The GenaCheck® COVID-19 Rapid Self-Test is a visually read lateral flow immunoassay test intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment. Positive results do not rule out co-infection with other respiratory pathogens. This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider. The performance characteristics for SARS-CoV-2 were established from March 2024 to October 2024 when SARS-CoV-2 Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected.
Device Story
Lateral flow immunoassay for qualitative detection of SARS-CoV-2 nucleocapsid protein antigen. Input: anterior nasal swab specimen; processed in lysis buffer to disrupt viral particles. Specimen applied to test cassette; migrates via capillary action across nitrocellulose membrane. Binding pad contains anti-SARS-CoV-2 antibodies labeled with latex bead marker. Test region (T) coated with monoclonal anti-SARS-CoV-2 antibodies; control region (C) coated with goat anti-mouse IgG. Immune complexes form visible pink lines in T and C regions for positive results; only C region for negative. Used in home setting by lay users. Results interpreted visually by user within 10-30 minutes. Negative results are presumptive and require serial testing. Positive results indicate presence of viral antigen but do not rule out co-infection. Assists users in identifying potential COVID-19 infection to guide follow-up care.
Clinical Evidence
Prospective multi-center study (n=643 evaluable) conducted March-October 2024. Compared lay-user self-collected anterior nasal swabs against high-sensitivity RT-PCR comparator. Results: PPA 93.3% (95% CI: 87.4%-96.6%); NPA 98.7% (95% CI: 97.3%-99.4%). Performance stratified by days since symptom onset (1-5 days) showed consistent agreement. Usability study (n=120) and readability study (n=69) confirmed lay users can correctly perform the test and interpret results.
Indicated for symptomatic individuals aged 2 years or older for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigen in anterior nasal swab specimens. For non-prescription home use.
Regulatory Classification
Identification
For the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen in individuals 2 years and older.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K251753
B Applicant
Genabio Diagnostics Inc.
C Proprietary and Established Names
GenaCheck COVID-19 Rapid Self-Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QYT | Class II | 21 CFR 866.3984 - Over-The-Counter Test To Detect SARS-Cov-2 From Clinical Specimens | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain 510k clearance for the GenaCheck COVID-19 Rapid Self-Test
B Measurand:
SARS CoV-2 nucleocapsid protein antigen
C Type of Test:
Qualitative lateral flow immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The GenaCheck COVID-19 Rapid Self-Test is a visually read lateral flow immunoassay test intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19.
This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment.
Positive results do not rule out co-infection with other respiratory pathogens.
This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider.
The performance characteristics for SARS-CoV-2 were established from March 2024 to October 2024 when SARS-CoV-2 Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspect.
C Special Conditions for Use Statement(s):
OTC - Over The Counter
D Special Instrument Requirements:
Not applicable.
IV Device/System Characteristics:
A Device Description:
The GenaCheck COVID-19 Rapid Self-Test Kit is a latex particle based immunochromatography test. The test is designed to detect nucleocapsid protein antigen in self-collected anterior nasal swab specimens from individuals who show symptoms of COVID-19 infection within the first five (5) days of symptom onset. The GenaCheck COVID-19 Rapid Self-Test Kit is validated for use from direct specimens testing without transport media.
K251753 - Page 2 of 19
{2}
The test package is composed of the following components:
Test Cassette within a Foil Pouch
- Lysis Buffer
- Anterior Nasal Swab
Quick Reference Instructions
The test cassette contains a test strip (binding pad) in a plastic housing. The binding pad is a nitrocellulose membrane with two regions: a test region (T) and control region (C). A schematic of the test device is shown in Figure 1.

Figure 1: Schematic of Genabio Genacheck COVID-19 Test Cassette

# B Principle of Operation:
To perform the test, an anterior nasal swab is obtained by the lay user. Swabs can be self-collected by the user (age $\geq 14$ years) or collected by an adult from another individual (age $\geq 2$ years). For sample processing, the swab is first inserted into the lysis buffer during which the buffer disrupts the virus particles in the specimen, exposing internal viral nucleocapsid antigens. Three (3) drops of the lysed specimen are added to the sample well on the test cassette, which migrate from the sample well over the binding pad via capillary action.
The binding pad is coated with anti-SARS-CoV-2 antibodies which are labeled with a latex bead marker. The test region is coated with monoclonal anti-SARS-CoV-2 antibodies. The control region is coated with goat anti-mouse IgG antibodies. During testing, the anti-SARS-CoV-2 antibodies labeled with the latex bead marker form immune complexes with the antigen protein of the virus in the specimen, if present. The immune complexes move along the membrane via capillary action and are captured by the anti-SARS-CoV-2 monoclonal antibodies coated in the test region to form a visible pink colored line. The anti-SARS-CoV-2 antibodies or immune
K251753 - Page 3 of 19
{3}
complexes continue to move forward and specifically bind to the goat anti-mouse antibody coated in the control region to form a visible line with pink color.
## C Interpretation of Results
### 1. Positive Result

The Control line (C line) and Test line (T line) both appear as pink-colored lines in the window. A positive test result is interpreted as protein antigen from the virus that causes COVID-19 was detected in the specimen. The individual is likely positive for COVID-19. There is a very small chance that this test can give a positive result that is incorrect (a false positive).
### 2. Negative Result

A negative test means that antigen from the virus that causes COVID-19 is not detected in the sample. Negative results do not rule out SARS-CoV-2 infection. Individuals without symptoms that test negative should be tested again within at least 24 hours and no more than 48 hours between tests. All negative results are considered presumptive, and confirmation with a molecular assay, if necessary for patient management, may be performed.
K251753 - Page 4 of 19
{4}
K251753 - Page 5 of 19
# 3. Invalid Result


If no line appears in the Control (C) area, the test result is invalid regardless of the presence or absence of a line in the Test (T) area. The test should be repeated with a new GenaCheck COVID-19 Rapid Self-Test.
# V Substantial Equivalence Information:
A Predicate Device Name(s):
Flowflex Plus COVID-19 Home Test
B Predicate 510(k) Number(s):
K233373
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K251753 | K233373 |
| --- | --- | --- |
| Device Trade Name | GenaCheck COVID-19 Rapid Self-Test | Flowflex Plus COVID-19 Home Test |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The GenaCheck COVID-19 Rapid Self-Test is a visually read lateral flow immunoassay test intended for the rapid, qualitative detection of SARS-CoV-2 virus nucleocapsid protein antigen directly in anterior nasal swab specimens from individuals with signs and symptoms of COVID-19. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. | The Flowflex Plus COVID-19 Home Test is a visually read lateral flow immunoassay device intended for the rapid, qualitative detection of SARS CoV-2 nucleocapsid protein antigens directly in anterior nasal (nares) swab specimens from individuals with signs and symptoms of COVID-19 within the first 6 days from symptom onset. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. |
{5}
K251753 - Page 6 of 19
| | adults testing individuals aged 2 years or older.
All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS-CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment.
Positive results do not rule out co-infection with other respiratory pathogens.
This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider.
The performance characteristics for SARS-CoV-2 were established from March 2024 to October 2024 when SARS-CoV-2 Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected. | The Flowflex Plus COVID-19 Home Test does not differentiate between SARS-CoV and SARS-CoV-2.
All negative results are presumptive. Symptomatic individuals with an initial negative test result must be re-tested once between 48 and 72 hours after the first test using either an antigen test or a molecular test for SARS CoV-2. Negative results do not preclude SARS-CoV-2 infections or other pathogens and should not be used as the sole basis for treatment.
Positive results do not rule out co-infection with other respiratory pathogens.
This test is not a substitute for visits to a healthcare provider or appropriate follow-up and should not be used to determine any treatments without provider supervision. Individuals who test negative and experience continued or worsening COVID-19 like symptoms, such as fever, cough and/or shortness of breath, should seek follow up care from their healthcare provider.
The performance characteristics for SARS-CoV-2 were established from June 2022 to April 2023 when the COVID-19 variant Omicron was dominant. Test accuracy may change as new SARS-CoV-2 viruses emerge. Additional testing with a lab-based molecular test (e.g., PCR) should be considered in situations where a new virus or variant is suspected. |
| --- | --- | --- |
| Target Analyte | SARS CoV-2 nucleocapsid antigen | Same |
| Specimen type | Anterior nasal swab samples | Same |
| Result interpretation | Visually read | Same |
| Test principle | Lateral flow immunoassay | Same |
| Result type | Qualitative | Same |
| Intended Use Setting | OTC | Same |
{6}
K251753 - Page 7 of 19
| General Device Characteristic Differences | | |
| --- | --- | --- |
| Time to Result | 10-30 min | 15-30 min |
VI Standards/Guidance Documents Referenced:
| Document | Title | Publisher | Applicable Study |
| --- | --- | --- | --- |
| ISO 11135:2014/A1:2018 | Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices - | ISO | Sterility |
| ISO 10993-10:2021 | Biological evaluation of medical devices - Part 10: Tests for skin sensitization | ISO | Biocompatibility |
| ISO 10993-23:2021 | Biological evaluation of medical devices - Part 23: Tests for irritation | ISO | Biocompatibility |
| ISO 10993-7:2008 | Biological evaluation of medical devices - Part 7: Ethylene Oxide Sterilization Residuals | ISO | Sterility |
| ISO 10993-5:2009 | Biological evaluation of medical devices - Part 5: Tests for In vitro Cytotoxicity | ISO | Biocompatibility |
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
a. Multi-Lot Precision
A precision study was conducted to evaluate variability across days, operators, and device lots. The study used a single internal site with two trained operators and three lots of test kits. Each day, three sample levels (Negative, 1.5 X LoD, and 3 X LoD) were tested with two replicates per run, per operator, and per kit lot. Two runs were conducted each day (at least four hours apart, typically morning and afternoon) by each operator with each lot. This exact testing scheme was followed for 12 days, with the same three sample levels tested across the same three lots by the same two operators in two runs per day, yielding a total of 864 tests (288 per level tested).
{7}
Samples with three concentrations of UV-inactivated SARS-CoV-2 virus material (USA-WA1/2020) were prepared by spiking the virus material into negative clinical matrix (NCM), resulting in the following sample types:
Negative Sample
Low Positive Sample at $1.5\mathrm{X}$ LoD
Positive Sample at $3\mathrm{X}$ LoD
For the testing, a $50~\mu \mathrm{L}$ volume of each sample was applied to dry nasal swabs. After blinding and randomization, samples were processed according to the Instructions for Use of the candidate device.
The results of this study are summarized in Table 1. All results matched the expected outcomes for the virus concentrations in the samples (all negative samples tested negative, all low positive and moderate positive samples tested positive). No variability was observed across different conditions, operators, lots, or days.
Table 1: Lot to Lot Precision Study Results
| Analyte Concentration | Lot 1 | | Lot 2 | | Lot 3 | | Total Lot-to-Lot Precision | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n/N | % Agmt | n/N | % Agmt | n/N | % Agmt | n/N | % Agmt | 95% CI |
| Negative | 96/96 | 100% | 96/96 | 100% | 96/96 | 100% | 288/288 | 100% | 98.7-100% |
| 1.5X LoD | 96/96 | 100% | 96/96 | 100% | 96/96 | 100% | 288/288 | 100% | 98.7-100% |
| 3X LoD | 96/96 | 100% | 96/96 | 100% | 96/96 | 100% | 288/288 | 100% | 98.7-100% |
# b. Supplemental Precision Study
This study was specifically conducted to further evaluate potential differences between distinct lots, because the initial precision study resulted in $100\%$ agreement across all sources of variation and did not allow a conclusive assessment of lot-to-lot variability of the test. The study was performed using a negative sample, one low positive sample at $0.8\mathrm{X}$ LoD (i.e., below the concentration tested in the initial study, and near the analyte's $\mathrm{C_{95}}$ concentration) and one positive sample at $3\mathrm{X}$ LoD.
Testing was conducted using the same protocol as the preliminary study, with the exception that the NCM and 3 X LoD samples were only tested with one replicate per run, whereas the 0.8 X LoD samples were tested twice per run. For the negative and the 3 X LoD, samples the results were in $100\%$ agreement with the expected result. Consistent with the analyte concentration of the low positive sample, the 0.8X LoD sample resulted in percent agreements less than $95\%$ , with similar results for all lots. The results are summarized below in Table 2.
K251753 - Page 8 of 19
{8}
Table 2: Supplemental Lot-to-Lot Precision Study Results
| Analyte Concentration | Lot 1 | | Lot 2 | | Lot 3 | | Total Lot-to-Lot Precision | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n/N | % Agmt | n/N | % Agmt | n/N | % Agmt | n/N | % Agmt | 95% CI |
| Negative | 48/48 | 100% | 48/48 | 100% | 48/48 | 100% | 144/144 | 100% | 97.4-100% |
| 0.8X LoD | 77/96 | 80.2% | 80/96 | 83.3% | 81/96 | 84.4% | 238/288 | 82.6% | 77.8-86.6% |
| 3X LoD | 48/48 | 100% | 48/48 | 100% | 48/48 | 100% | 144/144 | 100% | 97.4-100% |
# 2. Linearity:
Not applicable. This device is a qualitative, visually interpreted assay.
# 3. Analytical Specificity/Interference:
# a. Cross Reactivity Microbial Interference
The analytical specificity of the GenaCheck COVID-19 Rapid Self-Test was evaluated by testing various microorganisms and viruses, for cross-reactivity (in the absence of SARS-CoV-2) and microbial interference (in the presence of SARS-CoV-2 at $3\mathrm{X}$ LoD, or $4.57\times 10^{3}$ $\mathrm{TCID}_{50} / \mathrm{mL}$ ). Respiratory viral pathogens were tested at a final concentration of $\geq 1\times 10^{5}$ units/mL, or at the highest available concentrations. Common respiratory bacterial and yeast pathogens were tested at a final concentration of $\geq 1\times 10^{6}$ units/mL, or at the highest available concentrations. Each organism was assessed for potential cross-reactivity with the test antibodies by adding $50~\mu \mathrm{L}$ of each sample directly to the test swab, followed by processing in accordance with the test's IFU. Three replicates per microorganism were tested in this study. Microbial interference testing was conducted similarly, with SARS-CoV-2 (USA-WA1/2020) co-spiked into the samples at $3\mathrm{X}$ LoD.
The results of these studies are summarized in Table 3. Neither cross-reactivity nor microbial interference was observed for any of the tested microorganisms at the concentrations used in the study.
Table 3: Cross Reactivity/Microbial Interference Study Results
| Microbe Added | Concentration Tested | Cross Reactivity (Positive/Tested) | Microbial Interference (Positive/Tested) |
| --- | --- | --- | --- |
| Adenovirus Type 1 | 2.23E+05 TCID50/mL | 0/3 | 3/3 |
| Adenovirus Type 2 | 2.13E+06 TCID50/mL | 0/3 | 3/3 |
| Adenovirus Type 7A | 1.58E+05 TCID50/mL | 0/3 | 3/3 |
| Bordetella pertussis | 2.90E+08 CFU/mL | 0/3 | 3/3 |
| Candida albicans | 1.21E+07 CFU/mL | 0/3 | 3/3 |
| Chlamydia pneumoniae | 4.33E+06 IFU/mL | 0/3 | 3/3 |
| Enterovirus Type 68 | 2.23E+05 TCID50/mL | 0/3 | 3/3 |
| Haemophilus influenzae | 9.68E+06 CFU/mL | 0/3 | 3/3 |
K251753 - Page 9 of 19
{9}
| Microbe Added | Concentration Tested | Cross Reactivity (Positive/Tested) | Microbial Interference (Positive/Tested) |
| --- | --- | --- | --- |
| Human coronavirus 229E | 1.58E+05 TCID50/mL | 0/3 | 3/3 |
| Human coronavirus NL63 | 8.00E+04 TCID50/mL | 0/3 | 3/3 |
| Human coronavirus OC43 | 7.00E+05 TCID50/mL | 0/3 | 3/3 |
| Human Metapneumovirus | 1.40E+05 TCID50/mL | 0/3 | 3/3 |
| Influenza A H1N1 | 1.85E+08 CEID50/mL | 0/3 | 3/3 |
| Influenza A H3N2 | 6.50E+06 FFU/mL | 0/3 | 3/3 |
| Influenza B Victoria | 1.58E+05 TCID50/mL | 0/3 | 3/3 |
| Influenza B Yamagata | 1.90E+06 TCID50/mL | 0/3 | 3/3 |
| Legionella pneumophila | 6.50E+06 CFU/mL | 0/3 | 3/3 |
| MERS-coronavirus | 1.40E+05 TCID50/mL | 0/3 | 3/3 |
| Mycobacterium tuberculosis avirulent | 3.03E+06 CFU/mL | 0/3 | 3/3 |
| Mycoplasma pneumoniae | 2.50E+07 CFU/mL | 0/3 | 3/3 |
| Parainfluenza virus 1 | 2.00E+05 TCID50/mL | 0/3 | 3/3 |
| Parainfluenza virus 2 | 1.40E+05 TCID50/mL | 0/3 | 3/3 |
| Parainfluenza virus 3 | 7.00E+05 TCID50/mL | 0/3 | 3/3 |
| Parainfluenza virus 4 | 2.39E+05 TCID50/mL | 0/3 | 3/3 |
| Pooled Nasal Wash | N/A | 0/3 | 3/3 |
| Respiratory syncytial virus Type A | 3.50E+05 TCID50/mL | 0/3 | 3/3 |
| Respiratory syncytial virus Type B | 2.29E+05 TCID50/mL | 0/3 | 3/3 |
| Rhinovirus 1A | 7.05E+04 TCID50/mL | 0/3 | 3/3 |
| Rhinovirus B14 | 1.31E+05 TCID50/mL | 0/3 | 3/3 |
| SARS-CoV | 1.25E+05 PFU/mL | 0/3 | 3/3 |
| Staphylococcus aureus | 2.60E+08 CFU/mL | 0/3 | 3/3 |
| Staphylococcus epidermidis | 9.00E+07 CFU/mL | 0/3 | 3/3 |
| Streptococcus pneumoniae | 1.81E+07 CFU/mL | 0/3 | 3/3 |
| Streptococcus pyogenes | 7.50E+07 CFU/mL | 0/3 | 3/3 |
| HKU1 clinical sample 1 | Ct value: 31.8 | 0/3 | 3/3 |
| HKU1 clinical sample 2 | Ct value: 27.4 | 0/3 | 3/3 |
| HKU1 clinical sample 3 | Ct value: 23.1 | 0/3 | 3/3 |
# b. Exogenous Endogenous Interference Studies
Thirty (30) potentially interfering substances were evaluated using the GenaCheck COVID-19 Rapid Self- Test. Each substance was tested in triplicate, both in the absence and presence of SARS-CoV-2 at $3\mathrm{X}$ LoD $(4.57\times 10^{3}\mathrm{TCID}_{50} / \mathrm{mL})$ . The results of this
K251753 - Page 10 of 19
{10}
study are summarized in Table 4. Among all the substances tested, hand sanitizer cream lotion was the only one to cause a false-positive result when tested at concentrations of $5\%$ w/v or above. All other substances yielded expected results, indicating they do not impact the performance of the GenaCheck COVID-19 Rapid Self-Test at the concentrations specified.
Table 4:Exogenous/ Endogenous Interference Study Results
| Interfering Substances Tested | Test Concentration | SARS-CoV-2 Negative Sample (Positive/Tested) | SARS-CoV-2 Positive Sample (Positive/Tested) |
| --- | --- | --- | --- |
| Throat Lozenges (Benzocaine, Menthol) | 3 mg/mL | 0/3 | 3/3 |
| Sore Throat Spray (Phenol) | 5% v/v | 0/3 | 3/3 |
| Mucin (Purified mucin protein) | 2.5 mg/mL | 0/3 | 3/3 |
| Whole Blood (Human) | 2.5% v/v | 0/3 | 3/3 |
| Leukocytes | 2×10^6 cells/mL | 0/3 | 3/3 |
| Throat Spray (Zinc) | 15% v/v | 0/3 | 3/3 |
| Nasal Drops (Phenylephrine) | 15% v/v | 0/3 | 3/3 |
| Nasal Spray (Cromolyn) | 15% v/v | 0/3 | 3/3 |
| Nasal Spray (Oxymetazoline) | 15% v/v | 0/3 | 3/3 |
| Nasal Spray (Saline) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Beclomethasone) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Dexamethasone) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Flunisolide) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Triamcinolone) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Budesonide) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Mometasone furoate) | 15% v/v | 0/3 | 3/3 |
| Nasal corticosteroids (Fluticasone propionate) | 15% v/v | 0/3 | 3/3 |
| Nasal Spray (Zicam, Luffa oppeculata, Galphimia glauca, Histaminum hydrochloricum) | 15% v/v | 0/3 | 3/3 |
| Alkalol | 15% v/v | 0/3 | 3/3 |
| Oseltamivir Phosphate (Tamiflu) | 5 mg/mL | 0/3 | 3/3 |
| Remdesivir | 10 mg/mL | 0/3 | 3/3 |
| Molnupiravir | 10 mg/mL | 0/3 | 3/3 |
| Mupirocin | 10 mg/mL | 0/3 | 3/3 |
| Nasal gel | 5% v/v | 0/3 | 3/3 |
| Body Lotion (Dimethicone) | 0.5% w/v | 0/3 | 3/3 |
| Hand Sanitizer (62% Ethyl alcohol) | 5% v/v | 0/3 | 3/3 |
| Hand Sanitizer Cream Lotion* (Benzalkonium chloride) | 5% w/v | 3/3 | 3/3 |
| Hand Sanitizer (80% Ethyl alcohol) | 15% v/v | 0/3 | 3/3 |
| Hand Soap | 10% w/v | 0/3 | 3/3 |
| Biotin | 3,500 ng/mL | 0/3 | 3/3 |
K251753 - Page 11 of 19
{11}
* Hand Sanitizer Cream Lotion was found to not interfere with testing at 1.67% w/v, giving negative results in 3 out of 3 replicates tested.
4. Assay Reportable Range:
Not applicable, qualitative assay.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
a. Controls
The GenaCheck COVID-19 Rapid Self-Test has a built-in internal procedural control. A built-in internal control is needed to indicate the test is working and is automatically run onboard within the development time of each test run. This procedural control consists of goat anti-mouse IgG antibodies and generates a colored line at the "C" marked control region of the test window. The control line should always appear to demonstrate that the test procedure was performed properly, reagents are working as intended, and for a result to be valid.
b. Stability
i. Shelf-life
The shelf-life of the GenaCheck COVID-19 Rapid Self-Test under the intended storage conditions (2-30°C) was assessed in a real-time stability study using three (3) lots of test kits stored at 2-8°C and 30°C and 85% Relative Humidity (RH). The test kits were assessed with negative samples (NCM) and contrived positive samples prepared at 3 X LoD (4.57E+03 TCID₅₀/mL) using SARS-CoV-2 (USA-WA1/2020). Replicate samples were generated by spiking 50μL of virus diluted in NCM onto dry swabs. For each lot, five replicates per time point per storage condition were tested. Data collected to date in this study (baseline, 14 days, 4 months, 7 months, 10 months, 14 and 18 months) supports a test kit shelf-life of 14 months when stored at the intended storage temperature of 2-30°C at the time of clearance.
ii. Shipping Stability:
In order to evaluate the stability of the test kit under high temperature shipping conditions, kits from three lots were stored at: 60°C with 85% ±5% relative humidity for periods from 1 to 8 days. The stored test kits were tested in five (5) replicates daily through the entire test system including the sample application and processing step for each storage condition. All negative samples tested negative; all positive samples (3 X LoD) tested positive over the 8-day time course.
In order to determine the effect of variable extreme temperature shipping conditions on the performance of the test, kits from three (3) different lots were subjected to three freeze thaw cycles and stored at or 55°C for periods of 7, 21, or 35 days. All negative samples tested negative; all positive samples (3 X LoD) tested positive in all storage conditions.
iii. Sample Stability:
Samples in the OTC environment will not undergo storage as the IFU instructs the user to immediately proceed from sample collection to the testing steps.
K251753 - Page 12 of 19
{12}
6. Detection Limit:
# a. Limit of Detection (LoD)
The Limit of Detection (LoD) of the GenaCheck COVID-19 Rapid Self-Test was determined by evaluating different dilutions of heat-inactivated SARS-CoV-2 (isolate USA-WA1/2020) in NCM. The LoD was determined as the lowest virus concentration that was detected $>95\%$ of the time (e.g., concentration at which at least 19 out of 20 replicates tested positive). The limit of detection was established in two phases.
# i. Preliminary LoD Study:
Five 10-fold serial dilutions were prepared from heat-inactivated SARS-CoV-2 virus into NCM. Five replicates of each serial dilution were tested to determine the preliminary LoD concentration of the device. The lowest concentration with $3/3$ positive results from each lot was considered the preliminary LoD of the virus strain. For each replicate, $50~\mu \mathrm{l}$ of the virus dilution was applied to the swab and the swab was processed according to the test's Instructions for Use. The results are summarized below in Table 5 below.
Table 5: Preliminary LoD Study Summary
| Concentration (TCID50/mL) | Positive Results (n/N) |
| --- | --- |
| 4.57E+06 (Stock) | 3/3 |
| 4.57E+05 | 3/3 |
| 4.57E+04 | 3/3 |
| 4.57E+03 | 3/3 |
| 4.57E+02 | 0/3 |
| 4.57E+01 | 0/3 |
# ii. Confirmatory LoD Study:
The preliminary LoD concentration was tested with a total of twenty (20) replicates for each of 3 kit lots with at least two 2-fold dilutions tested below to demonstrate that the concentration above the LoD were $100\%$ positive and the concentration below the LoD were $< 95\%$ positive. The results are summarized in Table 6 below.
Table 6: Confirmatory LoD Study Summary
| Concentration (TCID50/mL) | Positive results (n/N) | | |
| --- | --- | --- | --- |
| | Kit Lot 1 | Kit Lot 2 | Kit Lot 3 |
| 4.57E+03 | 20/20 | 20/20 | 20/20 |
| 1.52E+03 | 20/20 | 20/20 | 20/20 |
| 5.08E+02 | 0/20 | 2/20 | 0/20 |
The limit of detection for the GenaCheck COVID-19 Rapid Self-Test was determined to be $1.52 \times 10^{3} \mathrm{TCID}_{50} / \mathrm{mL}$ of sample and was achieved by all tested lots. This is equivalent to $7.6 \times 10^{1} \mathrm{TCID}_{50} / \mathrm{swab}$ .
K251753 - Page 13 of 19
{13}
K251753 - Page 14 of 19
b. International Standard for SARS-CoV-2 Antigen (NIBSC code: 21/368)
The LoD of the GenaCheck COVID-19 Rapid Self-Test was also determined by evaluating different dilutions of SARS-CoV-2 antigen (NIBSC code: 21/368) in negative clinical matrix (NCM). Two vials of SARS-CoV-2 Antigen (NIBSC 21/368) were reconstituted in 250 µL ultra-pure water each and then combined for testing one (1) lot of the GenaCheck COVID-19 Rapid Self-Test. The LoD was determined as the lowest virus concentration that was detected ≥95% of the time (i.e., concentration at which at least 19 out of 20 replicates tested positive). The limit of detection was established in two (2) phases.
i. Preliminary LoD Study
The SARS-CoV-2 antigen stock solution (concentration 2 x 10⁴ IU/mL) was diluted 1:5, 1:10, and 1:20 into NCM. An additional 2-fold series of dilutions was prepared from the 1:20 dilution to a level of 62.5 IU/mL (1:320). Three (3) replicates were tested for each dilution level. For each replicate, 50 µL of virus dilution was applied to a swab and the swab was processed according to the IFU. The lowest concentration with 3/3 positive results was considered the preliminary LoD (i.e., 250 IU/mL). The results are summarized below in Table 7.
Table 7: Preliminary LoD for SARS-CoV-2 Antigen (NIBSC code: 21/368)
| Concentration of the International Standard (IU/mL) | Positive results (n/N) |
| --- | --- |
| 20,000 | 3/3 |
| 4,000 | 3/3 |
| 2,000 | 3/3 |
| 1,000 | 3/3 |
| 500 | 3/3 |
| 250 | 3/3 |
| 125 | 0/3 |
| 62.5 | 0/3 |
ii. Confirmatory LoD Study
The NIBSC 21/368 material was tested at one 3-fold concentration above and one 3-fold concentration below the preliminary LoD concentration in 20 replicates for each. To confirm the preliminary concentration as the LoD, at least 19 of 20 replicates (95%) should be positive. The results of the confirmatory testing are summarized below in Table 8.
Table 8: Confirmatory LoD for SARS-CoV-2 Antigen (NIBSC code: 21/368)
| Concentration of the International Standard (IU/mL) | Positive results (n/N) |
| --- | --- |
| 750 | 20/20 |
| 250 | 20/20 |
| 83.3 | 0/20 |
The LoD for the GenaCheck COVID-19 Rapid Self-Test using the International Standard for SARS-CoV-2 antigen (NIBSC code: 21/368) in nasal matrix was confirmed
{14}
to be 250 IU/mL (12.5 IU/swab).
7. Assay Cut-Off:
Not applicable. This is a qualitative, visually interpreted assay.
8. High Dose Hook Effect
No high-dose hook effect was observed with a concentration of 4.57 x 10⁶ TCID₅₀/mL of heat-inactivated SARS-CoV-2 virus tested the GenaCheck COVID-19 Rapid Self-Test which corresponded to ~3000X LoD.
9. Inclusivity (Analytical Reactivity)
The analytical reactivity of the GenaCheck COVID-19 Rapid Self-Test was demonstrated by testing 10 additional human SARS-CoV-2 variants listed in Table 9 below. Each variant was serially diluted 3-fold in negative clinical matrix (NCM) and tested in triplicate. The minimum detectable level of each variant (lowest concentration at which all triplicates tested positive) is summarized in Table 9 below.
Table 9: Inclusivity Study Results
| SARS-CoV-2 Variant | Minimum Detectable Level (TCID_{50}/mL) |
| --- | --- |
| SARS-CoV-2 Lineage B.1.1.7; Alpha Variant | 1.18E+04 |
| SARS-CoV-2 Lineage B.1.351; Beta Variant | 1.88E+03 |
| SARS-CoV-2 Lineage P.1 Brazil; Gamma variant | 1.56E+03 |
| SARS-CoV-2 Lineage B.1.617.2; Delta Variant | 1.30E+03 |
| SARS-CoV-2 Lineage XBB; Omicron Variant | 7.35E+03 |
| SARS-CoV-2 Lineage BA.2.3; Omicron Variant | 1.44E+02 |
| SARS-CoV-2 Lineage BA.5; Omicron Variant | 8.15E+02 |
| SARS-CoV-2 Lineage B.1.1.529; Omicron Variant | 4.80E+01 |
| SARS-CoV-2; Hong Kong/VM20001061/2020 | 5.19E+02 |
| SARS-CoV-2; Italy-INMI1 | 3.93E+03 |
B Comparison Studies:
1. Method Comparison with Predicate Device:
See Section C. Clinical Studies
2. Matrix Comparison:
Not applicable, the test uses anterior nasal swabs only.
K251753 - Page 15 of 19
{15}
# C Clinical Studies:
# 1. Clinical Sensitivity:
The performance of the GenaCheck COVID-19 Rapid Self-Test was compared to an FDA-cleared SARS-CoV-2 molecular assay in a prospective clinical study completed at nine (9) sites in the United States. Samples were collected by lay users from themselves or collected for a household member. A total of 678 subjects who were currently experiencing symptoms associated with COVID-19 were enrolled into the study. Of those, 643 were within 6 days of symptom onset (i.e., 5 days post symptom onset) and evaluable based upon the inclusion and exclusion criteria for analysis. The demographics of the enrolled clinical study subjects are summarized in Table 10 below.
Table 10: Clinical Study Demographics (All Enrolled Subjects)
| Demographic characteristic | n (%) |
| --- | --- |
| Age Group | |
| 2-13 years | 103 (16.0%) |
| 14-21 years | 52 (8.1%) |
| 22-64 years | 403 (62.7%) |
| ≥65 years | 85 (13.2%) |
| Gender | |
| Male | 249 (38.7%) |
| Female | 394 (61.3%) |
| Education Level (Lay User) | |
| Grade 12 or GED | 261 (40.6%) |
| Trade/tech/vocational | 136 (21.1%) |
| Some college, no degree | 105 (16.3%) |
| Bachelor's degree | 97 (15.1%) |
| Graduate or professional | 37 (5.8%) |
| Not reported | 7 (1.1%) |
An analysis based upon the comparator's Ct values estimates that the study cohort included $30.8\%$ low positive samples. The GenaCheck COVID-19 Rapid Self-Test detected SARS-CoV-2 with a Positive Percent Agreement (PPA) of $93.3\%$ and a Negative Percent Agreement (NPA) of $98.7\%$ in symptomatic individuals as compared to a highly sensitive FDA 510(k) cleared molecular SARS-CoV-2 assay. Results are provided in Tables 11 (overall) and 12 (broken down by DPSO).
K251753 - Page 16 of 19
{16}
Table 11: Clinical Study Results
| GenaCheck Self-Test | Comparator Method | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 112 | 7 | 119 |
| Negative | 8 | 516 | 524 |
| Total | 120 | 523 | 643 |
| Positive Percent Agreement (PPA) | 93.3% (112/120) (95% CI: 87.4%-96.6%) | | |
| Negative Percent Agreement (NPA) | 98.7% (516/523) (95% CI: 97.3%-99.4%) | | |
Table 12: Clinical Study Results by Days Past Symptom Onset (DPSO)
| DPSO | TN1 | FP2 | TP3 | FN4 | Total | PPA | NPA |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 28 | 2 | 4 | 0 | 34 | 100% (51.0-100%) | 93.3% (78.7-98.2%) |
| 2 | 147 | 3 | 39 | 3 | 192 | 92.9% (81.0-97.5%) | 98.0% (94.3-99.3%) |
| 3 | 161 | 1 | 40 | 4 | 206 | 90.9% (78.8-96.4%) | 99.4% (96.6-99.9%) |
| 4 | 114 | 1 | 18 | 1 | 134 | 94.7% (75.4-99.1%) | 99.1% (95.2-99.8%) |
| 5 | 66 | 0 | 11 | 0 | 77 | 100% (74.1-100%) | 100% (94.5-100%) |
| Total | 516 | 7 | 112 | 8 | 643 | 93.3% (87.4-96.6%) | 98.7% (97.3%-99.4%) |
$^{1}$ TN: true negative; $^{2}$ FP: false positive; $^{3}$ TP: true positive; $^{4}$ FN: false negative
# 2. Usability Study:
A usability study was conducted to assess the lay users' ability to understand the instructions for use to adequately execute the GenaCheck COVID-19 Rapid Self-Test workflow accordingly. The study took place at a single clinical site, where lay users followed the IFU to complete the test and subsequently answered questionnaires about the test kit and testing process. The lay user's test process was observed and noted by a research staff member.
A total of 120 lay users (46% male, 54% female) with diverse ages and educational backgrounds participated in the study, including 106 who performed self-tests and 14 who tested children aged 2-13 years. All participants completed user comprehension questionnaires, which were divided into three sections: general questions about the test, questions about using the instruction documents, and questions about performing the test steps.
A total of thirteen (13) predefined tasks were rated by the observer for each study subject participating in the usability study. The usability study obtained an overall success rate of $99.6\%$ (1555/1560 tasks) with almost all critical tasks exhibiting a success rate of $100\%$ . Additionally, $89\%$ of participants reported favorable responses (very easy and easy) and $11\%$ reported neutral responses (OK), regarding their ability to understand and follow the test's IFU.
K251753 - Page 17 of 19
{17}
K251753 - Page 18 of 19
# 3. Lay User Readability Study
The purpose of this study was to determine whether lay users could interpret test results correctly, especially with mock low positive samples from the GenaCheck COVID-19 Rapid Self-Test.
A total of 69 lay users with diversity in ages and educational background participated in the Readability Study. Twenty nine percent of the lay users exhibited some form of visual impairment. Each lay user was asked to interpret one of two panels of 4 mock test devices with three different concentrations (Negative, $1.5\mathrm{X}$ LoD, $1.5\mathrm{X}$ LoD, $5\mathrm{X}$ LoD) or (Negative, Negative, $1.5\mathrm{X}$ LoD, $5\mathrm{X}$ LoD).
The results are summarized in Table 13. Out of 276 mock device readings, one negative mock device was interpreted as invalid, and four low-positive mock devices (1.5 X LoD) were incorrectly read as negative. All moderate positive mock devices (5 X LoD) were interpreted correctly. Of the 5 incorrectly interpreted devices, $100\%$ were read by study participants with vision impairment. Therefore, it is recommended that users with conditions affecting their vision, such as far-sightedness, glaucoma, or color blindness, are encouraged to seek assistance to interpret results accurately (e.g., reading glasses, additional light source, or another person). A related warning statement will be included in the labelling documents.
Table 13: Lay User Readability Study Results
| Age Group | Negative (n/N) | Low positive (1.5X LoD) (n/N) | Positive (5X LoD) (n/N) | Total (n/N) |
| --- | --- | --- | --- | --- |
| 14-19 | 3/3 | 3/3 | 2/2 | 8/8 |
| 20-30 | 26/26 | 25/25 | 17/17 | 68/68 |
| 31-50 | 53/53 | 49/49 | 34/34 | 136/136 |
| 51-54 | 6/6 | 6/6 | 4/4 | 16/16 |
| ≥55 | 16/17 | 15/19 | 12/12 | 43/48 |
| Total | 104/105 | 98/102 | 69/69 | 271/276 |
| % Agreement with Expected Results | 99.0% | 96.1% | 100.0% | 98.2% |
# D Clinical Cut-Off:
Not applicable, qualitative assay
# E Expected Values/Reference Range:
Not applicable. A patient is expected to be negative for SARS CoV-2
{18}
K251753 - Page 19 of 19
## F Other Supportive Study/Device Information
### 1. Flex Studies
To assess the robustness of the GenaCheck COVID-19 Rapid Self-Test, flex studies were conducted that assessed all major aspects of the test procedure (e.g., sample volume, reading time, extraction buffer volume, swab elution time, and procedure) and variability of environmental test conditions that the test may be subjected to when in use (e.g., device orientation, lighting, various temperature, and humidity stress conditions). Testing was performed with contrived positive nasal swabs generated by diluting heat inactivated SARS-CoV-2 virus into NCM at 2 X LoD. The studies support that the test is robust in the intended use condition with an insignificant risk of erroneous result.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.