Fenom Flo™ FeNO monitoring system is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and decreases in response to anti-inflammatory treatment. Fenom Flo™ measures fractional exhaled nitric oxide (FeNO) according to guidelines established by the American Thoracic Society. Measurement of FeNO by Fenom Flo™ is a non-invasive quantitative method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Flo™ is suitable for adults and children ages seven years and older. Fenom Flo™ should be used in a point-of-care healthcare setting under professional supervision. Fenom Flo™ should not be used in critical care, emergency care, or anesthesiology.
Device Story
Handheld device measures fractional exhaled nitric oxide (FeNO) in human breath; utilizes digital electrochemical sensor technology. Patient performs breath maneuver; exhales into disposable mouthpiece. Device electronics/pneumatics process sample; electrochemical potential difference at sensor electrodes proportional to NO concentration. POC App (Windows 10/11) displays FeNO scores in ppb; provides test initiation, results history, quality control, and patient tutorial. Used in point-of-care settings (clinics, hospitals) by trained healthcare professionals. Output assists clinicians in assessing airway inflammation and monitoring therapeutic effect of anti-inflammatory pharmacological therapy in asthma patients. Benefits include non-invasive, quantitative tracking of treatment response.
Clinical Evidence
P10 study conducted at 13 U.S. clinical sites. Evaluated changes in FeNO levels in subjects with uncontrolled, non-steroid-treated asthma before and after corticosteroid therapy. Demonstrated statistically significant and clinically meaningful changes in FeNO levels, concordant with improvements in major asthma efficacy variables in subjects aged 7 years and older. Study adhered to ATS guidelines and ISO 14155:2020.
Technological Characteristics
Handheld POC device; electrochemical sensor technology (amperometric). Measures FeNO 5-300 ppb. Rechargeable lithium-ion battery. Single-use disposable mouthpiece with anti-bacterial/anti-viral filter. Operating conditions: 15-35°C, 20-90% RH, up to 6550 ft altitude. Standards: IEC 60601-1, IEC 60601-1-2, ISO 14971, AAMI TIR57.
Indications for Use
Indicated for adults and children ages 7+ to measure fractional exhaled nitric oxide (FeNO) as an adjunct to clinical assessment of asthma, specifically to monitor FeNO concentration decreases following anti-inflammatory pharmacological therapy. Contraindicated for children under 7, patients unable to follow instructions, and use in critical care, emergency care, or anesthesiology.
Regulatory Classification
Identification
A breath nitric oxide test system is a device intended to measure fractional nitric oxide in human breath. Measurement of changes in fractional nitric oxide concentration in expired breath aids in evaluating an asthma patient's response to anti-inflammatory therapy, as an adjunct to established clinical and laboratory assessments of asthma. A breath nitric oxide test system combines chemiluminescence detection of nitric oxide with a pneumotachograph, display, and dedicated software.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Breath Nitric Oxide Test System.” See § 862.1(d) for the availability of this guidance document.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K251674
B Applicant
MGC Diagnostics Corporation
C Proprietary and Established Names
Fenom Flo™ FeNO Monitoring System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MXA | Class II | 21 CFR 862.3080 - Breath Nitric Oxide Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Breath Nitric Oxide
C Type of Test:
Quantitative, electrochemical sensor
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Fenom Flo™ FeNO monitoring system is a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. FeNO is increased in some airway inflammatory processes, such as asthma, and decreases in response to anti-inflammatory treatment. Fenom Flo™ measures fractional exhaled nitric oxide (FeNO) according to guidelines established by the American Thoracic Society.
Measurement of FeNO by Fenom Flo™ is a non-invasive quantitative method to measure the decrease in FeNO concentration in asthma patients that often occurs after treatment with anti-inflammatory pharmacological therapy as an indication of therapeutic effect in patients with elevated FeNO levels. FeNO measurements are to be used as an adjunct to established clinical assessments. Fenom Flo™ is suitable for adults and children ages seven years and older.
Fenom Flo™ should be used in a point-of-care healthcare setting under professional supervision. Fenom Flo™ should not be used in critical care, emergency care, or anesthesiology.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Fenom Flo™ may not be used by children under the age of 7 years, including infants, or by patients who are unable to understand and execute the instructions given by healthcare providers, as measurement requires patient cooperation.
Fenom Flo™ should not be used in critical care, emergency care, or in anesthesiology.
All subjects should refrain from eating or drinking for at least 60 minutes before the FENOM test. Recent intake of nitrate rich food, such as Arugula, spinach, lettuce, radish, beetroot, chinese cabbage, turnips, cabbage, green beans, leek, spring onion, cucumber, carrots, potatoes, garlic, sweet pepper, green pepper, can lead to increased FeNO levels.
Smoking reduces exhaled NO levels. Fenom results obtained from subjects who smoke should only be considered after considering the subject’s smoking history and the potential impact on NO levels.
D Special Instrument Requirements:
Fenom Flo™ FeNO Monitoring System
IV Device/System Characteristics:
A Device Description:
K251674 - Page 2 of 11
{2}
Fenom Flo™ is a point-of-care (POC) breath analyzer that quantitatively measures fractional exhaled nitric oxide (NO) in expired human breath.
The device is a hand-held unit that houses a rechargeable lithium-ion battery, a nitric oxide (NO) sensor, and integrated pneumatics and electronics. The patient interfaces with the device through the mouthpiece which is attached to the handpiece. The mouthpiece is a single patient use, disposable component that contains an anti-bacterial/anti-viral filter. To measure the fraction of exhaled FeNO in human exhaled breath using the Fenom Flo™, the patient performs a breath maneuver by grasping the handpiece and exhaling into it. A graphical user interface (GUI) display assists the patient in keeping their breath flow rate within acceptable limits.
## B Principle of Operation:
The candidate device utilizes digital electrochemical sensor technology to measure fractional exhaled nitric oxide (FeNO). The NO sensor produces an electrochemical potential difference between the electrodes of the sensor and is proportional to the amount of NO in the breath sample. The Fenom Flo device provides a digital readout of the fractional concentration of fractional exhaled nitric oxide (FeNO) in parts per billion (ppb).
## C Instrument Description Information:
1. Instrument Name:
Fenom Flo™ FeNO Monitoring System
2. Specimen Identification:
There is no mechanism to identify the specimen, as it is analyzed at the point and time of collection. Results are stored in the order they are performed, cataloged by date and time. No patient information is entered or stored on the device.
3. Specimen Sampling and Handling:
The operator obtains a breath sample by having the subject exhale into the device.
4. Calibration:
The device is calibrated by the manufacturer; no calibration is required by operators.
5. Quality Control:
The device utilizes internal and external quality control procedures. For external quality control, two measurements (a positive and a negative control) are required to validate device performance. Quality control should be performed each day of device use.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
K251674 - Page 3 of 11
{3}
Fenom Pro
# B Predicate 510(k) Number(s):
K213611
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K251674 | K213611 |
| --- | --- | --- |
| Device Trade Name | Fenom Flo™ | Fenom Pro |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Same | For the measurement of fractional exhaled nitric oxide (FeNO) in human breath. |
| Sample Type | Same | Exhaled human breath |
| Sensor type | Same | Amperometric Sensor Technology |
| Measurement Mode | Same | 6-second and 10-second breath maneuver |
| General Device Characteristic Differences | | |
| Measurement range | 5-300 ppb NO | 10-200 ppb NO |
| Limit of Detection | 5 ppb | 10 ppb |
| Analysis Time | Approximately 27 seconds | Approximately 30 seconds |
# VI Standards/Guidance Documents Referenced:
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD).[Including Amendment 2 (2021)]
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances-Requirement and tests.
- ANSI AAMI ISO 14971: 2019 Medical devices - Applications of risk management to medical devices
- ISTA 3A2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
K251674 - Page 4 of 11
{4}
- CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- AAMI TIR57:2016 Principles for medical device security - Risk management.
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
**Analytical precision**
The sponsor performed studies to evaluate the within-device precision and repeatability of the candidate device, based on CLSI EP05-A3. Data were collected using 5 candidate devices, over 5 operating days, 2 sessions per day, 4 runs per session with 2 replicates for each concentration, at concentrations of 10, 25, 75 and 200 ppb, by multiple operators. The repeatability and within-device precision over the five days was determined for each concentration.
Precision was evaluated for the 10-second mode and 6-second mode separately and the results are summarized in the tables below:
**Precision Study Summary (6-second test)**
| Device | Repeatability | | | | Within-device Precision | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD (ppb) | SD (ppb) | CV(%) | CV(%) | SD (ppb) | SD (ppb) | CV(%) | CV(%) |
| | 10 ppb | 25 ppb | 75 ppb | 200 ppb | 10 ppb | 25 ppb | 75 ppb | 200 ppb |
| 200102 | 2.02 | 1.09 | 1.9% | 0.7% | 2.28 | 0.97 | 1.7% | 0.6% |
| 200111 | 1.13 | 1.64 | 1.9% | 1.2% | 1.13 | 1.42 | 1.7% | 1.0% |
| 200112 | 1.72 | 0.85 | 1.8% | 0.6% | 1.54 | 1.61 | 1.7% | 0.6% |
| 200113 | 1.34 | 1.0 | 1.6% | 1.0% | 1.26 | 1.16 | 1.5% | 0.9% |
| 200114 | 1.21 | 0.93 | 1.4% | 0.5% | 1.12 | 0.89 | 1.3% | 0.5% |
**Precision Study Summary (10-second test)**
| Device | Repeatability | | | | Within-device Precision | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD (ppb) | SD (ppb) | CV(%) | CV(%) | SD (ppb) | SD (ppb) | CV(%) | CV(%) |
| | 10 ppb | 25 ppb | 75 ppb | 200 ppb | 10 ppb | 25 ppb | 75 ppb | 200 ppb |
| 200102 | 1.56 | 0.96 | 2.2% | 1.0% | 1.84 | 1.09 | 2.5% | 1.0% |
| 200111 | 1.05 | 1.55 | 2.5% | 1.1% | 0.95 | 1.75 | 2.4% | 1.4% |
| 200112 | 0.90 | 0.92 | 1.3% | 1.3% | 0.87 | 1.84 | 2.5% | 1.5% |
| 200113 | 0.99 | 1.0 | 1.2% | 1.4% | 1.75 | 0.99 | 1.1% | 2.1% |
| 200114 | 0.99 | 0.98 | 1.2% | 1.4% | 1.75 | 0.99 | 1.1% | 2.1% |
**Clinical precision**
The clinical precision of the candidate device was evaluated as part of a clinical study which included a total of 13 point of care sites. Across all sites, 22 healthcare providers (operators)
K251674 - Page 5 of 11
{5}
assisted with the data collection. Each of the 94 subjects provided duplicate samples using both the 6-second and 10-second measurement modes at each of two visits. The clinical precision results per mode and visit are summarized in the tables below:
10-second mode – visit 1
| Median Concentrations | Number of subjects | Within Subject Mean SD (ppb) | 95% CI for SD (ppb) | Within Subject Mean %CV | 95% CI for %CV |
| --- | --- | --- | --- | --- | --- |
| ≤ 10 ppb | 0 | N/A | N/A | N/A | N/A |
| >10 - ≤20 ppb | 1 | 1.414 | 0.000, 0.000 | 7.86 | 0.00, 0.00 |
| >20 - ≤30 ppb | 6 | 1.414 | 0.832, 2.843 | 5.02 | 2.95, 10.09 |
| >30 - ≤40 ppb | 15 | 3.111 | 2.220, 4.652 | 9.13 | 6.52, 13.66 |
| >40 - ≤50 ppb | 13 | 1.904 | 1.326, 2.947 | 4.23 | 2.95, 6.55 |
| > 50 ppb | 59 | 4.231 | 3.556, 5.116 | 4.41 | 3.70, 5.33 |
10-second mode – visit 2
| Median Concentrations | Number of subjects | Within Subject Mean SD (ppb) | 95% CI for SD (ppb) | Within Subject Mean %CV | 95% CI for %CV |
| --- | --- | --- | --- | --- | --- |
| ≤ 10 ppb | 5 | 0.990 | 0.553, 2.172 | 11.38 | 6.35, 24.96 |
| >10 - ≤20 ppb | 23 | 1.076 | 0.818, 1.476 | 6.73 | 5.12, 9.24 |
| >20 - ≤30 ppb | 21 | 1.179 | 0.885, 1.644 | 4.55 | 3.41, 6.34 |
| >30 - ≤40 ppb | 10 | 2.121 | 1.406, 3.532 | 5.99 | 3.97, 9.98 |
| >40 - ≤50 ppb | 9 | 3.064 | 1.987, 5.274 | 6.77 | 4.39, 11.65 |
| > 50 ppb | 26 | 3.291 | 2.541, 4.422 | 3.68 | 2.84, 4.95 |
6-second mode – visit 1
| Median Concentrations | Number of subjects | Within Subject Mean SD (ppb) | 95% CI for SD (ppb) | Within Subject Mean %CV | 95% CI for %CV |
| --- | --- | --- | --- | --- | --- |
| ≤ 10 ppb | 0 | N/A | N/A | N/A | N/A |
| >10 - ≤20 ppb | 0 | N/A | N/A | N/A | N/A |
| >20 - ≤30 ppb | 7 | 1.010 | 0.618, 1.903 | 3.68 | 2.25, 6.94 |
| >30 - ≤40 ppb | 15 | 1.650 | 1.177, 2.467 | 4.72 | 3.37, 7.06 |
| >40 - ≤50 ppb | 11 | 1.607 | 1.085, 2.602 | 3.71 | 2.51, 6.01 |
| > 50 ppb | 60 | 3.677 | 3.095, 4.439 | 4.02 | 3.38, 4.85 |
6-second mode – visit 2
| Median Concentrations | Number of subjects | Within Subject Mean SD (ppb) | 95% CI for SD (ppb) | Within Subject Mean %CV | 95% CI for %CV |
| --- | --- | --- | --- | --- | --- |
| ≤ 10 ppb | 3 | 0.707 | 0.327, 2.153 | 7.44 | 3.44, 22.66 |
| >10 - ≤20 ppb | 24 | 1.591 | 1.216, 2.167 | 10.07 | 7.70, 13.72 |
| >20 - ≤30 ppb | 19 | 1.228 | 0.909, 1.746 | 4.81 | 3.56, 6.83 |
| >30 - ≤40 ppb | 12 | 1.827 | 1.254, 2.887 | 5.35 | 3.68, 8.46 |
| >40 - ≤50 ppb | 9 | 1.728 | 1.121, 2.975 | 3.89 | 2.52, 6.69 |
| > 50 ppb | 27 | 2.933 | 2.276, 3.918 | 3.41 | 2.64, 4.55 |
K251674 - Page 6 of 11
{6}
K251674 - Page 7 of 11
2. Linearity:
The sponsor performed studies to evaluate the linearity performance of the candidate device. Nitric oxide was mixed to create simulated breath gas to obtain nine NO concentration levels ranging from 5-300 ppb (5, 10, 15, 30, 50, 100, 150, 200, and 300 ppb). Five replicates were obtained at each level, and ten candidate devices were evaluated in both the 6-second and the 10-second modes. The range of slope, intercept, and R2 values obtained are listed in the tables below.
| | Range of Slopes | Range of Intercepts | R² |
| --- | --- | --- | --- |
| 6-second mode | 0.99 – 1.01 | -0.67 – 0.23 | 1.00 |
| 10-second mode | 0.99 – 1.02 | -0.93 – -0.12 | 1.00 |
Effect of extreme temperature and relative humidity
The sponsor performed a study to evaluate the effects of extreme temperature and humidity conditions on the performance of the candidate device. Two devices were tested in an environmentally controlled chamber at the four corners of temperature and humidity extremes (15-35°C and 20-90%RH) at three NO concentrations (15, 75, and 200 ppb) in air with 5 replicates at each combination of temperature and humidity:
| Test Case | Temperature (°C) | Relative Humidity (%RH) |
| --- | --- | --- |
| low temperature/low humidity | 15° | 20% |
| high temperature/low humidity | 35° | 20% |
| low temperature/high humidity | 15° | 90% |
| high temperature/high humidity | 35° | 90% |
The results support the claimed operating conditions for the candidate device of 15°C-35°C and 20-90% RH.
3. Analytical Specificity/Interference:
Interference from endogenous compounds and exogenous substances
The sponsor performed a study to evaluate potential interference from endogenous compounds and exogenous substances. A substance was defined as non-interfering if the response at the tested concentration was within ± 4 ppb NO.
The following endogenous compounds were tested at the concentrations listed below and showed no interference:
| Substance | Concentration Tested |
| --- | --- |
| Acetaldehyde | 923 ppb |
| Acetone | 102 ppm |
| Acetonitrile | 500 ppm |
| Ammonia | 100 ppm |
| Carbon Dioxide | 8% by volume |
| Carbon Monoxide | 249 ppm |
{7}
K251674 - Page 8 of 11
| Substance | Concentration Tested |
| --- | --- |
| Ethanol | 1000 ppm |
| Hydrogen | 480 ppb |
| Hydrogen Peroxide | 500 ppm |
| Hydrogen Sulfide | 5 ppm |
| Isoprene | 1.5 ppm |
| Nitrogen Dioxide | 1 ppm |
| Oxygen | 100% by volume |
## Exogenous substances
The sponsor also performed a study at a single site by testing FeNO levels after use of seven common substances that are orally consumed or utilized and could potentially interfere with the device. Each subject performed one baseline measurement before the exogenous substance was introduced, one measurement 10 minutes post exposure or consumption, and one measurement 60 minutes post exposure or consumption. All subjects refrained from eating or drinking for 60 minutes prior to the baseline testing. The results of the study are shown below. The device labeling recommends that no food or beverage be consumed, and no smoking be done, for at least one hour before taking a FeNO measurement.
### Exogenous Interference Results – 6 Second Tests
| Exogenous Gas Substance | n | Mean Difference (ppb) | 95% Confidence Intervals (ppb) |
| --- | --- | --- | --- |
| Mint without Menthol | 14 | 2.2 | [16, 26] |
| Mint with Menthol | 14 | 1.9 | [17, 28] |
| Toothpaste | 14 | -0.3 | [16, 29] |
| Mouthwash with Alcohol | 14 | -2.0 | [14, 25] |
| Alcohol Free Mouthwash | 14 | -3.0 | [15, 29] |
| Caffeinated Soda | 15 | -1.0 | [18, 29] |
| Caffeine Free Soda | 13 | 2.5 | [16, 29] |
### Exogenous Interference Results – 10 Second Tests
| Exogenous Gas Substance | n | Mean Difference (ppb) | 95% Confidence Interval (ppb) |
| --- | --- | --- | --- |
| Mint without Menthol | 19 | 1.2 | [17, 25] |
| Mint with Menthol | 17 | -0.8 | [18, 29] |
| Toothpaste | 17 | 0.3 | [18, 33] |
| Mouthwash with Alcohol | 16 | -2.1 | [14, 24] |
| Alcohol Free Mouthwash | 16 | 0.4 | [18, 29] |
| Caffeinated Soda | 16 | -0.8 | [17, 27] |
| Caffeine Free Soda | 16 | 0.7 | [15, 26] |
{8}
K251674 - Page 9 of 11
## Effect of altitude
The sponsor performed a study to evaluate the effect of altitude on the candidate device. Three devices were tested using both 6-second and 10-second modes at 25, 50, and 200 ppb NO in seven replicates each at 330 feet (the nominal condition) and at 6550 feet. The results were found acceptable to support that the candidate device functions as intended at altitudes up to the claimed altitude of 6550 feet.
4. **Assay Reportable Range:**
The results of the sponsor’s detection limit and linearity studies support the claimed measuring range of 5-300 ppb FeNO.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
*Calibration stability*
The manufacturer performs calibration for each Fenom Pro device. No calibration is required by the user.
6. **Detection Limit:**
The sponsor provided data collected with both the 6-second and 10-second modes that supported the claimed detection limit of 5 ppb.
7. **Assay Cut-Off:**
Not applicable.
8. **Accuracy (Instrument):**
Please see Comparison Studies section below.
9. **Carry-Over:**
The sponsor performed a study to evaluate the potential for carryover from a high concentration FeNO sample to a subsequent sample. Five devices were evaluated using both the 6-second and 10-second modes and the FeNO concentrations applied to assess carryover ranged from approximately 10 to 200 ppb. The results showed no carryover effect.
## B Comparison Studies:
1. **Method Comparison with Predicate Device:**
Not applicable. A clinical study was conducted to validate the clinical performance of the candidate device.
2. **Matrix Comparison:**
{9}
Not applicable. The assay can be run using breath samples only.
## C Clinical Studies:
1. **Clinical Sensitivity:**
See Other Clinical Supportive Data
2. **Clinical Specificity:**
See Other Clinical Supportive Data
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
The sponsor provided the results of a study incorporating both the 6-second and 10-second tests to evaluate the clinical accuracy of the candidate device. A total of 94 patients (53 adults 18 years of age and older and 41 children ages 7 – 17) participated in the study where measurements for FeNO, spirometry, and asthma control questionnaires (ACQ) were completed at baseline (Visit 1) and two weeks later (Visit 2) after therapeutic agents were administered. The study included data from a total of 13 sites, and 22 healthcare providers (operators) assisted with the data collection.
The American Thoracic Society (ATS) defines elevated FENO as >25 ppb for adults and >20 ppb for children. The initial (visit 1) FENO inclusion criteria for this study were >30 ppb for adults and >25 ppb for children.
A meaningful change in F150 FENO 10-s is defined as >20% for initial FENO values >50 ppb and >10 ppb for initial FENO values <50 ppb. Among pediatric subjects, 80% showed a meaningful decline in FENO. Among adult subjects, 80.0% showed a meaningful decline in FeNO. Overall results showed a mean FeNO change of -30.15 ppb (-37.94%) with a mean SD of 34.90 ppb.
The decline in FeNO after 2 weeks of corticosteroid therapy resulted in the following changes in subjective and objective asthma measures.
- ACQ: Mean ACQ score fell by 54.5% after corticosteroids
- FEV1: There was a mean FEV1 change of 6.8% after corticosteroids
## D Clinical Cut-Off:
Not applicable
## E Expected Values/Reference Range:
K251674 - Page 10 of 11
{10}
The sponsor states the following in the labeling: The fractional NO concentration in expired breath can be measured by the Fenom Flo device according to guidelines for NO measurement established by the American Thoracic Society (ATS).
F Other Supportive Instrument Performance Characteristics Data:
Not applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K251674 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.