K251694 · Tianjin Huahong Technology Co., Ltd. · FMK · Jun 26, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251694
Device Name
safety lancet
Applicant
Tianjin Huahong Technology Co., Ltd.
Product Code
FMK · General, Plastic Surgery
Decision Date
Jun 26, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4850
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The safety lancet is intended for capillary blood sampling.
Device Story
Single-use safety lancet designed for capillary blood sampling; consists of needle core, button, housing, protective cap, and spring. User twists off protective cap, presses button to activate spring-loaded mechanism; needle tip penetrates skin and automatically retracts into housing to prevent sharps injury. Used by healthcare professionals, patients, or lay persons in clinical or home settings. Provides a sterile, single-use method for obtaining small blood samples; benefits patient by reducing risk of accidental needlestick injuries during blood collection.
Clinical Evidence
No clinical study included. Bench testing only. Simulated clinical use study performed on 500 samples per FDA guidance and ISO 23908 to evaluate safety mechanism; all samples met pre-established criteria. Biocompatibility testing (ISO 10993 series) confirmed no cytotoxicity, sensitization, irritation, systemic toxicity, or pyrogenicity.
Technological Characteristics
Single-use, spring-activated safety lancet. Materials: stainless steel needle, silicone oil, ABS, PS, PP, and calcium powder. Sterilization: radiation (SAL 10^-6). Gauge range: 16G to 38G. Design includes integral sharps injury prevention feature. Biocompatibility per ISO 10993.
Indications for Use
Indicated for capillary blood sampling in patients of all ages, intended for use by healthcare professionals or lay persons.
Regulatory Classification
Identification
The regulation covers four types of blood lancets: (a) Single use only blood lancet with an integral sharps injury prevention feature; (b) Single use only blood lancet without an integral sharps injury prevention feature; (c) Multiple use blood lancet for single patient use only; and (d) Multiple use blood lancet for multiple patient use. All are defined as disposable or reusable devices comprised of a blade attached to a base used to puncture the skin to obtain a drop of blood for diagnostic purposes.
Special Controls
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
*Single use only blood lancet without an integral sharps injury prevention feature* —(1)*Identification.* A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
*Multiple use blood lancet for single patient use only* —(1)*Identification.* A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
*e.g.,* blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
*e.g.,* cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
*1* ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
*2* ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
*Multiple use blood lancet for multiple patient use* —(1)*Identification.* A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
*Classification.* Class III (premarket approval).(3)
*Date PMA or notice of completion of a PDP is required:* A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 26, 2025
Tianjin Huahong Technology Co., Ltd.
Ningning Wang
Registered Engineer
A01, Plant B, No. 278 Hangkong Road
Tianjin Pilot Free Trade Zone (Air Port Industrial Park)
Tianjin, 300308
China
Re: K251694
Trade/Device Name: safety lancet
Regulation Number: 21 CFR 878.4850
Regulation Name: Blood Lancets
Regulatory Class: Class II
Product Code: FMK
Dated: June 1, 2025
Received: June 2, 2025
Dear Ningning Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251694 - Ningning Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251694 - Ningning Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Jang -S
Digitally signed by
James H. Jang -S
Date: 2025.06.26
07:31:57 -04'00'
James Jang, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
safety lancet
Page 7 of 28
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251694 | ? |
| Please provide the device trade name(s). | | ? |
| safety lancet | | |
| Please provide your Indications for Use below. | | ? |
| The safety lancet is intended for capillary blood sampling. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary - K251694
# 510(k) summary
## I Submitter
Tianjin Huahong Technology Co., Ltd.
A01, Plant B, No.278, Hangkong Road, Tianjin Pilot Free Trade Zone (Air Port Industrial Park), 300308 Tianjin, China
Establishment Registration Number: 3009498536
Contact person: Ms. Ningning Wang
Registered Engineer
Tel.: +86-13021381776
E-mail: ningning.wang@hh-technology.com
Preparation date: 1 June 2025
## II Proposed Device
Trade Name of Device: Safety Lancet (XXXVI)
Common name: Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature
Regulation Number: 21 CFR 878.4850
Regulatory Class: Class II
Product code: FMK
Review Panel: General & Plastic Surgery
## III Predicate Devices
510(k)Number: K240806
Trade name: Safety Lancet
Classification: Class II
Product Code: FMK
Manufacturer: Tianjin Huahong Technology Co., Ltd.
{5}
510(k) Summary - K251694
## IV Device description
The safety lancet is single use medical device, which is designed to collect capillary blood sample.
The intended users include Healthcare personnel, patients and lay persons.
Model XXXVI, the safety lancets consist of needle core, button, housing, protective cap and spring.
The sterile part of the safety lancet is the needle tip.
The sterile barrier is the needle sleeve and sterilized to a SAL of 10⁻⁶ by radiation sterilization.
It is intended for single use only. The shelf-life of the product is 5 years.
## V Indication for use
The safety lancet is intended for capillary blood sampling.
## VI Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate device are listed in below table 1:
{6}
510(k) Summary - K251694
Table 1 General Comparison of Safety Lancet
| Item | Predicate device (K240806) | Proposed device | Comments on Similarities/Differences | Safety/Effectiveness Statement |
| --- | --- | --- | --- | --- |
| Product name | Safety Lancet | No Change | N/A | No new concerns.
The product name remains unchanged |
| Product Code | FMK | No Change | N/A | No new concerns.
The product code remains the same, reflecting no change in device functionality. |
| Regulation No. | 21 CFR § 878.4850 | No Change | N/A | No new concerns.
The regulation number remains consistent. |
| Class | II | No Change | N/A | No new concerns.
The device remains in Class II, indicating no change in risk level. |
| Prescription/over-the-counter use | Over-The-Counter Use | No Change | N/A | No new concerns.
The over-the-counter use remains consistent. |
| Indication for use | The safety lancet is intended for capillary blood sampling. | No Change | N/A | No new concerns.
The Indication for use remains consistent. |
| Applicable user | Healthcare professional or lay person | No Change | N/A | No new concerns.
The applicable user remains consistent. |
{7}
510(k) Summary - K251694
| Reuse durability | Single use | No Change | N/A | No new concerns.
The single use remains consistent. |
| --- | --- | --- | --- | --- |
| Sterilization method and SAL | Sterilized by Radiation SAL=10^{-6} | No Change | N/A | No new concerns.
The sterilization method and SAL remains consistent. |
| Manufacturing aspects | The injection molding of the needle core is completed in the purification workshop, the steel needle is wrapped in the needle core, sterilization of the needle core is outsourced, the subsequent process is the assembly and packaging of the safety lancet | No Change | N/A | No new concerns.
The manufacturing aspects remains consistent. |
| Design and Functionality aspects | The needle tip is sterile, and the plastic part of the needle core serves as a sterile barrier for the needle tip. When using, the cap needs to be twisted off first, after pressing the button, the safety needle is activated, and the needle tip penetrates the skin under the action of the spring, Subsequently, under the action of the spring, the needle tip retracts back into the safety needle. | No Change | N/A | No new concerns.
The design and functionality aspects remains consistent. |
| Gauge range | 1.50±0.02mm (16G)
1.40±0.02mm (17G)
1.20±0.01mm (18G)
1.07±0.01mm (19G) | No Change | N/A | No new concerns.
The gauge range remains consistent. |
{8}
510(k) Summary - K251694
| | 0.91±0.01mm (20G) | | | |
| --- | --- | --- | --- | --- |
| | 0.82±0.01mm (21G) | | | |
| | 0.72±0.01mm (22G) | | | |
| | 0.64±0.01mm (23G) | | | |
| | 0.57±0.01mm (24G) | | | |
| | 0.51±0.01mm (25G) | | | |
| | 0.46±0.01mm (26G) | | | |
| | 0.41±0.01mm (27G) | | | |
| | 0.36±0.01mm (28G) | | | |
| | 0.34±0.01mm (29G) | | | |
| | 0.31±0.01mm (30G) | | | |
| | 0.26±0.01mm (31G) | | | |
| | 0.24±0.01mm (32G) | | | |
| | 0.21±0.01mm (33G) | | | |
| | 0.19±0.01mm (34G) | | | |
| | 0.18±0.01mm (35G) | | | |
| | 0.17±0.01mm (36G) | | | |
| | 0.16±0.01mm (37G) | | | |
| | 0.15±0.01mm (38G) | | | |
| Shelf-life | 5 years | No Change | N/A | No new concerns.
The shelf-life remains consistent. |
| Materials of parts in contact with human body | Needle: stainless steel, silicone oil
Housing, Button: ABS, PS, PP | Needle: stainless steel, silicone oil | added the protective cap | No new concerns.
Just added the protective cap, no change in material. |
{9}
510(k) Summary - K251694
| | and Calcium powder | Housing, Button, Protective cap: ABS, PS, PP and Calcium | | |
| --- | --- | --- | --- | --- |
| Biocompatibility | Conforms to the requirements of ISO 10993 series standards. | powder
No Change | N/A | No new concerns.
The biocompatibility remains consistent. |
| Performance requirements | Remain no change | No Change | N/A | No new concerns.
Performance requirements remain unchanged, ensuring consistent functionality. |
| Label/Labeling | Complied with 21 CFR part 801 | No Change | N/A | No new concerns.
The labeling remains consistent. |
{10}
510(k) Summary - K251694
## VII Non-Clinical Testing
To support substantial equivalence to the predicate device, Tianjin Huahong Technology Co., Ltd. completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met.
## Verification and Validation Activities - Design Changes:
Model: Additional model were introduced to meet clinical needs, Model XXXVI have only cosmetic and dimensional changes from those in K240806.
## Testing Performed:
Performance Testing: The safety lancet underwent performance testing (report HH-JS-RP-2024-10), The testing confirmed that the safety lancet meet the performance criteria outlined.
Biocompatibility Testing:
The biocompatibility evaluations were conducted in accordance with the 2023 FDA Guidance document Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process". The tests include the following tests:
The formulation, processing, sterilization, geometry in the previously approved safety lancet products (K220370) are the same, and the nature and duration of contact with the patient are also the same.
| Item | Test method | Test results |
| --- | --- | --- |
| In Vitro Cytotoxicity | ISO 10993-5: 2009 | No Cytotoxicity |
| Skin Sensitization | ISO 10993-10: 2010 | No Skin sensitization |
| Intracutaneous reactivity | ISO 10993-10: 2010 | No irritation |
| Acute Systemic Toxicity | ISO 10993-11: 2017 | No Acute SystemicToxicity |
| Pyrogenicity | ISO 10993-11: 2017 | no thermogenic reaction |
## Simulated Clinical Use
A simulated clinical use study was performed on 500 device samples each for the safety lancet according to FDA Guidance, Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 and ISO 23908 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.
## Test Results:
All verification and validation tests passed without deviations, confirming that the safety lancet meet the necessary design specifications and regulatory requirements. The tests demonstrated that the product modifications did not introduce any new risks related to safety or effectiveness when compared to the predicate device.
Page 7
{11}
510(k) Summary - K251694
## Conclusions:
Based on the results of the non-clinical verification and validation activities, it can be concluded that the modified safety lancet are substantially equivalent to the predicate device K240806. The cosmetic and dimensional changes, have been thoroughly evaluated and verified to meet all applicable performance and safety standards. Therefore, the device is as safe and effective as the predicate device for its intended use, and no new risks have been introduced.
## VIII Clinical Testing
No clinical study is included in this submission.
## IX Conclusion
Any minor differences in the technological characteristics of the proposed device, when compared to the predicate devices have been successfully evaluated through appropriate safety and performance testing which demonstrates that the proposed device, when compared to the predicate device, does not raise any new questions of safety and effectiveness. Therefore, the proposed device have been determined to be substantially equivalent to the predicate devices.
Page 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.