The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media. The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature.
Device Story
Device is a single-lumen neurovascular long sheath with a dilator and rotating hemostasis valve (RHV). Input: manual insertion by trained physician into neurovasculature. Transformation: sheath provides a conduit for interventional devices (e.g., guidewires) and contrast media delivery. Output: physical access to neurovascular lesion sites. Used in clinical settings by neurointerventionalists. Benefits: facilitates navigation of therapeutic devices to target sites in the neurovasculature. Device features a hydrophilic coating on the outer surface and a fluoropolymer inner lumen to reduce friction during device advancement.
Clinical Evidence
Bench testing only. Testing performed per FDA guidance for peripheral PTA and specialty catheters and ISO 10555-1. Includes dimensional verification, simulated use, lubricity, coating integrity, and appearance. Biocompatibility testing conducted per ISO 10993 series, including cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, and hemocompatibility. In vivo thrombogenicity study demonstrated performance comparable to the predicate.
Technological Characteristics
Single-lumen catheter; 8Fr outer diameter; 6.8Fr inner diameter. Materials: hydrophilic polymer outer coating, fluoropolymer (PTFE) inner lumen. Sterilization: Ethylene Oxide (SAL 10^-6). Supplied with dilator and rotating hemostasis valve. Shelf life: 3 years. Complies with ISO 10555-1 and ISO 10993.
Indications for Use
Indicated for patients requiring percutaneous intravascular procedures in the neurovasculature, specifically to guide interventional devices to a lesion or procedural site and for the injection of contrast media.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 20, 2025
ASAHI INTECC Co., Ltd.
% Esther Kim
Quality System and Regulatory Affairs Associate
ASAHI INTECC USA, Inc.
3002 Dow Avenue, Suite 212
Tustin, California 92780
Re: K251560
Trade/Device Name: FUBUKI XF-R Neurovascular Long Sheath
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP
Dated: May 21, 2025
Received: May 21, 2025
Dear Esther Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251560 - Esther Kim
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251560 - Esther Kim
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251560 | |
| Device Name FUBUKI XF-R Neurovascular Long Sheath | |
| Indications for Use (Describe) The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media. The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
FUBUKI XF-R Neurovascular Long Sheath
# 510(k) Summary
(as required by 21 CFR § 807.92)
# ASAHINTECC CO.,LTD.
Global Headquarters and R&D Center
3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan
TEL: +81-561-48-5551 FAX: +81-561-48-5552
http://www.asahi-intecc.co.jp/
# FUBUKI XF-R Neurovascular Long Sheath
510(k) K251560
| Date Prepared: | 21 MAY, 2025 |
| --- | --- |
| Applicant: | ASAHI INTECC CO., LTD.
3-100 Akatsuki-cho,
Seto, Aichi 489-0071 Japan |
| Contact: | Esther Kim
Quality System and Regulatory Affairs Associate
ASAHI INTECC USA, Inc.
3002 Dow Avenue, Suite 212
Tustin, CA 92780
Office: (949) 756-8901 ext. 508
e-mail: aiu_ra@asahi-intecc-us.com |
| Device Name: | FUBUKI XF-R Neurovascular Long Sheath |
| Device Classification: | Class II, 21 CFR § 870.1250 |
| Regulation Name: | Percutaneous Catheter |
| Product Code: | QJP – Catheter, Percutaneous, Neurovasculature |
| Predicate Device: | FUBUKI XF Neurovascular Long Sheath (K213589) |
# Intended Use/Indications For Use
The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media.
The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature.
# Device Description
The FUBUKI XF-R Neurovascular Long Sheath (FUBUKI XF-R) consists of a long sheath and a dilator. The long sheath is a single lumen neurovascular catheter designed for introduction of interventional devices, such as guidewires and other therapeutic devices. The long sheath consists of three sections: (1) a shaft, (2) a protector and (3) a connector. The distal portion of the shaft consists of a soft tip and a soft tube. The proximal part of
ASAHI INTECC CO., LTD.
{5}
FUBUKI XF-R Neurovascular Long Sheath
the shaft is covered by the protector (strain relief) and the connector is bonded to the proximal end of the shaft.
The subject device is provided sterile, by ethylene oxide, and is intended for single use only by physicians who have been adequately trained in neurointerventional procedures.
The outer surface of the long sheath is coated with a hydrophilic polymer and the inner lumen of the shaft (excluding the connector portion) is lined with a fluoropolymer layer to facilitate movement of the guidewire and other devices.
The dilator consists of two parts: (1) a shaft and (2) a connector.
The FUBUKI XF-R is supplied with a dilator and rotating hemostasis valve (RHV) packed in a sterile package.
## Comparison with Predicate Device
The subject FUBUKI XF-R and the predicate FUBUKI XF have the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principle.
The following technological differences exist between the subject and predicate devices:
- Hydrophilic coating length of the long sheath
- Addition of a 95cm model and exclusion of the 110cm model for the long sheath
- Exclusion of the Angled tip long sheath model
- Addition of a 117cm dilator
These differences do not raise new or different questions of safety and effectiveness. The results of the non-clinical testing provide reasonable assurance of substantial equivalence to the predicate device.
ASAHI INTECC CO., LTD.
Page 2 of 5
{6}
FUBUKI XF-R Neurovascular Long Sheath
Comparison of Intended Use and Technological Characteristics to Predicate Device
| Device Name | Subject Device
FUBUKI XF-R Neurovascular Long Sheath | Predicate Device
FUBUKI XF Neurovascular Long Sheath |
| --- | --- | --- |
| 510(k) Number | K251560 | K213589 |
| Manufacturer | ASAHI INTECC CO., LTD. | |
| Regulation Number | 21 CFR § 870.1250 | |
| Regulation Name | Catheter, Percutaneous, Neurovasculature | |
| Regulatory Class | II | |
| Product Code | QJP | |
| Intended Use and Indications for Use | The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media.
The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature. | This product is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. This product is also intended to be used for injection of contrast media.
This product is intended for use only in the neurovasculature. |
| Long Sheath | | |
| Outer Diameter | 2.70mm (0.106", 8Fr) | |
| Inner Diameter | 2.28mm (0.090", 6.8Fr) | |
| Effective Length | 80cm, 90cm, 95cm, 100cm | 80cm, 90cm, 100cm, 110cm |
| Tip Shape | Straight | Straight, Angled |
| Coating | Hydrophilic | |
| Construction | Soft tip, soft tube and shaft, protector, connector | |
| Coating Length | 20cm | 8cm |
| Inner Lumen | PTFE | |
| Dilator | | |
| Effective Length | 102cm, 112cm, 117cm, 122cm | 102cm, 112cm, 122cm |
| Outer Diameter | 2.24mm (0.088") | |
| Inner Diameter | 0.91mm (0.036") | |
| Other Information | | |
ASAHI INTECC CO., LTD.
{7}
FUBUKI XF-R Neurovascular Long Sheath
| Accessory | Hemostasis Valve (Rotating Hemostasis Valve) |
| --- | --- |
| Sterilization Method | Ethylene Oxide Gas |
| Sterilization Level | SAL 10^{-6} |
| Shelf Life | 3 years |
| Single Use | Yes |
## Non-Clinical Bench Testing / Performance Data
The following non-clinical bench testing was performed for the FUBUKI XF-R Neurovascular Long Sheath to support the substantial equivalence determination. The FUBUKI XF-R met all acceptance criteria and performed similarly to the predicate FUBUKI XF. Testing was performed per FDA guidance “Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions” and the recommendations in ISO 10555-1 “Intravascular Catheters – Sterile and Single-Use Catheters – Part 1: General Requirements”.
- Dimensional Verification
- Simulated Use
- Lubricity
- Coating Integrity / Particulate Evaluation
- Appearance
## Biocompatibility
The FUBUKI XF-R was tested per ISO 10993 series and FDA guidance, “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”.” All testing performed met the requirements as specified within the applicable standard.
| Test | Test Method Summary | Conclusions |
| --- | --- | --- |
| Cytotoxicity
MEM Elution Test | Per ISO 10993-5 | Non-Cytotoxic |
| Sensitization
Kligman Maximization Test | Per ISO 10993-10 | Non-Sensitizing |
| Intracutaneous Reactivity
Intracutaneous Injection Test | Per ISO 10993-23 | Non-Irritant |
| Acute Systemic Toxicity
System Injection Test | Per ISO 10993-11 | Non-Toxic |
| Material Mediated Pyrogenicity
Rabbit Pyrogen Test | Per ISO 10993-11 | Non-Pyrogenic |
| Hemocompatibility
Rabbit Blood Hemolysis
Test
(Direct and Indirect Contact) | Per ISO 10993-4 | Non-Hemolytic |
| Hemocompatibility
Complement Activation Test
(SC5b-9)
(Direct Contact) | Per ISO 10993-4 | Non-Activator |
ASAHI INTECC CO., LTD.
{8}
FUBUKI XF-R Neurovascular Long Sheath
| Hemocompatibility
Partial Thromboplastin Time Test
(PTT)
(Direct Contact) | Per ISO 10993-4 | Non-Activator |
| --- | --- | --- |
| Thrombogenicity
In Vivo Thrombogenicity Study | Per ISO 10993-4 | Comparable
thromboresistance to
the predicate device. |
Sterilization and Shelf Life
The changes to the subject device do not impact sterilization and shelf life as compared to the predicate device. The shelf-life of the subject device was supported by device performance testing following 3-year accelerated aging.
Conclusion
The FUBUKI XF-R Neurovascular Long Sheath has the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, and operating principles as the predicate device. The differences between the subject and predicate device do not raise different or new questions of safety or effectiveness. The non-clinical tests demonstrate that the subject device performs as intended and that the FUBUKI XF-R Neurovascular Long Sheath is substantially equivalent to the predicate FUBUKI XF Neurovascular Long Sheath (K213589).
ASAHI INTECC CO., LTD.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.