K251570 · Vipamat Sarl · IQC · Jun 23, 2025 · Physical Medicine
Device Facts
Record ID
K251570
Device Name
Hippocampe Marathon and Trail
Applicant
Vipamat Sarl
Product Code
IQC · Physical Medicine
Decision Date
Jun 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3880
Device Class
Class 2
Indications for Use
The Hippocampe Marathon and Trail is indicated to provide mobility to persons restricted to a seated position on various terrains, including footpaths and trails.
Device Story
Manual, three-wheeled, occupant or care-provider propelled wheelchair; designed for everyday use and recreational activities like running/trail running. Frame constructed from bent aluminum tubing; features adjustable seat back, headrest, push bar, and footrest. Occupant sits in seat supported by fire-resistant upholstery. Care provider operates disc brakes via push bar levers for motion control and parking. Device disassembles for storage by detaching armrests and folding seat back. Used in indoor/outdoor environments; traverses smooth/rough/uneven ground, small obstacles, and slopes. Provides mobility to individuals with restricted mobility; enables recreational access for patients.
Clinical Evidence
Bench testing only. No clinical data. Testing included static stability (ISO 7176-1), brake effectiveness (ISO 7176-3), dimensions/maneuvering (ISO 7176-5), seating dimensions (ISO 7176-7), static/impact/fatigue strength (ISO 7176-8), test dummies (ISO 7176-11), coefficient of friction (ISO 7176-13), labeling (ISO 7176-15), ignition resistance (NF ISO 7176-16), and set-up procedures (NF ISO 7176-22).
Technological Characteristics
Three-wheeled manual wheelchair; bent aluminum tubing frame. Features: adjustable seat back, headrest, push bar, armrests. Braking: disc brakes (motion and parking). Upholstery: fire/ignition resistant (NF ISO 7176-16). Dimensions/strength tested per ISO 7176 series. No electronics or software.
Indications for Use
Indicated for persons restricted to a seated position requiring mobility assistance on various terrains, including footpaths and trails. Maximum weight capacity is 220 lbs.
Regulatory Classification
Identification
A special grade wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. It is intended to be used in all environments for long-term use, e.g., for paraplegics, quadraplegics, and amputees.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 23, 2025
Vipamat Sarl
% Sarah Robbins
Senior Quality Manager
Rook Quality Systems LLC
1155 Mount Vernon Hwy
Suite 800
Dunwoody, Georgia 30338
Re: K251570
Trade/Device Name: Hippocampe Marathon and Trail
Regulation Number: 21 CFR 890.3880
Regulation Name: Special Grade Wheelchair
Regulatory Class: Class II
Product Code: IQC
Dated: May 16, 2025
Received: May 22, 2025
Dear Sarah Robbins:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251570 - Sarah Robbins
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251570 - Sarah Robbins
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
Hippocampe Marathon and Trail
Page 10 of 42
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251570 | ? |
| Please provide the device trade name(s). | | ? |
| Hippocampe Marathon and Trail | | |
| Please provide your Indications for Use below. | | ? |
| The Hippocampe Marathon and Trail is indicated to provide mobility to persons restricted to a seated position on various terrains, including footpaths and trails. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K251570 | 510(k) Summary | Prepared on: 2025-06-10 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Vipamat Sarl | |
| Applicant Address | 3 Rue Gustave Eiffel Ploemeur 56270 France | |
| Applicant Contact Telephone | +33297862487 | |
| Applicant Contact | Mr. Vincent Leroy | |
| Applicant Contact Email | contact@vipamat.com | |
| Correspondent Name | Rook Quality Systems LLC | |
| Correspondent Address | 1155 Mount Vernon Hwy Suite 800 Dunwoody GA 30338 United States | |
| Correspondent Contact Telephone | 803-338-6041 | |
| Correspondent Contact | Ms. Sarah Lacey Robbins | |
| Correspondent Contact Email | Sarah.Robbins@rookqs.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Hippocampe Marathon and Trail | |
| Common Name | Special grade wheelchair | |
| Classification Name | Wheelchair, Special Grade | |
| Regulation Number | 890.3880 | |
| Product Code(s) | IQC | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K051709 | Hippocampe Wheelchair | IQC |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Hippocampe Marathon and Trail is an occupant or care provider propelled manual operated wheelchair. The Hippocampe Marathon and Trail is designed for everyday use as well as recreational activity such as running and trail running, in the following environments: Indoors and outdoors; Over smooth, rough, and uneven ground; Over soft and hard surfaces; Over small obstacles, like door jams; and Up, down and across slopes. The Hippocampe Marathon and Trail is a three-wheel triangular wheelchair comprised of bent aluminum tubing frame that are bent, joined, fastened, and/or coupled with connectors to form a frame. The frame consists primarily of a lower frame, adjustable height seat | | |
{5}
back frame, adjustable headrest, push bar and push bar frame, tubing clamp connectors, footrest, arm rests, rear axle assembly, and front fork and axle assembly. Once assembled, the wheelchair frame is fixed. The recline of the seat back frame can be adjusted with depth adjustment screws at the base of the seat back frame. The upholstery is attached to the frame using nylon straps and buckles and is fire resistant. A rigid plastic open basket-like structure located in the middle of the lower frame and behind the front fork assembled provides the occupant a footrest and has a removable layer of foam. On the push bar are control levers for disc brakes of the rear wheels which are operated by the occupant's care provider. The disc brakes also act as parking brakes and are activated by the care provider.
The Hippocampe Marathon and Trail can be easily assembled and disassembled for stowage or storage by detaching the lower frame of the armrest and folding the seat back frame forwards. The wheelchair is available in three sizes: medium, large and extra-large. The maximum weight capacity of Hippocampe Marathon and Trail is 220 lbs.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Hippocampe Marathon and Trail is indicated to provide mobility to persons restricted to a seated position on various terrains, including footpaths and trails.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Indications for Use and Intended use
The Hippocampe Marathon and Trail is indicated to provide mobility to persons restricted to a seated position up to a weight capacity of up to 220 lb. The predicate device has a similar intended use, with the difference being a higher weight capacity of 286 lb. Both the subject and predicate device have the same indicated use of providing mobility to persons restricted to a seated position indoor and outdoor use, on various terrain, surfaces, obstacles and slopes. The standard version of the predicate device has a wider scope of indication for use, including terrain such as sand, snow, and in water. The more limited indication of use of terrain of the subject device for footpaths and trails does not raise any questions in the safety or performance of the subject device. The indications for use of the subject device are substantially equivalent to the predicate device. The difference in weight capacity does not raise any questions in the safety or performance of the subject device.
## Technological Comparison
21 CFR 807.92(a)(6)
The subject device Hippocampe Marathon and Trail wheelchair and predicate device Hippocampe Wheelchair (K051709) are both unpowered three-wheeled design mechanical wheelchairs comprised of aluminum tubing fastened together.
Device Characteristics
Frame
The subject and predicate device share the same three-wheel frame design, with armrests that are detachable, adjustable headrests, height adjustable backrest, and pushbar.
The subject device has a lower weight capacity of 220 lbs compared to the 286 lbs of the predicate device. The weight capacity is indicated on the subject device's label and within the subject device's IFU.
The subject device and predicate devices can be disassembled for storage and transportation by unfastening the armrests and pivoting the seat back frame forwards
Brakes
The subject device also has disc brakes that are operable by the care provider for slowing down the wheelchair while in motion and as parking brakes while the predicate device only has parking brakes. Braking performance of both devices have been evaluated in accordance with ISO 7176-3.
Upholstery
Both the subject and predicate devices have upholstery that fasten to the wheelchair frame to provide support for the occupant when sitting. The subject device's seating upholstery is supported by a rigid frame while the primary predicate device is a sling-style. Both the upholsteries of the subject and primary predicate devices are fire/ignition resistant.
Care Provider Propulsion
Both the subject and predicate devices are intended to be able to have the occupant pushed by a care provider. The subject device and predicate devices have substantially designs which utilize the wheelchair frame or push bar to enable care provider pushing.
Accessories
The predicate device has accessories that enable usage on a wider range of terrain compared to the subject device, such as on sand, snow, and in water. The subject device has optional trail wheels which are similar to the standard wheels of the predicate device.
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
{6}
Non-clinical validation testing and evaluations were performed to demonstrate the safety and effectiveness of the device. The following tests were performed:
- ISO 7176-1 Third edition 2014-10-01 - Wheelchairs - Part 1: Determination of static stability
- ISO 7176-3 Third edition 2012-12-15 - Wheelchairs - Part 3: Determination of effectiveness of brakes
- ISO 7176-5 Second edition 2008-06-01 - Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
- ISO 7176-7 First Edition 1998-05-15 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- ISO 7176-8 Second edition 2014-12-15 - Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
- ISO 7176-11 Second edition 2012-12-01 Wheelchairs - Part 11: Test dummies
- ISO 7176-13 First edition 1989-08-01 - Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
- ISO 7176-15 First edition 1996-11-15 - Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
- NF ISO 7176-16:2014 Wheelchairs - Part 16: Resistance to ignition of postural support devices
- NF ISO 7176-22 Second edition 2014-09-01 Wheelchairs - Part 22: Set-up procedures
Not applicable
The bench testing performed demonstrates that the Hippocampe Marathon and Trail is substantially equivalent in safety, effectiveness, and performance to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.