K251522 · Bone Solutions, Inc. · MQV · Oct 13, 2025 · Orthopedic
Device Facts
Record ID
K251522
Device Name
Mg OSTEOINJECT™; Mg OSTEOREVIVE™; Mg OSTEOCRETE™
Applicant
Bone Solutions, Inc.
Product Code
MQV · Orthopedic
Decision Date
Oct 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Mg OSTEOINJECT™ is intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. Mg OSTEOINJECT™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis. Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process. Mg OSTEOINJECT™ is intended to be placed into bony voids either before or after final fixation. Mg OSTEOINJECT™ is resorbed and replaced with bone during the healing process. Mg OSTEOINJECT™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously. Mg OSTEOREVIVE™ is intended for bony voids or defects of the extremities, posterolateral spine, intervertebral disc space, and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. Mg OSTEOREVIVE™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis. Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process. Mg OSTEOREVIVE™ is intended to be placed into bony voids either before or after final fixation. Mg OSTEOREVIVE™ is resorbed and replaced with bone during the healing process. Mg OSTEOREVIVE™ must be used with morselized autograft and/or allograft bone in the posterolateral spine. When used in intervertebral body fusion procedures Mg OSTEOREVIVE™ must be used with morselized autograft and/or allograft bone with an intervertebral body fusion device cleared by FDA for use with a bone void filler. Mg OSTEOREVIVE™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously. Mg OSTEOCRETE™ is intended for bony voids or defects of the extremities, posterolateral spine, intervertebral disc space, and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. Mg OSTEOCRETE™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis. Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process. Mg OSTEOCRETE™ is intended to be placed into bony voids either before or after final fixation. Mg OSTEOCRETE™ is resorbed and replaced with bone during the healing process. Mg OSTEOCRETE™ must be used with morselized autograft and/or allograft bone in the posterolateral spine. When used in intervertebral body fusion procedures Mg OSTEOCRETE™ must be used with morselized autograft and/or allograft bone with an intervertebral body fusion device cleared by FDA for use with a bone void filler. Mg OSTEOCRETE™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
Device Story
Magnesium-based synthetic bone void filler; supplied as powder component and buffered saline mixing solution. Intraoperative mixing triggers acid-base reaction forming cohesive, moldable, drillable, radiopaque paste. Applied to bony voids; adheres to bone during curing. Acts as temporary support medium; resorbed and replaced by bone during healing. Used in clinical settings by surgeons. Benefits include osteoconductive scaffold for bone healing in non-structural defects.
Clinical Evidence
No clinical data included. Evidence consists of non-clinical bench testing (chemical, physical, sterilization, biocompatibility) and animal model data supporting pediatric use safety (serum magnesium levels). Bacterial endotoxin testing (LAL) met specifications.
Technological Characteristics
Magnesium-based synthetic bone void filler. Components: powder and buffered saline. Acid-base reaction curing. Properties: moldable, drillable, resorbable, adhesive/cohesive, radiopaque, osteoconductive. Sterile, single-use. Volumes: 3 cc to 15 cc.
Indications for Use
Indicated for bony voids/defects of extremities, pelvis, posterolateral spine, and intervertebral disc space (not intrinsic to stability) in adults and pediatric patients ≥ 6 years old. Used for benign bone cysts/tumors, surgically created defects, or traumatic injury. Adjunct to rigid hardware in extremities/pelvis. Requires autograft/allograft in spine/fusion procedures.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 13, 2025
Bone Solutions, Inc.
% Kevin Thomas
VP & Director of Regulatory Affairs
PaxMed International, LLC
1925 Palomar Oaks Way
Suite 210
Carlsbad, California 92008
Re: K251522
Trade/Device Name: Mg OSTEOINJECT™; Mg OSTEOREVIVE™; Mg OSTEOCRETE™
Regulation Number: 21 CFR 888.3045
Regulation Name: Resorbable Calcium Salt Bone Void Filler Device
Regulatory Class: Class II
Product Code: MQV, MBP, OIS
Dated: September 12, 2025
Received: September 15, 2025
Dear Kevin Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251522 - Kevin Thomas
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251522 - Kevin Thomas
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert M. Stefani -S
Digitally signed by
Robert M. Stefani -S
Date: 2025.10.13 19:24:49
-04'00'
For: Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number: K251522
## Device Name
Mg OSTEOINJECT™, Mg OSTEOREVIVE™; Mg OSTEOCRETE™
Indications for Use (Describe)
## Mg OSTEOINJECT™
Mg OSTEOINJECT™ is intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone.
Mg OSTEOINJECT™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis.
Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process.
Mg OSTEOINJECT™ is intended to be placed into bony voids either before or after final fixation.
Mg OSTEOINJECT™ is resorbed and replaced with bone during the healing process.
Mg OSTEOINJECT™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
## Mg OSTEOREVIVE™
Mg OSTEOREVIVE™ is intended for bony voids or defects of the extremities, posterolateral spine, intervertebral disc space, and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone.
Mg OSTEOREVIVE™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis.
Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process.
Mg OSTEOREVIVE™ is intended to be placed into bony voids either before or after final fixation.
Mg OSTEOREVIVE™ is resorbed and replaced with bone during the healing process.
Mg OSTEOREVIVE™ must be used with morselized autograft and/or allograft bone in the posterolateral spine.
When used in intervertebral body fusion procedures Mg OSTEOREVIVE™ must be used with morselized autograft and/or allograft bone with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Mg OSTEOREVIVE™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
## Mg OSTEOCRETE™
Mg OSTEOCRETE™ is intended for bony voids or defects of the extremities, posterolateral spine, intervertebral disc space, and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone.
Mg OSTEOCRETE™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis.
Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process.
Mg OSTEOCRETE™ is intended to be placed into bony voids either before or after final fixation.
Mg OSTEOCRETE™ is resorbed and replaced with bone during the healing process.
Mg OSTEOCRETE™ must be used with morselized autograft and/or allograft bone in the posterolateral spine.
When used in intervertebral body fusion procedures Mg OSTEOCRETE™ must be used with morselized autograft and/or allograft bone with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Mg OSTEOCRETE™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
{4}
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
Page 12 of 12
{5}
510(k) Summary
Page 1 of 3
Mg OSTEOINJECT™, Mg OSTEOREVIVE™, Mg OSTEOCRETE™
K251522
# 510(k) Summary
# K251522
# Mg OSTEOINJECT™, Mg OSTEOREVIVE™, Mg OSTEOCRETE™
# Bone Solutions, Inc.
October 10, 2025
# ADMINISTRATIVE INFORMATION
Manufacturer Name
Bone Solutions, Inc.
5712 Colleyville Boulevard, Suite 210
Colleyville, Texas 76034
Telephone +1 817-809-8850
Official Contact
Drew Diaz, CEO
Representative/Consultant
Kevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
Rebecca E. Kattan, PhD
PaxMed International, LLC
1925 Palomar Oaks Way, Suite 210
Carlsbad, CA 92008
Telephone +1 858-792-1235
Email kthomas@paxmed.com
flarson@paxmed.com
rkattan@paxmed.com
# DEVICE NAME AND CLASSIFICATION
Trade/Device Name
Mg OSTEOINJECT™, Mg OSTEOREVIVE™, Mg OSTEOCRETE™
Common Name
Filler, bone void, calcium compound
Regulation Number
21 CFR 888.3045
Regulation Name
Resorbable calcium salt bone void filler device
Regulatory Class
Class II
Product Code
MQV
Secondary Product Codes
MBP, OIS
Classification Panel
Orthopedic
Reviewing Office
Office of Health Technology 6 (Orthopedic Devices)
Reviewing Division
Division of Health Technology 6 C
(Restorative, Repair and Trauma Devices)
# PREDICATE DEVICE INFORMATION
Primary Predicate Device
K242372, Mg OSTEOREVIVE™, Mg OSTEOCRETE™, Bone Solutions Inc.
Reference Devices
K234013, Mg OSTEOINJECT™, Mg OSTEOREVIVE™, Mg OSTEOCRETE™, Bone Solutions Inc.
K231528, Bonalive Orthopedics granules, Bonalive Biomaterials, Ltd.
{6}
510(k) Summary
Page 2 of 3
Mg OSTEOINJECT™, Mg OSTEOREVIVE™, Mg OSTEOCRETE™
K251522
# INDICATIONS FOR USE STATEMENT
## Mg OSTEOINJECT™
Mg OSTEOINJECT™ is intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone.
Mg OSTEOINJECT™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis.
Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process.
Mg OSTEOINJECT™ is intended to be placed into bony voids either before or after final fixation.
Mg OSTEOINJECT™ is resorbed and replaced with bone during the healing process.
Mg OSTEOINJECT™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
## Mg OSTEOREVIVE™
Mg OSTEOREVIVE™ is intended for bony voids or defects of the extremities, posterolateral spine, intervertebral disc space, and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone.
Mg OSTEOREVIVE™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis.
Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process.
Mg OSTEOREVIVE™ is intended to be placed into bony voids either before or after final fixation.
Mg OSTEOREVIVE™ is resorbed and replaced with bone during the healing process.
Mg OSTEOREVIVE™ must be used with morselized autograft and/or allograft bone in the posterolateral spine.
When used in intervertebral body fusion procedures Mg OSTEOREVIVE™ must be used with morselized autograft and/or allograft bone with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Mg OSTEOREVIVE™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
## Mg OSTEOCRETE™
Mg OSTEOCRETE™ is intended for bony voids or defects of the extremities, posterolateral spine, intervertebral disc space, and pelvis that are not intrinsic to the stability of the bony structure. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), may be surgically created osseous defects or osseous defects created by traumatic injury to the bone.
Mg OSTEOCRETE™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis.
Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process.
Mg OSTEOCRETE™ is intended to be placed into bony voids either before or after final fixation.
Mg OSTEOCRETE™ is resorbed and replaced with bone during the healing process.
Mg OSTEOCRETE™ must be used with morselized autograft and/or allograft bone in the posterolateral spine.
When used in intervertebral body fusion procedures Mg OSTEOCRETE™ must be used with morselized autograft and/or allograft bone with an intervertebral body fusion device cleared by FDA for use with a bone void filler.
Mg OSTEOCRETE™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
# SUBJECT DEVICE DESCRIPTION
This submission includes three (3) devices with separate trade names bundled into the single 510(k) application. The purpose of this application is to expand the indications to include specific language for use in pediatric patients ≥ 6 years old. The subject devices are a magnesium-based synthetic bone void filler that is moldable, drillable, resorbable, adhesive/cohesive, radiopaque, and osteoconductive. The subject devices consist of a powder component (magnesium-based compound) and a mixing solution (buffered saline). Once the components are mixed intraoperatively prior to implantation, an acid-base reaction occurs to form a cohesive paste. Once the product is placed into the bony void, the paste will adhere to the adjacent bone during the curing process. The devices are provided sterile to the end user for single-use only in various sizes from 3 cc to 15 cc.
{7}
510(k) Summary Page 3 of 3
Mg OSTEOINJECT™, Mg OSTEOREVIVE™, Mg OSTEOCRETE™ K251522
# PERFORMANCE DATA
Non-clinical testing data according to the guidance documents Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device (issued June 2003) and Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (issued January 2024) were referenced from K242372 and K234013. The non-clinical testing data leveraged to demonstrate substantial equivalence included: chemical composition, physical properties, sterilization, sterile barrier shelf life, product shelf life, and biocompatibility. Performance testing data also leveraged demonstrated that the subject device is drillable, and may be used as an adjunct to conventional rigid hardware during the surgical procedure (only when used in the extremities and pelvis). Animal testing data to support the indications for use as a bone void filler also were referenced in K242372 and K234013.
Additional non-clinical data provided to support the indication for use in pediatric patients included a toxicological risk assessment, and an animal model to demonstrate that serum magnesium levels in animals implanted with the subject device did not significantly increase over time, and were not significantly different from control animals.
Bacterial endotoxin testing has been performed to ensure the device meets pyrogen limit specifications. The Limulus amebocyte lysate (LAL) test, kinetic turbidimetric method, was performed according to USP <85> Bacterial Endotoxins Test. The LAL testing met the limit acceptance criterion of ≤ 20 EU/device, based upon the recommendations for implanted devices in the FDA guidance document Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, (issued January 2024), Section V, A, 4.d.
No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICES
The subject device, the primary predicate device K242372, and the reference devices K234013 and K231528 have the same intended use, the same product classification, product codes (MQV for all, MBP for K242372, and OIS for K242372 and K234013), and have similar Indications for Use statements, with the additions to the subject device Indications for Use described above. Although the subject device, the primary predicate device, and the reference devices have slightly different Indications for Use language, these differences in language do not change the intended use as a bone void filler. The subject device, the primary predicate device K242372 and the reference device K234013 are made of identical materials, provided in the same range of physical dimensions (volumes), are packaged in identical materials, and are sterilized using identical methods.
Differences between the subject device and the reference device K231528 include the exact indications for use language, the mineral components, and the range of sizes (volumes). These minor differences do not raise new issues of safety or effectiveness, and therefore, do not impact substantial equivalence.
# CONCLUSION
The subject device, the primary predicate device K242372, and the reference device K234013 have the same intended use, have identical technological characteristics, and are made of identical materials. The subject device, the primary predicate device K242372, and the reference device K234013 encompass the same range of physical dimensions (volumes), are packaged in identical materials and are sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the primary predicate device and the reference devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.