← Product Code [DQA](/productcode/DQA) · K251751

# Spo2 Sensor CSS032D (K251751)

_Shenzhen Ykd Technology Co., Ltd. · DQA · Dec 19, 2025 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251751

## Device Facts

- **Applicant:** Shenzhen Ykd Technology Co., Ltd.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Dec 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

YKD SpO2 sensor (Models: CSS032D) is indicated for continuous non-invasive monitoring of arterial hemoglobin oxygen saturation (SpO2) and pulse rate (PR) in adult patients (>40kg) within medical institutions. It is for fingertip use only and prohibited in emergency medical transport, home care, long-term monitoring, mobile patients, motion, and low perfusion conditions.

## Device Story

SpO2 sensor (CSS032D) consists of a fingertip probe connecting to a pulse oximeter monitor; measures arterial hemoglobin oxygen saturation and pulse rate. Principle of operation: 2-wavelength relative optical absorption (red 660-666nm, infrared 880-950nm) to detect absorption characteristics of deoxygenated vs oxygenated hemoglobin. Used in medical institutions by clinicians; provides real-time SpO2 and PR data to monitor patient oxygenation status. Benefits include non-invasive, continuous assessment of patient physiological state. Device is a reusable accessory for compatible patient monitors.

## Clinical Evidence

Clinical study conducted on 13 healthy adult volunteers (18-33 years, 3 male/10 female, Fitzpatrick skin types II-V) per ISO 80601-2-61. Primary endpoint: SpO2 accuracy (Arms) vs. CO-oximeter. Results: Overall Arms 1.96% (70-100% range). Subgroup analysis performed by skin color (Light, Medium, Dark) and gender. All results support substantial equivalence.

## Technological Characteristics

Reusable fingertip sensor; materials: FEP and Silicone. Sensing principle: 2-wavelength relative optical absorption (Red 660-666nm, IR 880-950nm). Powered by host monitor. SpO2 range 70-100% (Accuracy ±3% Arms); PR range 30-250bpm (Accuracy ±3 digits). Complies with IEC 60601-1, IEC 60601-1-2, and ISO 80601-2-61.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Spo2 Sensor (model: N543-01, U103-01, U103S-01, U403-01, U403S-01, U503-01, U543-01, U410-03, U403-07, U410-02, U403-08, U403-125, U403-06, U410-09, U403S-91) ([K202851](/device/K202851.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 19, 2025

Shenzhen YKD Technology Co., Ltd.
% Youshan Gong
RA Specialist
Feiying Drug &amp; Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center, No. 3101-90
Qianhai Road
Shenzhen, Guangdong 518052
China

Re: K251751
Trade/Device Name: Spo2 Sensor CSS032D
Regulation Number: 21 CFR 870.2700
Regulation Name: Oximeter
Regulatory Class: Class II
Product Code: DQA
Dated: November 23, 2025
Received: November 25, 2025

Dear Youshan Gong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251751 - Youshan Gong
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}

K251751 - Youshan Gong
Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251751  |   |
|  Device Name Spo2 Sensor (Model: CSS032D)  |   |
|  Indications for Use (Describe) YKD SpO2 sensor (Models: CSS032D) is indicated for continuous non-invasive monitoring of arterial hemoglobin oxygen saturation (SpO2) and pulse rate (PR) in adult patients (>40kg) within medical institutions. It is for fingertip use only and prohibited in emergency medical transport, home care, long-term monitoring, mobile patients, motion, and low perfusion conditions.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

Shenzhen YKD Technology Co., Ltd.
510(k)s – 510(k) Summary

# 510(k) Summary

“510(k) Summary” as required by 21 CFR Part 807.92.

Date Prepared: 2025-12-19

## I. Submitter

Shenzhen YKD Technology Co., Ltd.
6F, Building 6, Huaxinruiming Industrial Area, Langrong RD, Xinwei Village, Dalang
Community, Dalang Street, Longhua District, Shenzhen, China
Tel.: 0755-28179760

Long Zhao
Management representative
Tel: +86-15219483193
E-mail: 819622792@qq.com

## II. Device

Name of Device: SpO2 sensor
Model(s): CSS032D
Regulation Name: Oximeter
Regulatory Class: II
Product Code: DQA
Regulation Number: 21 CRF 870.2700

## III. Predicate Device and Reference Device

### Predicate Device

|  Manufacturer | Predicate Device | 510(k) Number  |
| --- | --- | --- |
|  JKH USA, LLC | Spo2 Sensor (model: N543-01, U103-01, U103S-01, U403-01, U403S-01, U503-01, U543-01, U410-03, U403-07, U410-02, U403-08, U403-125, U403-06, U410-09, U403S-91) | K202851  |

## IV. Device Description

The SpO2 sensor consists of a probe attached to the patient's finger. The sensor shall be connected to its corresponding pulse oximeter monitor with a data acquisition system which is used to calculate and display oxygen saturation levels and heart rate conditions.

Oxygenation of blood is measured by detecting the infrared and red light absorption characteristics of deoxygenated hemoglobin and oxygenated hemoglobin. The saturation

{5}

Shenzhen YKD Technology Co., Ltd. 510(k)s -510(k) Summary

value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).

As an accessory of the legally marketed oximeters or patient monitors in the United States, the proposed device SpO2 Sensor is intended for hospital use and continuous monitoring of functional arterial oxygen saturation and pulse rate.

# V. Indications for Use

YKD SpO2 sensor (Models: CSS032D) is indicated for continuous non-invasive monitoring of arterial hemoglobin oxygen saturation (SpO2) and pulse rate (PR) in adult patients  $(&gt;40\mathrm{kg})$  within medical institutions. It is for fingertip use only and prohibited in emergency medical transport, home care, long-term monitoring, mobile patients, motion, and low perfusion conditions.

# VI. Materials

|  Model | Contacted Component Name | Construction Materials  |
| --- | --- | --- |
|  CSS032D | Soft-tip | FEP, Silicone  |

# VII. Comparison of Technological Characteristics With the Predicate Device

The SpO2 sensor has the same intended use as the predicate device. The technological characteristics such as light wavelength, SpO2 measurement rang, SpO2 accuracy, PR measurement range and PR accuracy, are identical to the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use.

Therefore, the SpO2 sensor may be found substantially equivalent to its predicate device.

|  Comparison Elements | Subject Device | Predicate Device | Remark  |
| --- | --- | --- | --- |
|  510(k) Number | K251751 | K202851 | /  |
|  Trade name | Spo2 Sensor | Spo2 Sensor | Same  |
|  Model | CSS032D | N543-01, U103-01, | /  |

{6}

Shenzhen YKD Technology Co., Ltd.
510(k)s –510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Remark  |
| --- | --- | --- | --- |
|   |  | U103S-01, U403-01, U403S-01, U503-01, U543-01, U410-03, U403-07, U410-02, U403-08, U403-125, U403-06,U410-09, U403S-91 |   |
|  Manufacturer | Shenzhen YKD Technology Co., Ltd. | JKH USA, LLC | /  |
|  Regulation number | 21 CFR 870.2700 | 21 CFR 870.2700 | Same  |
|  Product code | DQA | DQA | Same  |
|  Device classification | Class II | Class II | Same  |
|  Indication for use/ Intended use | YKD SpO2 sensor (Models: CSS032D) is indicated for continuous non-invasive monitoring of arterial hemoglobin oxygen saturation (SpO2) and pulse rate (PR) in adult patients (>40kg) within medical institutions. It is for fingertip use only and prohibited in emergency medical transport, home care, long-term monitoring, mobile patients, motion, and low perfusion conditions. | SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 40 kg, pediatric patients weighing 10-50 kg, and neonatal patients weighing no less than 3 kg. | Same  |
|  Prescription or OTC | Prescription | Prescription | Same  |
|  Clinical study | Clinical studies were conducted to verify the accuracy of subject device. The clinical studies were conducted per following ISO 80601-2-61 and FDA | The subject and predicate devices are exactly the same. Since the sensors are identical, no further clinical testing is necessary. | Difference
Note 1  |

Page 3 of 9

{7}

Shenzhen YKD Technology Co., Ltd.
510(k)s –510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Remark  |
| --- | --- | --- | --- |
|   | Guidance.
Clinical test results were obtained in human adult volunteers to validate the accuracy of SpO2 Sensors versus arterial oxygen saturation (SaO2) as determined by CO-oximeter. The volunteer population in the studies composed of healthy men and women from age 18 to 33, with variations of skin pigmentations and per FDA’s guidance for Pulse Oximeters, three darkly pigmented subjects are included in the clinical study. |  |   |
|  Energy source | Powered by compatible devices | Powered by compatible devices | Same  |
|  Measurement Method | 2-wavelength Relative Optical Absorption | 2-wavelength Relative Optical Absorption | Same  |
|  Light Emitting | Red light LED with wavelength 660-666nm
Infrad-light LED with wavelength 880-950nm | Red: 660-666nm, infrared: 880-950nm | Same  |
|  Sensor Material | Perfluoro(ethylene-propylene) plastic (FEP), Silicone | ABS, PVC, TPU, Silicone, 3M medical tape and sponge | Difference
Note 2  |
|  Distal connector Design | Soft tip | finger clip , soft tip, textile adhesive and sponge non-adhesive | Difference
Note 3  |
|  SpO2 Measurement Rang | 70-100% | 70-100% | Same  |

Page 4 of 9

{8}

Shenzhen YKD Technology Co., Ltd. 510(k)s –510(k) Summary

|  Comparison Elements | Subject Device | Predicate Device | Remark  |
| --- | --- | --- | --- |
|  SpO2 Accuracy | SpO2 Range (70%-100%) --- Accuracy ±3% Arms | ±3% | Same  |
|  PR Measurement Range | 30-250bpm | 30-250bpm | Same  |
|  PR Accuracy | Accuracy ±3 digits | ±3 | Same  |
|  Usage | Reusable | Reusable and disposable | Same  |
|  Applied population | Adult patients weighing greater than 40 kg. | Adult patients weighing greater than 40 kg, pediatric patients weighing 10-50 kg, and neonatal patients weighing no less than 3 kg. | Same  |
|  Measurement part | Fingers | Fingers | Same  |
|  Electrical safety | IEC 60601-1
IEC 60601-1-2 | IEC 60601-1
IEC 60601-1-2 | Same  |
|  Product performance | ISO 80601-2-61 | ISO 80601-2-61 | Same  |

Note 1:

The subject device conducts the clinical study to verify the accuracy, and the accuracy of SpO2 and PR is same as predicate device.

Clinical studies were conducted to verify the accuracy of subject device. The clinical studies were conducted per following ISO 80601-2-61 and FDA Guidance.

Clinical test results were obtained in human adult volunteers to validate the accuracy of SpO2 Sensors versus arterial oxygen saturation (SaO2) as determined by CO-oximeter. The volunteer population in the studies composed of healthy men and women from age 18 to 33, with variations of skin pigmentations and per FDA’s guidance for Pulse Oximeters, three darkly pigmented subjects are included in the clinical study.

Note 2:

{9}

Shenzhen YKD Technology Co., Ltd. 510(k)s –510(k) Summary

Though the sensor material of subject device is little different from the predicate device, the material-silicone also used in predicated device. The SpO2 sensor directly contacts with the finger of patient. The material of contacted components of all models are FEP, silicone, which outlined in Section B of Attachment G in Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" released by the FDA on September 8, 2023.

We provide the required information be included in the premarket submission for device types within the scope of the policy outlined in Section D of Attachment G to state we can meet the requirement of this guidance.

**Note 3:**

Though the Distal connector Design of subject device is different from the predicate device, the textile adhesive and sponge non-adhesive is used for disposable SpO2 sensor of predicated device. The subject device only contain the reusable SpO2 sensor and the distal connector design of reusable SpO2 sensor is same as predicate device, so these difference will not affect the safety and effectiveness of subject device.

**Conclusions**

SpO2 sensor is substantially equivalent to the predicate device.

**VIII. Performance Data**

The following performance data were provided in support of the substantial equivalence determination.

**1) Biocompatibility Testing**

&gt; The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process", as recognized by FDA.

**2) Electrical Safety and EMC**

Electrical safety and EMC testing was performed to, and passed, the following standards:

- IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-12 Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical

{10}

Shenzhen YKD Technology Co., Ltd. 510(k)s -510(k) Summary

electrical equipment and medical electrical systems intended for use in the emergency medical services environment

## 3) Product performance

&gt; IEC 80601-2-61 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment

## IX. Clinical data

Clinical studies were conducted to verify the accuracy of subject device. The clinical studies were conducted per following ISO 80601-2-61 and FDA Guidance.

Clinical test results were obtained in human adult volunteers to validate the accuracy of SpO2 Sensors versus arterial oxygen saturation (SaO2) as determined by CO-oximeter. The volunteer population in the studies composed of healthy men and women from age 18 to 33, with variations of skin pigmentations and per FDA’s guidance for Pulse Oximeters, three darkly pigmented subjects are included in the clinical study.

The range of skin pigmentation: II~V from Fitzpatrick scale

The number of tested subjects: The clinical trial was finished with 13 subjects, including 3 males and 10 females, 3 with dark skin and 10 with light skin, ranging in age from 18-33 years old.

The characteristic of subjects distinguished by gender

|  Index | Female (10) |   |   |   | Male (3)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean | SD | Min | Max | Mean | SD | Min | Max  |
|  Age (years old) | 23.70 | 1.83 | 21 | 26 | 26.33 | 7.64 | 18 | 33  |
|  Height (cm) | 157.30 | 5.23 | 150 | 165 | 184.00 | 19.08 | 172 | 206  |
|  Weight (kg) | 58.15 | 8.21 | 47 | 71.5 | 91.33 | 34.43 | 64 | 130  |

Basic characteristics of subjects - by skin color

|  Index | Light (5) |   |   |   | Medium (5) |   |   |   | Dark (3)  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean | SD | Min | Max | Mean | SD | Min | Max | Mean | SD | Min | Max  |
|  Age (years old) | 24.00 | 1.41 | 22 | 26 | 23.20 | 5.72 | 18 | 33 | 26.33 | 1.53 | 25 | 28  |

{11}

Shenzhen YKD Technology Co., Ltd. 510(k)s -510(k) Summary

Table 39 DUT (CSS032D) adult population test data analysis results - by skin color

|  Application | skin color | Index/blood oxygen range | [70-100%] | [90 - 100%] | [80 - 90%] | [70 - 80%]  |
| --- | --- | --- | --- | --- | --- | --- |
|  Adult | Light | Number | 5 | 5 | 5 | 5  |
|   |   |  Datasets | 124 | 54 | 36 | 34  |
|   |   |  Bias | -0.9 | -0.55 | -0.04 | -0.98  |
|   |   |  Sres | 1.55 | 1.14 | 1.44 | 1.99  |
|   |   |  Arms | 1.84 | 1.79 | 1.54 | 2.18  |
|   |  Medium | Number | 5 | 5 | 5 | 5  |
|   |   |  Datasets | 119 | 58 | 36 | 25  |
|   |   |  Bias | -0.7 | -0.42 | -0.49 | -0.71  |
|   |   |  Sres | 1.6 | 1 | 1.54 | 2.28  |
|   |   |  Arms | 1.79 | 1.37 | 1.75 | 2.54  |
|   |  Dark | Number | 3 | 3 | 3 | 3  |
|   |   |  Datasets | 74 | 33 | 16 | 25  |
|   |   |  Bias | 0.49 | -0.62 | -0.06 | -0.63  |
|   |   |  Sres | 2.28 | 1.36 | 2.31 | 2.92  |
|   |   |  Arms | 2.38 | 1.36 | 2.5 | 3.2  |

Table 40 DUT (CSS032D) adult population test data analysis results - by gender

|  Application | gender | Index/blood oxygen range | [100-70%] | [90 - 100%] | [80 - 90%] | [70 - 80%]  |
| --- | --- | --- | --- | --- | --- | --- |

{12}

Shenzhen YKD Technology Co., Ltd. 510(k)s -510(k) Summary

|  Adult | Male | Number | 3 | 3 | 3 | 3  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  Datasets | 69 | 36 | 18 | 15  |
|   |   |  Bias | -0.79 | -0.52 | -0.49 | -0.45  |
|   |   |  Sres | 1.39 | 0.81 | 1.34 | 2  |
|   |   |  Arms | 1.73 | 1.57 | 1.55 | 2.23  |
|   |  Female | Number | 10 | 10 | 10 | 10  |
|   |   |  Datasets | 248 | 109 | 70 | 69  |
|   |   |  Bias | -0.42 | -0.65 | -0.08 | -0.72  |
|   |   |  Sres | 1.94 | 1.35 | 1.86 | 2.46  |
|   |   |  Arms | 2.02 | 1.52 | 1.9 | 2.7  |

Table 41 DUT (CSS032D) adult population test data analysis results - by applicable population

|  Application | Index/blood oxygen range | [100-70%] | [90 - 100%] | [80 - 90%] | [70 - 80%]  |
| --- | --- | --- | --- | --- | --- |
|  Adult | Number | 13 | 13 | 13 | 13  |
|   |  Datasets | 317 | 145 | 88 | 84  |
|   |  Bias | -0.5 | -0.86 | -0.38 | 0.00  |
|   |  Sres | 1.84 | 1.27 | 1.78 | 2.41  |
|   |  Arms | 1.96 | 1.54 | 1.83 | 2.62  |

# X. Conclusions

Based on the above analysis, non-clinical tests and clinical tests performed, it can be concluded that the subject device SpO2 sensor is as safe, as effective, and performs as well as the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K251751](https://fda-staging.innolitics.com/device/K251751)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
