Single Use Aspiration Needle (NA-201SX-4021); Single Use Aspiration Needle (NA-201SX-4022)
K251664 · Olympus Medical Systems Corporation · KTI · Jul 29, 2025 · Ear, Nose, Throat
Device Facts
Record ID
K251664
Device Name
Single Use Aspiration Needle (NA-201SX-4021); Single Use Aspiration Needle (NA-201SX-4022)
Applicant
Olympus Medical Systems Corporation
Product Code
KTI · Ear, Nose, Throat
Decision Date
Jul 29, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The Single Use Aspiration Needle NA-201SX-4021 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope. The Single Use Aspiration Needle NA-201SX-4022 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope.
Device Story
Single-use aspiration needles (NA-201SX-4021, NA-201SX-4022) designed for ultrasound-guided fine needle aspiration (FNA) of tracheobronchial lesions. Used in clinical settings by physicians performing bronchoscopy with ultrasound endoscopes. Device consists of a needle (21G or 22G) with an adjustable sheath, inserted through the endoscope working channel. Physician navigates to the target lesion under ultrasound visualization, deploys the needle to aspirate tissue samples, and retracts the needle for specimen collection. Provides diagnostic tissue samples to assist in clinical decision-making and patient management. Sterile, ethylene oxide-processed, disposable device.
Clinical Evidence
Bench testing only. Performance testing included insertion/withdrawal, piercing, ultrasound visibility, needle slider reliability, depth limitation, tensile strength (needle-to-luer and sheath-to-handle), and structural integrity (breakage/coil separation). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility) and sterility/shelf-life validation (3 years) were performed. All tests met predefined acceptance criteria.
Indicated for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in patients requiring tissue sampling via ultrasound endoscope.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Single Use Aspiration Needle (NA-U401SX-4025N) (K190239)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 29, 2025
Olympus Medical Systems Corporation
Jillian Connery
Program Manager Regulatory Affairs
2951 Ishikawa-cho
Hachioji-shi, Tokyo 192-8507
Japan
Re: K251664
Trade/Device Name: Single Use Aspiration Needle (NA-201SX-4021); Single Use Aspiration Needle (NA-201SX-4022)
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: KTI
Dated: May 30, 2025
Received: May 30, 2025
Dear Jillian Connery:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251664 - Jillian Connery
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251664 - Jillian Connery
Page 2
Sincerely,
John S. Bender-S
2025.07.29 09:06:53 -04'00'
for Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Single Use Aspiration Needle
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251664 | ? |
| Please provide the device trade name(s). | | ? |
| Single Use Aspiration Needle (NA-201SX-4021);
Single Use Aspiration Needle (NA-201SX-4022) | | |
| Please provide your Indications for Use below. | | ? |
| The Single Use Aspiration Needle NA-201SX-4021 is intended to be used for ultrasound guided fine needle
aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an
ultrasound endoscope.
The Single Use Aspiration Needle NA-201SX-4022 is intended to be used for ultrasound guided fine needle
aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an
ultrasound endoscope. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
OLYMPUS
Traditional 510(k)
Single Use Aspiration Needle
NA-201SX-4021 & NA-201SX-4022
# SINGLE USE ASPIRATION NEEDLES
# 510(k) SUMMARY
510(k) Summary: K251664
# 1. COMPANY INFORMATION
- Applicant
OLYMPUS MEDICAL SYSTEMS CORPORATION
2951 Ishikawa-cho, Hachioji-shi
Tokyo 192-8507, Japan
FDA Establishment Registration #: 8010047
- Official Correspondent
Jillian Connery
c/o Olympus Surgical Technologies of the Americas
800 West Park Drive
Westborough, MA 01581
Cell: 404-542-5854
Email: jillian.connery@olympus.com
- Manufacturing Site
Olympus Vietnam Company Limited
8 Street
Long Thanh Industrial Zone, Tam An Commune
Long Thanh District Dong Nai, VN 810000
FDA Establishment Registration #: 3008040402
- Date Prepared: 25-July-2025
# 2. PRODUCT INFORMATION
- Trade Name: Single Use Aspiration Needle
- Models: NA-201SX-4021, NA-201SX-4022
- Common Name: Aspiration Needle
- Classification Name: Bronchoscope (flexible or rigid) and accessories
- Product Code: KTI
- Regulation Number: 21 CFR 874.4680
- Regulation Name: Bronchoscope (flexible or rigid) and accessories
- Device Class: II
510(k) Summary
Page 1 of 6
{5}
OLYMPUS
Traditional 510(k)
Single Use Aspiration Needle
NA-201SX-4021 & NA-201SX-4022
## 3. PREDICATE DEVICE
- Trade Name: Single Use Aspiration Needle
- Model: NA-U401SX-4025N
- 510(k) Number: K190239
## 4. REFERENCE DEVICE
- Trade Name: ViziShot 2 FLEX
- Model: NA-U403SX-4019
- 510(k) Number: K163469
## 5. DEVICE DESCRIPTION
The Olympus Single Use Aspiration Needle NA-201SX-4021 and NA-201SX-4022 are designed for use with ultrasound endoscopes for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. These devices are single-use, disposable, sterile (Ethylene Oxide) needles available in two models with different specifications:
- NA-201SX-4021:
- Working Length: 700mm
- Maximum Insertion Portion Diameter: 1.9mm
- Needle Width: 21G
- Normal Needle Length: 20mm
- Maximum Needle Length: 40mm
- NA-201SX-4022:
- Working Length: 700mm
- Maximum Insertion Portion Diameter: 1.8mm
- Needle Width: 22G
- Normal Needle Length: 20mm
- Maximum Needle Length: 40mm
Both models feature an adjustable sheath length and are intended for single use.
## 6. INDICATIONS FOR USE
The Single Use Aspiration Needle NA-201SX-4021 is intended to be used with ultrasound endoscopes for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree.
The Single Use Aspiration Needle NA-201SX-4022 is intended to be used with ultrasound endoscopes for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree.
510(k) Summary
Page 2 of 6
{6}
OLYMPUS
Traditional 510(k)
Single Use Aspiration Needle
NA-201SX-4021 & NA-201SX-4022
# 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The subject devices share the same intended use and technological characteristics as the predicate device, including design, materials, and operational features. Key similarities and differences are outlined below:
- Similarities:
- Same product code, classification, regulation number, common name, regulation name, and review panel.
- Same mechanism of tissue sampling (aspiration), ultrasound scanning direction, sheath composition, general design, syringe type, connector type, compatible endoscopes, method of sterilization, shelf life, energy used, visualization, and patient contact materials.
- Differences:
- Needle width: NA-201SX-4021 (21G), NA-201SX-4022 (22G) vs. Predicate Device (25G).
- Maximum insertion portion diameter: NA-201SX-4021 (1.9mm), NA-201SX-4022 (1.8mm) vs. Predicate Device (1.5mm).
Refer to Table 1 for Comparison of Technological Characteristics:
| DESCRIPTION | SUBJECT DEVICE
NA-201SX-4021 / NA-201SX-4022 | PREDICATE DEVICE
NA-U401SX-4025N
(K190239) | COMPARISON |
| --- | --- | --- | --- |
| Product Code | KTI | KTI | Same |
| Classification | Class II | Class II | Same |
| Regulation Number | 21 CFR 874.4680 | 21 CFR 874.4680 | Same |
| Common Name | Aspiration Needle | Aspiration Needle | Same |
| Indications for Use | The Single Use Aspiration Needle
NA-201SX-4021 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope. | This instrument has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. | Same |
510(k) Summary
{7}
OLYMPUS
Traditional 510(k)
Single Use Aspiration Needle
NA-201SX-4021 & NA-201SX-4022
| | The Single Use Aspiration Needle NA-201SX-4022 is intended to be used for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in combination with an ultrasound endoscope. | | |
| --- | --- | --- | --- |
| Working Length | 700 mm | 700 mm | Same |
| Needle Gauge | 21G (4021), 22G (4022) | 25G | Different – verified through bench testing |
| Max Insertion Diameter | 1.9 mm (4021), 1.8 mm (4022) | 1.5 mm | Different – verified through bench testing |
| Needle Length (Normal/Max) | 20 mm / 40 mm | 20 mm / 40 mm | Same |
| Sheath Length | Adjustable | Adjustable | Same |
| Mechanism of Sampling | Aspiration | Aspiration | Same |
| Ultrasound Scanning Direction | Parallel to insertion | Parallel to insertion | Same |
| Compatible Endoscope Channel Diameter | Ø 2.2 mm | Ø 2.0–2.2 mm | Similar - Subject device is optimized for 2.2 mm channels only; performance equivalence to broader predicate compatibility was demonstrated through bench testing. |
| Compatible Endoscope Working Length | 600 mm / BF | 600 mm / BF | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Shelf Life | 3 years | 3 years | Same |
| Single Use / Reusable | Single Use | Single Use | Same |
| Connector Type | Luer lock | Luer lock | Same |
| Patient Contact Materials | SUS304, NiTi, PTFE | SUS304, NiTi, PTFE | Same |
| Visualization | Ultrasound-guided | Ultrasound-guided | Same |
510(k) Summary
Page 4 of 6
{8}
OLYMPUS
Traditional 510(k)
Single Use Aspiration Needle
NA-201SX-4021 & NA-201SX-4022
# 8. PERFORMANCE DATA
Performance testing was conducted to demonstrate the safety and effectiveness of the Single Use Aspiration Needle NA-201SX-4021 and NA-201SX-4022. The following tests were performed and passed the acceptance criteria:
- Insertion performance into the endoscope – all results met acceptance criteria for smooth insertion and effective tissue penetration
- Piercing performance of the needle – all results met acceptance criteria for effective tissue penetration
- Ultrasound visibility of the needle – all results met acceptance criteria for visibility of the needle under ultrasound imaging.
- Needle extraction and retraction performance – The device met all predefined criteria for safe and effective needle extraction and retraction
- Needle aspiration performance – The device successfully aspirated target material under test conditions
- Withdrawal performance from the endoscope – The needle was withdrawn smoothly and safely from the endoscope
- Needle slider performance – The slider mechanism functioned reliably for needle deployment and retraction
- Limitation of needle depth – The device effectively limited needle extension to within specified parameters
- Needle-to-luer joint pull strength – The joint withstood required tensile forces without failure
- Sheath-to-handle joint pull strength – The connection between sheath and handle remained secure under stress
- Needle breakage strength – The needle demonstrated sufficient structural integrity under load
- Coil separation – The coil remained intact and did not separate under test conditions
# 9. BIOCOMPATIBILITY
Biocompatibility testing was conducted in accordance with ISO 10993-1 and relevant FDA guidance. The NA-201SX series was evaluated for cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility, and aged cytotoxicity. All tests met acceptance criteria, confirming that the device is biologically safe and biocompatible for its intended use. These results demonstrate that the subject device does not introduce new biocompatibility risks and supports a determination of substantial equivalence to the predicate device.
510(k) Summary
Page 5 of 6
{9}
OLYMPUS
Traditional 510(k)
Single Use Aspiration Needle
NA-201SX-4021 & NA-201SX-4022
# 10. STERILIZATION AND SHELF LIFE
Sterilization testing confirmed that the NA-201SX-4021 and NA-201SX-4022 devices meet endotoxin limits and achieve a sterility assurance level of 10⁻⁶ using ethylene oxide. Shelf-life testing validated a three-year shelf life through accelerated aging and simulated distribution, with all packaging integrity and product performance tests passing acceptance criteria. These results support the device’s continued safety and substantial equivalence to the predicate device.
# 11. SUBSTANTIAL EQUIVALENCE
Based on the intended use, technological characteristics, performance testing, and comparison to the predicate device, the Olympus Single Use Aspiration Needles NA-201SX-4021 and NA-201SX-4022 are substantially equivalent to the predicate device NA-U401SX-4025N.
# 12. CONCLUSION
Based on the results of the comparison of the indications for use, technological characteristics, and performance testing of the Subject and Predicate devices, the Subject Olympus SINGLE USE ASPIRATION NEEDLES device is substantially equivalent to the Predicate device NA-U401SX-4025N.
510(k) Summary
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.