SteriHub™ Disinfecting Device and Protective Cover
K251592 · 1World Vista Medical · QBP · Feb 11, 2026 · General Hospital
Device Facts
Record ID
K251592
Device Name
SteriHub™ Disinfecting Device and Protective Cover
Applicant
1World Vista Medical
Product Code
QBP · General Hospital
Decision Date
Feb 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The SteriHub™ Disinfecting Device and Protective Cover is intended for active disinfection of needleless connectors. The SteriHub Disinfecting Device and Protective Cover provides a physical barrier to contamination for up to 7 days, if not removed. The SteriHub may be used in the home or healthcare facility.
Device Story
SteriHub is a single-use disinfecting cap for needleless connectors; contains 70% IPA. Device consists of HDPE housing with internal thermoplastic/LDPE core/element. Operation: user screws device onto connector; twists back and forth 10 times (mechanical scrub); IPA disinfects surfaces. Device remains on connector as protective barrier for up to 7 days. Used in home or healthcare facilities by clinicians or trained patients. Output: disinfected needleless connector; physical barrier against recontamination. Benefits: reduces risk of contamination; provides consistent disinfection via mechanical scrub.
Clinical Evidence
Bench testing only. In vitro antimicrobial efficacy testing demonstrated >4 log reduction for C. albicans, C. glabrata, E. coli, P. aeruginosa, S. aureus (MRSA), and S. epidermidis (MRSE) at T=0 and T=6 months. Comprehensive non-clinical testing included mechanical performance (torque, axial load, leakage), particulate testing (USP 788), alcohol ingress, and full ISO 10993 biocompatibility suite (cytotoxicity, hemolysis, sensitization, irritation, systemic toxicity, pyrogenicity).
Technological Characteristics
HDPE housing, thermoplastic element, LDPE core; 70% IPA disinfectant. Dimensions/form factor: cap for needleless connectors. Sterilization: Gamma irradiation (SAL 1x10⁻⁶). Mechanical scrub principle of operation. No software/electronics.
Indications for Use
Indicated for patients requiring devices with needleless connectors; intended for active disinfection of needleless connectors and as a physical barrier to contamination for up to 7 days. For use by healthcare providers and trained home users.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 11, 2026
1World Vista Medical
% Laurie Lewandowski
Consultant, Regulatory and Quality
Honkanen Consultanting, Inc.
738 Saddle Wood Dr.
Eagan, Minnesota 55123
Re: K251592
Trade/Device Name: SteriHub™ Disinfecting Device and Protective Cover
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular administration set
Regulatory Class: Class II
Product Code: QBP
Dated: January 12, 2026
Received: January 12, 2026
Dear Laurie Lewandowski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251592 - Laurie Lewandowski
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251592 - Laurie Lewandowski
Page 3
Sincerely,
DAVID WOLLOSCHECK - S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
SteriHub Disinfecting Device and Protective Cover
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251592 | ? |
| Please provide the device trade name(s). | | ? |
| SteriHub™ Disinfecting Device and Protective Cover | | |
| Please provide your Indications for Use below. | | ? |
| The SteriHub™ Disinfecting Device and Protective Cover is intended for active disinfection of needleless connectors.
The SteriHub Disinfecting Device and Protective Cover provides a physical barrier to contamination for up to 7 days, if not removed. The SteriHub may be used in the home or healthcare facility. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
1WorldVista Medical
"Saving Lives by Design"
# K251592 - 510(k) SUMMARY
## 1. SUBMITTER INFORMATION
**Submitter:**
1WorldVista Medica, Inc.
Steven McQuillan
17956 E. Paria Canyon Drive,
Rio Verde, AZ 85263
USA
Email: steve@1wvmed.com
**Primary Contact:**
Laurie Lewandowski
Consultant, Honkanen Consulting, Inc.
738 Saddle Wood Drive
Eagan, MN 55123
Telephone: 612-770-4038 (cell)
Email: laurie@1wvmed.com
**Date Prepared:**
February 11, 2026
## 2. DEVICE INFORMATION
**Proprietary Name:** SteriHub™ Disinfecting Device and Protective Cover
**Regulation Number:** 21 CFR 880.5440
**Regulation Name:** Intravascular Administration Set
**Common/Usual Name:** Device Disinfectant Cap
**Regulatory Class:** Class II
**Product Code:** QBP
## 3. PREDICATE DEVICE INFORMATION
**Proprietary Name:** PureHub Disinfecting Cap
**Regulation Number:** 21 CFR 880.5440
**Regulation Name:** Intravascular Administration Set
**Common/Usual Name:** Device Disinfectant Cap
**Regulatory Class:** Class II
**Product Code:** QBP
**510K Number:** K193190
{5}
2
# 4. REFERENCE DEVICE INFORMATION
Proprietary Name: Site-Scrub IPA Device
Regulation Number: Unclassified / pre-amendment
Common/Usual Name: Pad, Alcohol, Device Disinfectant
Regulatory Class: Class II
Product Code: LKB
510K Number: K112791
# 5. DEVICE DESCRIPTION
The SteriHub™ Disinfecting Device and Protective Cover (SteriHub) is a single use disinfecting tool to be used on needleless connectors. The SteriHub, Model Number is SH-001, sterile cap contains 70% Isopropyl Alcohol (IPA) and disinfects the needleless connector via an active mechanical friction (scrub) while applying IPA. The SteriHub device also functions as a protective cover against contamination and recontamination of needleless connectors for up to 7 days.
The device consists of HDPE housing that contains an element and core that are filled with 70% IPA. The device is sealed with a foil lid. Disinfection takes place by screwing the SteriHub onto connector, scrubbing by twisting back and forth ten (10) times, like juicing an orange. The combination of scrubbing and IPA application provides an active disinfection.
# 6. INDICATIONS FOR USE
The SteriHub™ Disinfecting Device and Protective Cover is intended for active disinfection of needleless connectors.
The SteriHub Disinfecting Device and Protective Cover provides a physical barrier to contamination for up to 7 days, if not removed. The SteriHub may be used in the home or healthcare facility.
# 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| Attribute / Device Characteristic | SteriHub Disinfecting Device and Cover (Subject Device) | PureHub Disinfecting™ Cap (K193190) Predicate |
| --- | --- | --- |
| Intended Use | To disinfect needleless connectors prior to access and to act as a physical barrier for up to 7 days if not removed. | Identical |
| Indications for Use | The SteriHub™ Disinfecting Device and Protective Cover is intended for active disinfection of needleless connector.
The SteriHub Disinfecting Device and Protective Cover provides a physical barrier to contamination for up to 7 days, if not removed.
The SteriHub may be used in the home or healthcare facility. | BD PureHub™ Disinfecting Caps are intended to be used as a disinfecting cleaner for swabbable needle-free luer connectors prior to access and to act as a physical barrier between line accesses.
BD PureHub™ Disinfecting Cap will disinfect the needle-free luer connector one (1) minute after application and act as a physical barrier for up to seven (7) days, if not removed. |
{6}
| Attribute / Device Characteristic | SteriHub Disinfecting Device and Cover (Subject Device) | PureHub Disinfecting™ Cap (K193190) Predicate |
| --- | --- | --- |
| User Population | Healthcare providers and home users trained by health care providers | General Use |
| Use Conditions | Home and health care facility | Identical |
| Use | Needleless connectors | Identical |
| Target Population | Patients that require devices with needleless connectors | Identical |
| Principle of Operation | Active (physical) scrub disinfection with 70% IPA of needleless connectors.
The SteriHub fits the needleless connector maintaining a disinfecting barrier for up to 7 days. | Passive, soak, 70% IPA disinfection
The PureHub fits the needleless connector maintaining a disinfecting barrier for up to 7 days. |
| Components | • Ribbed Housing - HDPE
• Element -Thermoplastic
• Core - LDPE
• 70% IPA solution
• Foil Lid | • Ribbed Housing - HDPE
• Sponge – Polyester Urethane
• 70% IPA solution
• Foil lid |
| Disinfectant Active Ingredient | 70% Isopropyl Alcohol | Identical |
| Disinfecting Protocol* | 10 twists back and forth
Active disinfection | Immediate - 1 minute dwell time
Passive disinfection |
| Physical Barrier | After disinfecting, leave on up to 7 days | Identical |
| Features | Element and core are designed to scrub the septum, rim, and threads of needleless connectors with 70% IPA. | Element designed to soak the needleless connectors with 70% IPA. |
| Sterility | Gamma Irradiation | Identical |
| SAL | 1x10⁻⁶ | Identical |
| Target Organisms | • Candida albicans,
• Candida glabrata,
• Escherichia coli,
• Pseudomonas aeruginosa,
• Staphylococcus aureus (MRSA),
• Staphylococcus epidermidis (MRSE) | • Candida albicans,
• Candida glabrata,
• Escherichia coli,
• Pseudomonas aeruginosa,
• Staphylococcus aureus (MRSA),
• Staphylococcus epidermidis (MRSE)
• Acinetobacter baumannii |
* The SteriHub Disinfecting Device and Protective Cover provides an active disinfecting protocol like the reference device, Site Scrub (K112791) - twist back and forth a minimum of 10 seconds, dry 5 seconds.
The following summarizes why the identified differences in technological characteristics between the subject and predicate devices do not raise new questions of safety and effectiveness
{7}
- Indications for Use Wording Differences:
Both devices disinfect needleless connectors using 70% IPA and provide a barrier for up to 7 days. Wording differences clarify use environment and disinfection description but do not change intended use, function, or risk profile and therefore do not raise new questions of safety and effectiveness.
- User Population Description:
The subject device specifies healthcare providers and trained home users, which is consistent with the predicate terminology of general use and supported by usability validation. This does not change risk and does not raise new questions of safety and effectiveness.
- Principle of Operation — Active Scrub vs Passive Dwell:
Both devices use 70% IPA to disinfect connectors; the subject uses mechanical scrubbing while the predicate uses dwell time. IPA scrub disinfection is an established approach and performance testing confirms effectiveness, so this difference does not raise new questions of safety and effectiveness.
- Disinfecting Protocol — Twisting Action vs Dwell:
The subject device requires a labeled twisting step that was supported by performance and microbiological testing and is consistent with legally marketed scrub-based disinfecting devices. This difference in user interaction does not change intended function and does not raise new questions of safety and effectiveness.
- Internal Element and Material Design — Thermoplastic/LDPE vs Sponge:
The subject device uses a molded element/core rather than a sponge to deliver 70% IPA and provide mechanical friction during disinfection. The design difference has been tested in performance, microbiological, biocompatibility and particulate testing that support safe use and effective disinfection and therefore, does not raise new questions of safety and effectiveness.
- Scrub Feature vs Soak Feature:
Mechanical scrubbing versus soaking represents an alternate IPA delivery method to the same connector surfaces, with verified performance results. This does not raise new questions of safety and effectiveness.
- Target Organism Panel Differences:
The organism tested substantially overlap, with the subject device being tested against the most common organisms representing clinically relevant species. Testing demonstrates effective disinfection. The target panel differences do not alter mechanism or intended use and do not raise new questions of safety and effectiveness.
8. DISCUSSION
The intended use of the subject and predicate device are the same; both disinfect a needleless connector using 70% IPA and act as a physical barrier for up to 7 days. The disinfecting method in the subject device consists of scrubbing with 70% IPA while the predicate uses a disinfecting one-minute dwell with 70% IPA. The reference device, SiteScrub K112791, disinfects by scrubbing with 70% IPA similar to the subject device.
The differences in technological characteristics do not raise different questions of safety and effectiveness, and performance testing demonstrates that the subject device performs as intended and is as safe and effective as the predicate devices.
9. NON-CLINICAL TESTING
The following tests were performed on the SteriHub Disinfecting Device and Protective Cover at T=0 and T=6 months.
In Vitro Antimicrobial Efficacy (immediate disinfection and post 7 days)
- Candida albicans,
{8}
- Candida glabrata,
- Escherichia coli,
- Pseudomonas aeruginosa,
- Staphylococcus aureus (MRSA),
- Staphylococcus epidermidis (MRSE)
Test results showed >4 log count reduction in all scenarios.
## Testing at T=0 and T=aged
- Resistance to separation from unscrewing
- Physical Attributes
- Alcohol volume
- Component fit (housing / element, element/core)
- Dimensions
- Torque On/Off Force
- Override
- Air and Liquid Leakage
- Stress Cracking
- Resistance to Axial Load
- Visual – Device and Packaging
- Particulate Testing (USP 788)
- Package Integrity
- Usability Testing
Alcohol Ingress testing was conducted and demonstrated that across three different needless connector types and at two different temperatures, 20-25°C and 37°C, alcohol ingress into the connector was less 14 mmole / L.
## Biocompatibility Testing was performed
- Cytotoxicity: MEM Extraction (ISO 10993-5, 2009)
- Hemolysis: Direct and Indirect (ISO 10993-4 2017)
- Sensitization: Guinea Pig Maximization Test (ISO10993-10)
- Irritation: Intracutaneous Reactivity Test (ISO10993-23, 2021)
- Acute Systemic Toxicity (ISO 10993-11, 2017)
- Subacute Toxicity (ISO 10993-11, 2017)
- Material Mediated Pyrogenicity (ISO10993-11, 2017)
- LAL Endotoxin USP <85>
All devices met all acceptance criteria.
## 10. CONCLUSION
The 1WorldVista Medical SteriHub Disinfecting Devices and Protective Covers are substantially equivalent in terms of the indications for use, technological characteristic, performance testing and comparison to the cited predicate, and do not present any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.