K251534 · Foshan Kycare Medical Equipment Co., Ltd. · CAW · Dec 29, 2025 · Anesthesiology
Device Facts
Record ID
K251534
Device Name
Oxygen Concentrator (J10A)
Applicant
Foshan Kycare Medical Equipment Co., Ltd.
Product Code
CAW · Anesthesiology
Decision Date
Dec 29, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oxygen Concentrator provide supplemental oxygen to patients who require supplemental O2, by separating nitrogen from room air, by way of a molecular sieve, it can be used in the home or health care facility. It is not intended to sustain or support life.
Device Story
Portable oxygen concentrator; utilizes pressure swing adsorption (PSA) with molecular sieve to separate nitrogen from room air; produces 90-96% enriched oxygen at continuous flow rates of 1-10 L/min. Used in home or healthcare facilities; operated by patients or healthcare providers. Device monitors oxygen purity, pressure, and flow; provides visual/auditory alarms for low purity, pressure loss, or sensor failure. Output affects clinical decision-making by ensuring consistent oxygen delivery to patients with respiratory needs. Benefits include portable, continuous supplemental oxygen therapy.
Clinical Evidence
No clinical data. Bench testing only. Performance supported by electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 18562-2/3), and usability validation (IEC 62366-1).
Technological Characteristics
Materials: Synthetic Zeolite sieve bed. Principle: Pressure swing adsorption. Energy: AC 100-120V, 60Hz. Dimensions: 620 x 370 x 345mm. Connectivity: None (no external interfaces). Software: Microprocessor-based control for alarms and flow monitoring. Safety: Class II, Type BF. Standards: IEC 60601-1, IEC 60601-1-11, ISO 80601-2-69, IEC 60601-1-8.
Indications for Use
Indicated for adult patients (above 18 years old) requiring supplemental oxygen. Not intended to sustain or support life.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
December 29, 2025
Foshan KYCARE Medical Equipment Co., Ltd.
% Cassie Lee
Manager
Share Info (Guangzhou) Medical Consultant Ltd.
No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road
Huangpu District
Guangzhou, Guangdong 510530
China
Re: K251534
Trade/Device Name: Oxygen Concentrator (J10A)
Regulation Number: 21 CFR 868.5440
Regulation Name: Portable Oxygen Generator
Regulatory Class: Class II
Product Code: CAW
Dated: November 27, 2025
Received: November 28, 2025
Dear Cassie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K251534 - Cassie Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K251534 - Cassie Lee
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JAMES J. LEE -S
for Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Oxygen Concentrator
Page 9 of 39
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251534 | ? |
| Please provide the device trade name(s). | | ? |
| Oxygen Concentrator ( J10A) | | |
| Please provide your Indications for Use below. | | ? |
| The Oxygen Concentrator provide supplemental oxygen to patients who require supplemental O2, by separating nitrogen from room air, by way of a molecular sieve, it can be used in the home or health care facility. It is not intended to sustain or support life. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary of K251534
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
1. Submitter's Information
Company Name: Foshan KYCARE Medical Equipment Co., Ltd.
Address: Room 401, Building 29, Liandong Yougu Park, No 3, Dongsan District, Jiansha Road, Danzao Town, Nanhai District, Foshan City, Guangdong Province, China
Contact Person (including title): Leona Zhou (Regulatory Affair Manager)
Tel: 086 18666388285
Fax: /
Post code: 528216
E-mail: zhouliming@kycaremedical.com
Application Correspondent:
Contact Person: Ms. Cassie Lee
Share Info (Guangzhou) Medical Consultant Ltd.
Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China
Tel : +86 20 8266 2446
Email : 382198657@qq.com
2. Subject Device Information:
Type of 510(k): Traditional
Common Name: Portable Oxygen Concentrator
Classification Name: Generator, Oxygen, Portable
Trade Name: Oxygen Concentrator
Model Name: J10A
Review Panel: Anesthesiology
Product Code: CAW
Regulation Number: 868.5440
Regulatory Class: II
3. Predicate Device Information
Sponsor: FOSHAN CARE MEDICAL TECHNOLOGY CO., LTD.
Trade Name: Oxygen Concentrator
Model Name: ZY-10AB/115
Classification Name: Generator, Oxygen, Protable
510(k) Number: K240840
Review Panel: Anesthesiology
Product Code: CAW
Regulation Number: 868.5440
Regulatory Class: II
4. Device Description
The Oxygen Concentrator is a portable concentrator which could utilize a molecular sieve and pressure swing adsorption to produce the oxygen. The device could separate nitrogen by absorbing through the molecular sieve when the room air enters the device, and allows the enriched oxygen to be collected. The device could provide approximately 90~96% of oxygen to patients on a continuous flow basis at a rate of 1L/min to 10L/min. The nasal cannula is not sold with the device, users need to purchase the accessories by themselves that are legally marketed. The products have not undergone biocompatibility assessment of condensate precipitates according to ISO18562-4, and it is not recommended to use humidification bottles.
The maximum altitude the subject device can operate without degradation of concentration is 2000m.
1 / 4
{5}
The predicate device ZY-10AB/115 is the original device of subject device J10A, just different in the model name and the information of the 510k sponsor.
# 5. Intended Use / Indications for Use
The Oxygen Concentrator provide supplemental oxygen to patients who require supplemental O2, by separating nitrogen from room air, by way of a molecular sieve, it can be used in the home or health care facility. It is not intended to sustain or support life.
# 6. Intended patient population
Population: adult
Age: above 18 years old
Health status: adult patients requiring supplemental oxygen.
7. Comparison to predicate devices
| Elements of Comparison | Subject Device | Predicate Device | Verdict |
| --- | --- | --- | --- |
| Company | Foshan KYCARE Medical Equipment Co., Ltd. | FOSHAN CARE MEDICAL TECHNOLOGY CO., LTD. | -- |
| Trade Name | Oxygen Concentrator | Oxygen Concentrator | -- |
| Model | J10A | ZY-10AB/115 | -- |
| Classification Name | Generator, Oxygen, Portable | Generator, Oxygen, Portable | Same |
| 510(k) Number | K251534 | K240840 | -- |
| Product Code | CAW | CAW | Same |
| Intended Use / Indications for Use | The Oxygen Concentrator provide supplemental oxygen to patients who require supplemental O2, by separating nitrogen from room air, by way of a molecular sieve, it can be used in the home or health care facility. It is not intended to sustain or support life. | The Oxygen Concentrator provide supplemental oxygen to patients who require supplemental O2, by separating nitrogen from room air, by way of a molecular sieve, it can be used in the home or health care facility. It is not intended to sustain or support life. | Same |
| Work principle | Pressure swing adsorption principle, filtering by molecular sieve | Pressure swing adsorption principle, filtering by molecular sieve | Same |
| Working mode | Continuous | Continuous | Same |
| Environments of Use | Home or health care facility | Home or health care facility | Same |
| Intended patient population | Population: adultAge: above 18 years oldHealth status: adult patients requiring supplemental oxygen. | Population: adultAge: above 18 years oldHealth status: adult patients requiring supplemental oxygen. | Same |
| Performance Comparison | | | |
| Power supply | AC 100-120V, 60Hz | AC 100-120V, 60Hz | Same |
| Max.Input power | 550VA | 550VA | Same |
{6}
3 / 4
| Elements of Comparison | Subject Device | Predicate Device | Verdict |
| --- | --- | --- | --- |
| Oxygen concentration | 93% ± 3% | 93% ± 3% | Same |
| Oxygen flow | 1-10L/min | 1-10L/min | Same |
| Outlet pressure | 40kPa-80kPa | 40kPa-80kPa | Same |
| Noise | ≤60dB (A) under max. flowrate | ≤60dB (A) under max. flowrate | Same |
| Electrical classification | Class II Type BF | Class II Type BF | Same |
| Alarm | Low Oxygen purity; Low & high pressure; Power loss alarm; Oxygen Sensor failure; Flow rate less than 0.4L/Min; Display Circuit Board failure | Low Oxygen purity; Low & high pressure; Power loss alarm; Oxygen Sensor failure; Flow rate less than 0.4L/Min; Display Circuit Board failure | Same |
| Oxygen concentration warning | <82%, indicator yellow light and medium Blinking (0.4Hz-0.8Hz), auditory signal: Di-Di-Di 16±1 seconds interval and display error code “E2”; <73%±3%, indicator red light and fast Blinking (0.4Hz-0.8Hz), auditory signal: Di-Di-Di---Di-Di 4±1 seconds interval and display error code “E1” | <82%, indicator yellow light and medium Blinking (0.4Hz-0.8Hz), auditory signal: Di-Di-Di 16±1 seconds interval and display error code “E2”; <73%±3%, indicator red light and fast Blinking (0.4Hz-0.8Hz), auditory signal: Di-Di-Di---Di-Di 4±1 seconds interval and display error code “E1” | Same |
| Operating system | Time cycle/ Pressure swing adsorption | Time cycle/ Pressure swing adsorption | Same |
| Software/Hardware | Analog and digital electronics with microprocessor | Analog and digital electronics with microprocessor | Same |
| LCD display | Running time; Accumulating timing | Running time; Accumulating timing | Same |
| Accessories | manual | manual | Same |
| Dimensions | 620 x 370 x 345mm | 620 x 370 x 345mm | Same |
| Safety Comparison | | | |
| Sieve Bed | Synthetic Zeolite | Synthetic Zeolite | Same |
| Biocompatibility of materials contacting user | ISO 18562-2:2017, ISO 18562-3:2017 | ISO 18562-2:2017, ISO 18562-3:2017 | Same |
| EMC | Comply with IEC 60601-1-2:2014+A1:2020 | Comply with IEC 60601-1-2:2014+A1:2020 | Same |
| Electric safety | Comply with IEC 60601-1-11:2015/AMD1:2020, ISO 80601-2-69:2014, IEC 80601-1-8:2020 IEC 60601-1:2005/AMD2:2020 | Comply with IEC 60601-1-11:2015/AMD1:2020, ISO 80601-2-69:2014, IEC 80601-1-8:2020 IEC 60601-1:2005/AMD2:2020 | Same |
7. Test Summary
7.1 Non-Clinical Tests Performed
{7}
4 / 4
1) Electrical safety, and electromagnetic compatibility Test
Electrical safety and EMC testing were conducted on the proposed device. The device was shown to comply with IEC 60601-1-2:2020 and IEC TR 60601-4-2:2016 for electromagnetic compatibility, IEC 60601-1:2020, IEC 60601-1-8:2012, IEC 60601-1-11:2020 and ISO 80601-2-69:2020 for electrical safety.
2) Biocompatibility Test
Biocompatibility testing was performed on the subject device components in accordance with FDA guidance document entitled "Use of international Standard ISO 18562-1, "Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare Applications- Part 1: Evaluation and Testing within a Risk Management Process." The device was shown to comply with ISO 18562-2 and ISO 18562-3.
3) Software verification and validation
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
4) Usability validation
Usability evaluation was conducted and documented according to the recognized consensus standards of IEC 62366-1 and IEC 60601-6. The result of the usability demonstrated all the users in testing did not show any critical errors and represented the performance of using can be smoother by extra practices without specific usability problems. The device has been found to be reasonably safe and effective for the intended users, uses and uses environments.
5) Cybersecurity
The subject device no external interfaces, according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", no need cybersecurity evaluation.
7.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
8. Date of the summary prepared: November 27, 2025
9. Final Conclusion
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device K240840.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.