Air Compression Therapy Device (ST-504); Air Compression Therapy Device (ST-505); Air Compression Therapy Device (ST-506); Air Compression Therapy Device (ST-507)
K251662 · Shenzhen Future Electronic Co., Ltd. · IRP · Nov 13, 2025 · Physical Medicine
Device Facts
Record ID
K251662
Device Name
Air Compression Therapy Device (ST-504); Air Compression Therapy Device (ST-505); Air Compression Therapy Device (ST-506); Air Compression Therapy Device (ST-507)
Applicant
Shenzhen Future Electronic Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Nov 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Air Compression Therapy Device is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Compression Therapy Device simulates kneading and stroking of tissues by using an inflatable garment.
Device Story
Powered inflatable tube massager; includes controller, lithium-ion battery, and inflatable sleeves. Operates by inflating/deflating air chambers sequentially from distal to proximal limbs to simulate kneading/stroking. Features include adjustable pressure (0-165mmHg), heating (up to 45°C), and vibration. Used in home/OTC settings by patients. User interface includes buttons and display for pressure, mode, heat/vibration intensity, and battery status. Benefits include temporary muscle pain relief and circulation improvement. Safety features include manual stop button and mechanical safety valve.
Clinical Evidence
Bench testing only. Testing included pressure, temperature, and vibration accuracy, mechanical safety valve testing, seam strength, and failure mode testing. Biocompatibility testing (ISO 10993-5, -10, -23) and electrical safety/EMC testing (IEC 60601-1, -1-2, -1-11) were performed.
Indicated for temporary relief of minor muscle aches/pains and temporary increase in circulation in healthy individuals. Used via inflatable garment to simulate kneading/stroking of tissues.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
Leg Massager (Models:RP-ALM070H,RP-ALM071H) (K232965)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 13, 2025
Shenzhen Future Electronic Co., Ltd.
% Xiongxiu Zhou
Official Correspondent
Shenzhen Reanny Medical Devices Management Consulting Co. Ltd
Room 1509, Jingting Building, Dongzhou Community,
Guangming Street, Guangming District
Shenzhen, Guangdong 518107
China
Re: K251662
Trade/Device Name: Air Compression Therapy Device (ST-504); Air Compression Therapy Device (ST-505); Air Compression Therapy Device (ST-506); Air Compression Therapy Device (ST-507)
Regulation Number: 21 CFR 890.5650
Regulation Name: Powered Inflatable Tube Massager
Regulatory Class: Class II
Product Code: IRP, IRT, IRO
Dated: October 14, 2025
Received: October 14, 2025
Dear Xiongxiu Zhou:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251662 - Xiongxiu Zhou
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251662 - Xiongxiu Zhou
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Air Compression Therapy Device
Page 10 of 53
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251662 | ? |
| Please provide the device trade name(s). | | ? |
| Air Compression Therapy Device (ST-504);
Air Compression Therapy Device (ST-505);
Air Compression Therapy Device (ST-506);
Air Compression Therapy Device (ST-507) | | |
| Please provide your Indications for Use below. | | ? |
| The Air Compression Therapy Device is indicated for the temporary relief of minor muscle aches and pains
and for temporary increase in circulation to the treated areas in people who are in good health. The Air
Compression Therapy Device simulates kneading and stroking of tissues by using an inflatable garment. | | |
| Please select the types of uses (select one or both, as
applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Shenzhen Future Electronic Co.,Ltd.
510(k) Summary
# K251662
# 510(k) Summary
Type of 510(k) submission: Traditional
Date prepared: Nov. 13, 2025
Submitter's information
| Submitter | Shenzhen Future Electronic Co.,Ltd. |
| --- | --- |
| Address | 201, building B, No.30 Youtian Road, Anliang Community,Yuanshan Street, Longgang District, Shenzhen, China,518115 |
| Contact Person | Zhou Xiongxiu |
| Email | zhouxiongxiu@futuresz.com.cn |
| Tel | 0755-29987244 |
| Fax | 0755-29987244 |
Application Correspondent
| Company | Shenzhen Reanny Medical Devices Management Consulting Co.,Ltd. |
| --- | --- |
| Address | Room 1509, Jingting Building, Dongzhou Community, Guangming Street, Guangming District, Shenzhen Guangdong, China, 518107 |
| Contact person | Zhou Xiongxiu |
| Title | General Manager |
| Email | reanny@reanny.com |
| Tel | 0755-29987244 |
Device Identification
| Trade Name | Air Compression Therapy Device (Models: ST-504, ST-505, ST-506, ST-507) |
| --- | --- |
| Regulation Number | 21 CFR 890.5650 |
| Classification Name | Massager, Powered Inflatable Tube |
| Device Classification | Class II |
| Panel | Physical Medicine |
| Product Code | IRP, IRT, IRO |
| Previous Submissions | None |
{5}
Shenzhen Future Electronic Co.,Ltd.
510(k) Summary
# Indications for use
The Air Compression Therapy Device is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Compression Therapy Device simulates kneading and stroking of tissues by using an inflatable garment.
# Device Description
The Air Compression Therapy Device is a powered inflatable tube massager. It is mainly composed of controller or main unit (including Lithium-ion Battery Pack), sleeve and Type-C USB charging cable. The Air Compression Therapy Device simulates kneading and stroking of tissues by using an inflatable garment. The devices are powered by an internal IEC 62133-2 compliant lithium-ion battery. The user interface on the Air Compression Therapy Device is a series of buttons with a small display screen to display the air pressure level, massage mode, heating intensity level, vibration intensity level, battery level display and massage area display.
The air compressor inflate the air into the air chamber, the increased air pressure in air chamber act on the limbs and cause the limb tissue to be pressed. Then deflating the air to relax the limb tissue. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient.
The air is inflated / deflated orderly from distal to proximal limbs. These action simulating kneading and stroking of limb tissues to achieve the purpose of temporary relief of minor muscle aches and pains and temporary increase in circulation to the treated areas.
## Predicate Device Information
| Sponsor | Shenzhen Ruiyi Business Technology Co.,Ltd. |
| --- | --- |
| Trade/Device Name | Leg Massager (Models:RP-ALM070H,RP-ALM071H) |
| 510(k) Number | K232965 |
| Regulation Number | 21 CFR 890.5650 |
| Sponsor | Therabody, Inc. |
| --- | --- |
| Trade/Device Name | JetBoots PRO Plus |
| 510(k) Number | K241256 |
| Regulation Number | 21 CFR 890.5650 |
{6}
Shenzhen Future Electronic Co.,Ltd.
510(k) Summary
# Basic technological characteristics. New devices VS Predicate devices
Table 1A:Summary of comparison
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Manufacturer | Shenzhen Future Electronic Co.,Ltd. | Shenzhen Ruiyi Business Technology Co.,Ltd. | Therabody,Inc. | N/A |
| 510(K) number | Pending | K232965 | K241256 | N/A |
| Product name | Air Compression Therapy Device | Leg Massager(Models:RP-ALM070H,RP-ALM071H) | JetBoots PRO Plus | N/A |
| Models | ST-504,ST-505,ST-506,ST-507 | RP-ALM070H,RP-ALM071H | JetBoots PRO Plus | N/A |
| Product regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | Same |
| Classification name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Same |
| Regulation class | 2 | 2 | 2 | Same |
| Indications for Use (IFU) | The Air Compression Therapy Device is indicated for the | Leg Massager (Models:RP-ALM070H,RP-ALM071H) is | JetBoots PRO Plus is an air compression therapy device | Note 1 Similar. |
{7}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Air Compression Therapy Device simulates kneading and stroking of tissues by using an inflatable garment. | intended for home to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas in people who are in good health. | intended to provide graduated pressure to the legs. JetBoots Plus is indicated for the temporary relief of minor muscle aches and pains, and for a temporary increase in blood circulation to the treated area in people who are in good health. JetBoots PRO Plus simulates kneading and stroking of tissues by using an inflatable garment | The subject device and reference device has heating mode and vibration mode as auxiliary functions. The predicated device only has heating mode |
| OTC or Rx | OTC | OTC | OTC | Same |
| Classification Product Code | IRP,IRT,IRO | IRP,IRT | IRP, ILY, IRO | Similar |
| Pressure range | 0~165mmHg | 0~240mmHg | 20-100mmHg,steps of 5mmHg | Note 3 Different |
4 /
{8}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Inflation time | 10-240s | 3-30s | Information not available in 510(K) summary | Note 3
Different |
| Deflation time | 5-10s | 1-5s | Information not available in 510(K) summary | Note 3
Different |
| Treatment time | 20minutes±30s | 20 minutes | 10min-60min,steps of 5min for Pneumatic Compression
10min-45min for Infrared LED | Same |
| Mode of compression | Sequential | Sequential | Sequential | Same |
| Power source | No adaptor provided
5 VDC via an IEC 60601-1
compliant power supply(100-240 VAC input)
Integrated rechargeable battery | 100~240V 50/60Hz | Power Adaptor: AC Input:100-240V AC, 50/60Hz,
DC Output: 15.0V, 4.8A, 72W Or Internal Battery | Note 2
Similar |
| Power consumption | 10W | 24W | 72W | Note 2
Different |
| Size and | ST-504:340*840mm | Thighs: | Compression boots sleeves attached to consoles. | Note 4 |
5 /
{9}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| appearance of sleeves(leg part) | ST-505:610*321mm
ST-506:300*480mm
ST-507:730*550mm | One
size:11*16*24.9inch+8*33.5inch
=279.4mm*406.4mm*632.46mm+20
3.2mm*850.9mm | Sleeves(extended) only:
S:80cm(L)*37.5cm(H)=800mm(L)*375mm(H)
M:95cm(L)*41.0cm(H)=950mm(L)*410mm(H)
L:1050cm(L)*41.0cm(H)=10500 mm(L)*410mm(H) | Different |
6 /
{10}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Photo | ST-504:

ST-505:

ST-506: | RP-ALM070H

RP-ALM071H
 | Information not available in
510(K) summary | Note 4
Different |
7 /
{11}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | | | | |
| Housing materials | Molded PC+ABS enclosure | Molded ABS enclosure | Molded PC+ABS Enclosure | Same |
8 /
{12}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Number of chambers | ST-504:5 chambers
ST-505:2 chambers
ST-506: 5 chambers
ST-507: 2 chambers | 3 chambers | 4 chambers | Note 3
Different |
| Work mode | Sequence mode: it helps relaxing muscles with various combination massage movements mainly using a single chamber inflation; Circulation mode: it helps relaxing muscles through a variety of combined massage movements, mainly by inflating two chambers simultaneously; Whole mode:it has combined functions, which includes multiple combination massage actions | RP-ALM070H:
Six models (3 combine massage modes and 3 Separate Massage Modes)
C1: Massage full legs.
C2: Massage feet and calves.
(It can be used individually, and don't need to connect the air hose of thighs wraps.)
C3:Massage feet, calves, thighs by turn. It will be turn off When you press the Button C again. | Sequential,ISO,or Static Flow Cycles
Sequential mode that applies a directional massage, starting at the base of the treated area,and progresses upwards towards the torso and then releases. | Note 5
Different |
9 /
{13}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | such as single, two, and three chambers inflation.
Work flow:①~⑤Indicate the airbag number for the current inflation step,⑥indicating all airbags are deflated.
ST-504:

ST-505:

ST-506: | T: Massages thighs
C: Massages calves
F: Massages feet. It will be turn off when you press the Button S again.
RP-ALM071H:
Five models(3 combine massage modes and 2 Separate Massage Modes):
Combine 1: Massage feet and calves
Combine 2: Massage from calves to feet
Combine 3: Massage from feet to calves
Press it again to turn off combine message function. | Sequential cycle mode

ISO mode that applies a directional massage to a smaller, user-selected area. The first chamber inflates, and after a few seconds, the second chamber starts to inflate until both chambers reach the set pressure. The both chambers deflate, and after a pause the process starts again. | |
{14}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | ST-507:
Sequence mode
ST-504:①→⑩→②→⑩→③→
⑩→④→⑩→⑤→⑩→①→⑩→
①②→⑩→①②③→⑩→①②③
④→⑩→①②③④→⑩
ST-505:①→⑩→②→⑩→①→
⑩→②→⑩→①→①②→⑩→① | F: Massage feet.
C: Massage calves. Press it again to turn off separate massage function. | ISO cycle mode
Flow cycles Progress1: first inflate Chamber 1 to target pressure, then hold & release, then go to Progress 2: first inflated Chamber 1 to target pressure, then inflate Chamber 2 and hold Chamber 1. When Chamber 2 reach the target pressure, hold chamber 1&2 for specified time, then release | |
11 /
{15}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | →①②→⑩→①②→⑩→①→⑩
→①②→①→⑩→①②→②→⑩
→①②→⑩
ST-506: ①②→③→④→⑤→⑩
→①②→①②③→③④→④⑤→⑤
⑤①②→①②③→③④→④⑤→⑤
⑤→⑩→①②→①②③→①②③
④→①②③④⑤→⑩→①②③④
⑤→→③④⑤→④⑤→⑤→⑩→①②③④⑤→⑩
ST-507:①→⑩→②→⑩→①→⑩
→②→⑩→①→①②→⑩→①
→①②→⑩→①②→⑩→①→⑩
→①②→①→⑩→①②→②→⑩
→①②→⑩
Circulation mode
ST-504:①→①②→①②③→①② | | totally, then go to Progress 3.
Progress 3: first inflated Chamber 1 to target pressure, then inflate Chamber 2 and hold Chamber 1. When Chamber 2 reach the target pressure, hold chamber 1 & 2, & start to inflate Chamber 3, when Chamber 3 reach to the target pressure, the hold chamber 1,2,3 for specified time then release totally.The go to Progress 4.
Progress 4: first inflated Chamber 1 to target pressure, then inflate Chamber 2 and hold Chamber 1. When Chamber 2 reach the target pressure, hold chamber 1&2, & start to inflate Chamber 3, when Chamber 3 reach to the target pressure, then hold chamber 1,2,3,& start to inflate Chamber 4, when Chamber 4 reach to the target pressure, then hold chamber | |
12 /
{16}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | ③④→①②③④⑤→①
ST-505:①→②→①→②→①→②
→①→①→①②→①→②→①→
①→①②→①
ST-506: ①②→①②③→①②③
④→③④⑤→④⑤→⑤→①→①
②→①②③→①②③④→①②③
④⑤→③④⑤→④⑤→⑤→①→①
①②→①→①②③→①→①②③
④→①→①②③④⑤→①→①②
③④⑤→①②③④→①②③→①
②→①
ST-507: ①→②→①→②→①→
②→①→①→①②→①→②→①
→①→①②→①
Whole mode
ST-504:Sequence | | 1,2,3 & 4 for specified time then release totally. Then go back to Progress 1 again.
Static Cycle
The chambers inflate one at the time starting at chamber 1 while maintaining negative gradient of compression along the leg. The pressure is limited to 20mmHg to 30mmHg. Once inflated, the chambers do not deflate during the treatment. While the garment is inflated, it comes in contact with the legs to optimize the treatment of vibration and infrared LED light by ensuring optimal contact with the leg. | |
{17}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | mode+Circulation mode
ST-505:Sequence
mode+Circulation mode
ST-506: ①②→③→④→⑤→⑩
→①②→①②③→③④→④⑤→
⑤①②→①②③→③④→④⑤→
⑩→①②→①②③→①②③④→
①②③④⑤→①②③④→①②③
→①②→⑩→①②③④⑤→③④
⑤→④⑤→⑤→⑩→①②③④⑤
→⑩
ST-507: Sequence
mode+Circulation mode | | Static Cycle | |
| Safety feature | Button on display allows user to stop therapy session at any time | Button on display allows user to stop or pause therapy session at any time | Information not available in 510(K) summary | Same |
{18}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Technology/Device Function | Compressor and valve system which sequentially inflates inflatable chambers | Compressor and valve system which sequentially inflates inflatable chambers | Compressor and valve system which sequentially inflates cells of appliance, Light Emitting Diodes | Same |
| Operating environment | Temperature: 5°C~35°C
Relative Humidity:15-90%, non-condensing
Atmospheric pressure:70kPa-106kPa | Temperature:5-40°C,
Humidity:15%-90% | Information not available in 510(K) summary | Similar |
| Transportation and Storage environment | Temperature: -10°C~45°C
Relative Humidity:5-90%, non-condensing
Atmospheric pressure:70kPa-106kPa | Temperature:-25-70°C,
Humidity:15%-90% non-condensing
Atmospheric pressure:75kPa-106kPa | Information not available in 510(K) summary | Similar |
| Heat therapy | Not exceeding 45°C
H1: 36°C±2°C | Not exceeding 45°C | Maximum Device Surface | Same |
15 /
{19}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| | H2: 38°C±2°C
H3: 40°C±2°C | | Temperature:43.2°C
Highest Measured Skin Temperature During Treatment: 36.8°C | |
| Number of patients that can be treated at one time | One | One | One | Same |
| Static or Intermittent Pressure | Both | Both | Information not available in 510(K) summary | Same |
| Power down | Available | Available | Information not available in 510(K) summary | Same |
| Heating Mechanism | Thermotherapy | Thermotherapy | Infrared LED light therapy | Same |
| User Interface | Touch Screen | Touch Screen | Information not available in 510(K) summary | Same |
16 /
{20}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Types of garments | ST-504:Sleeve for thighs, knees, calves and feet
ST-505:Sleeve for calves
ST-506:Sleeve for knees, calves and feet
ST-507:Sleeve for thighs, knees and calves | Sleeves for thighs, calves, and feet | Sleeves for Legs and feet | Similar |
| Patient Contacting Material | All encompassed with a Nylon with a Polyurethane laminate material | All encompassed with a Nylon with a Polyurethane laminate material | Polyether Nylon Fabric | Same |
| Multi-Patient Use and Single Patient Use Wraps | Multi-Patient Use | Multi-Patient Use | Information not available in 510(K) summary | Same |
| Sterility | Non-sterile only | Non-sterile only | Non-sterile only | Same |
| Expected life of garments | Based on frequency of use and continued functional performance | Based on frequency of use and continued functional performance | Information not available in 510(K) summary | Same |
17 /
{21}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
| Device | Subject device | Predicate device | Reference Device | Comparison |
| --- | --- | --- | --- | --- |
| Standards | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Note 6 |
| | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | Minor different. |
| | IEC 60601-1-11 | IEC 60601-1-11 | ANSI C63,18 | |
| | IEC TS 60601-4-2 | IEC TS 60601-4-2 | IEC 60601-1-6 | |
| | ISO 10993-5 | ISO 10993-5 | IEC 60601-2-83 | |
| | ISO 10993-10 | ISO 10993-10 | IEC 62133-2 | |
| | ISO 10993-23 | | IEC 62304 | |
| | IEC 62133-2 | | IEC 62471 | |
| | | | IEC 60601-2-57 | |
| | | | ISO 10993-5 | |
| | | | ISO 10993-10 | |
| | | | ISO 10993-23 | |
18 /
{22}
Shenzhen Future Electronic Co.,Ltd.
510(k) Summary
# Comparison discussion
## Note 1:
The subject device has the intended use as same as the predicate device (K232965) and reference device (K241256). According to the 510(k) summary of predicate device (K232965), it has air compression and heating therapy only, while the reference device (K241256) has air compression, Infrared LED heat therapy and therapeutic vibration function. The type of use of the subject device is same to the predicate device and reference device. The temperature of heat therapy of the devices are not exceeding 45°C, which complied to the safety and effectiveness.
## Note 2:
The output of power source are different in current and power consumption among the devices. The subject device is powered by an internal rechargeable battery, with a Type-C USB charging cable only, no external DC Power Adaptor is provided by manufacturer. The predicate device (K232965) is powered by an external DC Power Adapter. The JetBoots PRO Plus boots (reference device, K241256) are powered by an internal rechargeable battery or by an external DC Power Adapter, plugged into a wall electrical outlet. The batteries are located inside each console and can be charged before the device operates. They are all complied with IEC 60601-1, IEC 60601-1-2 and IEC60601-1-11, the safety and effectiveness of the subject device is verified via tests, so the differences do not affect the safety and effectiveness.
## Note 3:
From the comparison table above, the device design among the predicate device, reference device and subject device are mainly equivalent, the specifications of subject device are not completely same among them. The Pressure range, Inflation time, Deflation time, number of chambers, working mode of subject device are similar or a bit different to the predicate device and reference device. The key parameters for the safety and effectiveness of the air compression therapy device are pressure range and temperature of thermal therapy. The safety and effectiveness of the subject device and reference device are verified via tests according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, non-clinical performance bench testing, so the differences do not affect the safety and effectiveness.
## Note 4:
The "Size and appearance of sleeves (leg part)", is belonging to basic physical characteristics. Although it is a bit different from the predicate device, it will not affect the main function and the intended use of the device. They all also comply with IEC 60601-1 requirements. Besides, the subtle change of the physical characteristics will not affect the critical functions or normal use, and not raise any safety or effectiveness issues.
{23}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
Note 5:
Although the subject device provides 3 working modes, which are not completely the same as the predicate device with six modes or five modes, the predicate device only shows the differences in treatment area when switching the modes. The treatment area of subject device are varied due to different models, thus all the models of subject device can cover the treatment area presented by the predicate device. The subject device has similar working flow compared to the reference device. The treatment pressure range of subject device and predicate are the same under different modes, thus different modes would not impact safety and effectiveness.
Note 6:
The "standards" of subject device is same with predicate device but minor different to the reference device. The subject device and predicate device have no LED Strips, thus no safety requirement for IEC 62471, IEC 60601-2-83 and IEC 60601-2-57. Therefore, this difference will not raise any safety or effectiveness issue.
20 /
{24}
Shenzhen Future Electronic Co., Ltd.
510(k) Summary
# Summary of the technological characteristics of the device
The device meets all the applicable technical requirements of:
- IEC 60601-1-11: 2015 - Medical electrical equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-1-2: 2014 +A1:2020 - Medical electrical equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC TS 60601-4-2-Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- ISO 10993-5: 2009 - Biological Evaluation of Medical Device - Part 5: Tests for in vitro Cytotoxicity
- ISO 10993-10: 2021 - Biological Evaluation of Medical Devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021- Biological Evaluation of Medical Devices - Part 23: Tests for irritation
- IEC 62133-2:2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications Part 2: Lithium systems
The following additional performance bench testing was conducted:
- Accuracy for Pressure Testing
- Accuracy for Temperature Testing
- Accuracy for Vibration Testing
- Mechanical safety valve testing
- Seam Strength Testing
- Failure Mode Testing
# Conclusion
Air Compression Therapy Device is substantially equivalent to the Predicate Device and the Reference Device in Indications for Use and technological characteristics. Though minor differences exist among the Proposed, Predicate, and Reference devices, these do not raise questions of safety and effectiveness. Therefore, Air Compression Therapy Device is as safe, as effective, and performs as well as the Predicate and Reference Devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.