K251684 · Alma Lasers, Inc. · OAP · Jun 5, 2026 · Physical Medicine
Device Facts
Record ID
K251684
Device Name
Alma TED+ System
Applicant
Alma Lasers, Inc.
Product Code
OAP · Physical Medicine
Decision Date
Jun 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Alma TED+ System is indicated to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) I-4, II-1, II-2, or frontal patterns of hair loss and in males who have Norwood Hamilton Classifications of IIa to V and who both have Fitzpatrick Skin Typers I-IV.
Device Story
Low-level laser therapy (LLLT) device; uses 650nm laser light to promote hair growth. System comprises console with LCD touchscreen, footswitch/hand switch, and IMPACT 650 applicator containing 7 laser diodes. User maintains distance to scalp via detachable guiding tip. Console micro-controller manages output parameters; audible/visual indicators provide operational feedback. Includes SmartClinic software for usage monitoring via cloud (AWS). Used in clinical settings; operated by healthcare professionals. Output affects hair growth stimulation; benefits patients with androgenetic alopecia.
Clinical Evidence
Bench testing only. Performance testing assessed output power of the IMPACT 650 applicator compared to the predicate. Compliance with safety and performance standards verified, including IEC 60601-1, IEC 60825-1, IEC 60601-1-2, ISO 10993-1, and IEC 62304.
Technological Characteristics
Low-level laser therapy (LLLT) system; 650nm visible red light; 7 laser diodes (5mW each). Class 3R laser. Components: console, footswitch, IMPACT 650 applicator with detachable tip. Air-cooled. Connectivity: Cloud-based (AWS) via SmartClinic software. Standards: IEC 60601-1, IEC 60825-1, IEC 60601-1-2, ISO 10993-1, ISO 14971, IEC 62366-1, IEC 62304.
Indications for Use
Indicated for females with Ludwig (Savin) I-4, II-1, II-2, or frontal hair loss patterns and males with Norwood Hamilton Classifications IIa to V, all with Fitzpatrick Skin Types I-IV, to treat Androgenetic Alopecia and promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 5, 2026
Alma Lasers, Inc
Jessica Rivera-Montejo
Director - Regulatory Affairs and Quality Assurance
485 Half Day Rd
#100
Buffalo Grove, Illinois 60089
Re: K251684
Trade/Device Name: Alma TED+ System
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP, OUG
Dated: May 5, 2025
Received: June 2, 2025
Dear Jessica Rivera-Montejo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K251684 - Jessica Rivera-Montejo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K251684 - Jessica Rivera-Montejo
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.06.05
17:10:23 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K251684
Device Name
Alma TED+ System
Indications for Use (Describe)
Alma TED+ System is indicated to treat Androgenetic Alopecia and promote hair growth infemales who have Ludwig (Savin) I-4, II-1, II-2, or frontal patterns of hair loss and in males who have Norwood Hamilton Classifications of IIa to V and who both have Fitzpatrick SkinTypers I-IV.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# **510(k) SUMMARY – K251684**
**Alma TED+ System**
# **Submitter's Name, Address, Telephone Number, and Contact Person**
Alma Lasers Inc.
485 Half Day Rd. Ste. 100
Buffalo Grove, IL 60089
Contact Person: Jessica Rivera-Montejo, Director Regulatory and Quality
**Date Prepared:** June 5, 2026
# **Name of Device**
Trade Name: Alma TED+ System
Classification Name: Laser Instrument, Surgical, Powered
Classification Regulation: 21 CFR 890.5500
Product Code: OAP, OUG
# **Predicate Device / Reference Device(s)**
Predicate: Laser Therapy Hair Growth Comb, Model: Lasercomb-001 & Lasercomb-002 (K230134) (Product Code(s): OAP, ISA
Reference: The Alma Soprano Titanium, K230371 (Product Code(s): OUG)
# **Device Description**
The Alma TED+ System is a Class II Medical Device that utilizes 650nm laser light to promote hair growth. The Alma TED+ has an applicator called the IMPACT 650 that utilizes the activation of the applicator performed by pressing the air footswitch or hand switch. Audible and visual indicators are designed to provide safe and effective operation for patients and operator by time counting, signals, and error messages when necessary.
Alma TED+ is a low-level laser therapy device that emits laser light designed to promote hair growth in women and men. The applicator contains 7 laser diodes, and the applicator provides distributed laser light while the tip guides the user to maintain a distance to the scalp.
Alma TED+ is based on the principle of low-level laser therapy laser diodes and a detachable guiding tip, for the user to maintain a distance to the scalp. Output parameters and other system features are controlled from the touch-screen control panel on the Alma TED+ console, which provides an interface to the system's micro-controller through an LCD touchscreen.
The activation of the applicator is performed by pressing the air footswitch or hand switch. Audible and visual indicators are designed to provide safe and effective operation for patients and operators by time counting, signals, and error messages when necessary. Air-
{5}
# 510(k) SUMMARY – K251684
# Alma TED+ System
cooling fans cool the applicator and the console elements. The system is grounded through the grounding conductor in the standard power cable and an internal grounding pin.
In addition, this submission will add the Smart Clinic Software. This software was previously cleared in the Alma Titanium K230371.
The system consists of the following major components:
- Alma TED+ System
- Footswitch
- Delivery device handpiece:
- IMPACT 650 Applicator
# Intended Use / Indications for Use
Alma TED+ System is indicated to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) I-4, II-1, II-2, or frontal patterns of hair loss and in males who have Norwood Hamilton Classifications of IIa to V and who both have Fitzpatrick Skin Typers I-IV.
Comparison of Technological Characteristics
| Parameter | Proposed Device | Predicate Device K230134 | Comparison to Predicates |
| --- | --- | --- | --- |
| Wavelength | 650nm±10nm | 650nm±10nm | Same |
| Type of laser | Visible red light-emitting diodes | Visible red light-emitting diodes | Same |
| Amount of laser diodes | 7 | 7 | Same |
| Energy per laser diode | 5 mW (per each laser diode) | 4.56 mW, 4.63 mW, 4.66 mW, 4.78 mW, 4.86 mW, 4.89 mW<5mW | Similar |
| Treatment Time [min] | 15 | 15 | Same |
| Class | 3R | 3R | Same |
| Treatment Frequency | 3 times per week, spaced out every other day, as little as 16 weeks | 3 times per week, spaced out every other day, as little as 16 weeks | Same |
{6}
# **510(k) SUMMARY – K251684**
**Alma TED+ System**
The Alma TED+ is nearly identical to the predicate device Laser Therapy Hair Growth Comb K230134. Sharing the same wavelength and the type of visible red light-emitting diodes with 7 laser diodes each. The treatment frequency for the subject Alma TED+ is identical to the predicate's 3 times per week, spaced out every other day, as little as 16 weeks. The only slight difference is the energy per laser diode, it is similar and within predicate range of equal to 5 mW per each laser diode or less.
The Alma TED+ will share the same SmartClinic software that was previously cleared in Alma's Soprano Titanium. This MDDS software collects information from the system regarding usage and error messages and sends that information to a cloud that resides on Amazon Web Services. The laser system owner can monitor one or several lasers for performance. The Smart clinic MDDS software does not control or affect the system software.
The SmartClinic was previously cleared in the Alma Titanium (K230371). There have been no changes to the previously cleared SmartClinic. It is identical to the software that was cleared in K230371.
# **Performance Data**
Performance testing was performed for the new IMPACT 650 Applicator with TED detachable tip. The primary difference between the subject device and the predicate device is the TED detachable tip. The company assessed the performance of the IMPACT 650 Applicator through testing the output power. The testing demonstrated that the applicator power output is substantially equivalent between the IMPACT 650 Applicator and the Laser Therapy Hair Growth Comb.
The following additional testing was performed to support substantial equivalence:
- IEC 60601-1:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60825-1:2014 Safety of laser products - Part 1: Equipment classification and requirements
- IEC 60601-1-2:2014:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -
# **Requirements and tests**
- ISO 15223-1:2021 Medical devices — Symbols to be used with information to be supplied by the manufacturer. Part 1: General requirements
- ISO 20417:2021 Medical devices — Information to be supplied by the manufacturer
- IEC 62304:2015 Medical Device Software: Software Life Cycle Process
{7}
# **510(k) SUMMARY – K251684**
**Alma TED+ System**
- ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 14971:2019 Medical devices – Application of risk management to medical devices
- IEC 62366-1:2020 Medical Devices: Part 1: Application of Usability Engineering to Medical Devices
- IEC 60601-1-6:2020 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
- ISO 13485:2016 Medical devices - Quality management systems - Requirements for regulatory purposes
In all instances, the Alma TED+ System and the IMPACT 650 Applicator functioned as intended and the results met the pre-defined acceptance criteria.
# **Conclusion**
The Alma TED+ System has the same intended use and indications for use as the predicate device (Laser Therapy Hair Growth Comb, Model: Lasercomb-001 & Lasercomb-002, K230134).
The primary technological difference between the subject device and the predicate device is the TED detachable tip, which has similar technological characteristics and principles of operation as the predicate device's Laser Therapy Hair Growth Comb. The difference in the tip being detachable does not raise different questions of safety or effectiveness because they do not change the overall intended therapeutic use of the laser system, or affect the fundamental scientific technology used to achieve the device's intended use. Furthermore, performance testing of the technological differences showed that the new IMPACT 650 Applicator is as safe and effective as the predicate device's Laser Therapy Hair Growth Comb.
Therefore, the Alma TED+ System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.