CareSuperb COVID-19/Flu A&B Antigen Combo Home Test
K251604 · Access Bio, Inc. · SCA · Aug 22, 2025 · Microbiology
Device Facts
Record ID
K251604
Device Name
CareSuperb COVID-19/Flu A&B Antigen Combo Home Test
Applicant
Access Bio, Inc.
Product Code
SCA · Microbiology
Decision Date
Aug 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3987
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CareSuperb™ COVID-19/Flu A&B Antigen Combo Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough, and/or shortness of breath, should seek follow up care from their healthcare provider. Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider for appropriate follow-up.
Device Story
Lateral flow immunochromatographic assay for qualitative detection of SARS-CoV-2 nucleocapsid, Influenza A nucleoprotein, and Influenza B nucleoprotein antigens. Input: anterior nasal swab specimen. Principle: adaptor-based lateral flow platform with conjugate wick filter; viral antigens bind to labeled detection antibodies; immune complexes migrate to nitrocellulose membrane; captured by immobilized antibodies forming visible colored lines (C19, A, B, and Control). Output: visual interpretation of colored lines on test cassette after 10 minutes. Used in home setting by lay users (self-test or adult-assisted). Results inform need for follow-up care or confirmatory molecular testing. Benefits: rapid, accessible respiratory virus screening.
Clinical Evidence
Prospective clinical study (N=1644 evaluable) compared the device to an FDA-cleared RT-PCR assay. SARS-CoV-2 PPA: 92.5% (86.4-96.0%), NPA: 99.6% (99.1-99.8%). Influenza A PPA: 85.6% (77.9-90.9%), NPA: 99.0% (98.4-99.4%). Influenza B PPA: 86.0% (72.7-93.4%), NPA: 99.7% (99.3-99.9%). Usability study (N=1795) showed 98.4% correct performance of critical tasks.
Technological Characteristics
Lateral flow immunochromatographic assay. Components: test cassette, extraction buffer, sterile nasal swab. Principle: capillary-driven antigen-antibody binding. Visual readout (colored lines). Storage: 2-30°C. Shelf-life: 13 months. Biocompatibility/sterility per ISO 10993 and ISO 11135.
Indications for Use
Indicated for symptomatic individuals aged 14+ (self-test) or 2+ (adult-assisted) with signs of respiratory infection. Negative results are presumptive; follow-up with molecular assay recommended if symptoms persist.
Regulatory Classification
Identification
A multi-analyte respiratory virus antigen detection test is an in vitro diagnostic device intended for the detection and/or differentiation of respiratory viruses directly from respiratory clinical specimens. The device is intended to be performed at the site of sample collection, does not involve sample storage and/or transport.
Predicate Devices
WELLlife COVID-19/Influenza A&B Home Test / WELLlife COVID-19/Influenza A&B Antigen Test (K243256)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K251604
B Applicant
Access Bio, Inc.
C Proprietary and Established Names
CareSuperb COVID-19/Flu A&B Antigen Combo Home Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| SCA | Class II | 21 CFR 866.3987 - Multi-Analyte Respiratory Virus Antigen Detection Test | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain 510(k) clearance for the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test
B Measurand:
Nucleoprotein antigen from Influenza A/B and nucleocapsid antigen from SARS-CoV-2.
C Type of Test:
Qualitative lateral flow immunoassay.
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K251604 - Page 2 of 28
## B Indication(s) for Use:
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigens directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough, and/or shortness of breath, should seek follow up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider for appropriate follow-up.
## C Special Conditions for Use Statement(s):
OTC - Over The Counter
## D Special Instrument Requirements:
Not applicable. There is no associated instrumentation as the test is visually.
## IV Device/System Characteristics:
### A Device Description:
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test is a rapid lateral flow immunochromatographic assay for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B antigens from anterior nasal swab specimens. Each kit includes:
- Single-use test cassette in a foil pouch with desiccant
- Assay buffer dropper vial (250 µL)
- Sterile anterior nasal swab
- Quick Reference Instructions (QRI)
The test device is packaged in a sealed aluminum foil pouch containing a single-use test cassette and a desiccant pack. Each cassette is labeled with "COVID-19/Flu" above the result window for identification. The plastic cassette housing contains an integrated adaptor that includes a silver-foiled cap and sample port. The conjugate wick filter within the adaptor is dispensed with labeled detecting antibodies. See Figure below.
{2}

## B Principle of Operation:
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of SARS-CoV-2 nucleocapsid antigen and influenza A and B nucleoprotein antigens in anterior nasal swab specimens.
To initiate testing, the user collects an anterior nasal specimen using the sterile nasal swab provided in the test kit. For individuals aged 14 years or older, the specimen samples are self-collected; for individuals younger than 14 years, an adult collects the sample. The swab is then inserted into the sample port of the test cassette and the extraction reagent in the dropper vial is added to the sample port for the sample extraction to occur exposing the viral antigens. The viral antigens present in the sample bind to their respective labeled antibodies dispensed in the conjugate wick filter. These antigen-antibody complexes migrate to the test strip encased in a plastic cassette and across the membrane through capillary action and are captured by either anti-SARS-CoV, anti-influenza A, or anti-influenza B antibodies causing color at the defined lines, respectively: "C19" (blue line), "A" (red line), and "B" (blue line). The control line incorporates anti-chicken IgY antibody and is designated as "Cont", causing a purple-colored line to appear on the membrane. The complexes are interpreted visually between 10 to 15 minutes based on the presence or absence of the test line.
## C Interpretation of Results:
### Positive Result
#### Influenza A Positive
One purple-colored line next to "Cont" and one red-colored line next to "A" indicates Influenza A positive result (Influenza A antigen detected)
#### Influenza B Positive
One purple-colored line next to "Cont" and one blue-colored line next to "B" indicates Influenza B positive result (Influenza B antigen detected)

K251604 - Page 3 of 28
{3}
K251604 - Page 4 of 28
# COVID-19 Positive
One purple-colored line next to "Cont" and one blue-colored line next to "C19" indicates COVID-19 positive result (COVID-19 antigen detected)

# Coinfection Positive Result
Any lines next to "A", "B", and/or "C19" with a control line next to "Cont" indicate Influenza A, B, and/or COVID-19 positive results. (Influenza A, B, and/or COVID-19 antigen detected)

# Negative Result
One purple-colored line ONLY next to "Cont" indicates a negative result, with no test line next to "A", "B", and "C19". This means Influenza A, B, and COVID-19 antigen has not been detected.

# Invalid Result
No purple-colored line next to "Cont" indicates an invalid result. A new test is needed to get a valid result.

# V Substantial Equivalence Information:
## A Predicate Device Name(s):
WELLlife COVID-19 / Influenza A&B Home Test; WELLlife COVID-19 / Influenza A&B AntigenTest
## B Predicate 510(k) Number(s):
K243256
## C Comparison with Predicate(s):
| Device & Predicate Device(s): | K251604 | K243256 |
| --- | --- | --- |
| Device Trade Name | CareSuperb COVID-19/Flu A&B Antigen Combo Home Test | WELLlife COVID-19/Influenza A&B Home Test / WELLlife COVID-19/Influenza A&B |
{4}
K251604 - Page 5 of 28
| | | Antigen Test |
| --- | --- | --- |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough, and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider for appropriate follow-up. | WELLlife COVID-19 / Influenza A&B Home Test:
The WELLlife COVID-19 / Influenza A&B Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.
Positive results do not rule out co-infection with other respiratory pathogens, and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.
WELLlife COVID-19 / Influenza A&B Antigen Test:
The WELLlife COVID-19 / Influenza A&B Antigen Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 |
{5}
K251604 - Page 6 of 28
| | | years or older testing themselves, or adults testing aged 2 years or older.
All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider.
Negative results do not rule out infection with influenza, SARS-CoV-2, or other pathogens.
Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare providers.
Positive results do not rule out co-infection with other respiratory pathogens.
Test results should not be used as the sole basis for treatment or other patient management decisions. |
| --- | --- | --- |
| Test Principle | Lateral flow immunoassay | Same |
| Assay Type | Qualitative | Same |
| Intended Use Population | Self-testing for symptomatic individuals aged 14 years or older and adults testing individuals aged 2 years and older within 4 days post symptom onset. | Same |
| Assay Target | Influenza A nucleoprotein antigen, Influenza B nucleoprotein antigen and/or SARS-CoV-2 nucleocapsid antigen | Same |
| Specimen Type | Direct anterior nasal swab specimen | Same |
| Result Interpretation | Visually read | Same |
| Control | Internal control | Same |
| Storage Condition | 2-30°C | Same |
| General Device Characteristic Differences | | |
| Development Time | 10-15 minutes | 10-20 minutes |
| Intended Users | Over-the-counter lay users | Over-the-counter lay users or professional use |
VI Standards/Guidance Documents Referenced:
| Document | Title | Publisher | Applicable Study |
| --- | --- | --- | --- |
| Special Controls under 21 CFR 866.3987 (multi-analyte respiratory virus antigen detection test) | Special controls under 21 CFR 866.3987 | FDA/CDRH | All Studies |
{6}
| Document | Title | Publisher | Applicable Study |
| --- | --- | --- | --- |
| ISO11135:2014 | Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices | ISO | Sterility |
| ISO 10993-7 | Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals | ISO | Sterility |
| 10993-10: Third Edition 2009-06-01 | Biological Evaluation of Medical Devices – Part 5: Tests for in vitro cytotoxicity | ISO | Biocompatibility |
| 10993-10: Third Edition 2010-08-01 | Biological Evaluation of Medical Devices – Part 10: Tests for irritation and skin sensitization | ISO | Biocompatibility |
| 10993-23: First Edition 2021-01 | Biological Evaluation of Medical Devices – Part 23: Tests for irritation | ISO | Biocompatibility |
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
The precision study for the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test was evaluated in two different studies conducted at single internal site each using three (3) lots of test kits and two (2) operators. Both studies are described below separately, and results are presented in the Table 1 together.
Study 1 was conducted using test samples prepared at three (3) different concentrations of heat inactivated SARS-CoV-2 B.1.1.529, live influenza A (Flu A): H1N1pdm09/A/Indiana/02/2020, and live influenza B (Flu B): Victoria/New Hampshire /01/2021 spiked into pooled negative swab matrix (PNSM) to prepare the following test sample panel members:
- Negative sample of each analyte
- Co-spike of each analyte (1xLoD)
- Co-spike of each analyte (3xLoD)
Samples were blinded and randomized before allotting them to the operators. 50μL of each sample was applied to dry nasal swabs and processed per the IFU of the candidate device. All panel members were tested in triplicate with 3 device lots, each in 2 runs per day for each of 2 operators, and the study was conducted for 10 days (i.e., 1 site x 3 lots x 2 operators x 2 runs per day with 3 replicates each x 10 days). 360 results were obtained for each panel member. All replicates prepared at 1xLoD demonstrated above 95% detection across the operators, lots, days and runs test. All replicates prepared at 3xLoD demonstrated 100% agreement with the expected results.
Study 2 was specifically conducted to further evaluate potential variability between lots and was performed using a negative sample without any of the analytes, and low positive samples
K251604 - Page 7 of 28
{7}
prepared at the $0.75\mathrm{x}$ LoD for each analyte (i.e., below the concentration tested in study 1 and near the analytes' C95 concentration):
Negative sample for each analyte
- Co-spike of each analyte as described in study 1 (0.75xLoD)
Two operators tested the samples above in a randomized and blinded manner each using three different lots of the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test in a total of 2 runs/operators and across 3 non-consecutive days (i.e., 3 replicates $\times$ 2 runs/day $\times$ 3 days $\times$ 2 operators $= 36$ sample replicates/lot).
The results for Study 2 demonstrated $100\%$ agreement for negative results and results for 0.75 LoD detection did not indicate significant variability between lots for each analytes tested.. The results from study 1 and study 2 are summarized below.
Table 1. Summary of precision study 1 and study 2
| Analyte concentration | Analyte | # of positive replicates/# of total replicates | | | Percent Positive | 95% CI |
| --- | --- | --- | --- | --- | --- | --- |
| | | Lot 1 | Lot 2 | Lot 3 | | |
| Negative | SARS-CoV-2 | 0/156 | 0/156 | 0/156 | 0% | 97.6-100% |
| | Flu A | 0/156 | 0/156 | 0/156 | 0% | 97.6-100% |
| | Flu B | 0/156 | 0/156 | 0/156 | 0% | 97.6100% |
| 0.75x LoD | SARS-CoV-2 | 32/36 | 29/36 | 31/36 | 85.2% | 77.2-90.1% |
| | Flu A | 32/36 | 31/36 | 33/36 | 88.9% | 81.6-93.5% |
| | Flu B | 32/36 | 32/36 | 34/36 | 90.7% | 83.8-94.9% |
| 1x LoD | SARS-CoV-2 | 120/120 | 120/120 | 117/120 | 99.2% | 97.6-99.7% |
| | Flu A | 120/120 | 120/120 | 117/120 | 99.2% | 97.6-99.7% |
| | Flu B | 120/120 | 120/120 | 119/120 | 99.7% | 98.4-100% |
| 3x LoD | SARS-CoV-2 | 120/120 | 120/120 | 120/120 | 100% | 98.9-100% |
| | Flu A | 120/120 | 120/120 | 120/120 | 100% | 98.9-100% |
| | Flu B | 120/120 | 120/120 | 120/120 | 100% | 98.9-100% |
# 2. Linearity:
Not Applicable - the device is a qualitative assay with binary visually-read results.
# 3. Analytical Specificity/Interference:
# Cross Reactivity and Microbial Interference:
Cross reactivity and microbial interference studies were conducted to determine if other respiratory pathogens/microbial flora that may be present in nasal swab samples could cause a false positive test result or interfere with the detection of a true positive result and cause a false negative result. A panel of viruses, bacteria, fungi, and pooled nasal wash were used for these studies.
For the cross-reactivity study, dilutions of the organisms were prepared in pooled nasal swab matrix (PNSM) and tested in triplicates in the absence of SARS-CoV-2, influenza A, and
K251604 - Page 8 of 28
{8}
influenza B. No cross-reactivity was observed with the organisms tested for any of the 3 analytes (Table 2).
For the microbial interference study, dilutions of the panel organisms were prepared in PNSM, in the presence of low levels (3x LoD) of heat inactive SARS-CoV-2 (SARS-CoV-2 B.1.1.529, omicron), live influenza A (H1N1; A/Indiana/02/2020), and live influenza B (Victoria; B/New Hampshire/01/2021) and tested in triplicate. No microbial interference was observed for any of the 3 analytes tested (Table 2).
Table 2. Cross-reactivity and microbial interference study results
| Microorganism | Conc. | Analyte | Test Results | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Cross reactivity | | Microbial interference | |
| | | | n/N1 | Result Agreement (%)2 | n/N1 | Result Agreement (%)2 |
| Adenovirus 1 | 1.58 X 105CID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Adenovirus 7 | 1.6 X 105TCID50/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Enterovirus71 | 1.6 X 106TCID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Human coronavirus (OC43) | 1.4 X 105TCID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Human coronavirus (229E) | 1.4 X 104TCID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Human coronavirus (NL63) | 8 X 104TCID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Human metapneumovirus (hMPV) | 2.8 X 105TCID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| MERS-Coronavirus, Irradiated Lysate | 1.78 X 105CID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Parainfluenza virus type 1 | 7.4 X 105TCID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Parainfluenza virus type 2 | 1.58 X 105CID50/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| | 1.58 X 105 | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
K251604 - Page 9 of 28
{9}
K251604 - Page 10 of 28
| Parainfluenza virus type 3 | CID_{50}/mL | Flu A | 0/3 | 100% | 3/3 | 100% |
| --- | --- | --- | --- | --- | --- | --- |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Parainfluenza virus type 4 | 5.0 X 10^{5} TCID_{50}/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Respiratory Syncytial virus | 1.58 X 10^{5} CID_{50}/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Rhinovirus | 1.58 X 10^{5} CID_{50}/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| SARS-Coronavirus | 1.0 X 10^{5} PFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Human coronavirus HKU1^{+} (clinical specimen) | 0.1X Stock | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Bordetella pertussis | 1.0 X 10^{7} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Chlamydophila pneumoniae | 1.4 X 10^{6} IFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Haemophilus influenzae | 1.6 X 10^{6} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Legionella pneumophila | 1.35 X 10^{6} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Mycoplasma pneumoniae | 6.5 X 10^{5} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Streptococcus pneumoniae | 1.0 X 10^{7} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Streptococcus pyogenes | 1.0 X 10^{7} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Staphylococcus aureus | 1.0 X 10^{7} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Staphylococcus epidermidis | 1.0 X 10^{7} CFU/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Candida albicans | 1.0 X 10^{7} | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
{10}
K251604 - Page 11 of 28
| | CFU/ml | Flu A | 0/3 | 100% | 3/3 | 100% |
| --- | --- | --- | --- | --- | --- | --- |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Pooled human nasal wash | N/A | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
¹ # of positive results/# of replicates.
² Agreement with the expected result.
* 10-fold dilution of the stock Human Coronavirus HKU1 (HCoV-HKU1) clinical sample was tested in triplicates in the presence and absence of SARS-CoV-2. The Ct value of the undiluted sample was 11.7.
## Competitive Interference:
A competitive inhibition study was conducted to evaluate the potential for a high concentration of one target analyte to interfere with the detection of another target analyte at low concentration. Testing was performed in triplicate with different combinations of low (3x LoD) and high concentrations (≥ 10⁵ CEID₅₀/mL or TCID₅₀/mL) of SARS-CoV-2, influenza A, and influenza B. The study used inactivated SARS-CoV-2 but live influenza A and B virus strains. There was no competitive interference observed.
Table 3: Competitive inhibition study results
| Testing Panel | Viral targets in sample | | | Results (# pos / total reps) | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Influenza A | Influenza B | SARS-CoV-2 | Influenza A | Influenza B | SARS-CoV-2 |
| 1 | High | Low | Negative | 3/3 | 3/3 | 0/3 |
| 2 | High | Negative | Low | 3/3 | 0/3 | 3/3 |
| 3 | High | Low | Low | 3/3 | 3/3 | 3/3 |
| 4 | Low | High | Negative | 3/3 | 3/3 | 0/3 |
| 5 | Negative | High | Low | 0/3 | 3/3 | 3/3 |
| 6 | Low | High | Low | 3/3 | 3/3 | 3/3 |
| 7 | Low | Negative | High | 3/3 | 0/3 | 3/3 |
| 8 | Negative | Low | High | 0/3 | 3/3 | 3/3 |
| 9 | Low | Low | High | 3/3 | 3/3 | 3/3 |
| 10 | High | High | Low | 3/3 | 3/3 | 3/3 |
| 11 | High | Low | High | 3/3 | 3/3 | 3/3 |
| 12 | Low | High | High | 3/3 | 3/3 | 3/3 |
| 13 | High | High | Negative | 3/3 | 3/3 | 0/3 |
| 14 | High | Negative | High | 3/3 | 0/3 | 3/3 |
| 15 | Negative | High | High | 0/3 | 3/3 | 3/3 |
| 16 | High | High | High | 3/3 | 3/3 | 3/3 |
| 17 | High | Negative | Negative | 3/3 | 0/3 | 0/3 |
| 18 | Negative | High | Negative | 0/3 | 3/3 | 0/3 |
| 19 | Negative | Negative | High | 0/3 | 0/3 | 3/3 |
## Endogenous/Exogenous Substances Interference:
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test was evaluated for performance in the presence of potentially interfering substances that might be present in respiratory specimens. Potentially interfering substances were prepared and diluted in PNSM to the recommended concentration. Virus negative PNSM specimens were evaluated in triplicate to confirm that the potentially interfering substances were not cross-reactive with the test. Positive samples were prepared in PNSM at 3x LoD using each analyte individually and in combination together (co-spiked) to confirm that these substances do not interfere
{11}
with detection of SARS-CoV-2, influenza A, and influenza B. No cross-reactivity or interference was observed among the substances tested.
Table 4. Interfering substances study results
| Microorganism | Conc. | Analyte | Test Results | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | No analyte | | With Analyte | |
| | | | n/N1 | Result Agreement (%)2 | n/N1 | Result Agreement (%)2 |
| Acetaminophen | 10 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Acetyl salicylic acid | 15 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Beclomethasone | 5 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Benzocaine | 5 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Budesonide | 2 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Chlorpheniramine maleate | 5 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Dexamethasone | 1 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Dextromethorphan HBr | 2 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Diphenhydramine HCl | 5 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Flunisolide | 5 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Fluticasone | 1 mg/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Guaiacol Glyceryl Ether | 20 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Histamine Dihydrochloride | 10 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
K251604 - Page 12 of 28
{12}
K251604 - Page 13 of 28
| Menthol | 10 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| --- | --- | --- | --- | --- | --- | --- |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Mometasone | 1 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Molnupiravir | 1mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Mucin (Bovine submaxillary glands) | 2.5 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Mupirocin | 1 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Phenylpropanol-amine | 5 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Remdesivir | 1 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Tamiflu | 5 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Tobramycin | 1 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Trimcinolone | 1 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Zanamivir | 1 mg/ml | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Sore throat spray (Phenol) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Zicam Oral Mist (Zincum aceticum, Zincum gluconicum) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Nasal Spray (Phenylephrine HCl) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
{13}
K251604 - Page 14 of 28
| NasalCrom Nasal Spray (Cromoly sodium) | | Flu B | 0/3 | 100% | 3/3 | 100% |
| --- | --- | --- | --- | --- | --- | --- |
| Vicks Sinex Nasal (Oxymspertaayzoline HCl) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Alkalol AllergyRelief (Galphimiaglauca, Luffa operculate, Sabadilla) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Zicam Allergy Relief (Galphimia glauca, Histaminum hydrochloricum, Luffa operculate, Sulphur) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Hand sanitizer | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Hand soap | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Whole blood | 2.5% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Buffy coat | 2.5% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Allergy spray (Mometasone Furoate) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Budesonide Nasal Spray (Budesonide/Gluco corticoid) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Nasacort Allergy 24HR (Triamcinolone acetonide) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Allergy Relief Nasal Spray (Fluticasone Propionate) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Saline Nasal Spray (Sodium chloride and preservatives) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Alkalol Saline Nasal Spray | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
{14}
K251604 - Page 15 of 28
| (Sodium chloride, Menthol, Thymol, Camphor, Benzoin, Resin Extract, Oils of Eucalyptus, Wintergreen, Spearmint, fir needle, Cinnamon & preservatives) | | Flu A | 0/3 | 100% | 3/3 | 100% |
| --- | --- | --- | --- | --- | --- | --- |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Leukocytes | 5.0 X 10^{6} cells/mL | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
| Zinc Throat Spray (Thera Zinc Throat spray) | 15% v/v | SARS-CoV-2 | 0/3 | 100% | 3/3 | 100% |
| | | Flu A | 0/3 | 100% | 3/3 | 100% |
| | | Flu B | 0/3 | 100% | 3/3 | 100% |
¹ # of positive results/# of replicates.
² Agreement with the expected result.
## Biotin Interference:
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test uses streptavidin/biotin technology for antibody immobilization for the detection of influenza A. A study on biotin interference was performed to evaluate the effect of high biotin (vitamin B7) levels on test results. Biotin serial dilutions (5,000 to 0 ng/mL) were assessed in triplicate with both analyte-negative and analyte-positive samples (3x LoD co-spiked SARS-CoV-2, Influenza A, and Influenza B). There were no false positives detected at any concentration for any of the three (3) analytes. However, false negative results were detected for influenza A at concentrations of 3,750 ng/mL and 5,000 ng/mL.
Table 5. Biotin interference study results
| Biotin Concentration (ng/mL) | # of Positive / # of Replicates | | | Negative (Analyte-absent) |
| --- | --- | --- | --- | --- |
| | SARS-CoV-2 | Influenza A | Influenza B | |
| 5,000 | 3/3 | 0/3 | 3/3 | 0/3 |
| 3,750 | 3/3 | 0/3 | 3/3 | 0/3 |
| 2,500 | 3/3 | 3/3 | 3/3 | 0/3 |
| 1,250 | 3/3 | 3/3 | 3/3 | 0/3 |
| 625 | 3/3 | 3/3 | 3/3 | 0/3 |
| 312.5 | 3/3 | 3/3 | 3/3 | 0/3 |
| 0 | 3/3 | 3/3 | 3/3 | 0/3 |
## 4. Assay Reportable Range:
Not applicable - the device is a binary qualitative assay that is visually read.
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Controls
a. Internal Controls:
{15}
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test contains a built-in internal procedural control that is included in the test cassette. The internal control is part of the test strip membrane and is therefore, automatically run within the development time of each test. A purple-colored line appearing in the control region “Cont” is designed as an internal procedural control. The appearance of the procedural control line indicates that sufficient flow has occurred, and functional integrity of the test cassette has been maintained. If the procedural control line does not develop in 10 minutes, the test result is considered invalid and retesting with a new device is recommended.
b. External Controls:
Not applicable. This test is intended for non-prescription over the counter use.
**Stability**
a. Real Time Stability:
A real-time stability study was conducted to evaluate stability and determine the shelf-life of the unopened kit. To validate shelf life for 2-30°C storage, three unopened CareSuperb COVID-19/Flu A&B Antigen Combo Home Test kit lots were stored at 2°C and 30 °C. At defined intervals, an assessment of each lot was conducted with the following panel of test samples: negative clinical matrix, and co-spiked positive samples with inactivated SARS-CoV-2, live Flu A and/or Flu B viruses, each at 4x LoD.
Fifty (50) µL of each sample were applied to the swab and tested according to the Instructions for Use. Five replicates of each sample were tested at each time point.
Baseline testing was performed within one month of each manufactured lot. Subsequent testing was performed each month after baseline and up to 15 months. At the time of clearance, all study data have met the protocol defined acceptance criteria, and support storage of the test kits at 2-30°C for up to 13 months.
b. Shipping Stability:
Transport stability under simulated summer and winter shipping conditions was tested to evaluate worst-case shipping and handling conditions. Unopened test kits were stored for each condition for 10 days. Performance of unopened test kits was assessed by testing at baseline (T0) and every day post-distribution for up to 10 days, using the same sample panel described for the real-time stability study (above). Samples were tested in replicates of five for each of three device lots. Candidate test kits were stored at the designated temperature profiles described below and then tested with the test panel to evaluate performance. The following temperature profiles were assessed:
- To mimic summer shipping conditions, test kits were incubated in a chamber set to 60°C/85% RH for 10 days.
- To mimic winter shipping conditions, test kits were incubated at -20°C for 10 days.
All results were as expected for all time points and support shipping at high and low temperatures experienced around the USA during summer and winter months, respectively.
K251604 - Page 16 of 28
{16}
6. Detection Limit:
**Single Analyte Limit of Detection (LoD):**
A limit of detection (LoD) study was conducted to determine the lowest detectable concentration of two strains of SARS-CoV-2 (USA-WA1/2020, and B.1.1.529 omicron; both heat-inactivated), five strains of live influenza A (3 strains of H1N1 and 2 strains of H3N2), and four strains of live influenza B (2 strains of Victoria and 2 strains of Yamagata) at which at least 95% of all true positive replicates return a positive result. Testing was conducted on three lots of test devices.
A preliminary LoD was first determined by testing serial 10-fold dilutions of virus stocks diluted in PNSM in five replicates per device lot for a total of 15 replicates per dilution. A 50-μL sample of each virus diluted in PNSM was pipetted onto the dry swab. The swab was then tested per the instructions for use. The preliminary LoD of each virus was confirmed by testing an additional twenty samples/lot for each viral stock at the preliminary LoD concentration as well as 10-fold above and below. The last concentrations that produced 20-positive test results/lot from a 1:10 dilution was further evaluated using a 3-fold dilution series, in 20 replicates/lot for each level, to refine the LoD.
As per the acceptance criteria for confirmation of the LoD, at least 95% of the replicates (≥ 19/20) should be positive to be considered as the confirmed LoD. The confirmed LoDs observed were identical for the three lots tested for each virus strain. The results of the last two 10-fold dilution of preliminary LoD and all confirmatory LoD at 10-fold and 3-fold dilutions for each strain are included below. Final stated LoD for each virus is presented in the bold in the tables 5, 6 and 7.
Table 6. SARS-CoV-2 single analyte LoD results
| Virus Strain | Virus conc. | | Positive results/total replicates | |
| --- | --- | --- | --- | --- |
| | TCID_{50}/mL | TCID_{50}/Swab | Preliminary LoD | Confirmatory LoD |
| USA-WA1/2020 | 7.9 x 10^{3} | 3.9 x 10^{2} | 15/15 | 60/60 |
| | 7.9 x 10^{2} | 3.9 x 10^{1} | 15/15 | 60/60 |
| | 2.6 x 10^{2} | 1.3 x 10^{1} | - | 60/60 |
| | 8.8 x 10^{1} | 4.4 x 10^{0} | - | 20/60 |
| B.1.1.529, Omicron | 1.5 x 10^{3} | 7.5 x 10^{1} | 15/15 | 60/60 |
| | 1.5 x 10^{2} | 7.5 x 10^{0} | 15/15 | 60/60 |
| | 5 x 10^{1} | 2.5 x 10^{0} | - | 8/60 |
Table 7. Influenza A single analyte LoD results
| Virus strain | Virus concentration | | Positive results/total replicates | |
| --- | --- | --- | --- | --- |
| | CEID_{50}/mL | CEID_{50}/Swab | Preliminary LoD | Confirmatory LoD |
| A/Indiana/02/2020 (H1N1) pdm09 | 1.6 x 10^{8} | 8.0 x 10^{6} | 15/15 | 60/60 |
| | 1.6 x 10^{7} | 8.0 x 10^{5} | 15/15 | 60/60 |
| | 5.3 x 10^{6} | 2.6 x 10^{5} | - | 0/60 |
| A/Hawaii/66.2019 (H1N1) pdm09 | 7.4 x 10^{7} | 3.7 x 10^{6} | 15/15 | 60/60 |
| | 7.4 x 10^{6} | 3.7 x 10^{5} | 15/15 | 60/60 |
| | 2.4 x 10^{6} | 1.2 x 10^{5} | - | 0/60 |
| A/Brisbane/59/2007 (H1N1) | 9.7 x 10^{6} | 4.8 x 10^{5} | 15/15 | 60/60 |
| | 9.7 x 10^{5} | 4.8 x 10^{4} | 15/15 | 60/60 |
| | 3.2 x 10^{5} | 1.6 x 10^{4} | - | 0/60 |
| Virus strain | Virus concentration | | Positive results/total replicates | |
| | FFU/mL | FFU/Swab | CEIDD_{50}/mL | CEID_{50}/Swab |
| A/Hawaii/66.2019 (H1N1) pdm09 | 1.6 x 10^{8} | 8.0 x 10^{6} | 15/15 | 60/60 |
| | 1.6 x 10^{7} | 8.0 x 10^{5} | 15/15 | 60/60 |
| | 5.3 x 10^{6} | 2.6 x 10^{5} | - | 0/60 |
K251604 - Page 17 of 28
{17}
| Virus strain | Virus concentration | | Positive results/total replicates | |
| --- | --- | --- | --- | --- |
| | CEID50/mL | CEID50/Swab | Preliminary LoD | Confirmatory LoD |
| A/California/55/2020 (H3N2) | 3.5 x 10^6 | 1.7 x 10^5 | 15/15 | 60/60 |
| | 3.5 x 10^5 | 1.7 x 10^4 | 15/15 | 60/60 |
| | 1.1 x 10^5 | 5.8 x 10^3 | - | 60/60 |
| | 3.8 x 10^4 | 1.9 x 10^3 | - | 3/60 |
| A/Delaware/01/2021 (H3N2) | 3.1 x 10^5 | 1.5 x 10^4 | 15/15 | 60/60 |
| | 3.1 x 10^4 | 1.5 x 10^3 | 15/15 | 60/60 |
| | 1.0 x 10^4 | 5.1 x 10^2 | - | 0/60 |
Table 8. Influenza B single analyte LoD results
| Virus strain | Virus concentration | | Positive results/total replicates | |
| --- | --- | --- | --- | --- |
| | TCID50/mL | TCID50/Swab | Preliminary LoD | Confirmatory LoD |
| B/New Hampshire/01/2021 (Victoria) | 1.3 x 10^4 | 6.5 x 10^2 | 15/15 | 60/60 |
| | 1.3 x 10^3 | 6.5 x 10^1 | 15/15 | 60/60 |
| | 4.3 x 10^2 | 2.1 x 10^1 | - | 60/60 |
| | 1.4 x 10^2 | 7.2 x 10 | - | 3/60 |
| B/Michigan/01/2021 (Victoria) | 5.7 x 10^4 | 2.8 x 10^3 | 15/15 | 60/60 |
| | 5.7 x 10^3 | 2.8 x 10^2 | 15/15 | 60/60 |
| | 1.9 x 10^3 | 9.5 x 10^1 | - | 60/60 |
| | 6.3 x 10^2 | 3.1 x 10^1 | - | 3/60 |
| Virus strain | Virus concentration | | Positive results/total replicates | |
| | TCID50/mL | TCID50/Swab | CEIDD50/mL | CEID50/Swab |
| B/Oklohoma/10/2018 (Yamagata) | 7.6 x 10^5 | 3.8 x 10^4 | 15/15 | 60/60 |
| | 7.6 x 10^4 | 3.8 x 10^3 | 15/15 | 60/60 |
| | 2.5 x 10^4 | 1.2 x 10^3 | - | 60/60 |
| B/Indiana/17/2017 (Yamagata) | 1.0 x 10^6 | 5 x 10^4 | 15/15 | 60/60 |
| | 1.01 x 10^5 | 5 x 10^3 | 15/15 | 60/60 |
| | 3.3 x 10^4 | 1.6 x 10^2 | - | 60/60 |
| | 1.1 x 10^4 | 5.5 x 10^2 | - | 60/60 |
## Co-spiked multi-analyte LoD:
After single analyte LoDs were determined, co-spiked equivalency testing was conducted to characterize the performance of samples that contained all analytes at their respective 1x LoD concentrations. Based on individual analyte LoDs, 3-fold above and below and at 1x LoD of the single analyte was evaluated in co-spiked samples prepared by mixing viruses (one each of SARS-CoV-2, Flu A, and Flu B) in PNSM. Two panels were tested for the co-spike study as shown in the table below.
Overall, 20 replicates per lot were evaluated at each concentration for both panels by pipetting 50 µL of co-spiked sample at defined LoD concentration onto the dry swab and testing swabs with the device according to the instructions for use. Equivalency is confirmed separately for each analyte if ≥95% replicates are positive within 2x LoD of the individually tested analyte.
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test demonstrated co-spike equivalency for SARS-CoV-2, Influenza A and Influenza B at 1x single analyte LoD. This study supports the use of co-spiked samples in subsequent analytical studies.
K251604 - Page 18 of 28
{18}
Table 9. Summary of Co-spike equivalency LoD study results
| Fold LoD | Analytes | Conc. (Titer Unit/mL) | Conc. (Titer Unit/swab) | Test Results # of positives/# of replicates | Positive result agreement (%) |
| --- | --- | --- | --- | --- | --- |
| Panel 1 | | | | | |
| 3x | SARS-CoV-2 (B.1.1.529, omicron) | 4.5 x 102 | 2.2 x 101 | 60/60 | 100% |
| | Flu A (Indiana/02/2020 (H1N1) pdm09) | 2.9 x 106 | 1.4 x 105 | 60/60 | 100% |
| | Flu B Victoria (New Hampshire/01/2021) | 1.3 x 103 | 6.5 x 101 | 60/60 | 100% |
| 1x | SARS-CoV-2 (B.1.1.529, omicron) | 1.5 x 102 | 7.5 x 100 | 59/60 | 98.3% |
| | Flu A (Indiana/02/2020 (H1N1) pdm09) | 9.7 x 105 | 4.8 x 104 | 58/60 | 96.6% |
| | Flu B Victoria (New Hampshire/01/2021) | 4.3 x 102 | 2.1 x 101 | 59/60 | 98.3% |
| 1/3x | SARS-CoV-2 (B.1.1.529, omicron) | 5 x 100 | 2.5x 100 | 0/60 | 0% |
| | Flu A (Indiana/02/2020 (H1N1) pdm09) | 3.2 x 105 | 1.6 x104 | 0/60 | 0% |
| | Flu B Victoria (New Hampshire/01/2021) | 1.5 x 102 | 7.2 x100 | 0/60 | 0% |
| Panel 2 | | | | | |
| 3x | SARS-CoV-2 (B.1.1.529, omicron) | 4.5 x 102 | 2.2 x 101 | 60/60 | 100% |
| | Flu A (Hawaii/66.2019 (H1N1) pdm09) | 2.2 x 107 | 1.1 x 106 | 60/60 | 100% |
| | Flu B Victoria (Michigan/01/2021) | 5.7 x 103 | 2.8 x 102 | 60/60 | 100% |
| 1x | SARS-CoV-2 (B.1.1.529, omicron) | 1.5 x 102 | 7.5 x 100 | 59/60 | 98.3% |
| | Flu A (Hawaii/66.2019 (H1N1) pdm09) | 7.4 x 106 | 3.7 x 105 | 58/60 | 96.6% |
| | Flu B Victoria (Michigan/01/2021) | 1.9 x 103 | 9.5 x 101 | 59/60 | 98.3% |
| 1/3x | SARS-CoV-2 (B.1.1.529, omicron) | 5 x 100 | 2.5x 100 | 0/60 | 0% |
K251604 - Page 19 of 28
{19}
K251604 - Page 20 of 28
7. International Standard Material NIBSC code: 21/368 – Limit of Detection:
The LoD of the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test was also determined by evaluating different dilutions of the International Standard for SARS-CoV-2 antigen (NIBSC code: 32/368) in negative pooled nasal swab matrix. The International Standard for SARS-CoV-2 containing lyophilized SARS-CoV-2 antigen was reconstituted in ultra-pure water (for a final concentration of 20,000 IU/mL). The LoD was determined as the lowest virus concentration that was detected ≥95% of the time (i.e., concentration at which at least 19/20 replicates tested positive).
Five-fold serial dilutions were made from the International Standard for SARS-CoV-2 antigen into negative clinical matrix (pooled nasal swab matrix). Three (3) replicates were tested on one (1) lot of the test device for each dilution to determine the preliminary LoD concentration of the device. The lowest concentration with all concordant positive results was considered the preliminary LoD. For each replicate, 50 μL of virus dilution was applied to a swab and the swab was processed according to the IFU.
The preliminary LoD concentration was tested with an additional 20 replicates to confirm the LoD. Concentrations above and below the preliminary LoD were also tested with 20 replicates to further refine the LoD. Samples were prepared as for the preliminary LoD study above. To confirm the LoD, at least 19 of 20 replicates should be positive per lot. The results are summarized in Table below.
Table 10. Summary of LoD study for the international standard
| Conc. (IU/mL) | Conc. on dry swab (IU/swab) | # Positive replicates | |
| --- | --- | --- | --- |
| | | Preliminary LoD | Confirmatory LoD |
| 4000 | 200 | 3/3 | - |
| 800 | 40 | 3/3 | 20/20 |
| 160 | 8 | 3/3 | 19/20 |
| 32 | 1.6 | 0/3 | 0/20 |
| 6.4 | 0.32 | 0/3 | - |
| 1.28 | 0.064 | 0/3 | - |
The LoD for the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test using the 1st International Standard for SARS-CoV-2 antigen (NIBSC code: 21/368) in nasal matrix was determined to be 160 IU/mL (8 IU/swab).
8. High Dose Hook Effect:
A high-dose hook effect study was conducted to evaluate whether high levels of any of the target analytes in a sample could result in a false-negative test result. Individually spiked samples were prepared; where each sample was prepared at a high, or "stock," viral concentration with the following analytes: SARS-CoV-2 (heat-inactivated), influenza A
{20}
(live), and influenza B (live) viruses. Each sample was tested across a series of two-fold dilutions up to 3 dilutions. Fifty $(50)\mu \mathrm{L}$ sample was spiked onto swabs and swabs were processed in accordance with the instructions for use. Analytes were tested co-spiked, and all samples were tested in 5-replicates. No evidence of a high-dose hook effect was observed with the virus stocks and concentrations tested.
Table 11. High-dose hook effect study results
| Virus Strain | Testing concentration | Test results (# of positives/ # total replicates) | | |
| --- | --- | --- | --- | --- |
| | | SARS-CoV-2 | Influenza A | Influenza B |
| SARS-CoV-2 | | | | |
| USA-WA1/2020 | 3.9 x 105TCID50/mL | 5/5 | 0/5 | 0/5 |
| | 1.9 x 105TCID50/mL | 5/5 | 0/5 | 0/5 |
| | 9.8 x 104TCID50/mL | 5/5 | 0/5 | 0/5 |
| B.1.1.529, Omicron | 7.5 x 105TCID50/mL | 5/5 | 0/5 | 0/5 |
| | 3.7 x 105TCID50/mL | 5/5 | 0/5 | 0/5 |
| | 1.8 x 105TCID50/mL | 5/5 | 0/5 | 0/5 |
| Influenza A | | | | |
| A/Indiana/02/2020 (H1N1)pdm09 | 4.8 x 108CEID50/mL | 0/5 | 5/5 | 0/5 |
| | 2.4 x 108CEID50/mL | 0/5 | 5/5 | 0/5 |
| | 1.2 x 108CEID50/mL | 0/5 | 5/5 | 0/5 |
| A/Delaware/01/2021(H3N2) | 1.5 x 107FFU/mL | 0/5 | 5/5 | 0/5 |
| | 7.7 x 106FFU/mL | 0/5 | 5/5 | 0/5 |
| | 3.8 x 106FFU/mL | 0/5 | 5/5 | 0/5 |
| Influenza B | | | | |
| B/New Hampshire/01/2021(Victoria) | 6.5 x 105TCID50/mL | 0/5 | 0/5 | 5/5 |
| | 3.2 x 105TCID50/mL | 0/5 | 0/5 | 5/5 |
| | 1.6 x 105TCID50/mL | 0/5 | 0/5 | 5/5 |
| B/Indiana/17/2017(Yamagata) | 5.0 x 107TCID50/mL | 0/5 | 0/5 | 5/5 |
| | 2.5 x 107TCID50/mL | 0/5 | 0/5 | 5/5 |
| | 1.2 x 107TCID50/mL | 0/5 | 0/5 | 5/5 |
# 9. Inclusivity:
Analytical reactivity testing for the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test was performed to ensure that the device can adequately detect a variety of strains for the SARS-CoV-2, influenza A, and influenza B viruses. A selection of temporally, geographically, and genetically diverse SARS-CoV-2 and influenza strains were tested for inclusivity, including 21 Influenza A strains (10 H1N1 and 10 H3N2, and 1 H5N1), 10 Influenza B strains (5 Yamagata and 5 Victoria lineages), and 5 SARS-CoV-2 strains. A series of ten-fold dilutions of each virus strain was spiked into PNSM and tested to determine an approximate LoD of the test for each virus. The lowest concentration with $100\%$ positive replicates was identified and additional 2-fold dilutions above and below that approximate LoD were tested to demonstrate inclusivity. Based on the dilution series, the minimum detectable concentration was defined as the lowest concentration for which all five (5) replicates were detected. Results are summarized below and demonstrate that the test tests can detect the analytes across a range of viral strains.
Table 12. Inclusivity results
| Strain name | Titer Unit/mL | Titer unit/Swab |
| --- | --- | --- |
K251604 - Page 21 of 28
{21}
| SARS-CoV-2 | | |
| --- | --- | --- |
| BA2.12.1, Omicron | 6.3 x 102TCID50/mL | 3.1 x 101TCID50/swab |
| BA. 2.3, Omicron | 1.1 x 102TCID50/mL | 5.8 x 100TCID50/swab |
| BA.2.75.5, Omicron | 8.5 x 101TCID50/mL | 4.2 x 100TCID50/swab |
| BA.4.6, Omicron | 5.7 x 102TCID50/mL | 2.8 x 101TCID50/swab |
| JN.1.4 Omicron | 2.1 x 102TCID50/mL | 1.0 x 101TCID50/swab |
| Influenza A | | |
| Wisconsin/588/2019 (H1N1)pdm09 | 1.4 x 103FFU/ml | 7.0 x 101FFU/Swab |
| Dominican Republic/7293/2013 (H1N1)pdm09 | 1.2 x 103TCID50/mL | 6.2 x 101TCID50/Swab |
| Massachusetts/15/2013 (H1N1)pdm09 | 8.0 x 105CEID50/mL | 4.0 x 104CEID50/Swab |
| Bangladesh/3002/2015 (H1N1)pdm09 | 1.3 x 104CEID50/mL | 6.5 x 102CEID50/Swab |
| Michigan/45/2015 (H1N1)pdm09 | 7.8 x 102CEID50/mL | 3.9 x 101TCID50/Swab |
| Iowa/53/2015 (H1N1)pdm09 | 1.4 x 106CEID50/mL | 7.2 x 104CEID50/Swab |
| St. Petersburg/61/2015 (H1N1pdm09) | 4.6 x 105CEID50/mL | 2.3 x 104CEID50/Swab |
| Hong Kong/H090-761-V1(0)/2009 (H1N1)pdm09 | 4.0 x 102TCID50/mL | 2.0 x 101TCID50/Swab |
| Victoria/4897/2022 (H1N1)pdm09 | 1.0 x 106.5EID50/mL | 5.0 x 104.5EID50/Swab |
| Victoria/2570/2019 (H1N1)pdm09 | 1.0 x 105.3EID50/mL | 5.0 x 103.3EID50/Swab |
| New York/21/2020 (H3N2) | 1.30 x 105FFU/ml | 6.5 x 105FFU/Swab |
| Michigan/173/2020 (H3N2) | 1.9 x 105FFU/ml | 9.7 x 105FFU/Swab |
| Tasmania/503/2020 (H3N2) | 6.5 x 104FFU/ml | 3.2 x 103FFU/Swab |
| Texas/50/2012 (H3N2) | 8.7 x 103TCID50/mL | 4.3 x 102TCID50/Swab |
| Switzerland/9715293/2013 (H3N2) | 6.0 x 105CEID50/mL | 3.0 x 104CEID50/Swab |
| Hong Kong/4801/2014 (H3N2) | 9.6 x 105CEID50/mL | 4.8 x 104CEID50/Swab |
| Singapore/INFIMH-16-0019/2016 (H3N2) | 5.5 x 104CEID50/mL | 2.7 x 103CEID50/Swab |
| Perth/16/2009 (H3N2) | 1.1 x 105CEID50/mL | 5.5 x 103CEID50/Swab |
| Darwin/9/2021 (H3N2) | 1.0 x 104.3EID50/ml | 5.0 x 102.3EID50/Swab |
| Georgia/02/2022 (H3N2) | 5.0 x 105.5EID50/mL | 2.5 x 104.5EID50/Swab |
| bovine/Ohio/824OSU-439/2024 (H5N1) | 3.1 x 102TCID50/mL | 1.5 x 101TCID50/Swab |
| Influenza B | | |
| Texas/43/2019 (Victoria Lineage) | 5.0 x 102TCID50/mL | 2.5 x 101TCID50/Swab |
| Washington/02/2019 (Victoria Lineage) | 2.1 x 106CEID50/ml | 1.0 x 105CEID50/Swab |
| Brisbane/60/2008 (Victoria Lineage) | 5 x 103CEID50/mL | 2.5 x 102CEID50/Swab |
| Austria/1359417/2021 (Victoria Lineage) | 5.0 x 104.5EID50/mL | 2.5 x 103.5EID50/Swab |
| Netherlands/10894/2022 (Victoria Lineage) | 1.0 x 104.7EID50/mL | 5.0 x 102.7EID50/Swab |
| Phuket/3073/2013 (Yamagata Lineage) | 2.7 x 104CEID50/mL | 5.5x 102CEID50/Swab |
| Wisconsin/10/2016 (NA 1221V) (Yamagata Lineage) | 3.2 x 105TCID50/mL | 1.6 x 104TCID50/Swab |
| Texas/06/2011 (Yamagata Lineage) | 1.6 x 105CEID50/mL | 8.0 x 103CEID50/Swab |
| Phuket/3073/2013 (Yamagata Lineage) | 1.0 x 103.8EID50/mL | 5.0 x 101.8EID50/Swab |
| Norway/2134/2019 (Yamagata Lineage) | 2.5 x 104.5EID50/mL | 1.2 x 103.5EID50/Swab |
# 10. Assay Cut-Off:
N/A as this is a qualitative visually read assay without numeric raw data.
K251604 - Page 22 of 28
{22}
# B Comparison Studies:
1. Method Comparison with Predicate Device:
See Section C (Clinical Studies) below.
2. Matrix Comparison:
The CareSuperb COVID-19/Flu A&B Antigen Combo Home Test is only intended for use with direct anterior nasal swab specimens. As no other sample types are claimed for use with the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test, a matrix comparison study is not applicable.
# C Clinical Studies:
1. A prospective lay person clinical study was conducted from November 2023 to March 2025 to assess the performance of the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test when compared to a $510(\mathrm{k})$ -cleared SARS-CoV-2 RT-PCR assay with an extraction step. The study prospectively enrolled symptomatic subjects at 13 CLIA-waived clinical study sites (representative of lay-user environment) located in the United States. Enrolled subjects were aged 2 years or older who exhibited symptoms of respiratory infection consistent with COVID-19 or influenza at the time of collection.
Testing was performed in a simulated home environment. Two swabs were collected in random order: a nasopharyngeal (NP) swab for comparator, and an anterior nasal (AN) swab for candidate test. NP swabs for comparator testing were collected by an operator and placed into transport tubes containing universal transport media, refrigerated, and shipped to a central lab for testing with a highly sensitive RT-PCR comparator assay. AN swabs were either self-collected by a lay user aged $\geq 14$ years, or collected by an adult (parent/ guardian) from individuals aged 2 to $< 14$ years. Samples were then immediately tested with the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test according to the instructions.
A total of 1799 participants were enrolled, with 155 samples deemed unevaluable and removed from the final performance analysis: 85 participants were excluded due to being over 4 days post symptom onset, and 68 participants were excluded for protocol deviation, invalid results on either the candidate test or comparator test, resulting in 1644 evaluable study participants. Of these 1644 subjects, 27 samples were excluded from both influenza A/B analysis due to transit delays of comparator samples. Therefore, the total of 182 samples were excluded from both influenza A/B analysis. In addition, one more sample was excluded from influenza B analysis as one subject did not record an influenza B result but had valid results for COVID-19 and influenza A, therefore, total 183 samples were excluded from influenza B analysis.
Table 13. Demographics of evaluable subjects
| Characteristic | Self-Collecting (N=1447) | Lay-user/ Tester Collection (N=197) | Overall (N=1644) |
| --- | --- | --- | --- |
| Age | | | |
| Mean (SD) | 42.0 (17.2) | 7.6 (3.7) | 37.8 (19.8) |
| Median [Min, Max] | 41 [13, 90] | 7.5 [2, 17] | 37 [2, 90] |
| Age Group | | | |
| Age 18-24 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 25-34 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 35-44 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 45-54 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 55-64 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 65-74 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 75-84 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 85-94 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
| Age 95-104 | 23.0 (10.0) | 7.0 (3.0) | 23.0 (10.0) |
K251604 - Page 23 of 28
{23}
K251604 - Page 24 of 28
| 2-13 | 1 | 192 | 193 |
| --- | --- | --- | --- |
| 14-24 | 267 | 5 | 272 |
| 25-64 | 1025 | 0 | 1025 |
| >64 | 154 | 0 | 154 |
| Sex at Birth | | | |
| Female | 842 | 104 | 946 |
| Male | 605 | 93 | 698 |
| Ethnicity | | | |
| Hispanic or Latino (of any race) | 635 | 74 | 709 |
| Not Hispanic or Latino | 806 | 116 | 922 |
| Unknown | 6 | 7 | 13 |
| Race | | | |
| American Indian or Alaskan Native | 6 | 1 | 7 |
| Asian | 91 | 13 | 104 |
| Black or African American | 208 | 16 | 224 |
| Native Hawaiian or Other Pacific Islander | 14 | 2 | 16 |
| More Than One Race | 31 | 11 | 42 |
| Prefer Not to Answer | 1 | 0 | 1 |
| Unknown | 16 | 12 | 28 |
| White | 1080 | 142 | 1222 |
Results obtained with the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test were compared to the results obtained with highly sensitive RT-PCR comparator tests giving rise to the following performance estimates:
## SARS-CoV-2 Performance:
Table 14. Clinical Performance Compared to SARS-CoV-2 Molecular Assay
| CareSuperb COVID-19/Flu A&B Antigen Combo Home Test | FDA-cleared Molecular Assay | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 111 | 6 | 117 |
| Negative | 9 | 1518 | 1527 |
| Total | 120 | 1524 | 1644 |
| Positive Percent Agreement | 92.5% (111/120) (95% CI:86.4%-96.0%) | | |
| Negative Percent Agreement | 99.6% (1518/1524) (95% CI:99.1%-99.8%) | | |
Table 15. SARS-CoV-2 performance by DPSO
| DPSO | Candidate positive | Comparator positive | PPA |
| --- | --- | --- | --- |
| Day 0 | - | 0 | Not tested |
| Day 1 | 16 | 18 | 88.2% |
| Day 2 | 44 | 47 | 93.5% |
| Day 3 | 35 | 36 | 97.2% |
| Day 4 | 16 | 19 | 84.2% |
{24}
K251604 - Page 25 of 28
# Influenza A Performance:
Table 16. Clinical Performance Compared to influenza A Molecular Assay
| CareSuperb COVID-19/Flu A&B Antigen Combo Home Test | FDA-cleared Molecular Assay | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 95 | 15 | 110 |
| Negative | 16 | 1491 | 1507 |
| Total | 111 | 1506 | 1617 |
| Positive Percent Agreement (PPA) | 85.6% (95/111) (95% CI:77.9%-90.9%) | | |
| Negative Percent Agreement (NPA) | 99.0% (1491/1506) (95% CI:98.4%-99.4%) | | |
# Influenza B Performance:
Table 17. Clinical Performance Compared to influenza B Molecular Assay
| CareSuperb COVID-19/Flu A&B Antigen Combo Home Test | FDA-cleared Molecular Assay | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 37 | 5 | 42 |
| Negative | 6 | 1568 | 1574 |
| Total | 43 | 1573 | 1616 |
| Positive Percent Agreement (PPA) | 86.0% (37/43) (95% CI:72.7%-93.4%) | | |
| Negative Percent Agreement (NPA) | 99.7% (1568/1573) (95% CI:99.3%-99.9%) | | |
# Clinical Sensitivity:
Please refer to Section VII.C (Clinical Studies) above for the clinical validation.
# Clinical Specificity:
Please refer to Section VII.C (Clinical Studies) above for the clinical validation.
# 2. Usability Study:
Concurrent with the clinical study, the sponsor evaluated usability of the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test and user comprehension of the test's Quick Reference Instructions (QRI). Of the 1799 enrolled subjects, 1795 individuals were included in the usability and user comprehension study, irrespective of their evaluability for analysis in the clinical study, while four subjects were excluded due to not meeting the inclusion criteria. Each subject filled out a usability questionnaire as part of their visit and interpreted mock test results in the user comprehension (readability) study. Demographics of the enrolled subjects are summarized in table below.
# a. Human Factor Assessment
All subjects in the study received the QRI prior to performing the test. For the human factors assessment, the study personnel or a healthcare provider observed and evaluated the subject/tester's ability to correctly perform the test and interpret the results by evaluating test tasks during the testing process as shown below in table format. The study personnel did not provide any training, or assistance to the subjects/testers. Overall, $98.4\%$ critical task and $90.8\%$ non-critical tasks associated with sample collection and
{25}
running the CareSuperb COVID-19/Flu A&B Antigen Combo Home Test were performed correctly.
Table 18. Critical and non-critical task performed correctly.
| Steps | | Tasks Performed Correctly | Total Number of Tasks | Percentage of Tasks Performed Correctly |
| --- | --- | --- | --- | --- |
| Critical | Swabbed one nostril 5 rotations and for 15 seconds, then second nostril for 5 rotations and 15 seconds. Placed the swab into the sample port before the vial contents added. Added the entire contents of the vial into the sample port. Rotate swab approximately 10x in the sample port before discarding swab. Removed the swab from the sample port and discarded. Read the test between 10-15 minutes. | 10,582 | 10,754 | 98.40% |
| Non-critical | Washed or sanitized hands. Used a timer. | 3,257 | 3,586 | 90.83% |
| Sum | | 13,839 | 14,340 | 96.51% |
Table 19. Evaluation of user test tasks performance.
| Task | Yes | No |
| --- | --- | --- |
| Washed or sanitized hands? | 84.29% (1513/1795) | 15.71% (282/1795) |
| Swabbed one nostril 5 rotations and for 15 seconds, then second nostril for 5 rotations and 15 seconds? | 96.88% (1739/1795) | 3.12% (56/1795) |
| Placed the swab into the sample port before the vial contents added?* | 98.55% (1767/1793) | 1.45% (26/1793) |
| Added the entire contents of the vial into the sample port?* | 99.22% (1778/1792) | 0.78% (14/1792) |
| Rotate swab approximately 10x in the sample port before discarding swab | 98.21% (1760/1792) | 1.79% (32/1792) |
| Removed the swab from the sample port and discarded?* | 98.66% (1768/1792) | 1.34% (24/1792) |
| Read the test between 10-15 minutes?* | 98.88% (1770/1790) | 1.12% (20/1790) |
| Did the subject use a timer?* | 97.38% (1744/1791) | 2.62% (47/1791) |
| Does the observer agree with Subject/Tester results?* | 98.71% (1763/1786) | 1.29% (23/1786) |
*9 subjects did not complete all steps. These subjects were re-tested and completed all steps successfully, however that data is not included in the table above.
K251604 - Page 26 of 28
{26}
The results of the usability/user comprehension questionnaire demonstrated that the instructions were clear and easy to follow, samples could be collected and processed easily. The overall evaluation of the lay user experience did not raise any concerns regarding the usability of the investigational device.
## b. Readability Assessment
A readability study was conducted with a total of 50 lay users. Out of 50 subjects, 21 subjects had some vision impairment (7 near-sighted, 10 far-sighted, and 7 astigmatism). Subjects interpreted two mock panels of investigational test results and thereafter, completed a labeling and comprehension questionnaire. The mock panel contained results for various analyte combinations and analyte concentrations. Mock samples were prepared at concentrations of negative, 1-2x LoD, and 3-5x LoD. All 50 lay users interpreted the mock results in a blinded and randomized fashion. Each lay user interpreted ten random mock samples to avoid a training effect. Overall, the study demonstrated that lay user participants can accurately read and interpret the test as 94.8% of the samples were interpreted correctly across all analyte concentration.
## D. Clinical Cut-Off:
The test is a qualitative test with a binary positive/negative signal and there is no clinical cut-off for the test.
## E. Expected Values/Reference Range:
When the test is valid, it produces binary values, positive or negative for SARS-CoV-2, influenza A and influenza B antigens.
## F. Flex Studies:
To assess the robustness and risk for false results of the test when deviating from the IFU/QRI test steps, flex studies were conducted that assessed all major aspects of the test procedure (extraction buffer volume, reading time, procedure involving sample port [swab insertion, rotation, and removal], delay in sample processing) and variability of environmental test conditions that the test may be subjected to when in use (non-level surface, lighting, disturbance during use, temperature and humidity stress conditions).
Flex study design incorporated 3 operators who tested a test panel comprised of negative samples (PNSM only), low positive samples (2x co-spike LoD with all analytes in PNSM), in replicates of five (5)/operator, upon varying the conditions during testing with the candidate device. The strains used for testing were SARS-CoV-2 B.1.1.529, omicron, Influenza A H1N1, and Influenza B Victoria. Samples were blinded and randomized for testing.
The flex study that assessed the timing of reading results from test initiation noted that reading the device at 4 minutes or 30 minutes after test initiation leads to false positive or false negative results. The package insert instructs users to not interpret the test result before 10 minutes or after 15 minutes.
Overall, the studies support that the test is robust when used as instructed with an insignificant risk of erroneous result.
K251604 - Page 27 of 28
{27}
VIII Proposed Labeling:
The labeling supports finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K251604 - Page 28 of 28
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.