The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups.
Device Story
Profoject™ Enteral Feeding Syringe is a manual, single-use device for dispensing and measuring fluids for oral or enteral administration. It consists of a barrel, plunger, and plunger stopper. The device connects to enteral access devices via a male ENFit connector. Available in sizes 0.5mL to 60mL, it is supplied sterile (EO gas, SAL 10⁻⁶) or non-sterile. Used in clinical or home settings by practitioners or supervised laypersons to deliver nutrition or medication. The device ensures accurate dosing (±10% for volumes ≥20% of capacity) and prevents misconnections through ENFit compatibility. It benefits patients by providing a standardized, safe method for enteral fluid delivery.
Clinical Evidence
No clinical data included. Substantial equivalence is supported by bench testing, including performance testing (ISO 20695, ISO 7886-1, ISO 80369 series), biocompatibility testing (ISO 10993-1, -5, -10, -23), and sterilization validation (ISO 11135, ISO 11737-1). Dose accuracy testing confirmed performance within ±10% limits.
Technological Characteristics
Materials: Polypropylene (PP) barrel/plunger, Polyisoprene rubber plunger stopper. Connectivity: Male ENFit connector (ISO 80369-3). Sterilization: Ethylene Oxide (EO) to SAL 10⁻⁶. Form factor: Manual syringe, 0.5mL to 60mL. Standards: ISO 20695, ISO 7886-1, ISO 80369-1/3/20, ISO 10993-1/5/10/23, ISO 11135, ISO 11737-1, ASTM F1886, F1140, F88, F1929.
Indications for Use
Indicated for oral or enteral fluid delivery in all age groups, including clinical and home care settings, operated by medical practitioners or supervised laypersons.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Oral/Enteral Syringe with ENFit connector (K211025)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 18, 2025
CMT Health PTE. Ltd.
Monica Ma
Sponsor/Applicant
150 Beach Road, #28-05, Gateway West
Singapore, 189720
Singapore
Re: K251585
Trade/Device Name: Profoject™ Enteral Feeding Syringe
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube and Accessories
Regulatory Class: Class II
Product Code: PNR
Dated: May 23, 2025
Received: May 23, 2025
Dear Monica Ma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251585 - Monica Ma
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251585 - Monica Ma
Page 3
Sincerely,
STEPHANIE COLE -S
for Anthony C. Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251585 | ? |
| Please provide the device trade name(s). | | ? |
| Profoject™ Enteral Feeding Syringe | | |
| Please provide your Indications for Use below. | | ? |
| The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is
intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or
home care settings by operators ranging from qualified medical practitioners to laypersons (under the
supervision of a qualified medical practitioners) in all age groups. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251585 Page 1 of 5
# 510(k) Summary
## 1. Submission Information
510(k) Number: K251585
Date: May 23, 2025
Type of 510(k) Submission: Traditional
Submitter: CMT HEALTH PTE. LTD.
150 BEACH ROAD, #28-05, GATEWAY WEST, SINGAPORE, 189720
Contact Person: Monica Ma
E-mail: ra@cmthealth.com
Tel: +65 6846 1379
## 2. Device Identification
Trade Name: Profoject™ Enteral Feeding Syringe
Common Name: Enteral Syringes With Enteral Specific Connectors
Regulation Name: Gastrointestinal tube and accessories
Product Code: PNR
Device Class: II
Regulation Number: 21CFR 876.5980
Review Panel: Gastroenterology/Urology
Indications for use: The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups.
Device Description: Profoject™ Enteral Feeding Syringe provided in a variety of sizes from 0.5mL to 60mL. It consists of plunger, plunger stopper, barrel, and used to deliver fluids into the body orally or connected to an enteral access device with male ENFit connector.
The proposed syringe is sterile or non-sterile. Sterile device was sterilized by Ethylene Oxide Gas to achieve a SAL of 10⁻⁶ and supplied sterility maintenance package. The shelf life of both sterile and non-sterile proposed devices has been validated as 5 years.
{5}
K251585 Page 2 of 5
## 3. Predicate Device Identification
K211025 - Oral/Enteral Syringe with ENFit connector
## 4. Non-Clinical Test Conclusion
Non-clinical verification of the Enteral Feeding Syringe has been conducted to evaluate their safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods, and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the effectiveness of the proposed device in accordance with relevant test methods cited below:
## Performance Testing
1. ISO 20695 First edition 2020-03 Enteral feeding systems - Design and testing
2. ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
3. ISO 80369-1 Second edition 2018-11 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
4. ISO 80369-3 First Edition 2016-07-01 Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications [Including AMENDMENT 1 (2019)].
5. ISO 80369-20 First edition 2015-05-15 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
6. ISO 11737-1 Third edition 2018-01 [Including AMD1:2021] Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
7. USP - NF <62> Microbiological Examination of Nonsterile Products: Tests for Specified Microorganisms
8. ASTM F1886/F1886M-16 (2024) Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
9. ASTM F1140/F1140M-13 (2020) Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
10. ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
11. ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
12. USP - NF <71> Sterility Tests
13. ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
14. ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
15. ISTA 3A: 2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
16. Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications - Guidance for Industry and Food and Drug Administration Staff
In addition, dose accuracy testing is conducted to demonstrate the enteral syringes are accurate to ±10% when the syringe is filled with a minimum dose of 20% of the overall syringe capacity.
{6}
K251585 Page 3 of 5
# Biocompatibility Testing
The biocompatibility evaluation for the Enteral Feeding Syringe was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed device is classified as a surface device, mucosal membrane contact.
1. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2. ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
3. ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
# 6. Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data was not including in this submission
{7}
K251585 Page 4 of 5
7. Substantially Equivalent Comparison Conclusion
| Parameters | | Proposed Device | Predicate Device | Remark |
| --- | --- | --- | --- | --- |
| 1 | 510(k) Number | K251585 | K211025 | -- |
| 2 | 510(k) Holder | CMT HEALTH PTE. LTD. | Ningbo Tianyi Medical Appliance Co., Ltd. | -- |
| 3 | Trade Name | Profoject™ Enteral Feeding Syringe | Oral/Enteral Syringe with ENFit connector | -- |
| 4 | Product Code | PNR | PNR | Same |
| 5 | Regulation Number | 21CFR 876.5980 | 21CFR 876.5980 | Same |
| 6 | Review Panel | Gastroenterology/Urology | Gastroenterology/Urology | Same |
| 7 | Device Class | II | II | Same |
| 8 | Indications for use | The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups | The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care settings by operators ranging from qualified medical practitioners to laypersons (under the supervision of a qualified medical practitioners) in all age groups | Same |
| 9 | Configuration | Plunger stopper; Plunger; Barrel with ENFit connector | Piston; Plunger; Barrel with ENFit connector; Tip cap | Similar Note 1 |
| 10 | Single Use | Single Use | Single Use | Same |
| 11 | Product Size (nominal volumes) | Low dose tip ENFit syringe: 0.5ml, 1ml, 3ml, 5ml, 6ml | Low dose tip ENFit syringe: 0.5ml, 1ml, 3ml, 6ml | Similar Note 2 |
| | | Standard ENFit syringe: 10ml, 12ml, 20ml, 30ml, 35ml, 50ml, 60ml | Standard ENFit syringe: 12ml, 20ml, 35ml, 60ml | |
| 12 | Product Performance | Complied with: ISO 80369-3 ISO 80369-20 ISO 7886-1 | Complied with: ISO 80369-3 ISO 80369-20 ISO 7886-1 | Same |
| 13 | Materials | Barrel: Polypropylene (PP) | Barrel: Polypropylene (PP) | Similar |
{8}
K251585 Page 5 of 5
| | | Plunger: Polypropylene (PP) and purple color additive
Plunger stopper: Polyisoprene rubber | Plunger: Polypropylene (PP) and white pigment
Piston: Silicone rubber
Tip Cap: Polypropylene (PP) and Orange pigment
Or polypropylene (PP) and purple pigment | Note 3 |
| --- | --- | --- | --- | --- |
| 14 | Biocompatibility | No Cytotoxicity
No Irritation
No Sensitization | No Cytotoxicity
No Irritation
No Sensitization | Same |
| 15 | Sterile | Sterile or non-sterile | Sterile or non-sterile | Same |
| 16 | Sterile Method | EO Sterilized | EO Sterilized | Same |
| 17 | Sterilization | EO (ethylene gas) to SAL=10^{-6} | EO (ethylene gas) to SAL=10^{-6} | Same |
| Differences between proposed device and predicate device, reference device: | | | | |
| **Note 1:**
The configuration of proposed device is different from predicate device, depending on the design and sales requirements of the product. According to ISO 20695, enteral syringes shall consist of at least the following: a) a graduated container; b) unless the enteral syringe is designed for gravity use, there shall be a means to create pressure (e.g. a plunger or a bulb); c) an outlet port. Tip cap is not a required component, it will not affect the safety and effectiveness of the proposed device. | | | | |
| **Note 2:**
The product size for proposed device is different from predicate device and this difference is just in infusion capacity and dose not effect indication for use, the physician can select by per patient’s condition. And through performance test reports that proposed device will not raise safety and effectiveness issues. | | | | |
| **Note 3:**
The materials of plunger and plunger stopper are different between the proposed device and predicate device. This difference does not raise any new questions of safety and effectiveness. The biocompatibility test of the proposed device was conducted to demonstrate that the proposed device met the biocompatibility requirements. | | | | |
## The Conclusions:
The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices identified in the submission. Thus the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.