K251727 · Touchbeauty Beauty & Health (Shenzhen) Co., Ltd. · GEX · Aug 29, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251727
Device Name
GLAM LED Facial Mask (TB-2386F)
Applicant
Touchbeauty Beauty & Health (Shenzhen) Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 29, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Red Light is intended to treat full face wrinkles. The Blue Light is intended to treat mild to moderate inflammatory acne. The Yellow Light is intended to treat wrinkles.
Device Story
GLAM LED Facial Mask is a wearable, home-use phototherapy device. It features an outer PET shell, inner PC shell, and silicone eye goggles. Device emits narrow-band visible light via LEDs: Blue (415nm), Yellow (590nm), and Red (625nm). User operates device by unfolding ear hooks and placing mask on face; activation is automatic. A touch switch on the right ear hook allows mode selection. Device operates in 15-minute cycles before entering standby mode. Light therapy is intended to treat facial wrinkles and mild to moderate inflammatory acne. Benefits include non-invasive, self-administered dermatological treatment. Healthcare providers do not operate the device; it is for patient self-use.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified via electrical safety (IEC 60601-1, IEC 60601-1-11), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), and performance standards (IEC 60601-2-57, IEC 60601-2-83). Biocompatibility confirmed per ISO 10993-5, -10, and -23.
Technological Characteristics
Wearable LED mask; PET outer shell, PC inner shell, ABS ear hooks, silicone goggles. Wavelengths: Blue 415nm, Yellow 590nm, Red 625nm (±10nm). Powered by external adapter. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-2-83, IEC 62471, ISO 10993-5/-10/-23. Software lifecycle per IEC 62304.
Indications for Use
Indicated for individuals seeking treatment for full-face wrinkles (Red and Yellow light) and mild to moderate inflammatory acne of the face (Blue light).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 29, 2025
Touchbeauty Beauty & Health (shenzhen) Co., Ltd
% Owen He
Consultant
Microkn Medical Technology Service (Shanghai) Co., Ltd.
Room 901, Huafa Center, 889 Pinglu Road, Jing'an District,
Shanghai, Shanghai 200040
China
Re: K251727
Trade/Device Name: GLAM LED Facial Mask (TB-2386F)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: June 5, 2025
Received: June 5, 2025
Dear Owen He:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251727 - Owen He
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K251727 - Owen He
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2025.08.29
12:23:12 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251727 | |
| Device Name GLAM LED Facial Mask(TB-2386F) | |
| Indications for Use (Describe) | |
| The Red Light is intended to treat full face wrinkles. The Blue Light is intended to treat mild to moderate inflammatory acne. The Yellow Light is intended to treat wrinkles. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) #: K251727
510(k) Summary
Prepared on: 2025-08-28
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | TOUCHBEAUTY BEAUTY & HEALTH (SHENZHEN) CO., LTD. |
| --- | --- |
| Applicant Address | 29/F, Block B, Tiley Central Plaza, No.195 Haide 3rd Road, Nanshan District, Shenzhen Guangdong China |
| Applicant Contact Telephone | 13537806505 |
| Applicant Contact | Mr. Shidong Qin |
| Applicant Contact Email | qinshidong@touchbeauty.com |
| Correspondent Name | Microkn Medical Technology Service (Shanghai) Co.,Ltd. |
| Correspondent Address | Room 901, Huafa Center, 889 Pinglu Road, Jing 'an District, Shanghai Shanghai Shanghai 200040 China |
| Correspondent Contact Telephone | 13641938412 |
| Correspondent Contact | Mr. Owen He |
| Correspondent Contact Email | fda@microkn.com |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | GLAM LED Facial Mask (TB-2386F) |
| --- | --- |
| Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification Name | Powered Laser Surgical Instrument |
| Regulation Number | 878.4810 |
| Product Code(s) | OHS&OLP |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K230293 | TheraFace Mask | OHS&OLP |
| K242700 | Radiant Renewal Skincare Lid | OHS&OLP |
# Device Description Summary
21 CFR 807.92(a)(4)
The GLAM LED light therapy mask is a home use wearable light emitting diode phototherapy device whose purpose is to produce an even, cool, narrow band of light for the treatment of full-face wrinkles and mild to moderate acne of the face. The outer shell of the mask is manufactured from Polyethylene terephthalate (PET). The inner shell is a clear Polycarbonate (PC). The Light emitting diodes are mounted behind the clear Polycarbonate. The LEDs generate the light. The ear hooks are made of Acrylonitrile butadiene styrene (ABS) and the silicone goggle protect the eyes from LED lights. Unfold the ear hooks and place the mask on your face, the mask will automatically activate the light therapy mode.
{5}
The LEDs produce blue, red and yellow light in the visible spectrum (Blue:415nm +/- 10nm, Yellow: 590nm +/-10nm, Red: 625nm +/-10nm.). The device works by emitting the specified wavelengths to treat full face of wrinkles or to treat mild to moderate inflammatory acne.
Press the touch switch on the right ear hook twice to select the light therapy mode you want to use. Each mode operates in a 15-minute cycle. After 15 minutes, the device automatically deactivates the light mode and enters the standby mode.
| **Intended Use/Indications for Use** | **21 CFR 807.92(a)(5)** |
| --- | --- |
| The Red Light is intended to treat full face wrinkles.
The Blue Light is intended to treat mild to moderate inflammatory acne.
The Yellow Light is intended to treat wrinkles. | |
| **Indications for Use Comparison** | **21 CFR 807.92(a)(5)** |
| The subject and predicate devices share the same indications for use. | |
| **Technological Comparison** | **21 CFR 807.92(a)(6)** |
| The Subject Device (GLAM LED Facial Mask, Model: TB-2386F) and the legally marketed Predicate Device (TheraFace Mask & Radiant Renewal Skincare Lid) are all Class II powered laser surgical instrument, sharing Product Codes OHS and OLP and falling under regulatory classification 21 CFR 878.4810. These devices all light emitting diode phototherapy device whose purpose is to produce band of light for the treatment of full-face wrinkles and mild to moderate acne of the face.
The power supply of Subject device and Predicate Devices are very similar but not identical, the IEC-60601-1 test demonstrated the safety of the power adapter. The slight difference will not raise safety and effective issue.
Key differences between Subject Device and Predicate Device lie in “Power Density” and “Treatment Time”. The “Standard Dose in Joules” can be calculated as the product of “Power Density” and “Treatment Time.” This value may be achieved using either a high power density over a short treatment time or a low power density over a longer treatment time. Although the power density and treatment time of the Subject device and the Predicate devices differ, the resulting standard dose in Joules is comparable and falls within an acceptable deviation range. The device has successfully passed testing in accordance with IEC 60601-2-57, confirming that its performance and safety profile remain unaffected. Thus, these variations do not pose any concerns regarding safety or effectiveness.
Non-clinical testing confirms biocompatibility compliance with ISO 10993-5/-10/-23(no cytotoxicity, sensitization or irritation), electrical safety (IEC 60601-1), EMC (60601-1-2), home use (IEC 60601-1-11) and performance (IEC 60601-2-57 & IEC 62471).
In conclusion, the Subject Device demonstrates substantial equivalence to Predicate in technical principles, clinical functionality, and safety thresholds, satisfying FDA 510(k) substantial equivalence criteria for market authorization. | |
| **Non-Clinical and/or Clinical Tests Summary & Conclusions** | **21 CFR 807.92(b)** |
| Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following testing items and standards:
Electromagnetic Compatibility (EMC) and Electrical Safety
IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
IEC TR 60601-4-2:2016, Medical electrical equipment- Part 4-2: Guidance and interpretation – Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
IEC 60601-1-11:2020 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
IEC 60601-2-83: 2023 Medical electrical equipment -Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
Biocompatibility test
For the components in contact with the human body, in vitro cytotoxicity test, skin irritation test, skin sensitization test and systemic toxicity test were conducted according to:
ISO 10993-5:2009, Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation | |
{6}
Photobiological Safety
The photobiological safety was tested according to IEC 62471:2006 - Photobiological safety of lamps and lamp systems
Software verification and validation testing
IEC 62304: 2006 (ed. 1.0) Medical Device Software - Software Life Cycle Processes
Clinical Testing
No animal or clinical study is included in this submission
The subject device is substantial equivalence as the predicate devices K230293 and K242700
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.