Flexiva Pulse Laser Fiber; Flexiva Pulse TracTip Laser Fiber; Flexiva Pulse ID Laser Fiber; Flexiva Pulse ID TracTip Laser Fibers
K251759 · Boston Scientific Corporation · GEX · Jul 9, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251759
Device Name
Flexiva Pulse Laser Fiber; Flexiva Pulse TracTip Laser Fiber; Flexiva Pulse ID Laser Fiber; Flexiva Pulse ID TracTip Laser Fibers
Applicant
Boston Scientific Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 9, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Flexiva Pulse and Flexiva Pulse TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use. Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use.
Device Story
Fiber optic laser delivery device; transmits Ho:YAG laser energy from external laser console to urological anatomy. Components: SMA-905 connector, ETFE-jacketed silica core fiber, polished flat or ball-shaped (TracTip) output tips. Flexiva Pulse ID models include RFID tag for data storage with compatible closed-system laser consoles. Used in endoscopic surgical procedures; operated by physicians. Energy exits distal fiber tip to fragment urinary calculi or ablate soft tissue. Benefits: minimally invasive surgical access to urinary tract for stone management and soft tissue treatment.
Clinical Evidence
Bench testing only. Design verification included bent transmission, fiber durability, connector temperature, and laser system output accuracy. Biocompatibility assessment concluded no risks associated with the material change.
Technological Characteristics
Fiber optic delivery system; silica core with ETFE jacket. Tip configurations: flat (242μm, 365μm, 550μm, 910μm) or ball-shaped (242μm). Connector: SMA-905. RFID tag for data storage (ID models). Sterilization: Ethylene Oxide (EO). Single-use.
Indications for Use
Indicated for urologic endoscopic procedures including vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi using Ho:YAG laser energy.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 9, 2025
Boston Scientific Corporation
Galit Gershon
Senior Regulatory Affairs Specialist
100 Boston Scientific Way - Urology Division
Marlborough, Massachusetts 01752
Re: K251759
Trade/Device Name: Flexiva Pulse Laser Fiber; Flexiva Pulse TracTip Laser Fiber; Pulse ID Laser Fiber; Flexiva Pulse ID TracTip Laser Fibers
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General and Plastic Surgery and In Dermatology
Regulatory Class: II
Product Code: GEX
Dated: June 8, 2025
Received: June 9, 2025
Dear Galit Gershon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251759 - Galit Gershon
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K251759 - Galit Gershon
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino -S
Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251759
Device Name
Flexiva Pulse Laser Fiber; Flexiva Pulse TracTip Laser Fiber; Flexiva Pulse ID Laser Fiber; Flexiva Pulse ID TracTip Laser Fibers
Indications for Use (Describe)
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Flexiva Pulse and Flexiva Pulse TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use.
Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Flexiva Pulse/ ID Single Use Laser Fibers
Special 510(k)
# 510(k) Summary for Flexiva Pulse/ ID Laser Fiber
Date Prepared: 08-June-2025
## A. Submitter
Boston Scientific Corporation
Urology Division
100 Boston Scientific Way
Marlborough, MA 01752
## B. Contacts
Galit Gershon
Sr. Regulatory Affairs Specialist
+972-52-8786899
Galit.gershon@bsci.com
## C. Device Name
**Trade Name**
Flexiva Pulse Laser Fiber; Flexiva Pulse TracTip Laser Fiber; Flexiva Pulse ID Laser Fiber; Flexiva Pulse ID TracTip Laser Fibers
**Common Name**
Laser Instrument, Surgical, Powered
**Classification**
Class II per 21 CFR 878.4810
**Classification Name**
Laser Surgical Instrument for use in general and plastic surgery and in dermatology
**Product Code**
GEX
**Product Code Name**
Powered Laser Surgical Instrument
## D. Predicate Devices
For the purposes of establishing substantial equivalence, the designs and technological characteristics of the proposed Flexiva Pulse/ ID Laser Fiber were compared to the following 510(k)-cleared device.
Table 11-1: Predicate Device for Establishing Substantial Equivalence
| Characteristics | Predicate Device |
| --- | --- |
| Device Trade Name | Flexiva Pulse Laser Fiber,
Flexiva Pulse TracTip Laser Fiber,
Flexiva Pulse ID Laser Fiber,
Flexiva Pulse ID TracTip Laser Fiber |
| Classification Name | Laser Surgical Instrument for use in general and plastic surgery and in dermatology |
Confidential & Proprietary Information of Boston Scientific Corporation
{5}
Flexiva Pulse/ ID Single Use Laser Fibers
Special 510(k)
Page 2 of 5
| Regulation Number | 21 CFR 878.4810 |
| --- | --- |
| Classification | Class II |
| Product Code | GEX |
| 510(k) Submitter/ Holder | Boston Scientific Corporation |
| 510(k) Clearance Number & Date | K210925 Cleared 28-April -2021 |
Confidential & Proprietary Information of Boston Scientific Corporation
{6}
Flexiva Pulse/ ID Single Use Laser Fibers
Special 510(k)
Page 3 of 5
# E. Device Description
Flexiva Pulse, Flexiva Pulse TracTip, Flexiva Pulse ID and Flexiva Pulse ID TracTip Single Use Laser Fibers are fiber optic laser energy delivery devices consisting of a SMA connector (Black Hole design), and an ETFE jacketed silica core fiber. Flexiva Pulse and Flexiva Pulse ID fibers are equipped with a polished, flat output tip (242μm, 365μm, 550μm and 910μm size) and the Flexiva Pulse TracTip and Flexiva Pulse ID TracTip fibers are equipped with a polished and reinforced ball-shaped output tip (242μm size).
These fibers may be used in a variety of laser-based surgical cases. For Flexiva Pulse ID laser fibers, an RFID (Radio-frequency identification) tag enables read/write data storage for compatible RFID-equipped laser systems (closed systems).
# F. Intended Use/Indication for Use
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Flexiva Pulse and Flexiva Pulse TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use.
Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use.
# G. Principles of Operation
The operating principle for the Flexiva Pulse/ ID Laser Fiber (submitted under K210925) remains unchanged.
Flexiva Pulse/ ID Laser Fibers deliver holmium (Ho:YAG) laser energy endoscopically from a laser console to soft tissues or urinary calculi. The energy travels within the fiber's glass core and exits through the distal end of the fiber. Upon impact, the holmium energy fragments urinary calculi and/or ablates soft tissue.
# H. Comparison of Key Technological/Performance Characteristics
The proposed Flexiva Pulse/ ID Laser Fiber have the same technological characteristics and fundamental design as the predicate Flexiva Pulse/ ID Laser Fiber (K210925).
Confidential & Proprietary Information of Boston Scientific Corporation
{7}
Flexiva Pulse/ ID Single Use Laser Fibers
Special 510(k)
Table 11-2: Flexiva Pulse/ ID Laser Fiber Comparison and Substantial Equivalence
| Characteristics | Predicate Flexiva Pulse/ ID Laser Fiber (K210925) | Proposed Flexiva Pulse/ ID Laser Fiber | Similarities / Differences |
| --- | --- | --- | --- |
| Indication for Use | Flexiva Pulse and Flexiva Pulse TracTip Laser Fiber:
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy.
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use. | Flexiva Pulse and Flexiva Pulse TracTip Laser Fiber:
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy.
Flexiva Pulse and Flexiva Pulse TracTip laser fibers are indicated for urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use. | Identical |
| | Flexiva Pulse ID and Flexiva Pulse ID TracTip Laser Fiber:
Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy.
Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are indicated for Urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use. | Flexiva Pulse ID and Flexiva Pulse ID TracTip Laser Fiber:
Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy.
Flexiva Pulse ID and Flexiva Pulse ID TracTip laser fibers are indicated for Urologic applications for which the laser systems are cleared, limited to endoscopic procedures involving vaporization, ablation, hemostasis, coagulation, excision, resection, incision of soft tissue, and lithotripsy of urinary calculi. The fiber is designed for use with a standard SMA-905 connector and has been cleared for surgical use. | |
| Reusability | Single Use | Single Use | Identical |
| Flexiva Pulse/ ID Laser Fiber is Supplied | Sterile | Sterile | Identical |
Confidential & Proprietary Information of Boston Scientific Corporation
{8}
Flexiva Pulse/ ID Single Use Laser Fibers
Special 510(k)
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Identical |
| --- | --- | --- | --- |
| Packaging | Tyvek pouch | Tyvek pouch | Identical |
| Sizes Offered (μm) | 242, 365, 550, and 910
242 μm TracTip configuration | 242, 365, 550, and 910
242 μm TracTip configuration | Identical |
| Overall Length | 242 μm: 3 meters
365, 500, 910 μm: 2.6 meters | 242 μm: 3 meters
365, 500, 910 μm: 2.6 meters | Identical |
I. Substantial Equivalence
A direct comparison of key characteristics demonstrates that the proposed Flexiva Pulse/ ID Laser Fibers are substantially equivalent to the predicate Flexiva Pulse/ ID Laser Fibers (K210925) in terms of intended use, technological characteristics, and performance characteristics. The proposed Flexiva Pulse/ ID Laser Fiber is as safe, as effective, and performs as well as the predicate devices (K210925).
J. Performance Testing
To evaluate the secondary coating resin material change introduced with the proposed Flexiva Pulse/ ID Laser Fiber, a Design Verification was executed to support the safe and effective use of the proposed Flexiva Pulse/ ID Laser Fiber. The new secondary coating resin was subjected to the following tests:
- Bent Transmission
- Fiber durability while firing
- Fiber connector temperature
- Laser System output accuracy
From a biocompatibility standpoint, the contact classification of the proposed Flexiva Pulse/ ID Laser Fiber is identical to the predicate Flexiva Pulse/ ID Laser Fiber (cleared under K210925). A biological evaluation assessment for the proposed Flexiva Pulse/ ID Laser Fiber was conducted and it was concluded there are no biocompatibility risks associated with the proposed Flexiva Pulse/ ID Laser Fiber.
K. Conclusion
Based on the intended use/indications for use, comparison of key technological characteristics, and performance testing presented in this premarket submission, it is concluded that the proposed Flexiva Pulse/ ID Laser Fibers are substantially equivalent to the predicate Flexiva Pulse/ ID Laser Fibers (cleared under K210925).
Confidential & Proprietary Information of Boston Scientific Corporation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.