K) Summary
MESI mTABLET SPIRO - K251777
## Biocompatibility
The patient contacting materials (Externally Communicating, Limited Duration Tissue and Surface Contact, Skin / Mucosa, were evaluated using: ISO 10993-5 (2009) – Cytotoxicity; ISO 10993-10 (2010) – Sensitization and Irritation; ISO 10993-18 (2020) – Chemical Characterization; ISO 18562-2 (2017) – Particulate Materials; ISO 18562-3 (2017) – Volatile Organic Compounds (VOC), with a toxicological risk assessment. The materials were found to be acceptable for their intended use.
## Electrical Safety and EMC
Electrical safety and EMC testing were conducted on the subject device. The system complies with AAMI ANSI ES 60601-1:2005 + A1:2012 – Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance and IEC 60601-1-2:2014 – Medical Electrical Equipment – Part 1-2: Electromagnetic Compatibility – Requirements and Tests. In addition, wireless coexistence testing was performed in accordance with ANSI/IEEE C63.27:2017, consistent with the device’s AAMI TIR69 risk assessment to evaluate the performance of the Bluetooth communication link in the presence of intentional and unintentional RF interference
## Software Verification and Validation Testing
Software verification and validation testing were conducted per FDA Guidance on Content of Premarket Submissions for Device Software Functions. The documentation level for the software for this device was considered “Basic”.
Cybersecurity testing and documentation conform to FDA Guidance on Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions.
## Spirometry Performance
Performance testing per ATS/ERS (2019) for peak flow and timed forced expired volume per ISO 26782:2009 – Respiratory Equipment – Spirometers Intended for the Measurement of Forced Expired Volumes in Humans was performed.
## Mechanical, Animal, and Clinical Testing
Reprocessing validation testing was conducted on the reusable aluminum housing in accordance with FDA’s Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling guidance and AAMI ST98. No animal or clinical testing was performed
## Substantial Equivalence Conclusion
The performance testing has demonstrated that the subject device has met the applicable standard performance requirements. Through a comparison of performance testing, design and features, and non-clinical testing, the subject device and predicate are found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.