EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System
Applicant
Philips Ultrasound, LLC
Product Code
IYN · Radiology
Decision Date
Oct 9, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging. The clinical environments where EPIQ Series Diagnostic ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure. The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging. The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Device Story
EPIQ and Affiniti Series are diagnostic ultrasound systems; input consists of ultrasonic echoes processed to generate images/flow analysis. New modification adds a custom electromagnetic interference (EMI) filter module to reduce electrical noise from external hospital environments (e.g., poor grounding/powerline noise). Filter ensures cleaner electrical signals to the system; prevents image artifacts/distortions. Operated by trained healthcare professionals (sonographers, physicians) in clinics/hospitals. Output is visual ultrasound imagery used for clinical diagnosis. Integration with Philips EchoNavigator allows live X-ray/echo guidance for cardiovascular procedures. Device benefits include improved image quality and reliability in electrically noisy environments.
Clinical Evidence
No clinical data required. Substantial equivalence demonstrated via design features, indications, fundamental technology, and non-clinical performance testing (EMC/safety).
Indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body, including abdominal, cardiac (adult/pediatric/fetal), vascular, cephalic, obstetric, gynecological, intraoperative, musculoskeletal, ophthalmic, urological, pediatric, small organ, transesophageal, transrectal, transvaginal, and lung applications. Used by trained healthcare professionals in clinics, hospitals, and point-of-care settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound System (K250177)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 9, 2025
Philips Ultrasound LLC
Swetha Paritala
Senior Regulatory Affairs Specialist
22100 Bothell Everett Highway
Bothell, Washington 98021
Re: K251651
Trade/Device Name: EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, OBJ, QIH
Dated: May 29, 2025
Received: September 11, 2025
Dear Swetha Paritala:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251651 - Swetha Paritala
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251651 - Swetha Paritala
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
EPIQ Series Diagnostic Ultrasound System
Page 1 of 2
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251651 | ? |
| Please provide the device trade name(s). | | ? |
| EPIQ Series Diagnostic Ultrasound System;
Affiniti Series Diagnostic Ultrasound System | | |
| Please provide your Indications for Use below. | | ? |
| EPIQ:
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow
analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult),
Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac),
intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic,
Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal
(Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode,
Power Doppler, and Harmonic Imaging.
The clinical environments where EPIQ Series Diagnostic ultrasound Systems can be used include clinics,
hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with
image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and
live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures
and operating instructions given in the product user information. Systems are to be operated only by
appropriately trained healthcare professionals for the purposes for which they were designed.
However, nothing stated in the user information reduces your responsibility for sound clinical judgement and
best clinical procedure.
Affiniti:
The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid
flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult),
Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac),
Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel,
Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode,
Power Doppler, and Harmonic Imaging. | | |
{4}
The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
EPIQ Series Diagnostic Ultrasound System
Page 2 of 2
{5}
PHILIPS
510(k) Summary
#K251651
This summary of safety and effectiveness information is submitted in accordance with 21 CFR § 807.92.
Date Prepared: January 10, 2025
I. Submitter
Manufacturer Name and Address
Philips Ultrasound LLC
22100 Bothell Everett Hwy
Bothell, WA 98021-8431 USA
Contact Information
SwethaRao Paritala
Senior Regulatory Affairs Specialist
22100 Bothell Everett Hwy
Bothell, WA 98021-8431 USA
Swetharao.partitala@philips.com
1-412-969-1010
Secondary Contact
Priscilla Herpai
Senior Regulatory Affairs Manager
22100 Bothell Everett Hwy
Bothell, WA 98021-8431 USA
Priscilla.Herpai@philips.com
1-610-533-7984
II. Device
Proprietary Name
EPIQ Series Diagnostic Ultrasound System
Affiniti Series Diagnostic Ultrasound System
Common Name
Diagnostic Ultrasound System and Transducers
Product Code; Regulation Description; Regulation Number
IYN; Ultrasonic Pulsed Doppler Imaging System; 21 CFR 892.1550 (Primary)
IYO; Ultrasonic Pulsed Echo Imaging System; 21 CFR 892.1560
ITX; Diagnostic Ultrasonic Transducer; 21 CFR 892.1570
OBJ; Diagnostic Intravascular Catheter; 21 CFR 870.1200
QIH; Medical Image Management and Processing System; 21 CFR 892.2050
Device Class
Class II
Review Panel
Radiology
Predicate Device
K250177; EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound System
III. Device Description
The purpose of this special 510(k) Pre-Market Notification is to introduce EPIQ and Affiniti Series Diagnostic Ultrasound Systems with Custom Electromagnetic Interference (EMI) filter. The proposed EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound Systems with new Custom EMI filter spare part, will reduce the electrical interferences induced by the external environments and feed the clean electrical signals to the ultrasound device, equivalent to predicate device, K250177.
1/7
{6}
PHILIPS
Ultrasound systems are very sensitive in nature. These electrical interference couplings result in Image distortions or artifacts in Images, using this Custom EMI filter will reduce such artifact effects appearing on Ultrasound systems.
## IV. Indications for Use
### EPIQ:
The intended use of EPIQ Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where EPIQ Series Diagnostic Ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
### Affiniti:
The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
2/7
{7}
PHILIPS
# V. Comparison of Technological Characteristics with the Predicate
The purpose of the submission is to introduce the EPIQ and Affiniti Series Diagnostic Ultrasound Systems with EMI Filter (optional spare part). The proposed EPIQ and Affiniti Series Diagnostic Ultrasound Systems are all Track 3 Devices and comply with the referenced standards as well as the FDA Ultrasound Guidance Document, Guidance for Industry and FDA Staff – Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, issued in February 2023. The subject devices are substantially equivalent to the predicate devices (K250177).
| Feature | EPIQ and Affiniti Series Diagnostic Ultrasound System EMI filter Proposed Device | EPIQ and Affiniti Series Diagnostic Ultrasound System (K250177) Predicate Device | Comparison |
| --- | --- | --- | --- |
| Indications for Use | EPIQ: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Affiniti: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. | EPIQ: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Affiniti: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. | EPIQ: Identical to predicate device.
Affiniti: Identical to predicate device. |
| Intended Users | Trained healthcare professionals | Trained healthcare professionals | Identical |
3/7
{8}
PHILIPS
| | Intended for sonographers, physicians, and biomedical engineers who operate and maintain your product.
Before use of the system and user information, the user must be familiar with ultrasound techniques. Sonography training and clinical procedures are not included in the User Manual or with the EPIQ Series Diagnostic Ultrasound System.
Before use of the system and user information, the user must be familiar with ultrasound techniques. Sonography training and clinical procedures are not included in the User Manual or with the Affiniti Series Diagnostic Ultrasound System. | Intended for sonographers, physicians, and biomedical engineers who operate and maintain your product.
Before use of the system and user information, the user must be familiar with ultrasound techniques. Sonography training and clinical procedures are not included in the User Manual or with the EPIQ Series Diagnostic Ultrasound System.
Before use of the system and user information, the user must be familiar with ultrasound techniques. Sonography training and clinical procedures are not included in the User Manual or with the Affiniti Series Diagnostic Ultrasound System. | |
| --- | --- | --- | --- |
| Intended User Environment | Clinics, hospitals, and clinical point-of-care for diagnosis of patients. | Clinics, hospitals, and clinical point-of-care for diagnosis of patients. | Identical |
| USA FDA Classification | Class II | Class II | Identical |
| Primary Product Code | IYN | IYN | Identical |
| Primary Regulation Number | 21 CFR 892.1550 | 21 CFR 892.1550 | Identical |
| Secondary Product Codes | ITX
IYO
OBJ*
QIH
*Applicable only to Philips EPIQ Series Diagnostic Ultrasound System, per clearance under K202216; Not applicable for Philips Affiniti Series Diagnostic Ultrasound System | ITX
IYO
OBJ*
QIH
*Applicable only to Philips EPIQ Series Diagnostic Ultrasound System, per clearance under K202216; Not applicable for Philips Affiniti Series Diagnostic Ultrasound System | Identical.
The EMI filter is not associated with product code QIH. |
4/7
{9}
PHILIPS
| Secondary Regulation Name | Diagnostic ultrasonic transducer
Ultrasonic pulsed echo imaging system
Diagnostic intravascular catheter
Automated Radiological Image Processing Software | Diagnostic ultrasonic transducer
Ultrasonic pulsed echo imaging system
Diagnostic intravascular catheter
Automated Radiological Image Processing Software | Identical |
| --- | --- | --- | --- |
| Secondary Regulation Number | 21 CFR 892.1560
21 CFR 892.1570
21 CFR 892.2050
21 CFR 870.1200 | 21 CFR 892.1560
21 CFR 892.1570
21 CFR 892.2050
21 CFR 870.1200 | Identical |
| Reusable | Yes | Yes | Identical |
| Duration of use | Limited (≤ 24 hours) | Limited (≤ 24 hours) | Identical |
| Marketing Name of Application | Custom EMI Filter Module | - | Introduction of the optional spare part
Custom EMI Filter Module |
| Image Artifacts | Image appears clean without any Image artifacts potentially caused by the external hospital environments such as bad non isolated grounding. | Visible imperfections or distortions in an Ultrasound image that are a result of external interference. These artifacts can be caused by external factors such as conductive noise induced by the electrical powerline or poor electrical grounding can affect the overall quality of the image. | The proposed device has better & cleaner Image over Predicate Device when image artifacts may arise due to external hospital environmental factors, such as poor or non-isolated grounding. Verification and Validation Testing has been conducted with Custom EMI Filter on the EPIQ & Affiniti Systems to support Clear Image without any Image artifacts. |
| User Interface | Yes | Yes | Custom EMI filter doesn't involve the user interface of the system |
| Software Performance | Yes | Yes | Custom EMI filter doesn't involve the |
5/7
{10}
PHILIPS
| | | software of the system. Its primary purpose is to filter electrical powerline noise or poor electrical grounding from the hospital's end, ensuring a clean electrical source for the EPIQ/Affiniti Ultrasound systems. |
| --- | --- | --- |
## VI. Safety Considerations
The subject device comply with the referenced standards as well as the FDA ultrasound guidance document, Guidance for Industry and FDA Staff – Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, issued on February 2023.
## VII. Non-Clinical Performance Data
The proposed modification to the EPIQ and Affiniti Series Diagnostic Ultrasound Systems includes the integration of a Custom EMI Filter to reduce electromagnetic interference from external sources. Non-clinical testing was conducted to evaluate the impact of this modification on device performance and safety.
### Nonclinical Testing Summary (21 CFR 807.92(b)(1))
#### Purpose of Testing:
To evaluate the impact of the EMI filter on the system's safety, performance, and electromagnetic compatibility (EMC), ensuring continued compliance with applicable standards and substantial equivalence to the predicate device.
#### Standards Applied:ss
- IEC 60601-1-2:2014 + AMD1:2020 – General EMC requirements for medical electrical equipment.
- IEC 60601-2-37:2007 + AMD1:2015 – Particular requirements for ultrasonic diagnostic equipment.
#### Tests Conducted:
| Test Type | Description | Outcome |
| --- | --- | --- |
| Radiated Emissions | Evaluated emissions from the device | Passed |
| Conducted Emissions | Assessed conducted noise on power lines | Passed |
| Electrostatic Discharge (ESD) | Tested immunity to static electricity | Passed |
| Radiated RF Immunity | Assessed resistance to RF fields (420–470 MHz) | Passed |
| Voltage Interruptions | Simulated power fluctuations | Passed |
| Proximity Field Testing | Evaluated performance near RF sources | Passed |
#### Results Summary:
- All EMC tests were successfully passed.
The nonclinical testing confirms that the integration of the Custom EMI Filter does not adversely affect the safety or effectiveness of the device. The system continues to meet performance
6/7
{11}
PHILIPS
requirements and complies with applicable EMC standards. These results support the claim of substantial equivalence to the predicate device.
## VIII. Clinical Data
The proposed EPIQ and Affiniti Series Diagnostic Ultrasound System did not require clinical data for determination of substantial equivalence since substantial equivalence was demonstrated based on the following attributes:
- Design features
- Indications for use
- Fundamental scientific technology
- Non-clinical performance testing
- Safety and effectiveness
## IX. Conclusion
For testing, all pre-determined acceptance criteria were met. Results of these tests show that the proposed EPIQ and Affiniti Series Diagnostic Ultrasound System with Custom EMI filter meets the intended use. The proposed devices and predicate device:
- Are indicated for diagnostic ultrasonic imaging.
- Have identical intended users and use environments.
- Have the same device classification, product codes.
- Have identical software.
- Are introducing no new transducers.
The differences between the proposed device and predicate device do not raise new questions of safety and/or effectiveness. The major differences include:
The Custom EMI Filter Module reduces electromagnetic interference (EMI) from external sources, providing cleaner power to the EPIQ and Affiniti Series Diagnostic Ultrasound Systems. By filtering out noise from hospital power supplies, it helps prevent image distortions and artifacts in EPIQ and Affiniti Series Diagnostic Ultrasound Systems.
Therefore, the proposed EPIQ and Affiniti Series Diagnostic Ultrasound System with Custom EMI filter is substantially equivalent or enhanced to the predicate EPIQ and Affiniti Series Diagnostic Ultrasound Systems in terms of safety and effectiveness.
#
7/7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.