K251603 · Rapid Medical Ltd. · NRY · Oct 6, 2025 · Cardiovascular
Device Facts
Record ID
K251603
Device Name
Tigertriever 17 Ultra Revascularization Device
Applicant
Rapid Medical Ltd.
Product Code
NRY · Cardiovascular
Decision Date
Oct 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tigertriever 17 Ultra Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
Device Story
Stentriever device for neurointerventional ischemic stroke treatment; removes thrombus from large intracranial vessels. Input: manual physician operation under fluoroscopic guidance. Mechanism: nitinol braided mesh expands to conform to vessel diameter when physician pulls handle slider; mesh penetrates and encapsulates clot. Output: restored blood flow. Used in neurointerventional settings by trained physicians. Benefits: mechanical clot removal for patients ineligible for or failing thrombolytic therapy.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, including simulated use in tortuous anatomical models, radial force measurement, durability, torque strength, tip flexibility, kink resistance, tensile strength, and particulate evaluation. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Materials: nitinol braided mesh, 304 stainless steel shaft, nitinol core wire (PTFE coated), 90% platinum/10% iridium markers. Principle: manual expansion of braided mesh via handle slider. Dimensions: 32 mm stent length, 224 cm overall length, 0.017 inch microcatheter compatibility. Sterilization: ethylene oxide (ISO 11135). Single-use.
Indications for Use
Indicated for patients experiencing ischemic stroke within 8 hours of symptom onset who are ineligible for or failed thrombolytic drug therapy, for the purpose of restoring blood flow by removing thrombus from a large intracranial vessel.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 6, 2025
Rapid Medical Ltd.
Ina Gutman
QA/RA Senior Director
Carmel Building, P.O. Box 337
Yokneam, 2069205
Israel
Re: K251603
Trade/Device Name: Tigertriever 17 Ultra Revascularization Device
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: NRY
Dated: September 4, 2025
Received: September 4, 2025
Dear Ina Gutman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251603 - Ina Gutman
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251603 - Ina Gutman
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251603 | |
| Device Name Tigertriever 17 Ultra Revascularization Device | |
| Indications for Use (Describe) The Tigertriever 17 Ultra Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Rapid Medical
# 510(k) Summary K251603
## Submission Sponsor
Rapid Medical Ltd.
Carmel Building, P.O. Box 337
Yokneam, 2069205
Israel
Company Phone No.: +972-72-250-3331
## Contacts
Ina Gutman, QA/RA Senior Director
Email: ina.gutman@rapid-medical.com
Ronen Eckhouse, CEO
Email: ronen@rapid-medical.com
## Date Prepared
August 26, 2025
## Device Identification
Trade/Proprietary Name: Tigertriever 17 Ultra Revascularization Device
Common/Usual Name: Catheter, Thrombus Retriever
Classification Name: Percutaneous Catheter
Regulation Number: 21 CFR 870.1250
Product Code: NRY
Device Class: II
Classification Panel: Neurology
## Legally Marketed Predicate and Reference Devices
Predicate Device: Tigertriever 17 Revascularization Device (K203592)
Reference Device: Solitaire X Revascularization Device (K203358)
## Indications for Use Statement
The Tigertriever 17 Ultra Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.
{5}
Rapid Medical
# Device Description
The Tigertriever 17 Ultra Revascularization Device (TRPP7266) is a line extension of the existing Tigertriever 17 Revascularization Device (TRPP7166) cleared under K203592. The Tigertriever 17 Ultra Revascularization Device is a stentriever that is comprised of an adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire and a handle. The shaft connects the mesh and the handle by the core wire that runs inside the shaft from the distal end of the mesh to the slider activation element in the handle. The mesh is expanded when the physician pulls the slider, since the wires of the mesh are completely radiopaque, the physician sees the mesh under fluoroscopy and controls it until it conforms to the vessel diameter. The design of the wire mesh is optimized to penetrate the clot and encapsulate it during retrieval. The Tigertriever 17 Ultra Revascularization Device is supplied sterile and is intended for single-use only by physicians trained in neurointerventional procedures and the treatment of ischemic stroke.
# Comparison of Technological Characteristics
The subject device has the same principles of operation and indication for use as the predicate device. The subject device has different technological characteristics compared to the predicate device, the Tigertriever 17 Revascularization Device, namely a longer mesh length and a slightly larger microcatheter compatibility. Both stentriever models are comprised of a nitinol braided mesh, stainless steel shaft, nitinol core wire, and a handle. The table below compares the indications for use, principles of operation, technological characteristics, and materials of the Tigertriever 17 Ultra Revascularization Device with that of the predicate device (the Tigertriever 17 Revascularization Device). The technological differences of the Tigertriever 17 Ultra Revascularization Device do not raise any new questions of safety or effectiveness compared to the predicate device.
| | Tigertriever 17 Ultra Revascularization Device (TRPP7266, Subject Device) | Tigertriever 17 Revascularization Device (TRPP7166, Predicate Device) |
| --- | --- | --- |
| 510(k) Number | K251603 | K203592 |
| Regulation | 21 CFR 870.1250 | 21 CFR 870.1250 |
| Product Code | NRY | NRY |
{6}
Rapid Medical
| | Tigertriever 17 Ultra
Revascularization Device
(TRPP7266, Subject Device) | Tigertriever 17 Revascularization
Device
(TRPP7166, Predicate Device) |
| --- | --- | --- |
| Indications for
Use | The Tigertriever 17 Ultra
Revascularization Device is
intended to restore blood flow by
removing thrombus from a large
intracranial vessel in patients
experiencing ischemic stroke within
8 hours of symptom onset. Patients
who are ineligible for thrombolytic
drug therapy or who failed
thrombolytic drug therapy are
candidates for treatment. | The Tigertriever Revascularization
Device is intended to restore blood
flow by removing thrombus from a
large intracranial vessel in patients
experiencing ischemic stroke within
8 hours of symptom onset. Patients
who are ineligible for intravenous
tissue plasminogen activator (IV t-
PA), or who fail IV t-PA therapy,
are candidates for treatment. |
| Anatomical
Location | Neurovasculature | Neurovasculature |
| Mode of
Operation | Manual expansion of the braided
distal portion into the clot using the
handle component | Manual expansion of the braided
distal portion into the clot using the
handle component |
| Design of Distal
Portion | Close end braided nitinol mesh,
manually expandable | Close end braided nitinol mesh,
manually expandable |
| Stent Length
(Unexpanded
Configuration) | 32 mm | 23 mm |
| Stent Distal OD
(Unexpended and
Expanded
Configurations) | Unexpanded configuration 0.5 mm
Expanded configuration 3 mm | Unexpanded configuration 0.5 mm
Expanded configuration 3 mm |
| Overall Length | 224 cm | 200 cm |
| Microcatheter
Compatibility | Microcatheter with a minimum
inner diameter of 0.017 inches | Microcatheter with a minimum
inner diameter of 0.0165 inches |
| Accessory
Devices | Two insertion loading sheaths are
provided, one is located on the device
mesh and one is provided within the
product package. | Three insertion loading sheaths are
provided, one is located on the
device mesh and two are provided
within product package. |
| Stent (Mesh)
Material | Nitinol | Nitinol |
| Markers
Material | 90% Platinum/ 10% Iridium | 90% Platinum/ 10% Iridium |
| Core Wire
Material | Nitinol core wire coated with Green
Lubriskin PTFE | Nitinol core wire coated with Green
Lubriskin PTFE |
3
{7}
Rapid Medical
| | Tigertriever 17 Ultra Revascularization Device (TRPP7266, Subject Device) | Tigertriever 17 Revascularization Device (TRPP7166, Predicate Device) |
| --- | --- | --- |
| Shaft Material | 304 Stainless Steel | 304 Stainless Steel |
| Insertion Loading Sheath Material | PTFE (Natural color) | PTFE (Natural color) |
| Sterilization Method | Ethylene oxide | Ethylene oxide |
| Single Use | Yes | Yes |
| Packaging | Placed into a Dispenser hoop, blister, Tyvek pouch, and Carton box | Placed into a Dispenser hoop, blister Tyvek pouch, and Carton box |
# Non-Clinical Performance Data
In order to demonstrate safety and effectiveness of the subject device and to show substantial equivalence to the predicate device, Rapid Medical Ltd. completed a number of non-clinical tests.
# Bench Tests
The device passed all performance bench testing in accordance with internal requirements, national standards and international standards as shown in the table below to support substantial equivalence of the device.
| Performance Bench Testing | | |
| --- | --- | --- |
| Test | Test Method Summary | Results |
| Simulated Use | Simulated use testing of the Tigertriever 17 Ultra Revascularization Device was performed in an anatomical model which simulated the tortuosity of the neurovasculature. Devices were delivered through the tortuous anatomical model to evaluate the effectiveness of the device at retrieval of firm and soft clots. | The subject device retrieved clots and restored flow in the test model. |
| Radial Force | The radial force of the subject device was measured within a range of lumen diameters applicable to the intended vasculature and compared with the radial forces measured for the predicate device. | The radial force of the subject device when tested in applicable lumen sizes is comparable to the predicate device. |
| Durability Testing | Damage was evaluated after delivery and withdrawal of the device beyond the number of passes and resheathings recommended in the instructions for use. | Devices tested demonstrated no damage after delivery and withdrawal testing. |
{8}
Rapid Medical
| Delivery, Deployment and Retrieval forces | The forces required to deliver, deploy, and retrieve the device were evaluated in an in vitro neurovascular model. | The subject device demonstrated acceptable performance with respect to delivery, deployment and retrieval. |
| --- | --- | --- |
| Torque Strength Testing | The ability of the device to endure turns without failure was evaluated. | All devices passed acceptance criteria. |
| Dimensions Test | Test samples were examined to verify that the dimensions of the Tigertriever 17 Ultra devices are in compliance with specifications. | All devices passed acceptance criteria. |
| Tip Flexibility | The tip deflection maximum force of the device was evaluated. | All devices passed acceptance criteria and were comparable to the predicate device. |
| Kink Resistance Testing | Kink resistance is tested along the device. | All devices passed acceptance criteria. |
| Tensile Strength | The tensile strength was tested for all joints. | All devices passed acceptance criteria. |
| Particulates Evaluation | The number and size of particulates following simulated use of the device in a neurovascular model were evaluated. | The particulate generated by the subject device was similar to the particulate generated by the predicate device. |
| Austenite Finish (Af) Temperature Evaluation | Data from predicate submission is used because of similarities in device design. | All devices passed acceptance criteria. |
| Corrosion Resistance | Data from predicate submission is used because of similarities in device design. | All devices passed acceptance criteria. |
# Biocompatibility
Biocompatibility testing conducted supports the biocompatibility of the Tigertriever 17 Ultra Revascularization Device in accordance with ISO 10993 and consisted of the following tests conducted on various models of the Tigertriever devices, which were considered worst-case in terms of overall design and included the same materials and manufacturing processes as the Tigertriever 17 Ultra Revascularization Device:
{9}
Rapid Medical
| Test | Test Description | Result Summary | Conclusion |
| --- | --- | --- | --- |
| Cytotoxicity | Cytotoxicity Study using MEM Elution | The test article extract showed no evidence of causing cell lysis or toxicity. The test article showed no cytotoxic potential. | Pass |
| Irritation (Intracutaneous Reactivity) | ISO Intracutaneous Irritation Study using extracts | Test article met the requirements of the test with no differences between each test article extract overall mean score and corresponding control extract mean score. | Pass |
| Sensitization | ISO Guinea Pig Maximization Sensitization Test | Test articles showed no evidence of causing delayed dermal contact sensitization. Test article is not considered a sensitizer. | Pass |
| Hemocompatibility – Hemolysis | ASTM Hemolysis Study – Extract and Direct Contact | The test articles in direct and indirect contact with blood were non-hemolytic. | Pass |
| Hemocompatibility – Complement Activation | SC5b-9 Complement Activation Assay | The test article was not considered to be a potential activator of the complement system. | Pass |
| Hemocompatibility – Thrombogenicity | Acute Pre-Clinical Evaluation of the Safety of Tigertriever device in a Swine Model Thrombogenicity | Data from predicate submission is used. | Pass |
6
{10}
Rapid Medical
| Pyrogenicity | USP Rabbit Pyrogen Study, Material Mediated | The total rise of rabbit temperatures during the observation period was within acceptable USP limits. The test article was judged as nonpyrogenic. | Pass |
| --- | --- | --- | --- |
| Acute Systemic Toxicity | ISO Acute Systemic Toxicity in Mice | No mortality or evidence of systemic toxicity from extracts injected into mice. | Pass |
All tests confirmed that the Tigertriever 17 Ultra Revascularization Device met biological safety requirements per the ISO 10993 standard.
## Pre-Clinical Animal Testing Data
No animal studies were conducted, as substantial equivalence of the Tigertriever 17 Ultra Revascularization Device to the predicate device is established through the results of non-clinical performance testing.
## Sterilization and Shelf Life
The subject device and its accessories are sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of $1 \times 10^{-6}$ in accordance with ISO 11135: 2014/AC: 2014 (“Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices”). The subject device and package shelf-life studies established that the subject device and packaging remain functional for the labeled expiration date.
## Clinical Performance Data
No clinical studies were conducted, as substantial equivalence of the Tigertriever 17 Ultra Revascularization Device to the predicate device is established through the results of non-clinical performance testing.
## Statement of Substantial Equivalence
The Tigertriever 17 Ultra Revascularization Device has the same intended use, indications for use, and similar technological characteristics as the predicate Tigertriever 17 Revascularization Device. The technological differences do not raise new questions regarding the safety or effectiveness. Performance data demonstrate that the subject Tigertriever 17 Ultra Revascularization Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.