← Product Code [GEX](/productcode/GEX) · K251699

# The Botticelli (Model: Athena) (K251699)

_Bellamia Technologies, Inc. · GEX · Oct 8, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251699

## Device Facts

- **Applicant:** Bellamia Technologies, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Oct 8, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

1470 nm Laser: Dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines. 1927 nm Laser: The Botticelli (Model: Athena) is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). 2910 nm Laser: The Botticelli (Model: Athena) with its accessories is intended for use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle), such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing Treatment of wrinkles; Scar revision (including acne scars).

## Device Story

The Botticelli (Model: Athena) is a cart-based, multi-laser surgical system (1470 nm diode, 1927 nm thulium fiber, 2910 nm erbium fiber) for dermatological and plastic surgery. The device uses a 450 nm blue aiming beam to visualize treatment areas. A handpiece with an internal telescopic mechanism focuses laser energy into microspots on target tissue. The system includes a 24-inch touchscreen, footswitch, plume evacuator, and skin cooling air chiller. Operated by physicians in clinical settings, the device provides controlled energy delivery for tissue coagulation and ablation. Feedback mechanisms ensure accurate optical positioning. The system benefits patients by enabling precise skin resurfacing, wrinkle reduction, and scar revision. Clinical decision-making is supported by the system's GUI, which displays energy delivery parameters to the provider.

## Clinical Evidence

Bench testing only. Evidence includes risk analysis per ISO 14971, laser safety testing per IEC 60825-1, electrical safety per IEC 60601-1, and electromagnetic compatibility per IEC 60601-1-2. Software verification and validation were performed per FDA guidance. System verification testing confirmed operation within specifications.

## Technological Characteristics

Cart-based Class IV laser system. Three modules: 1470 nm diode, 1927 nm thulium fiber, 2910 nm Er:YAG. Energy delivery via fiber-optic handpiece with telescopic focusing. Cooling: Water-to-air and air-to-air. Electrical: 220 VAC, 60 Hz, Class I Type BF. Connectivity: Touchscreen GUI, remote interlock. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1. Software: Microprocessor-controlled.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Medical diode laser systems (CHARISMA, REGAL) ([K234004](/device/K234004.md))
- LASEMD Laser System ([K171009](/device/K171009.md))
- JOULE ProFractional System ([K180508](/device/K180508.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 8, 2025

Bellamia Technologies, Inc.
Ahmed Mohammed
VP Quality and Regulatory
285 W Central Pkwy, Suite 1735
Altamonte Springs, Florida 32714

Re: K251699
Trade/Device Name: The Botticelli (Model: Athena)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: June 1, 2025
Received: June 2, 2025

Dear Ahmed Mohammed:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251699 - Ahmed Mohammed
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K251699 - Ahmed Mohammed
Page 3

Sincerely,

YAN FU-S
Digitally signed by
YAN FU-S
Date: 2025.10.08
23:50:41 -04'00'

for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251699  |   |
|  Device Name The Botticelli (Model: Athena)  |   |
|  Indications for Use (Describe) 1470 nm Laser: Dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines. 1927 nm Laser: The Botticelli (Model: Athena) is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). 2910 nm Laser: The Botticelli (Model: Athena) with its accessories is intended for use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle), such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing Treatment of wrinkles; Scar revision (including acne scars).  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

Page 1 of 9

# 510(k) Summary

510(k) number K251699

Device Trade Name The Botticelli (Model: Athena)

Device Common Name Laser Powered Surgical Device

Device Classification Name Powered Laser Surgical Instrument

510(k) Type Traditional

Applicant BellaMia Technologies, Inc.
285 W Central Pkwy,
Altamonte Springs, FL 32714

Applicant Contact
Name: Ahmed Mohammed
Phone: 763-439-4602
Email: amohammed@bellamiatechnologies.com

Date Prepared 07-Oct-2025

Medical Specialty General and Plastic Surgery

Regulation § 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

Product Code GEX (Class 2) – Powered Laser Surgical Instrument

Predicate Devices Medical diode laser systems (CHARISMA, REGAL) [K234004]
Regulation Number: 878.4810
Classification Product Code: GEX

LASEMD Laser System [K171009]
Regulation Number: 878.4810
Classification Product Code: GEX

JOULE ProFractional System SYSTEM [K180508]
Regulation Number: 878.4810
Classification Product Code: GEX

{5}

Page 2 of 9

# Device Description

The Botticelli Laser System is a cart-based device containing three individual lasers. The Botticelli is designed for ease of use and offers a small footprint. The system includes the 24" touch screen, on/off switch, foot switch, an emergency-stop button, remote interlock, and calibration port.

The device (the cart) houses the power consuming components, including the laser modules, power supplies, the scanner drivers, software, TEC cooling module, water pump, fans, software controls, and all other electrical control components. The cart also includes the plume evacuator and the skin cooling air chiller.

The 1470 nm laser subsystem is a diode laser that is coupled to a fiber to deliver treatment energy via a handpiece. A 450 nm visible blue laser diode aiming beam is used to visualize the location of the beam during laser treatment.

The 1927 nm laser subsystem is a thulium fiber laser. The system incorporates a diode pumped thulium fiber to generate laser emissions. The laser energy is delivered during treatment via the handpiece. A 450 nm visible blue laser diode aiming beam is used to visualize the location of the beam during laser treatment.

The 2910 nm laser system is an erbium fiber laser. The system incorporates a diode pumped erbium doped fluoride glass fiber to generate laser emission. The laser energy is delivered during treatment via the handpiece. A 450 nm visible blue laser diode aiming beam is used to visualize the location of the beam during laser treatment.

# Definitions

**Microbeam**: A single laser pulse characterized by a specific diameter and predetermined energy; it is synonymous with terms such as (Laser) Shot, Dot, and Spot.

**Laser Pattern Density**: The number of laser spots in a certain area divided by the maximum number of spots in the same area (as a percentage).

**Spot Fluence**: the laser energy density for a single laser spot

**Average Fluence**: the total fluence of the pattern area (calculated using the total number of spots, the energy of each spot, and the total area)

# Energy Delivery

The Botticelli system efficiently delivers energy through a handpiece equipped with laser fibers. This handpiece features an internal telescopic mechanism that precisely focuses the laser beam to the designated spot size at the target area. Additionally, it contains integrated electronics that facilitate communication with the system's computer, enabling a feedback mechanism that ensures accurate positioning of the optics.

# Microspots

The 1470 nm laser has three different spot sizes to choose from:

- 100 microns
- 150 microns
- 200 microns

The 1927 nm laser has three different spot sizes to choose from:

- 100 microns
- 150 microns
- 200 microns

{6}

The 2910 nm laser has two different spot sizes to choose from:
- 310 microns
- 430 microns

|  Device Specifications  |
| --- |
|  Device Dimensions  |
|  Width: 15 in (38 cm)  |
|  Length: 32 in (82 cm)  |
|  Height: 43 in (110 cm)  |
|  Weight: ~ 300 lb. (135 Kg)  |
|  Electrical Specifications  |
|  Rated Voltage: 220 VAC  |
|  Rated Frequency: 60 Hz  |
|  Rated Current: 15 A  |
|  Electrical Safety Classification: Class I, Type BF  |
|  Power Cord Length: 15 ft (4.6 m)  |
|  Footswitch Cable Length: 15 ft (4.6 m)  |
|  Certification: IEC 60601-1, IEC 60601-1-2  |
|  Laser Specifications  |
|  1470nm Laser Output Class IV; 42 W, Operation Range: 0.5W – 15W  |
|  1927nm Laser Output Class IV; 18.4 W, Operating Range: 5W  |
|  2910nm Laser Output Class IV; 10 W, Operating Range: 10W  |
|  Environmental Requirements  |
|  Ambient Operating Temperature: 50 °F (10 °C) to 95 °F (35 °C), must be above dew point.  |
|  Maximum Humidity: 90%, non-condensing  |
|  Minimum 18 in (46 cm) distance from walls, to ensure full circulation of cooling air.  |
|  Maximum Operating Altitude: 5,000 ft (1,524 m)  |
|  Shipping and Storage (Non-Operational) Requirements  |
|  Storage Temperature: 39.2 °F (-4 °C) to 113 °F (50 °C), must be above dew point.  |
|  Maximum Humidity: 90%, non-condensing  |

Page 3 of 9

{7}

Page 4 of 9

# Indications for Use

1470 nm Laser: Dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines.

1927 nm Laser: The Botticelli (Model: Athena) is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles).

2910 nm Laser: The Botticelli (Model: Athena) with its accessories is intended for use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle), such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing Treatment of wrinkles; Scar revision (including acne scars).

{8}

Technological Comparison

1470 nm Laser

|  Specification | This Application | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Ref. 510(k) | K251699 | K234004 | N/A  |
|  Product | The Botticelli Laser Platform
1470 nm Laser System | Medical diode laser systems
(CHARISMA, REGAL) | N/A  |
|  Product Code | GEX | GEX | Same  |
|  Indications for Use | Dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines. | CHARISMA:
Dermatological procedures requiring coagulation of soft tissue and skin resurfacing procedures. further indicated for treatment of benign pigmented lesions, such as, but not limited to lentigines (age spots), solar lentigos (sun spots), melasma, dyschromia, and for treatment of facial wrinkles and fine lines.
-Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue
-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux
-Further indicated for laser assisted lipolysis.
REGAL:
-Incision, excision, ablation, vaporization, hemostasis and/or coagulation of soft tissue
-Endovascular coagulation and endovenous occlusion of the greatest saphenous vein in patients with superficial vein reflux
-Further indicated for laser assisted lipolysis. | The subject device indications for use are a subset of the indications for use of the predicate device.  |
|  CDRH Laser Class | Class 4 | Class 4 | Same  |
|  Energy Source | Diode Laser, CW, Pulsed | GaAlAs diode laser | Same  |
|  Laser Wavelength | 1470nm | 1470nm±10nm | Same  |
|  Output Power | 0.5W to 15W | CHARISMA: 0.5W-15.0W (±10%) | Same  |
|  Display Screen | Yes | Yes | Same  |
|  Power Requirements | 220 VAC/15A, 60 Hz | ~100-240V, 50-60Hz, 350VA | Similar  |
|  Aiming Beam | 450 nm <2.5 mW | Diode laser of 650nm, power <2mW, adjustable brightness. | Similar  |

Page 5 of 9

{9}

|  Specification | This Application | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Delivery System | Fiber Optic | CHARISMA
contact: fibers of 200μm, 400μm, 600μm with SMA905 connector;
Non-contact: fibers | Similar  |
|  Cooling System | Water to Air | Air | Similar  |
|  Control System | Microprocessor | MCU (Micro Controller Unit) | Same  |
|  Pulse Width | 10ms-2.5s | CHARISMA:
10ms-2.5s (General Mode) | Same  |
|  Repetition Rate | 0.2 Hz – 50 Hz | CHARISMA:
0.2Hz-50Hz (General Mode) | Same  |
|  Energy Density per Microbeam | Up to 50 mJ | CHARISMA:
Up to 50mJ (General Mode) | Same  |
|  Output Mode | CW, single pulse, repeat pulse | CW, single pulse, repeat pulse | Same  |
|  Laser Type | Laser diode | Laser diode | Same  |

Page 6 of 9

{10}

Page 7 of 9

1927 nm Laser

|  Specification | This Application | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Ref. 510(k) | K251699 | K171009 | N/A  |
|  Product | The Botticelli Laser Platform
1927 nm Laser System | LASEMD Laser System | N/A  |
|  Product Code | GEX | GEX | Same  |
|  Indications for Use | The Botticelli (Model: Athena) is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). | The LASEMD Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of benign pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephelides (freckles). | Same  |
|  CDRH Laser Class | Class 4 | Class 4 | Same  |
|  Wavelength | 1927 nm | 1927 nm | Same  |
|  Laser Type | Thulium Laser | Thulium Laser | Same  |
|  Energy Source | 1927 nm | 1927 nm | Same  |
|  Spot Size | 100, 150, 200 μm | 100, 200 μm | Within Range  |
|  Pulse Repetition Rate | 50-300 Hz | 43.5 – 307.7 Hz | Within Range  |
|  Max Pulse Duration | 20 ms | 20 ms | Same  |
|  Tip Size (*) | 4 mm x 10 mm | 4 mm x 10 mm | Same  |
|  Max Pulse Energy | 20 mJ | 20 mJ | Same  |
|  Utilities | 220 VAC/15A, 60 Hz | 100-240 VAC, 50/60 Hz | Similar  |
|  Power | 5 W (1927 nm) | 5 W (1927 nm) | Same  |
|  Delivery System | Fiber and Handpiece | Fiber and Handpiece | Same  |
|  Emission Control | Footswitch | Footswitch | Same  |
|  Display Screen | Yes | Yes | Same  |
|  Cooling System | Air to Air | Air to Air | Same  |
|  Control System | Microprocessor | Microprocessor | Same  |
|  Energy Monitor | Display Indicates Energy Delivered to Tissue | Display Indicates Energy Delivered to Tissue | Same  |
|  Safety | Safety Eyewear and Remote Interlock Connector | Safety Eyewear and Remote Interlock Connector | Same  |

(*) The Botticelli 1927 nm has a rectangular tip (pattern) shape which is fixed to the 4 mm x 10 mm size.

{11}

2910 nm Laser

|  Specification | This Application | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Ref. 510(k) | K251699 | K180508 | N/A  |
|  Product | The Botticelli Laser Platform 2910 nm Laser System | JOULE ProFractional System | N/A  |
|  Product Code | GEX | GEX | Same  |
|  Indications for Use | The Botticelli (Model: Athena) with its accessories is intended for use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle), such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing Treatment of wrinkles; Scar revision (including acne scars). | The Joule 2940 nm System with its accessories is intended for use in soft tissue (skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes; organs, and glands) such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing Treatment of wrinkles; Epidermal nevi; Telangiectasia; Spider veins; Actinic chelitis; Keloids; Verrucae; Skin tags; Anal tags; Keratoses; Scar revision (including acne scars). | The subject device indications for use are a subset of the indications for use of the predicate device.  |
|  Prescription Use | Yes | Yes | Same  |
|  Laser Energy Source | Er: YAG | Er: YAG | Same  |
|  Laser Delivery | Fiber, Handpiece | Fiber, Handpiece, Articulated arm | Similar  |
|  Wavelength | 2910 nm | 2940 nm | Similar  |
|  Adjustable Spot Size (Pattern Size) | 1.3x1.3mm - 20x20mm | 1.3x1.3mm - 20x20mm | Same  |
|  Micro spot size | 310, 430 μm | 430 μm | Similar  |
|  Repetition Rate (*) | Up to 3 Hz | Up to 3 Hz | Same  |
|  Pulse Width/ Duration | 0.5 to 1.5 msec (or 1500 μsec) | 0.5 to 1.5 msec (or 1500 μsec) | Same  |
|  Energy per Microbeam (mJ/mb) | Up to 35 | Up to 70 | Within Range  |
|  Fluence | 0.15J/cm² to 8.4 J/cm² | - | Similar  |
|  Target Chromophore | Water | Water | Same  |
|  Delivery System | Fiber optic with handpiece | Articulated arm and/or fiber optic arm with handpiece | Similar  |
|  Cooling System | Water & Air | Water & Air | Same  |
|  Electrical Requirements | 220 VAC/15A, 60 Hz | 230 VAC/25A, 50/60 Hz | Similar  |
|  Software/GUI/ Touch Screen | Yes | Yes | Same  |
|  Energy Monitor | Display Indicates Energy Delivered to Tissue | Display Indicates Energy Delivered to Tissue | Same  |

(*) The number of 'Patterns' per second, different from Pulses Per Second

Page 8 of 9

{12}

Page 9 of 9

# Summary of Non-Clinical Testing

- Risk analysis activities, in compliance with the requirements of ISO 14971 Third Edition 2019-12. Medical devices - Application of risk management to medical devices
- IEC 60825-1:2014, Safety of laser products – Part 1: Equipment classification and requirements.
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance; IEC 60601-1:2005/AMD1:2012/AMD2:2020
- IEC 60601-1-2:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- Software verification and validations, in compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 2005 and Content of Premarket Submissions for Device Software Functions, June 2023.
- System Verification Testing

The tests confirmed that the subject device operates in alignment with its specifications and complies with international consensus standards and FDA guidance.

# Conclusion

The Botticelli Laser System is substantially equivalent to its predicate devices. The nonclinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K251699](https://fda-staging.innolitics.com/device/K251699)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com