Solidex® Ti-Links and Screws

K251515 · CreoDent Hudson Valley · NHA · Feb 10, 2026 · Dental

Device Facts

Record IDK251515
Device NameSolidex® Ti-Links and Screws
ApplicantCreoDent Hudson Valley
Product CodeNHA · Dental
Decision DateFeb 10, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

Solidex® Ti-Links and Screws is intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Solidex® Ti-Links is to be sent to a CREODENT HUDSON VALLEY validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories.

Device Story

Two-piece dental abutment system comprising titanium alloy base (Ti-Link) and CAD/CAM-designed zirconia superstructure; includes titanium fixation screw. Used in dental clinics/labs; operated by clinicians/technicians. Digital workflow utilizes intra-oral/desktop scanner data, CAD software (3Shape Abutment Designer), and CAM milling to fabricate patient-specific superstructures. Design library enforces parameters (wall thickness, post height, angulation) to ensure structural integrity. Superstructure cemented to Ti-Link in lab; final assembly sterilized before clinical placement. Provides support for prosthetic crowns; restores chewing function. Benefits include patient-specific fit and integration with existing digital dentistry workflows.

Clinical Evidence

No clinical data included. Evidence consists of bench testing: biocompatibility (ISO 10993-5), moist heat sterilization validation (ANSI/AAMI ST79), mechanical fatigue testing (ISO 14801), software verification of design constraints, and reverse engineering analysis for implant compatibility. MRI compatibility established via scientific rationale and literature review.

Technological Characteristics

Two-piece abutment system; Ti-6Al-4V ELI titanium alloy (ASTM F136). CAD/CAM zirconia superstructure (Y-TZP). Machined surface finish. Dimensions: 0.5-6.0mm gingival height, 0-20° angulation. Connectivity: Digital library integration with 3Shape CAD software. Sterilization: Moist heat (autoclave) post-cementation.

Indications for Use

Indicated for patients requiring prosthetic restoration support via endosseous dental implants in the maxillary or mandibular arch.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 10, 2026 Creodent Hudson Valley Calvin Shim Vice President 1769 Rt 52 Fishkill, New York 12524 Re: K251515 Trade/Device Name: Solidex® Ti-Links and Screws Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: May 2, 2025 Received: January 9, 2026 Dear Calvin Shim: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251515 - Calvin Shim Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K251515 - Calvin Shim Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Andrew I. Steen -S Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K251515 Device Name Solidex® Ti-Links and Screws **Indications for Use (Describe)** Solidex® Ti-Links and Screws is intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Solidex® Ti-Links is to be sent to a CREODENT HUDSON VALLEY validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories. **Compatible Implant Systems** | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter(mm) | 510(k) Number | | --- | --- | --- | --- | | Straumann Bone Level | 3.3 | 2.8 (NC) | K083550 | | | 4.1, 4.8 | 3.3 (RC) | K121131 | | Megagen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 | K122231 | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} CREODENT CREODENT HUDSON VALLEY 1769 Rt 52 Fishkill, NY 12524 USA # 510(k) Summary (K251515) This 510(k) Summary is being submitted in accordance with requirement of 21 CFR 807.92. ## Submitter : CREODENT HUDSON VALLEY 1769 Rt 52 Fishkill, NY 12524, USA Phone : +1.845.562.1555 ## Official Correspondent : CREODENT HUDSON VALLEY / Calvin Shim 1769 Rt 52 Fishkill, NY 12524, USA Phone : +1.845.562.1555 ## Device Information : Device Name : Solidex® Ti-Links and Screws Common Name : abutment, implant, dental, endosseous Regulation Name : Endosseous dental implant abutment Classification : II Primary Product Code : NHA Secondary Product Code : PNP Regulation Number : 872.3630 Date Prepared : Feb 10, 2026 ## Predicate Device - K240982, Terrats Medical SL, DESS Dental Smart Solutions ## Device Description Solidex® Ti-Links & Screws consists of a two-piece abutment, where the titanium base will be used to support a CAD/CAM designed superstructure (the second part of the two-piece abutment) that composes the final abutment. The system also includes a Solidex® Screw for fixation to the implant body. Solidex® Ti-Link are made of titanium alloy conforming to ASTM F136 standard specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications and are provided in various prosthetic platform diameters. Solidex® Screw is composed of titanium alloy per ASTM F136. Solidex® Ti-Links & Screws is designed for fabrication of a patient-specific CAD/CAM zirconia superstructure on which a crown may be placed. There are two-piece abutments for which the second part (or top half) is the zirconia superstructure. They also may be used for support of a crown directly on the Ti-Link. CAD/CAM zirconia superstructure fabrication for Ti-Link is by prescription on the order of the clinician. Solidex® Ti-Links & Screws all zirconia superstructures for use with the subject device Ti-Link will be manufactured using a digital dentistry workflow, either at a Validated Milling Center (VMC) or at the Point of Care (POC), by using scan files from intra-oral and lab (desktop) scanners, CAD/CAM software, ceramic material, and associated tooling and accessories. The digital dentistry workflow uses scan files from intra-oral and lab (desktop) scanners, CAD/CAM software, ceramic material, and associated tooling and accessories. The digital workflow includes the following products : - Ceramic material : Sagemax NexxZr T (K130991) - Cement : Panavia SA cement Universal (K183537) - Desktop scanner : 3Shape E series Lab Scanner (510(k) exempt under 21 CFR 872.3661) - Design software : 3Shape Abutment Designer Software (K151455) - Milling machine : PrograMill PM7 (510(k) exempt under 21 CFR 872.3661) - Milling software : PrograMill CAM (510(k) exempt under 21 CFR 872.3661) {5} CREODENT CREODENT HUDSON VALLEY 1769 Rt 52 Fishkill, NY 12524 USA - Digital Library / Templates (3 Shape) : The Solidex® Ti-Link CAD/CAM System shall only be used with the official “Solidex® Ti-Link Digital Library” and templates identified. The library is imported into the 3Shape software, and the design restriction parameters (minimum wall thickness, post height, maximum gingival height, etc.) are automatically applied. The official “Solidex® Ti-Link Digital Library” can be obtained from CREODENT HUDSON VALLEY by contacting support email (inquiry@creodental.com) or from CREODENT HUDSON VALLEY web-site. When imported into the 3Shape software, the design restriction parameters (minimum wall thickness, post height, maximum gingival height, etc.) and built-in protection features (“stops” and restriction shells) are automatically applied. The design parameters for the CAD/CAM zirconia superstructure to be used Solidex® Ti-Links & Screws: - Minimum wall thickness: 0.5 mm - Minimum abutment post height: 6.0 mm (Note) : The abutment post height is the dimension above the gingival height. : Feature maximum abutment post height are based on patient anatomy and clinician discretion. - Minimum gingival height: 0.5 mm - Maximum gingival height: 6.0 mm - Minimum Angulation: 0° - Maximum Angulation: 20° The zirconia superstructure must be cemented to the Solidex® Ti-Links using Panavia SA cement Universal in a dental laboratory. Solidex® Ti-Links & Screws is provided non-sterile therefore must be sterilized after the cementation of the zirconia superstructure on Solidex® Ti-Links. Solidex® Screws are designed to attach the abutment to the implant. ## Indications for Use Solidex® Ti-Links and Screws is intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Solidex® Ti-Links is to be sent to a CREODENT HUDSON VALLEY validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories. ### Compatible Implant Systems | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) | 510(k) Number | | --- | --- | --- | --- | | Straumann Bone Level | 3.3 | 2.8 (NC) | K083550 | | | 4.1, 4.8 | 3.3 (RC) | K121131 | | Megagen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 | K122231 | ## Substantial Equivalence [Predicate device] K240982, Terrats Medical SL, DESS Dental Smart Solutions [Substantial equivalence summary] The subject device is substantially equivalent in indications and design principles to the predicate device listed above. Provided at the end of this summary is a table comparing the Indications for Use and the technological characteristics of the subject device and the predicate device. The indications for use for the subject device is identical to that of the predicate device except for wording regarding product name and the validated milling center name. The differences in the indications for use between the subject and predicate device do not substantively alter their meaning. The design (Abutment Designs, Prosthesis Attachment, Restoration, Platform Size, and Abutment Angle) is substantially equivalent to that of predicate device. {6} CREODENT CREODENT HUDSON VALLEY 1769 Rt 52 Fishkill, NY 12524 USA The material of Ti-Links, Screw, and Material of Zirconia Superstructure is substantially equivalent to that of predicate device. Regarding surface treatment, the predicate device is surface-treated, whereas the subject device is not surface-treated and is made of a material that conforms to the ASTM F136 standard (Ti-6Al-4V ELI). The manufacturing and digital design workflow for subject device is identical to that of predicate device. Digital manufacturing workflow of the subject device and predicate device are similar. The difference is in the manufacturer of the equipment and software used. The differences in the digital manufacturing workflow between the subject and predicate device do not affect the determination of substantial equivalence. Since the features of the subject device are substantially equivalent to the above-mentioned devices, any minor differences in certain features compared to the predicate device are insignificant and do not affect the determination of substantial equivalence. [Comparison between Subject and Predicate Device] | FEATURE | Subject device | Predicate device | Comparison Discussion | | --- | --- | --- | --- | | 510(k) Number | K251515 | K240982 | Not applicable | | Company Name | CREODENT HUDSON VALLEY | Terrats Medical SL | Not applicable | | Trade Name | Solidex® Ti-Links and Screws | DESS Dental Smart Solutions | Not applicable | | Product Code | NHA, PNP | NHA, PNP | Identical | | Reason for Predicate Device | Not applicable | Intended Use, Indications for Use, Design (Abutment Designs, Prosthesis Attachment, Restoration, Platform Size, Abutment Angle), Material (Material for Ti-Links, Surface Treatment for Ti-Links, Material for Screw, Surface Treatment for Screw, Material for Zirconia Superstructure), Manufacturing(Manufacturing, Digital Design Workflow, Digital Manufacturing Workflow) | Not applicable | | Intended Use | Support of a prosthesis to restore chewing function | Support of a prosthesis to restore chewing function | Identical | | Indications for Use | Solidex® Ti-Links and Screws is intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Solidex® Ti-Links is to be sent to a CREODENT HUDSON VALLEY validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories. | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with DESS Bases or Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture, or to be designed and manufactured according to the digital dentistry workflow. The digital dentistry workflow integrates multiple components: scan files from intra-oral and lab (desktop) scanners, CAD software, CAM software, ceramic material, milling machine, and associated tooling and accessories. | This indications for use of the subject and predicate device are similar The indications for use of the subject and predicate device have slight differences in wording, for clarification only. For the subject and predicate device, there is a difference only in the product name and the validated milling center name. The differences in the indications for use | {7} CREODENT CREODENT HUDSON VALLEY 1769 Rt 52 Fishkill, NY 12524 USA | FEATURE | Subject device | Predicate device | Comparison Discussion | | --- | --- | --- | --- | | | | | between the subject and predicate device do not substantively alter their meaning. | | Design | | | | | Abutment Designs | CAD/CAM Bases | CAD/CAM Bases, CAD/CAM Blanks | Similar | | Prosthesis Attachment | Cement-retained, Screw re-tained | Cement-retained, Screw re-tained | Identical | | Restoration | Single-unit, Multi-unit | Single-unit, Multi-unit | Identical | | Platform Size | Ø2.8mm – Ø5.7mm | Ø2.52mm – Ø6.5mm | Platform size is covered by predicate device. | | Angulation of base component | 0° for Ti-Links | Max 30° for Blanks, 0° for TiBases | Similar | | Angulation of the final two piece abutment (base and top half combined) | Max 20° | Max 30°(for Pre-milled Blanks), 0° (straight only for Ti bases configurations) | Similar | | Material | | | | | Material of Ti-Links | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Identical | | Surface Treatment of Ti-Links | Machined | Anodization and a SelectGrip® surface | The subject device is not surface-treated and is made of material conforming to ASTM F136 standard (Ti-6Al-4V ELI). | | Material of Screw | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Identical | | Surface Treatment of Screw | Machined | DLC Coating | The subject device is not surface-treated and is made of material conforming to ASTM F136 standard (Ti-6Al-4V ELI). | | Material of Zirconia Superstructure | Zirconium Oxide (Y-TZP), K130991 | Zirconium Oxide (Y-TZP), K180703 | Similar (510(k) Clearance) | | Manufacturing | | | | | Manufacturing | Digital Dentistry Workflow & Validated Milling Center | Digital Dentistry Workflow & Validated Milling Center | Similar | | Digital Design Workflow | • 3Shape Intraoral Scanner Trios series • 3Shape E-series Lab Scanner • 3Shape Abutment Designer Software(3Shape A/S, K151455) | • 3Shape Intraoral Scanner Trios series • 3Shape E-series and D/R2000 Lab Scanner • 3Shape Abutment Designer Software(3Shape A/S, K151455) | Similar | | Digital Manufacturing Workflow | PrograMill Dry By Ivoclar with PrograMill CAM | VHF R5 By vhf camfacture AG with DentalCAM & DentalCNC | Digital Manufacturing Workflow of the subject and predicate device are similar. The difference is in | {8} CREODENT CREODENT HUDSON VALLEY 1769 Rt 52 Fishkill, NY 12524 USA | FEATURE | Subject device | Predicate device | Comparison Discussion | | --- | --- | --- | --- | | | | | the manufacturer of the equipment and software used. The differences in the digital manufacturing workflow between the subject and predicate device do not affect the determination of substantial equivalence. | ## Non-clinical test data The following tests were performed for the subject device : - Biocompatibility testing in accordance with ISO 10993-1:2018 : Cytotoxicity(Elution, MTT) in accordance with ISO 10993-5:2009 - User moist heat sterilization validation in accordance with ANSI/AAMI ST79:2017 with Amendments A1:2020, A2:2020, A3:2020, A4:2020, ISO 17665:2024, ISO 11138-1:2017, ISO 11138-3:2017, ISO 11737-1:2018/Amd 1:2021, ISO 11737-2:2019, and ISO 11140-3:2007 - Mechanical testing(fatigue test) in accordance with ISO 14801:2016 - Software verification included testing of restrictions that prevent design of components outside of the stated design parameters. In addition, the abutment design library was validated to demonstrate that the established design limitations and specifications are locked and cannot be modified by the user. - Reverse engineering analysis was conducted on OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility with the 3rd party implant systems. - MR Compatibility Statement : A non-clinical worst-case MRI review was performed to evaluate the subject device in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795). This MRI review was leveraged for the subject device. The results of the non-clinical testing confirm that the subject device meets the criteria of the applicable standards and is substantially equivalent to the predicate device. No clinical data were included in this submission. ## Conclusion Based on the information provided in this premarket notification, CREODENT HUDSON VALLEY concludes that the Solidex® Ti-Links and Screws is substantially equivalent to the above-mentioned currently cleared devices. End of Document
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...