The Mariana Minerva is intended to provide non-invasive ventilation, for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). It is intended for home and hospital use. The Mariana Nimbus is intended to provide non-invasive ventilation, for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). It is intended for home and hospital use.
Device Story
Mariana Minerva and Mariana Nimbus are non-invasive ventilators (NIV) for home and hospital use; utilize microprocessor-controlled brushless centrifugal blower, pressure/flow sensors to regulate air delivery; provide CPAP, S, ST, and iVAPS with AutoEPAP therapy modes. Device monitors/compensates for mask leak via VSync; integrates heated humidification; includes alarm module for safety. Operated by clinician or patient via LCD touch screen; provides therapeutic/technical data. Supplemental oxygen can be added. Bluetooth/cellular/SD card connectivity enables data transfer. Benefits include airway splinting for non-dependent, spontaneously breathing patients with respiratory impairment or OSA. Device does not support external batteries; intended for patients who can tolerate brief therapy interruptions.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including system verification and comparative side-by-side testing against the predicate device. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-70, ISO 80601-2-74, ISO 80601-2-79, ISO 10993-1, and ISO 18562-1 verified.
Indicated for patients >30 lb (13 kg) or >66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). Intended for home and hospital use.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 23, 2026
ResMed Pty Ltd
% Nicole Wilson
Director, Regulatory Product Strategy & Delivery Regulatory Affairs
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153, Australia
Re: K251661
Trade/Device Name: Mariana Minerva; Mariana Nimbus
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous Ventilator
Regulatory Class: Class II
Product Code: MNS, BTT
Dated: January 23, 2026
Received: January 23, 2026
Dear Nicole Wilson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251661 - Nicole Wilson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251661 - Nicole Wilson
Page 3
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251661 | |
| Device Name Mariana Minerva; Mariana Nimbus | |
| Indications for Use (Describe) The Mariana Minerva is intended to provide non-invasive ventilation, for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). It is intended for home and hospital use. The Mariana Nimbus is intended to provide non-invasive ventilation, for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). It is intended for home and hospital use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
ResMed
Traditional 510(k)
Mariana
# 510(k) Summary
[As required by 21 CFR 807.92(c)]
Date Prepared: February 2026
Company Name/Owner: ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista, NSW, 2153 Australia
Prepared and Submitted by: Ms Nicole Wilson
Director, Regulatory Product Strategy & Delivery
nicole.wilson@resmed.com.au
Official Contact: Ms Nicole Wilson
Director, Regulatory Product Strategy & Delivery
Regulatory Affairs
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153, Australia
Phone: +61 404 820 279
Email: nicole.wilson@resmed.com.au
Device Trade Name: Mariana
Device Common Name: Ventilator, continuous, non-life supporting
Classification and Classification Name: 21 CFR §868.5895, 73 MNS (Class II)
Continuous Ventilator
Product Code: MNS
Predicate Device: Juno VPAP ST-A (K161492)
Reference Device: N/A
Submission Reason: New device
Page 1
{5}
ResMed
Traditional 510(k)
Mariana
## Device Description:
The Mariana devices, that is Mariana Minerva and Mariana Nimbus, are non-invasive ventilation (NIV) devices that use a microprocessor controlled blower, along with pressure and flow sensors, to achieve pressure, flow and time regulation of air delivery. They have an integrated alarms module. The Mariana devices also include optional humidification, with air delivery to the patient via heated or non-heated breathing circuits. The devices provide both therapeutic (e.g. tidal volume) and technical data (e.g. system fault), and a user interface allowing adjustment of device parameters. The devices use an external AC power supply and allow the addition of low pressure supplemental oxygen.
## Indications for Use:
The Mariana Minerva is intended to provide non-invasive ventilation, for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). It is intended for home and hospital use.
The Mariana Nimbus is intended to provide non-invasive ventilation, for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA). It is intended for home and hospital use.
Substantial Equivalence:
| Characteristic | Predicate device: Juno VPAP ST-A K161492 | Subject devices: Mariana Minerva / Mariana Nimbus | Substantially Equivalent? |
| --- | --- | --- | --- |
| Regulatory Classification | 73MNS
Class II
21 CFR 868.5895 | 73MNS
Class II
21 CFR 868.5895 | YES |
| Intended Use | The Juno VPAP ST-A is intended to provide non-invasive ventilation | The Mariana is intended to provide non-invasive ventilation | YES |
| Indications for Use Statement | The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30 lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).
The iVAPS mode is indicated for patients weighing more than 66 lbs (30 kg). | The Mariana is indicated to provide noninvasive ventilation for patients weighing more than 30 lb (13 kg) or more than 60 lb (30 kg) in iVAPS mode with respiratory impairment or obstructive sleep apnoea (OSA).
The Mariana is intended for home and hospital use. | YES
The intended patient population between predicate and subject device is equivalent.
Respiratory impairment terminology included in subject device to better reflect patient population, alignment with ISO 80601-2-79 |
Page 2
{6}
ResMed
Traditional 510(k)
Mariana
| Characteristic | Predicate device: Juno VPAP ST-A K161492 | Subject devices: Mariana Minerva / Mariana Nimbus | Substantially Equivalent? |
| --- | --- | --- | --- |
| | The Juno VPAP ST-A is intended for home and hospital use. | | |
| Environment of Use | Home Healthcare Environment (including aircraft), Professional Healthcare Facilities | Home Healthcare Environment (including aircraft), Professional Healthcare Facilities | YES |
| Therapies | | | |
| Therapy Modes | • CPAP
• S, ST, T
• PAC
• iVAPS with AutoEPAP | • CPAP
• S, ST
• PAC
• iVAPS with AutoEPAP | YES
The iVAPS with AutoEPAP therapy algorithm implemented in the subject devices is identical to the algorithm cleared under K161492.
Subject device does not have Timed Mode to better reflect the needs of intended patient population (non-dependent, spontaneously breathing).
Removal of Timed Mode does not affect the AutoEPAP algorithm or its clinical functionality and does not impact substantial equivalence. |
| Therapy Features | • iBR
• Ti Control
• Rise time adjustment
• VSync | • iBR
• Ti Control
• Rise time adjustment
• VSync | YES |
| Vsync | Vsync monitors and compensates for unintentional mask leak, so that patients receive their prescribed therapy pressure. Working in sync with TiControl, Vsync ensures reliable breath detection even in conditions of high leak. | Vsync monitors and compensates for unintentional mask leak, so that patients receive their prescribed therapy pressure. Working in sync with TiControl, Vsync ensures reliable breath detection even in conditions of high leak. | YES |
| Pressure Range | • 4-20 cmH₂O (CPAP)
• 3-30 cmH₂O (bi-level)
• Fixed EPAP range: 3-25 cmH₂O
AutoEPAP range 3-25 cmH₂O | • 4-20 cmH₂O (CPAP)
• 3-30 cmH₂O (bi-level)
• Fixed EPAP range: 3-25 cmH₂O
AutoEPAP range 3-25 cmH₂O | YES |
Page 3
{7}
ResMed
Traditional 510(k)
Mariana
| Characteristic | Predicate device: Juno VPAP ST-A K161492 | Subject devices: Mariana Minerva / Mariana Nimbus | Substantially Equivalent? |
| --- | --- | --- | --- |
| Other Features | Ramp
Ramp Down
Smart Start / Smart Stop
Learn Targets
Mask Fit | Ramp
Ramp Down
Smart Start / Smart Stop
Learn Targets
Mask Fit
Test Drive
Acoustic Signal | YES
Test Drive is an optional, non-therapeutic feature intended for patient acclimatisation accessed separately from therapy. It does not modify the patient's prescribed therapy settings and is not required for device function or therapeutic operation. No impact to safety or efficacy.
Acoustic signal is an optional non-therapeutic feature that captures a series of short 0.2 second acoustic signals to identify important characteristics about a user's equipment. It is not required for device function or therapeutic operation. No impact to safety or efficacy. |
| Technology & Design | | | |
| System Components | Flow Generator
Integrated Humidifier
Air Tubing
BV filter (optional)
Mask | Flow Generator
Integrated Humidifier
Air Tubing
BV filter (optional)
Mask | YES |
| Operating Principle | Micro-processor controlled brush-less centrifugal blower as air source to provide splinting of patient airway | Micro-processor controlled brush-less centrifugal blower as air source to provide splinting of patient airway | YES |
| Materials | Various materials, including:
Polycarbonate (PC)
Polyurethane (PU)
Polypropylene (PP)
Thermoplastic
Silicone
Stainless steel
Copolyester
PC/ABS
TPE | Various materials, including:
Polycarbonate (PC)
Polyurethane (PU)
Polypropylene (PP)
Thermoplastic
Silicone
Stainless steel
Copolyester
PC/ABS
TPE
Polybutylene terephthalate (PBT)
Liquid crystal polymer (LCP)
Low Density Polyethylene
Polyolefin Polymer | YES
Substantially equivalent materials verified for device functionality, reprocessing and biocompatibility. |
Page 4
{8}
ResMed
Traditional 510(k)
Mariana
| Characteristic | Predicate device: Juno VPAP ST-A K161492 | Subject devices: Mariana Minerva / Mariana Nimbus | Substantially Equivalent? |
| --- | --- | --- | --- |
| | | Non Woven Textile (Mariana Nimbus) | |
| User Interface | LCD screen, dial & buttons | LCD touch screen & buttons | YES
Touch screen has no impact on safety or efficacy |
| Alarm module | Yes | Yes | YES |
| Supplemental Oxygen | Labelled for use with supplemental oxygen | Labelled for use with supplemental oxygen | YES |
| Power Supply | 100-240V, 50-60Hz | 100-240V, 50-60Hz | YES |
| DC-DC Converter for Vehicle Use | Yes | Yes | YES |
| External Battery Compatibility | Yes | No | YES
The subject device does not include external battery compatibility, which is appropriate for its intended patient population (non-dependent, spontaneously breathing). Mariana has no Essential Performance requirements related to continuous operation and does not include a Timed Mode; therefore, it is intended only for patients who can tolerate brief interruptions to therapy and who are not ventilator-dependent. |
| Data Connectivity | SD card, Cellular Wireless | SD card, Cellular Wireless, Bluetooth Wireless | YES
Bluetooth is an additional wireless technology option to provide data transfer capabilities. No impact on safety or efficacy. |
| Humidification | | | |
Page 5
{9}
ResMed
Traditional 510(k)
Mariana
| Characteristic | Predicate device: Juno VPAP ST-A K161492 | Subject devices: Mariana Minerva / Mariana Nimbus | Substantially Equivalent? |
| --- | --- | --- | --- |
| Humidifier | Integrated thermostatically controlled heated humidifier with detachable water chamber | Integrated thermostatically controlled heated humidifier with detachable water chamber | YES |
| Usage | Single patient use at home, multi patient multi use in the hospital | Single patient use at home, multi patient multi use in the hospital | YES |
| Operating Principle | Water is heated by a heater plate to create warm and moist air within the water chamber. Airflow from the device is passed through the heated water chamber, which moistens and warms the air. | Water is heated by a heater plate to create warm and moist air within the water chamber. Airflow from the device is passed through the heated water chamber, which moistens and warms the air. | YES |
| Humidifier output | > 10 mg/L BTPS | > 10 mg/L BTPS | YES |
| Heated Tube dimensions | Length 2 m Diameter 15 mm | Length 2 m Diameter 15 mm | YES |
| Heated Tube maximum power | 60 watt | 60 watt | YES |
| Connections | Custom connector to flow generator. Cuff to fit 22mm ISO taper to patient port. | Custom connector to flow generator. Cuff to fit 22mm ISO taper to patient port. | YES |
| Heated Tube temperature settings | 60-86 °F (16-30 °C) | 60-86 °F (16-30 °C) | YES |
| Heated Tube temperature cutout | 106 °F (41 °C) | 106 °F (41 °C) | YES |
| Alarm Functions | | | |
| Alarm conditions | • Non-vented Mask
• High Leak
• Blocked Tube
• Tube Disconnected
• Power Fail
• System Fault
• Apnea
• Low Minute Ventilation | • Non-vented Mask
• High Leak
• Blocked Tube
• Tube Disconnected
• Power Fail
• System Fault
• Apnea
• Low Minute Ventilation | YES
Subject device does not monitor or display SpO2 as this is handled by the compatible Bluetooth oximeter. |
Page 6
{10}
ResMed
Traditional 510(k)
Mariana
| Characteristic | Predicate device: Juno VPAP ST-A K161492 | Subject devices: Mariana Minerva / Mariana Nimbus | Substantially Equivalent? |
| --- | --- | --- | --- |
| | • Low SpO2 | • Tub/End cap disconnected | |
| Standards | | | |
| Standards Compliance | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-6
IEC 60601-1-8
IEC 60601-1-11
ISO 8185 | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-6
IEC 60601-1-8
IEC 60601-1-11
ISO 80601-2-70
ISO 80601-2-74
ISO 80601-2-79 | YES
Updated compliance to FDA recognized performance standards. Bench testing demonstrates device performs to specification. |
## Non-Clinical Data:
Verification bench testing for the Mariana devices comprises system verification and comparative side-by-side predicate testing. Verification confirmed the Mariana devices met the predetermined acceptance criteria as defined in the relevant compliance standards and as defined in the system verification protocols.
Comparative predicate testing supports the determination that the Mariana devices are substantially equivalent to the predicate device Juno VPAP ST-A (K161492). Verification bench testing included testing the performance of the therapy modes and therapy functions.
The Mariana devices were designed and tested in accordance with the applicable requirements in relevant FDA consensus standards including:
- IEC 60601-1 Edition 3.2:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014+AMD1:2021, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-6 Edition 3.2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-8:2006+AMD2:2021, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 60601-1-11:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical
Page 7
{11}
ResMed
Traditional 510(k) Mariana
electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-70, Medical Electrical Equipment – Part 2-70: Particular requirements for basic safety and essential performance of sleep apnoea breathing therapy equipment
- ISO 80601-2-74, Medical Electrical Equipment – Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment
- ISO 80601-2-79, Medical electrical equipment - Part 2-79: Particular requirements for the basic safety and essential performance of ventilatory support equipment for ventilatory equipment
- ISO 10993-1:2018 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 18562-1:2024 – Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process
## Clinical Data:
Clinical performance data is not required as the subject devices use established therapeutic technology and bench testing is sufficient to demonstrate substantial equivalence
## Substantial Equivalence Conclusion:
The Mariana devices have the same intended use and similar indications and technological characteristics as the predicate device. The differences in the technological characteristics between the predicate device and subject devices do not impact or raise new questions of safety or efficacy. Non-clinical performance data supports the determination that the subject devices are substantially equivalent to the predicate device.
Page 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.