← Product Code [GCJ](/productcode/GCJ) · K251743

# SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System (K251743)

_Surgical Principals, Inc. · GCJ · Sep 22, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251743

## Device Facts

- **Applicant:** Surgical Principals, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Sep 22, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery.

## Device Story

The LaproSac™ is a sterile, single-use laparoscopic specimen retrieval system. It consists of a flexible rip-stop nylon bag, optionally paired with a deployment mechanism comprising an introducer, actuation rod, and biasing arms. The device is inserted through a standard laparoscopic port (5-15mm). Upon deployment, the biasing arms automatically open the bag mouth to facilitate tissue or organ collection. A closure suture is used to secure the bag after specimen collection. The device is operated by surgeons in professional healthcare facilities (operating rooms). It enables safe containment and removal of tissue, reducing the risk of contamination or spillage during extraction. The device is non-latex.

## Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993 series), accelerated aging (ASTM F1980-21), sterilization validation (ISO 11135:2014), and performance testing (dimensional measurements, deployment/closure forces, seam strength per ASTM F88/F88M, and puncture resistance per MIL-STD-3010C).

## Technological Characteristics

Materials: Rip-stop nylon. Dimensions: 50-6000 mL volume; 5-15mm introducer/trocar diameter. Sensing/Actuation: Mechanical biasing arms for bag opening. Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide (ISO 11135) to SAL 10^-6. Standards: ISO 10993 (biocompatibility), ASTM F1980-21 (aging), ASTM F88/F88M (seam strength), MIL-STD-3010C (puncture resistance).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Espiner Tissue Retrieval System ([K111845](/device/K111845.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

September 22, 2025

Surgical Principals, Inc.
Timothy Wynne
President and CEO
1625 South Tacoma Way
Pierce, Washington 98409

Re: K251743
Trade/Device Name: SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: GCJ
Dated: August 27, 2025
Received: August 27, 2025

Dear Timothy Wynne:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251743 - Timothy Wynne
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K251743 - Timothy Wynne
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2025.09.22
16:33:44 -04'00'

Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K251743

Device Name
SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval System

Indications for Use (Describe)
The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

&gt; "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

Surgical Principals, Inc.
1625 South Tacoma Way
Tacoma, WA 98409
Office: 888-801-9251
customerservice@surgicalprincipals.com

510(k) K251743
510(k) SUMMARY

1. SUBMITTER'S CONTACT INFORMATION
Company: Surgical Principals Inc.
Address: 1625 South Tacoma Way, Tacoma, WA 98409
Contact Person: Timothy Wynne
Phone: 888-801-9251
Date Prepared: September 16, 2025

2. DEVICE NAME
Trade Name – SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems
Common Name – Retrieval Bag, Specimen Retrieval Bag
Regulation Number – 21 CFR 876.1500
Classification Name – Laparoscope, General &amp; Plastic Surgery
Product Code – GCJ
Device Classification – Class II
Classification Panel – General and Plastic Surgery

3. PREDICATE DEVICE
The SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems claim Substantial Equivalence to the Espiner Tissue Retrieval System (Sac or E-Sac) under 510(k) K111845.

4. DEVICE DESCRIPTION
The SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems are sterile single patient use devices, which comprise of a flexible nylon bag with and without a deployment mechanism. The bag consists of a large, easily accessible opening and a closure suture that facilitates closure of the specimen bag after the specimen(s) have been collected. The deployment mechanism consists for an introducer and actuation rod with biasing arms, which allows for insertion through an appropriately sized port and automatic opening of the bag upon deployment.

These devices are not made with natural rubber latex or synthetic derivatives of natural rubber latex.

5. INDICATIONS FOR USE
The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery.

Surgical Principals, Inc.
www.SurgicalPrincipals.com
Fax: 866-333-9524

{5}

Surgical Principals, Inc.
1625 South Tacoma Way
Tacoma, WA 98409
Office: 888-801-9251
customerservice@surgicalprincipals.com

# 6. SUBSTANTIAL EQUIVALENCE TABLE

|  Category | Subject Device | Predicate Device  |
| --- | --- | --- |
|  Device | SPI LaproSac™ Laparoscopic Single-Use Rip-Stop Nylon Specimen Retrieval Systems (510(k) K251743) | The Espiner Tissue Retrieval System (K111845)  |
|  Classification | Endoscope and Accessories (GCJ) | Same  |
|  Intended Use | The LaproSac™ Specimen Retrieval Bags is indicated for use in surgical procedures to capture organs or tissue to be removed from the body cavity during Laparoscopic Surgery. | The Espiner Tissue Retrieval System consists of a family of impervious sacs which are sterile single use device that can be used alone of with a dedicated introducer system for encapture and removal of an organ, tissue or fluid from the bag cavity during laparoscopic surgery.  |
|  Product Picture |  |   |
|  Design | The LaproSac™ Specimen Retrieval Bags are sterile single patient use devices, which comprise of a flexible nylon bag with and without a deployment mechanism. The bag consists of a large, easily accessible opening and a closure suture that facilitates closure of the specimen bag after the specimen(s) have been collected. The deployment mechanism consists for an introducer and actuation rod with biasing arms, which allows for insertion through an appropriately sized port and automatic opening of the bag upon deployment.
These devices are not made with natural rubber latex or synthetic derivatives of natural rubber latex. | The sac includes introduction tab(s) or tether and a drawstring (closure suture). The introduction tab(s) are used as a contact points for the instrument (tether for attachment to the dedicated introducer) for introduction through an appropriately sized cannula. The drawstring facilitates a secure closure of the sac.
The dedicated introducer comprises an introducer tube with handle, and push rod. The introducer allows for the bag to be inserted through an appropriately sized port. The push rod assembly comprises a loop on one end and biasing arms on the other, which allow the bag mouth to open upon deployment.  |
|  Bag Volume | 50 to 6000 mL | Unknown  |
|  Introducer / Trocar Diameter | 5 to 15mm | Same  |
|  Biocompatibility | Conforms to ISO 10993 series of FDA recognized consensus standards and associated FDA Guidance documents on application. | Unknown  |
|  Sterilization | Sterilized using Ethylene Oxide for single patient use in accordance with ISO 11135 to an SAL of 10^{-6}. | Sterilized using Ethylene Oxide  |
|  Prescription Use | Yes | Yes  |
|  Intended Environment | Professional Healthcare Facility (Surgical Room or Operating theatre) | Same  |

Surgical Principals, Inc.
www.SurgicalPrincipals.com
Fax: 866-333-9524

{6}

Surgical Principals, Inc.
1625 South Tacoma Way
Tacoma, WA 98409
Office: 888-801-9251
customerservice@surgicalprincipals.com

# 7. NONCLINICAL TESTS

Nonclinical testing has been conducted to verify that these devices met all design specifications and are substantially equivalent to the predicate device. Testing included the following:

- Biocompatibility Testing performed in accordance with the following:
- ISO 10993-1: 2018 – Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (Recognition No. 2-258)
- ISO 10993-5:2009 – Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (Recognition No. 2-245)
- ISO 10993-7:2008 – Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (Recognition No. 14-408)
- ISO 10993-10:2021 – Biological evaluation of medical devices - Part 10: Tests for skin sensitization (Recognition No. 2-296)
- ISO 10993-11:2017 – Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (Recognition No. 2-255)
- ISO 10993-23:2021 – Biological evaluation of medical devices - Part 10: Tests for irritation (Recognition No. 2-291)

- Aging Study per ASTM F1980-21 – Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices (Recognition No. 14-575)
- Ethylene Oxide Sterilization Validation per ISO 11135:2014 – Sterilization of health-care products - Ethylene Oxide (Recognition No. 14-529)

In addition, these devices have been compared to the predicate device through various performance studies designed to test the general device operation, function and integrity. Testing included the following:

- Dimensions: Dimensional measurements of the bag length, width/opening and volume.
- Deployment Operation: Measurement of the bag retraction, deployment, and introduction forces.
- Bag Operation: Measurement of the bag closure force, resulting reduction in the bag opening and force required to reopen the bag.
- Device Integrity (Destruction): Measurement of the tensile strength of the tether, buttonhole, and introduction tabs. Measurement of the bag seam strength per ASTM F88/F88M and puncture resistance per MIL-STD-3010C Method 2065.

# 8. CLINICAL TESTS

There were no clinical trials performed on these devices.

# 9. CONCLUSIONS

The subject device has equivalent indications for use as the predicate device. There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.

Surgical Principals, Inc.
www.SurgicalPrincipals.com
Fax: 866-333-9524

---

**Source:** [https://fda-staging.innolitics.com/device/K251743](https://fda-staging.innolitics.com/device/K251743)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).