The Navilas® Laser System is an ophthalmic laser integrated with a digital camera. The Navilas® Laser System is indicated for the treatment of ocular pathology in the anterior and posterior segment of the eye, with supporting imaging functions (capture, display, storage and annotation) of the retina of the eye via color and infrared imaging. The Navilas® Laser System is intended for use in the posterior segment to perform Retinal Photocoagulation, Panretinal photocoagulation, focal photocoagulation or grid photocoagulation for vascular and structural abnormalities of the retina and choroid including: - Proliferative and non-proliferative diabetic retinopathy - Macular edema - Choroidal neovascularization associated with wet age-related macular degeneration - Age-related macular degeneration - Central and branch retinal vein occlusion - Lattice degeneration - Retinal tears and detachments Intended for use in the treatment of ocular pathology in the anterior segment including: - Iridotomy - Iridoplasty - Trabeculoplasty
Device Story
Navilas Laser System 577sl is an ophthalmic laser photocoagulator with integrated digital fundus camera. Device captures color/infrared images of retina; physician uses touch-sensitive display to identify areas of interest and plan laser treatment. System supports pre-planned and ad-hoc laser delivery modes; uses motorized mirrors to direct 577nm semiconductor laser beam through ophthalmoscope. Laser operates in continuous wave or micro-second pulsing modes. Physician monitors aiming beam position on real-time display. Device stores images and treatment data for reporting. Used in clinical settings by physicians to treat vascular/structural retinal/choroidal abnormalities and anterior segment pathologies. Benefits include precise, pre-planned laser delivery and integrated imaging/documentation workflow.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including optical radiation hazard evaluation (ANSI Z80.36:2021), electromagnetic compatibility (IEC 60601-1-2:2020), human factors/usability engineering (IEC 62366-1, IEC 60601-1-6), and software verification/validation.
Technological Characteristics
Semiconductor laser (577nm); CW and MSP modes; power <1700mW. Delivery via scanning slit lamp with motorized mirrors. Integrated digital camera (color/IR). Connectivity via local network/USB. Software-controlled (embedded Windows). Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and ANSI Z80.36.
Indications for Use
Indicated for patients requiring treatment of ocular pathology in the anterior and posterior segments, including diabetic retinopathy, macular edema, wet age-related macular degeneration, retinal vein occlusion, lattice degeneration, retinal tears/detachments, and anterior segment conditions (iridotomy, iridoplasty, trabeculoplasty).
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Yellow Laser Photocoagulator System YLC-500 (K170302)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 29, 2025
OD-OS GmbH
Camila Mundt Ilgenfritz
Head of Regulatory Affairs
Warthestr. 21
Teltow, 14513
Germany
Re: K251772
Trade/Device Name: Navilas Laser System 577sl (156691)
Regulation Number: 21 CFR 886.4390
Regulation Name: Ophthalmic laser
Regulatory Class: Class II
Product Code: HQF
Dated: September 19, 2025
Received: September 25, 2025
Dear Camila Mundt Ilgenfritz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K251772 - Camila Mundt Ilgenfritz
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251772 - Camila Mundt Ilgenfritz
Page 3
Sincerely,
Elvin Y. Ng -S
Elvin Ng
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K251772
Device Name
Navilas Laser System 577sl
Indications for Use (Describe)
The Navilas® Laser System is an ophthalmic laser integrated with a digital camera. The Navilas® Laser System is indicated for the treatment of ocular pathology in the anterior and posterior segment of the eye, with supporting imaging functions (capture, display, storage and annotation) of the retina of the eye via color and infrared imaging.
The Navilas® Laser System is intended for use in the posterior segment to perform Retinal Photocoagulation, Panretinal photocoagulation, focal photocoagulation or grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
- Proliferative and non-proliferative diabetic retinopathy
- Macular edema
- Choroidal neovascularization associated with wet age-related macular degeneration
- Age-related macular degeneration
- Central and branch retinal vein occlusion
- Lattice degeneration
- Retinal tears and detachments
Intended for use in the treatment of ocular pathology in the anterior segment including:
- Iridotomy
- Iridoplasty
- Trabeculoplasty
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
OD-OS
OD-OS GmbH
510(k) SUMMARY
K251772
## 1. Submitter Information
510(k) Owner: OD-OS GmbH
Warthestraße 21
14513 Teltow, Germany
Phone: (+49) 3328 31282100
Contact Person: Camila Mundt Ilgenfritz
OD-OS GmbH
Head of Regulatory Affairs
Email: camila.ilgenfritz@od-os.com
Phone: (+49) 3328 31282148
Date Prepared: October 29th, 2025
## 2. Device Name and Classification
Device Trade Name: Navilas Laser System 577SL
Common Name: Photocoagulator
Regulation Name: Ophthalmic Laser (21 CFR 886.4390)
Device Classification: Class II
Product Code(s): HQF, Ophthalmic Laser (21 CFR §886.4390)
## 3. Predicate Device(s)
- Primary Predicate: Iridex PASCAL (Iridex, K223132)
Product code: HQF
- Predicate 2: Navilas Laser System 577s (OD-OS GmbH, K162191)
Product code: GEX, HKI, NFF, NFG
- Predicate 3: Yellow Laser Photocoagulator System YLC-500 (Nidek, K170302)
Product code: HQF
K251772 - Navilas Laser System 577SL
Page 1 of 11
{5}
OD-OS
OD-OS GmbH
# 4. Device Description
The Navilas Laser System 577SL is a laser photocoagulator with an integrated digital fundus camera. The Navilas Laser System 577SL combines imaging technologies (color and infra-red imaging) with established laser photocoagulation treatment methods by providing the physician a system for imaging and treatment planning prior to the photocoagulation procedure.
The Navilas Laser System 577SL is comprised of:
- A semiconductor laser source that operates at 577nm wavelength. The laser can operate with Continuous Wave (CW) or Micro-Second Pulsing (MSP).
- An integrated delivery system that directs the laser beam through an ophthalmoscope using motorized mirrors.
- A digital camera that provides continuous real-time imaging in color with white light illumination of the fundus, or in monochrome using infrared illumination.
- A software platform that is based on an embedded Windows operating system, that controls the startup, use, calibration, monitoring, and shutdown of the system.
The Navilas Laser System 577SL supports the user during multiple steps of a laser treatment procedure with digital imaging, image storage, planning and laser treatment options including: Digital imaging - Provided by a color image with white light, supporting mydriatic and non-mydriatic image acquisition (with and without dilated pupils), or a monochrome IR image. Images are presented using a digital display. An illumination mode is selected where images are acquired and either stored, exported or discarded after viewing on the touch sensitive digital display.
Image Storage - Captured images can be digitally stored in the Navilas Laser System 577SL database along with other patient related data to create a complete patient record for future reference. Images from other devices may also be imported and stored.
Planning - Areas identified on acquired or imported images by the user that are selected for future treatment consideration can be marked through the use of treatment planning tools available. The physician has the ability to highlight areas on acquired images (called Points of Interest). These locations are created and manipulated using the touch sensitive digital display.
Laser Treatment - Treatment options are also unchanged from the predicate device with Pre-planned and Ad-hoc Modes available on all Navilas laser models. Pre-positioning of the aiming beam onto locations which are selected by the physician during planning is also facilitated. The position of the aiming beam can be monitored on the real-time image that is displayed on the touch sensitive digital display.
Report generation - Information collected in the database includes images obtained before, during and after treatment. This information can be used for the generation of patient reports for documentation purposes.
The Navilas Laser System 577SL can emit a 577 nm wavelength beam for photocoagulation with power up to 1700 mW and pulse duration up to 4000 ms.
K251772 - Navilas Laser System 577SL
Page 2 of 11
{6}
OD-OS
OD-OS GmbH
# 5. Indications for Use
The Navilas® Laser System is an ophthalmic laser integrated with a digital camera. The Navilas® Laser System is indicated for the treatment of ocular pathology in the anterior and posterior segment of the eye, with supporting imaging functions (capture, display, storage and annotation) of the retina of the eye via color and infrared imaging.
The Navilas® Laser System is intended for use in the posterior segment to perform Retinal Photocoagulation, Panretinal photocoagulation, focal photocoagulation or grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
- Proliferative and non-proliferative diabetic retinopathy
- Macular edema
- Choroidal neovascularization associated with wet age-related macular degeneration
- Age-related macular degeneration
- Central and branch retinal vein occlusion
- Lattice degeneration
- Retinal tears and detachments
Intended for use in the treatment of ocular pathology in the anterior segment including:
- Iridotomy
- Iridoplasty
- Trabeculoplasty
# 6. Comparison of Technological Characteristics with the Predicate Device(s)
The Navilas Laser System 577SL is an ophthalmic laser that performs both retinal photocoagulation and laser trabeculoplasty. The Navilas 577SL is a variation of the Navilas Laser System 577s (Predicate 2) which was 510(k)-cleared under K162191 and modified step-wise over the years with minor changes documented via notes to file.
The Navilas 577SL is functionally identical to the current version of Navilas 577s, with the exception that the Navilas 577SL has a modified power range, an updated indications for use statement, and a few software improvements, including:
- Addition of two patterns options on planning function: macula grid pattern and double-spot titration
- Addition of a multiple snapshot acquisition option: allows for capturing a series of multiple snapshots with progressively increasing illumination intensity levels
- The proprietary *.npt import/export file format has been updated to allow capture, of additional patient-related meta data to enhance planning and back-up capabilities
- Minor adjustments to user interface for workflow efficiency associated with the changes listed above
K251772 - Navilas Laser System 577SL
Page 3 of 11
{7}
OD-OS
OD-OS GmbH
Although the Navilas Laser System 577SL and the predicate devices do not share all the same technological characteristics, these differences do not raise different types of questions of safety and effectiveness.
K251772 - Navilas Laser System 577SL
Page 4 of 11
{8}
OD-OS
OD-OS GmbH
| Characteristic | Subject Device
Navilas Laser System 577sl
(OD-OS)
K251772 | Primary Predicate
Iridex PASCAL
(Iridex)
K223132 | Predicate 2
Navilas Laser System 577s
(OD-OS)
K162191 | Predicate 3
Yellow Laser
Photocoagulator System
YLC-500
(Nidek)
K170302 | Comparison of Subject
Device to Predicate
Devices |
| --- | --- | --- | --- | --- | --- |
| Device
Classification | II | II | II | II | Same |
| Product Code | HQF | HQF | GEX
HKI, NFF, NFG | HQF | Same to Primary Predicate
and Predicate 3
NOTE: Navilas Laser System
577s (Predicate 2) was
originally assigned product
code GEX (powered laser
surgical instrument), which
is the code that was
historically assigned to
ophthalmic lasers. More
recently, ophthalmic lasers
are assigned to product
code HQF (ophthalmic
lasers), which is more
specific. |
| Regulation
Number | 21 CFR 886.4390 | 21 CFR 886.4390 | 21 CFR 878.4810 | 21 CFR 886.4390 | Same to Primary Predicate
and Predicate 3
NOTE: For Predicate 2 see
note above. |
K251772 - Navilas Laser System 577sl
Page 5 of 11
{9}
OD-OS
OD-OS GmbH
| Characteristic | Subject Device
Navilas Laser System 577sl
(OD-OS)
K251772 | Primary Predicate
Iridex PASCAL
(Iridex)
K223132 | Predicate 2
Navilas Laser System 577s
(OD-OS)
K162191 | Predicate 3
Yellow Laser Photocoagulator System
YLC-500
(Nidek)
K170302 | Comparison of Subject Device to Predicate Devices |
| --- | --- | --- | --- | --- | --- |
| Indication for Use | The Navilas® Laser System is an ophthalmic laser integrated with a digital camera. The Navilas® Laser System is indicated for the treatment of ocular pathology in the anterior and posterior segment of the eye, with supporting imaging functions (capture, display, storage and annotation) of the retina of the eye via color and infrared imaging..
The Navilas® Laser System is intended for use in the posterior segment to perform Retinal Photocoagulation, Panretinal photocoagulation, focal photocoagulation or grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
• Proliferative and non-proliferative diabetic retinopathy
• Macular edema
• Choroidal neovascularization associated with wet age-related macular degeneration
• Age-related macular degeneration | Intended for use in the treatment of ocular pathology in both the posterior and anterior segments.
Intended for use in the posterior segment to perform retinal photocoagulation, panretinal photocoagulation, focal photocoagulation and grid photocoagulation for vascular and structural abnormalities of the retina and choroid including:
• proliferative and non-proliferative diabetic retinopathy
• macular edema
• choroidal neovascularization associated with wet age-related macular degeneration
• age-related macular degeneration
• lattice degeneration
• retinal tears and detachments
Intended for use in the treatment of ocular pathology in the anterior segment including:
• Iridotomy | The Navilas® Laser System is an ophthalmic laser integrated with a digital camera.
The Navilas® Laser System is indicated for use:
• Retinal Photocoagulation for the treatment of Clinically Significant Diabetic Macular Edema (Focal or Grid Laser), Proliferative Diabetic Retinopathy (Panretinal Photocoagulation), Sub-retinal (Choroidal) Neovascularization (Focal Laser), Central and Branch Retinal Vein Occlusion (Scatter Laser Photocoagulation, Focal or Grid Laser), Lattice Degeneration, Retinal Tears and Detachments (Laser Retinopexy).
• For the imaging (capture, display, storage and manipulation) of the retina of the eye, including via color and infrared imaging; and for aiding in the diagnosis and treatment of ocular | The Yellow Laser Photocoagulator System YLC-500 is intended to be used in ophthalmic surgical procedures including retinal and macular photocoagulation, iridotomy and trabeculoplasty.
The Yellow Laser Photocoagulator System YLC-500 is intended to work in conjunction with the following delivery units in ophthalmic photocoagulation procedures: NIDEK SL-1800, SL-1600, ZEISS SL 130, HAAG BQ900, HEINE OMEGA 500 | Similar Navilas 577sl has two additional indications for use in comparison to Predicate 2, which are within scope of the Primary Predicate (non-proliferative diabetic retinopathy and age-related macular degeneration) |
K251772 - Navilas Laser System 577sl
Page 6 of 11
{10}
OD-OS
OD-OS GmbH
| Characteristic | Subject Device
Navilas Laser System 577sl
(OD-OS)
K251772 | Primary Predicate
Iridex PASCAL
(Iridex)
K223132 | Predicate 2
Navilas Laser System 577s
(OD-OS)
K162191 | Predicate 3
Yellow Laser
Photocoagulator System
YLC-500
(Nidek)
K170302 | Comparison of Subject Device to Predicate Devices |
| --- | --- | --- | --- | --- | --- |
| | • Central and branch retinal vein occlusion
• Lattice degeneration
• Retinal tears and detachments
Intended for use in the treatment of ocular pathology in the anterior segment including:
• Iridotomy
• Iridoplasty
• Trabeculoplasty | • Trabeculoplasty | pathology in the posterior segment of the eye.
• In Laser Trabeculoplasty for primary open angle glaucoma, as well as iridotomy and iridoplasty for closed angle glaucoma | | |
| | | | | | |
| Laser medium | Optically pumped semiconductor laser | Optically pumped semiconductor laser | Optically pumped semiconductor laser | Optically pumped semiconductor laser | Same |
| Wavelength(s) | 577 nm | 577 nm and 532 nm | 577 nm | 577 nm | Same |
| Laser Class | IV | IV | IV | IV | Same |
| Laser Output | - Continuous Wave
- Microsecond pulsing (MSP) | - Continuous Wave
- Microsecond pulsing | - Continuous Wave
- Microsecond pulsing (MSP) | - Continuous wave | Same as Predicate 1 and 2.
Similar to Predicate 3, when taking into account the treatment power and pulse duration ranges. |
| Treatment Power | <1700mW | 0 to 2000 mW | < 2000 mW | 50 to 1500 mW | Similar, range between Predicates 1/2 and Predicate 3.
Differences do not raise different types of questions of safety and effectiveness.
Performance testing was performed to verify and validate this change. |
K251772 - Navilas Laser System 577sl
Page 7 of 11
{11}
OD-OS
OD-OS GmbH
| Characteristic | Subject Device
Navilas Laser System 577sl
(OD-OS)
K251772 | Primary Predicate
Iridex PASCAL
(Iridex)
K223132 | Predicate 2
Navilas Laser System 577s
(OD-OS)
K162191 | Predicate 3
Yellow Laser
Photocoagulator System
YLC-500
(Nidek)
K170302 | Comparison of Subject
Device to Predicate
Devices |
| --- | --- | --- | --- | --- | --- |
| Treatment
Pulse Duration | CW: 10 – 4000 ms
MSP: 0.05 – 0.5ms | CW: 5 – 1000ms
MicroPulse: 0.05 – 1.0 ms | CW: 10 – 4000 ms
MSP: 0.05 – 0.5 ms | Single spot: 10 – 3000ms
Scan Mode: 10 – 50ms
Auto M Mode : 10 – 3000ms | Same as Predicate 2
Similar to Primary Predicate
and Predicate 3. |
| Repeat Mode | yes | yes | yes | Not publicly reported | Same |
| Pulse Interval
Time | 10 – 4000 ms
Short pulse: 0.05 – 2.0 ms | 125 – 1000ms
Micropulse: 0.1 – 10.0 ms | 10 – 4000 ms
Short pulse: 0.05 – 2.0ms | 5 – 1000 ms | Same as Predicate 2
Similar to Primary Predicate
and Predicate 3. |
| Retinal Spot
size | 50 – 500 μm | 50 – 300 μm | 50 – 500 μm | Slit lamp delivery unit:
50 – 990 μm
Scan slit lamp delivery unit:
50 – 500 μm, 100 – 500 μm in
scan mode
B.I.O. delivery:
212 to 664 μm | Same as Predicates 2 and 3.
Similar to Primary Predicate. |
| Visualization
Principle | Scanning slit lamp | Scanning Slit lamp | Scanning slit lamp | Slit lamp | Same |
| Patterns | Multiple patterns:
- single spot
- titration spots (single/double)
- freeform pattern
- macula grid
- rectangle
- circle
- arc | Multiple patterns:
- Single Spot
- Titration spots
- Array
- Hexagon
- Arc
- Triple Arc
- Wedge
- Triple Ring
- Line
- Octants
- Enhanced Octants
- PSLT 3 Row | Multiple patterns:
- single spot
- titration spots (single)
- freeform pattern
- rectangle
- circle
- arc | Multiple patterns:
- Single
- Square
- Equal Space
- Line
- Triangle
- Curve
- Triple Curve
- Circle
- Arc
- Triple Arc
- Arcade Grid
- Rectangle | Similar to predicate devices.
All devices offer multiple
patterns. The Navilas 577SL
includes two patterns that
are not available in the
predicate devices (double
titration spots and a Macula
Grid pattern), but this
difference does not raise
different types of questions
of safety and effectiveness. |
K251772 - Navilas Laser System 577sl
Page 8 of 11
{12}
OD-OS
OD-OS GmbH
| Characteristic | Subject Device
Navilas Laser System 577sl
(OD-OS)
K251772 | Primary Predicate
Iridex PASCAL
(Iridex)
K223132 | Predicate 2
Navilas Laser System 577s
(OD-OS)
K162191 | Predicate 3
Yellow Laser
Photocoagulator System
YLC-500
(Nidek)
K170302 | Comparison of Subject Device to Predicate Devices |
| --- | --- | --- | --- | --- | --- |
| | | | | | Software testing was completed to verify and validate these new patterns. |
| | | | | | |
| Aiming Laser Type | Semiconductor diode laser | Semiconductor diode laser | Semiconductor diode laser | Semiconductor diode laser | Same |
| Wavelength | 635 nm | 635 nm | 635 nm | 635 nm | Same |
| Maximum Avg. Power | Adjustable to less than 1mW | Adjustable to less than 1mW | Adjustable to less than 1mW | 0.3±0.1 mW | Same |
| Laser Class | II | II | II | II | Same |
| | | | | | |
| Imaging Modes | - Color (non-mydriatic and mydriatic color)
- Infra red | n/a | - Color (non-mydriatic and mydriatic color)
- Infra red | n/a | Same as Predicate 2 |
| Image Acquisition, Storage and Annotation | Yes | n/a | Yes | n/a | Same as Predicate 2, with additional option for multiple snapshots series |
| Treatment Planning Tools | Image Overlay, Points of Interest, Documentation | Documentation | Image Overlay, Points of Interest, Documentation | Documentation | Same as Predicate 2. Similar to Primary Predicate and Predicate 3. |
| Import/export function | Yes, shared folder on local network or USB | yes, export on USB | yes, shared folder on local network or USB | yes, export on local network or USB | Same as Predicate 2 with additional data now included on the proprietary npt file, e.g. multiple snapshots series |
K251772 - Navilas Laser System 577sl
Page 9 of 11
{13}
OD-OS
OD-OS GmbH
# 7. Summary of Non-Clinical Testing
Based on the risk assessment and design control requirements, the following verification and validation testing was performed:
- Optical radiation hazard evaluation and safety testing was conducted per:
- ANSI Z80.36:2021 American National Standard for Ophthalmics - Light Hazard Protection for Ophthalmic Instruments
- Electromagnetic compatibility (EMC) was conducted per:
- IEC 60601-1-2:2020 Medical electrical equipment – Part 1-2: General requirements for safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirements and tests.
- Human factors testing was conducted per:
- FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices"
- ANSI/AAMI IEC 62366-1 Medical devices – Part 1: Application of usability engineering to medical devices, and
- IEC 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
- Software verification and validation testing were conducted, and documentation was provided as recommended by FDA Guidance for the Content of Premarket Submissions for Software Device Software Functions, June 2023, requiring Enhanced Documentation Level.
- Device performance bench testing
The device also complies with:
- Laser safety:
- IEC 60601-2-22: 2007 (3rd Ed.) Medical electrical equipment Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1:2014 Safety of laser products – Part 1: Equipment classification and requirements
- Electrical safety:
- IEC 60601-1:2020 Medical electrical equipment – Part 1: General requirements for safety 1: collateral standard: Safety Requirements for Medical Electrical Systems.
Results of the non-clinical testing support a substantial equivalence determination. The Navilas Laser System 577SL is substantially equivalent to its predicate devices for the indications for use.
Clinical testing was not required for this product change.
K251772 - Navilas Laser System 577SL
{14}
OD-OS
OD-OS GmbH
# 8. Conclusions
The Navilas Laser System 577SL has the same intended use as the legally marketed predicate device(s) identified in this premarket notification. Some of the technological characteristics of the Navilas Laser System 577SL differ from those of the predicate device(s), but the differences do not raise new or different types of questions of safety or effectiveness.
The results of the non-clinical performance testing demonstrate that the Navilas Laser System 577SL functions as intended and meets all product design requirements and applicable standards. The non-clinical performance testing demonstrates that the Navilas Laser System 577SL meets all performance specifications and requirements and is substantially equivalent to the predicate devices, and is safe and effective.
K251772 - Navilas Laser System 577SL
Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.