← Product Code [NXC](/productcode/NXC) · K251758

# Serafin® (K251758)

_Tns Co., Ltd. · NXC · Sep 11, 2025 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251758

## Device Facts

- **Applicant:** Tns Co., Ltd.
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Sep 11, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.

## Device Story

Serafin® is a clear aligner system for orthodontic treatment of malocclusion. Input: 3D scanned image file of patient's dentition. Process: Dental practitioner uses 3D software to create a treatment plan; software generates incremental 3D models of teeth; models are 3D printed; thermoplastic polyurethane (Zendura FLX) is thermoformed over models and trimmed. Output: Series of transparent, removable appliances worn by patient (at least 20 hours/day) to apply continuous gentle force for tooth realignment. Context: Used in dental clinics; prescribed by dental practitioners. Benefit: Corrects malocclusion (overbite, underbite, crossbite, gaps, crowding) over 6-24 months. Output affects clinical decision-making by providing a staged, customized treatment path for tooth movement.

## Clinical Evidence

Bench testing only. Physical/chemical testing included appearance, dimension, water sorption/solubility, 3-point flexural strength, and hardness per ISO 20795-2. Biological evaluation included cytotoxicity, sensitization, irritation, systemic toxicity, and chemical characterization per ISO 10993 series (Parts 1, 5, 10, 11, 17, 18, 23) and ISO 7405. All results met preset criteria.

## Technological Characteristics

Material: Zendura FLX (thermoplastic urethane/copolyester, 3-layer structure). Principle: Thermoforming on 3D-printed models. Form factor: Transparent removable orthodontic appliances. Connectivity: None. Sterilization: Not stated. Software: 3D modeling/treatment planning software. Standards: ISO 20795-2 (orthodontic base polymers), ISO 10993 (biocompatibility).

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- Serafin® ([K232074](/device/K232074.md))
- Invisalign System ([K220287](/device/K220287.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

September 11, 2025

TNS Co., Ltd.
Byung Ju Kang
Regulatory Affairs Manager
19F, Unit 1907, Daeryung Post Tower 6th
298 Beotkkot-ro, Geumcheon-gu
Seoul, 08510
REPUBLIC OF KOREA

Re: K251758
Trade/Device Name: Serafin®
Regulation Number: 21 CFR 872.5470
Regulation Name: Orthodontic Plastic Bracket
Regulatory Class: Class II
Product Code: NXC
Dated: September 8, 2025
Received: September 8, 2025

Dear Byung Ju Kang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251758 - Byung Ju Kang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K251758 - Byung Ju Kang
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251758 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Serafin®  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) SUMMARY

K251758

Date: September 8, 2025

1. SUBMITTER
TNS Co., Ltd.
19F, Unit 1907, Daeryung Post Tower 6th
298 Beotkkot-ro, Geumcheon-gu, Seoul 08510
TEL: +82-2-785-2804
FAX: +82-2-785-2803
Contact Name: Won-gun Chang
Email: tns0109@tnsmile.kr

2. DEVICE
· Trade Name: Serafin®
· Common Name: Clear Aligner System
· Regulation Number 872.5470
· Class: 2
· Classification Product Code: NXC
· Classification Name: Orthodontic plastic bracket

3. PREDICATE DEVICE
Primary Predicate: Serafin® (K232074)
Additional Predicate: Invisalign System (K220287, Align Technology Inc.)

4. DEVICE DESCRIPTION
Serafin® is composed of series of transparent removable orthodontic appliances manufactured using thermoforming technology based on a 3D model of the patient's teeth, customized according to a Dental Practitioner's specific prescription, indicated for

{5}

treatment of dental malocclusion by applying continuous gentle force to realign teeth.

The time for each stage of aligner usage is set in the personal treatment plan, and this varies for each patient and the Dental Practitioner's choice. Each set of aligners needs to be worn for at least 2 weeks. Patients the total treatment time lasts about 6 to 24 months.

## 5. INDICATIONS FOR USE

Serafin®

- The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.

## 6. NON-CLINICAL TESTING

The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.

1) Physical and chemical test

- Guideline on Review and Approval for appliance, Orthodontic, Resin (Sheet Type)(2018.12.) (MFDS)
- Appearance, Dimension, Water sorption &amp; solubility, 3-point flexural strength, hardness
- Appearance, pH, Potassium permanganate reducing substances residue, Residue on evaporation, Heavy metals, Ultraviolet-visible spectrum
- ISO 20795-2 Dentistry – Base polymers – Part 2: Orthodontic base polymers

2) Biological evaluation

- ISO 10993-1 – Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 7405 – Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
- ISO 10993-17 – Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances
- ISO 10993-18 – Biological evaluation of medical devices - Part 18: Chemical characterization

{6}

- ISO 10993-5 – Biological evaluation of medical devices - Part 5: Cytotoxicity
- ISO 10993-10 - Biological evaluation of medical devices - Part 10: Skin sensitization, Acute oral mucosa irritation, Intracutaneous reactivity
- ISO 10993-11 - Biological evaluation of medical devices - Part 11: Acute systemic toxicity
- ISO 10993-23 - Biological evaluation of medical devices - Part 23: Tests for irritation
- ADA ANSI 41-2020 Evaluation of Biocompatibility of Medical Devices Used in Dentistry

7. SUBSTANTIAL EQUIVALENCE

|   | Subject device | Predicate Device | Discuss/Justify the Differences  |
| --- | --- | --- | --- |
|  510(k) Number | New | K220287 | -  |
|  Trade Name | Serafin® | Invisalign | -  |
|  Manufacturer | TNS Co., Ltd. | Align Technology Inc. | -  |
|  Common Name | Clear Aligner System | Aligner, Sequential | Same  |
|  Device Class | 2 | 2 | Same  |
|  Product Code | NXC | NXC | Same  |
|  Device Description | Serafin® is composed of series of transparent removable orthodontic appliances manufactured using thermoforming technology based on a 3D model of the patient's teeth, customized according to a Dental Practitioner's specific prescription, indicated for treatment of dental malocclusion by applying continuous gentle force to | Invisalign is a series of transparent removable orthodontic appliances, indicated for treatment of dental malocclusion. Invisalign consists of upper and lower parts in variable sizes and lengths customized for a patient. It is produced by using 3D printer based on 3D scanned image file sent by a Dental Practitioner. The aligners and attachment templates are made | Equivalent  |

{7}

|   | realign teeth.
The time for each stage of aligner usage is set in the personal treatment plan, and this varies for each patient and the Dental Practitioner's choice. Each set of aligners needs to be worn for at least 2 weeks. For patients, the total treatment time lasts about 6 to 24 months. | of thermoplastic polymer, with/without PFA system made of photopolymerizable methacrylate based resin. |   |
| --- | --- | --- | --- |
|  **Indications for use** | **The Serafin is intended for the orthodontic treatment of malocclusion in patients with mixed and permanent dentition.** | **The Invisalign System is intended for the orthodontic treatment of malocclusion in patients with primary, mixed, and permanent dentition.** | **Same intended use; dentition range aligned via additional predicate (K220287).**  |
|  Clinical condition | Overbite, underbite, crossbite, gap teeth, open bite, crowded teeth, baby and permanent teeth.
Limitations include severe cases where large movement of a tooth is necessary such as a tooth rotated over 20 degrees, large gap between teeth, teeth which require intrusion or extrusion | Overbite, underbite, crossbite, gap teeth, open bite, crowded teeth, baby and permanent teeth.
Limitations include severe cases where large movement of a tooth is necessary such as a tooth rotated over 20 degrees, large gap between teeth, teeth which require intrusion or extrusion | Same  |
|  **Patient Population** | **Children (mixed dentition); Adolescents/Adults (permanent dentition).** | **Children (primary/mixed); Adolescents/Adults (permanent).** | **Difference addressed by identifying Invisalign (K220287) as an additional predicate; no new questions of safety or effectiveness.**  |
|  Use Location | Dental intraoral Device | Dental intraoral Device | Same  |
|  OTC or | Rx only | Rx only | Same  |

{8}

|  Prescription (Rx) Device |  |  |   |
| --- | --- | --- | --- |
|  Design  |   |   |   |
|  Principle of Operation | Serafin® is series of transparent removable orthodontic appliances, indicated for treatment of dental malocclusion. It is produced by using 3D printer based on 3D scanned image file sent by a Dental Practitioner along with a prescription of a specific patient. Series of clear plastic appliances (Aligners) move the patients’ teeth in small increments from their original state to a final, treated state by applying continuous gentle force to teeth while the patient is wearing it. | The Invisalign System consists of a series of doctor-prescribed, thin, clear plastic removable orthodontic appliances (aligners) and proprietary 3-D software. The aligners gently move the patient’s teeth in small increments from their original state to a more optimal, treated state. | Equivalent  |
|  Appearance |  |  | Equivalent  |
|  3D Software | 3D Software: Produces a 3D model file of the digital scan. Create a treatment plan, which is reviewed by the dental practitioner to reject or request modifications. Approved treatment plan model is 3D printed. | 3D Software: Produces 3D model file of the PVS impression or the digital scan. Create a treatment plan which is reviewed by dental practitioner to reject or request modifications. Approved treatment plan model is 3D printed. | Equivalent  |

{9}

|  Materials | Thermoplastic polymer : Zendura FLX : Thermoplastic Urethane, Copolyester (3-layer structure) | Thermoplastic polymer SmartTrack™ (Multilayer structure) | Equivalent  |
| --- | --- | --- | --- |
|  Physical State | Transparent removable orthodontic appliances consist of upper and lower parts | Transparent removable orthodontic appliances consist of upper and lower parts | Same  |
|  Material property | Zendura FLX Demonstrates sufficient Ultimate Flexural Strength, Water Sorption, Water Solubility as per ISO 20795-2 : 2013. Intended, marketed to be used as dental aligner | SmartTrack™ is material engineered for use with Invisalign orthodontic treatment (Polyurethane) Intended to be used as dental aligner | Equivalent  |
|  Manufacturing process | Thermoforming on models, Trimming, Marking | Thermoforming on models, Trimming, Marking | Same  |
|  Biological  |   |   |   |
|  Tissue Contact | Surface device, permanent use (>30 days), Mucosal membrane | Surface device, permanent use (>30 days), Mucosal membrane | Same  |
|  Anatomical site | Intact teeth, gingival mucosa, interior vermilion (inner lip), and tongue | Intact teeth, gingival mucosa, interior vermilion (inner lip), and tongue | Same  |
|  Contact duration | at least 20 hrs or longer a day, up to 30 days | at least 20 hrs or longer a day, up to 30 days | Same  |
|  Biocompatibility | Biological risk assessment, Toxicological risk assessment as per ISO 10993-1, ISO 7405 | Biocompatibility per the ISO 10993 series standards | Equivalent  |

We identified an additional predicate (Invisalign System, K220287) that covers primary, mixed, and permanent dentition. The subject and predicates share the same intended use and technological characteristics; the absence of MAF in the subject device does not raise new questions of safety or effectiveness.

{10}

# 8. SUBSTANTIAL EQUIVALENCE DISCUSSION

The subject device shares the same indications for use and principle of operation as the predicate devices. They are intended for the same clinical conditions, patient populations, and use locations.

Predicate device includes an optional mandibular advancement feature (MAF), but subject device does not provide a MAF feature. The MAF option is a necessary function for 'Skeletal malocclusion', so it does not correspond to the indications for subject device.

Both devices comprise a series of transparent, removable orthodontic appliances, used for at least 20 hours a day over a period of several weeks. The manufacturing process begins with a 3D image file of a patient's dentition, which serves as an input to each manufacturer's software. Each software identifies the individual teeth requiring treatment. The software then generates a 3D model of the patient's dentition, incrementally modifying it towards alignment. This 3D model is then printed out using a 3D printer. Subsequently, a thermoplastic polyurethane material is heated and moulded onto the 3D printed model, known as a thermoforming process. Any excess material is eliminated by trimming process. The resulting plastic appliances, both for Serafin® and Invisalign, are designed to apply a gentle, continuous force on the malocclusion site to realign the teeth.

Correspondingly, the principles of operation, intended use, manufacturing process, and 3D modeling process for Serafin® and Invisalign are identical. Both devices utilize a similar thermoforming polymer as a material, which demonstrates chemical, biological, and functional equivalence. The materials are thermoplastic polymers with a biocompatible polyurethane copolymer and are intended for use as clear aligner material.

However, since the materials are not exactly the same, further biological assessments and pre-clinical performance tests have been carried. For biocompatibility, Serafin® was tested as per ISO 10993-1 in Biological Risk Assessment. In addition, Toxicological

{11}

Risk Assessment was conducted to monitor and mitigate the biological risk from unintended process residuals and degradation material under the light-curing process in 3D printing and the thermoforming process.

Regarding physical properties and clinical performance, Zendura FLX was tested according to “ISO 20795-2 Dentistry - Base polymers - Part 2: Orthodontic base polymers,” under a 37°C wet environment. Since orthodontic base polymers are categorized as Type 3 (thermoplastic materials), testing was conducted for Flexural Modulus, Flexural Strength, Tear Strength, Water Sorption, and Water Solubility. The Flexural Modulus did not meet standard requirement of 1500 MPa. However, this requirement is relevant to highly rigid or inflexible materials, which does not reflect the current requirements for thermoplastic aligners. Considering that Serafin® and Invisalign are engineered to apply a subtle force for realignment, this standard does not apply, and the value primarily serves as an informative detail about the device's physical properties. The other results conformed to the standard, indicating good toughness, an effective ability to move teeth, low water uptake, and low water solubility.

Consequently, it can be concluded that the subject devices are substantially equivalent to the predicate devices in terms of similarity, risk acceptability, and efficacy.

---

**Source:** [https://fda-staging.innolitics.com/device/K251758](https://fda-staging.innolitics.com/device/K251758)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
