← Product Code [DQD](/productcode/DQD) · K251494

# Eko Foundation Analysis Software with Transformers (EFAST) (K251494)

_Eko Health, Inc. · DQD · Aug 12, 2025 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K251494

## Device Facts

- **Applicant:** Eko Health, Inc.
- **Product Code:** [DQD](/productcode/DQD.md)
- **Decision Date:** Aug 12, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1875
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** AI/ML, Software as a Medical Device, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K251494 · Aug 12, 2025 | Eko Foundation Analysis Software with Transformers (EFAST) | Eko Health, Inc. | Retrospective de-identified heart sound recordings; Retrospective de-identified ECG recordings | Retrospective clinical datasets were used to evaluate the performance (sensitivity/specificity) of the EFAST algorithm for structural murmur classification and atrial fibrillation detection. | Retrospective study; Algorithm validation; Structural murmur classification; Atrial fibrillation detection |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective Structural Murmur Classification Study; Retrospective study of de-identified data; Follow-up/Duration: 15-second recordings | 615 unique subjects (175 pediatric, 440 adult); Sample Size: 2,460 heart sound recordings; Number of Sites: 3 | Not applicable for this study | Sensitivity and specificity for structural murmur classification |
| Retrospective ECG Rhythm Classification Study; Retrospective study of de-identified data; Follow-up/Duration: 15-second recordings | 314 unique subjects; Sample Size: 1,256 ECG recordings; Number of Sites: 2 | Not applicable for this study | Sensitivity and specificity for atrial fibrillation detection |

## Indications for Use

The Eko Foundation Analysis Software with Transformers (EFAST) is intended to support healthcare professionals in the evaluation of patients' heart sounds and electrocardiograms (ECGs). The software calculates heart rate, detects the presence of murmurs associated with Structural Heart Disease, and determines murmur timing, including the timing of S1, S2 heart sounds. When ECG is available, the software detects the presence of atrial fibrillation and normal sinus rhythm. The software is intended for use by or under the supervision of a healthcare professional and is not to be used as a sole means of diagnosis. The interpretations of heart sounds and ECG offered by the software are designed to support and not replace the healthcare professional's clinical judgment. The murmur detection is intended for use on both pediatric and adult patients, while the detection of atrial fibrillation is intended for use on adults (> 18 years).

## Device Story

Cloud-based SaMD providing clinical decision support for heart sound (PCG) and ECG analysis. Inputs: .WAV files of heart sounds and ECGs recorded via Eko Digital Stethoscopes, stored in AWS S3. Processing: Deep Neural Networks and signal processing algorithms analyze inputs to classify structural murmurs, determine S1/S2 timing, detect atrial fibrillation/sinus rhythm, and calculate heart rate. Output: JSON data structure returned via API for display on integrated software/apps. Used in clinical settings by healthcare professionals. Benefits: Assists clinicians in identifying structural heart disease and arrhythmias, supporting clinical decision-making without replacing professional judgment.

## Clinical Evidence

Retrospective studies on de-identified data. Structural murmur classification: 2,460 recordings from 615 subjects (28.5% pediatric, 71.5% adult); sensitivity 83.4%, specificity 86.0%. Heart sound timing: S1 sensitivity 98.58%, S2 sensitivity 94.81%. ECG rhythm classification: 1,256 recordings from 314 subjects; atrial fibrillation detection sensitivity 94.7%, specificity 94.1%. Heart rate calculation error <5%. Subgroup analysis showed no significant performance differences by sex or race; pediatric specificity for murmur detection was lower (76.0%) than adults (93.9%).

## Technological Characteristics

Cloud-based SaMD; utilizes Deep Neural Networks and signal processing. Inputs: PCG and ECG waveforms (.WAV). Connectivity: AWS S3 cloud storage, API-based integration. Software algorithm class: Deep Learning/ML. No physical materials (software-only).

## Regulatory Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Eko Murmur Analysis Software (EMAS) ([K213794](/device/K213794.md))

## Reference Devices

- Eko Analysis Software (EAS) ([K192004](/device/K192004.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

August 12, 2025

Eko Health, Inc.
Sam Huang
Director of Regulatory Affairs
2100 Powell Street
Suite 300
Emeryville, California 94608

Re: K251494
Trade/Device Name: Eko Foundation Analysis Software with Transformers (EFAST)
Regulation Number: 21 CFR 870.1875
Regulation Name: Stethoscope
Regulatory Class: Class II
Product Code: DQD, DQC, DPS, MWI
Dated: May 9, 2025
Received: May 15, 2025

Dear Sam Huang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251494 - Sam Huang
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K251494 - Sam Huang
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K251494

Device Name
Eko Foundation Analysis Software with Transformers (EFAST)

Indications for Use (Describe)
The Eko Foundation Analysis Software with Transformers (EFAST) is intended to support healthcare professionals in the evaluation of patients' heart sounds and electrocardiograms (ECGs). The software calculates heart rate, detects the presence of murmurs associated with Structural Heart Disease, and determines murmur timing including the timing of S1, S2 heart sounds. When ECG is available, the software detects the presence of atrial fibrillation and normal sinus rhythm.

The software is intended for use by or under the supervision of a healthcare professional and is not to be used as a sole means of diagnosis. The interpretations of heart sounds and ECG offered by the software are designed to support and not replace the healthcare professional's clinical judgment. The murmur detection is intended for use on both pediatric and adult patients, while the detection of atrial fibrillation is intended for use on adults (&gt; 18 years).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740

{4}

K251494

# GENERAL INFORMATION

Applicant:
Eko Health, Inc.
2100 Powell Street, Suite 300
Emeryville, CA 94608
Phone: 844-356-3384

Contact Person:
Sam Huang, PhD
Director of Regulatory Affairs
Eko Health, Inc.
2100 Powell Street, Suite 300
Emeryville, CA 94608

Date Prepared: May 13, 2025

# DEVICE INFORMATION

Trade/Proprietary Name: Eko Foundation Analysis Software with Transformers (EFAST)
Regulation number: 21 CFR 870.1875
Device Classification Name: Stethoscope
Regulatory Class: Class II
Product Code: DQD, DQC, DPS, MWI

# PREDICATE DEVICE

Primary Predicate: Eko Murmur Analysis Software (EMAS, K213794)
Reference Device: Eko Analysis Software (EAS, K192004)

# DEVICE DESCRIPTION

Eko Foundation Analysis Software with Transformers (EFAST) is a cloud-based Software as a Medical Device (SaMD) intended to provide clinical decision support to healthcare professionals (HCP) in the evaluation of patients' heart sounds (phonocardiogram, PCG) and electrocardiograms (ECGs). The software employs signal processing techniques and machine learning (Deep Neural Networks) to perform simultaneous analysis of recorded heart sounds and ECG data (when available), and identify the presence of murmurs associated with Structural Heart Disease, and determine murmur timing including the timing of S1, S2 heart sounds. When ECG is available, the software also detects the presence of atrial fibrillation and sinus rhythm.

K251494
Page 1 of 10

{5}

The software does not identify other arrhythmias. In addition, it calculates the heart rate of the patient.

The EFAST Software accepts input consisting of heart sounds and ECG waveforms recorded using Eko Digital Stethoscopes that are saved in .WAV file format and stored in the Eko Cloud, which utilizes the Amazon Web Services (AWS) Simple Storage Service (S3) for data storage and for easy communication with the EFAST analysis server. When an API Request is sent to the EFAST API, the corresponding recordings (PCG and/or ECG) are accessed by the EFAST software and they are analyzed by the EFAST software components. After analysis, the EFAST software returns a JSON format data structure with the algorithm results, which is saved back to the Eko Cloud and can be displayed on EFAST API integrated target software or devices (e.g. user-facing apps).

The EFAST Software consists of the following algorithm components:

- Heart Sound Signal Quality Classifier: Evaluates whether an audio recording contains heart sounds of sufficient quality for further clinical analysis
- Structural Murmur Classifier: Determines whether a good-quality heart sound recording contains a structural murmur or not
- Heart Sound Timing &amp; Murmur Timing: Determines the timing of S1 and S2 heart sounds and determines whether a structural murmur occurs during the systolic or diastolic phase
- ECG Rhythm Classifier: Analyzes ECG signals and categorizes them into one of four classifications: Sinus Rhythm, Atrial Fibrillation, Unspecified Rhythm, and Poor ECG Signal
- Heart Rate Calculation: Signal processing algorithm that utilizes ECG and PCG to calculate heart rate value in beats per minute

## INDICATIONS FOR USE

The Eko Foundation Analysis Software with Transformers (EFAST) is intended to support healthcare professionals in the evaluation of patients' heart sounds and electrocardiograms (ECGs). The software calculates heart rate, detects the presence of murmurs associated with Structural Heart Disease, and determines murmur timing, including the timing of S1, S2 heart sounds. When ECG is available, the software detects the presence of atrial fibrillation and normal sinus rhythm.

The software is intended for use by or under the supervision of a healthcare professional and is not to be used as a sole means of diagnosis. The interpretations of heart sounds and ECG offered by the software are designed to support and not replace the healthcare professional's clinical judgment. The murmur detection is intended for use on both pediatric and adult patients, while the detection of atrial fibrillation is intended for use on adults (&gt; 18 years).

K251494
Page 2 of 10

{6}

# SUBSTANTIAL EQUIVALENCE

The Eko Foundation Analysis Software with Transformers (EFAST) has the same intended use and fundamental technological characteristics as both the predicate (Eko Murmur Analysis Software, EMAS). Differences in ECG analysis, patient population, and technological characteristics do not raise new questions on safety or effectiveness compared to the predicate and reference devices. A substantial equivalence comparison table between the subject device and the predicate device is provided below.

Table 1: Substantial Equivalence Comparison Table

|  Feature | Subject Device: Eko Foundation Analysis Software with Transformers (EFAST) | Predicate Device: Eko Murmur Analysis Software (EMAS) (K213794) | Reference Device: Eko Analysis Software (EAS) (K192004)  |
| --- | --- | --- | --- |
|  Regulation Number and Name | 21 CFR 870.1875 Stethoscope | 21 CFR 870.1875 Stethoscope | 21 CFR 870.2300 Cardiac Monitor (Including Cardiotachometer And Rate Alarm)  |
|  Classification Product Code | DQD, DQC, DPS, MWI | DQD, DQC, DPS | MWI, DQD, DPS  |
|  Indications for Use | The Eko Foundation Analysis Software with Transformers (EFAST) is intended to support healthcare professionals in the evaluation of patients' heart sounds and electrocardiograms (ECGs). The software calculates heart rate, detects the presence of murmurs associated with Structural Heart Disease, and determines murmur timing including the timing of S1, S2 heart sounds. When ECG is | The Eko Murmur Analysis Software is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms (and ECG signals, when available). The software will automatically detect murmurs that may be present, and the murmur timing and character, including S1, S2, innocent heart murmurs, structural | The Eko Analysis Software is intended to provide support to the physician in the evaluation of patients' heart sounds and ECG's. The software analyzes simultaneous ECG and heart sounds. The software will detect the presence of suspected murmurs in the heart sounds. The software also detects the presence of atrial fibrillation and normal sinus rhythm from the ECG signal. In  |
|   | murmur timing, the timing of S1, S2, and the non-accidental heart murmurs is not specified. The Eko Analysis Software is intended to provide support to the physician in the evaluation of patients' heart sounds and ECG's. The software also detects the presence of suspected murmurs in the heart sounds. The software also detects the presence of atrial fibrillation and normal sinus rhythm from the ECG signal. In | murmur timing, the Eko Analysis Software is intended to provide support to the physician in the evaluation of patients' heart sounds and ECG's. The software also detects the presence of suspected murmurs in the heart sounds. The software also detects the presence of atrial fibrillation and normal sinus rhythm from the ECG signal. In | murmur timing, the Eko Analysis Software is intended to provide support to the physician in the evaluation of patients' heart sounds and ECG's. The software also detects the presence of atrial fibrillation and normal sinus rhythm from the ECG signal. In  |

K251494

{7}

K251494
Page 4 of 10
|  Feature | Subject Device: Eko Foundation Analysis Software with Transformers (EFAST) | Predicate Device: Eko Murmur Analysis Software (EMAS) (K213794) | Reference Device: Eko Analysis Software (EAS) (K192004)  |
| --- | --- | --- | --- |
|   | available, the software detects the presence of atrial fibrillation and normal sinus rhythm.

The software is intended for use by or under the supervision of a healthcare professional and is not to be used as a sole means of diagnosis. The interpretations of heart sounds and ECG offered by the software are designed to support and not replace the healthcare professional’s clinical judgment. The murmur detection is intended for use on both pediatric and adult patients, while the detection of atrial fibrillation is intended for use on adults (> 18 years). | heart murmurs, and the absence of a heart murmur.

The Eko Murmur Analysis Software is not intended as a sole means of diagnosis and is for use in environments where healthcare is provided by clinicians. The interpretations of heart sounds offered by the software are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. The interpretations are not diagnoses. The Eko Murmur Analysis Software is intended for use on pediatric and adult patients. | addition, it calculates certain cardiac time intervals such as heart rate, QRS duration and EMAT. The software does not distinguish between different kinds of murmurs and does not identify other arrhythmias.

It is not intended as a sole means of diagnosis. The interpretations of heart sounds and ECG offered by the software are only significant when used in conjunction with physician over-read and is for use on adults (> 18 years).  |
|  Prescription / OTC | Prescription Use Only | Prescription Use Only | Prescription Use Only  |
|  Patient Population | Murmur detection: Adult and pediatric patients

Atrial fibrillation detection: Adults (>18 years) patients | Adult and pediatric patients | Adults (>18 years) patients  |
|  Components | Software Only | Software Only | Software Only  |

{8}

|  Feature | Subject Device: Eko Foundation Analysis Software with Transformers (EFAST) | Predicate Device: Eko Murmur Analysis Software (EMAS) (K213794) | Reference Device: Eko Analysis Software (EAS) (K192004)  |
| --- | --- | --- | --- |
|  Interface | Callable application programming interface (API) | Callable application programming interface (API) | Callable application programming interface (API)  |
|  Display | No primary display | No primary display | No primary display  |
|  Murmur Detection | Yes | Yes | Yes  |
|  Structural Murmur Classification | Yes | Yes | No  |
|  Physiological Input Signals | Patients’ heart sounds and electrocardiograms (ECGs) | Patients’ heart sounds and electrocardiograms (ECGs, when available) | Patients’ heart sounds and electrocardiograms (ECGs)  |
|  Heart Sound Signal Quality Classification | Yes | Yes | Yes  |
|  Heart Sound Timing and Murmur Timing | S1 and S2 Systolic vs. Diastolic Murmur characterization | S1 and S2 Systolic vs. Diastolic Murmur characterization | N/A  |
|  ECG Rhythm Classification | Sinus Rhythm Atrial Fibrillation Unspecified Rhythm Poor ECG Signal | N/A | Normal Sinus Rhythm Atrial Fibrillation Unclassified ECG Poor ECG Signal  |
|  Heart Rate Calculation | Yes, ECG-based or PCG-based | N/A | Yes, ECG-based or PCG-based  |

K251494
Page 5 of 10

{9}

K251494
Page 6 of 10

# PERFORMANCE DATA - NONCLINICAL TESTING SUMMARY

The performance characteristics for the Eko Foundation Analysis Software with Transformers (EFAST) have been evaluated with the following non-clinical testing: software unit, integration, and system-level verification testing consistent with the IEC 62304 standard, and cybersecurity testing.

# PERFORMANCE DATA - ALGORITHM PERFORMANCE SUMMARY

Additionally, the EFAST algorithm was validated on proprietary datasets captured with the Eko CORE and CORE 500 Digital Stethoscope. The Eko CORE and CORE 500 Digital Stethoscope were used to capture 15-second duration heart sound and ECG recordings from volunteer patients.

- Structural Murmur Classification

A retrospective study of de-identified data was conducted with the EFAST algorithms to evaluate the structural murmur classification performance. Three sites within the US (n = 2) and Canada (n = 1) contributed data for 2,460 unique heart sound recordings from 615 unique subjects. All heart sound recordings were collected using the Eko CORE (52.7%) or Eko CORE 500 (47.3%) Digital Stethoscopes. 175 (28.5%) patients were pediatric (under 18 years old), and 440 (71.5%) were adults. 355 (57.7%) patients were male, and 260 (42.3%) were female. Race was documented for 458 (74.5%) patients and of that subset, the majority was White (412 [90.0%]), followed by Black or African American (22 [4.8%]), Other (15 [3.3%]), Asian (6 [1.3%]), Native Hawaiian or Other Pacific Islander (2 [&lt;1%]), and American Indian or Alaska Native (1 [&lt;1%]) and therefore representative of the intended use population.

All recordings were annotated by multiple cardiologists in respect to their quality and the presence of any murmur. Ground truth for structural murmur classification was obtained by pairing the cardiologist annotations with gold standard echocardiogram. For 259 of the 2,460 heart sound recordings, the cardiologists could not come to agreement, so the recordings were excluded from analysis.

Of the recordings that the expert cardiologists identified as good signal, 49.5% had a confirmed structural murmur and 50.5% had no structural murmur. Additionally, 96 heart sound recordings were annotated by expert cardiologists to obtain the timing of the S1 and S2 sounds.

Cardiologists were blinded to all subject demographic data and echocardiogram findings during annotation.

The following tables demonstrate the results of the primary performance analyses (Structural Murmur Classification and Heart Sound Timing).

{10}

|   | Sensitivity (%) | Specificity (%)  |
| --- | --- | --- |
|   |  EFAST | EFAST  |
|  Structural Murmur Classification | 83.4 (95% CI: 80.2 - 86.6) | 86.0 (95% CI: 82.2 - 89.8)  |

Table 2: EFAST Structural Murmur Classification Performance

|   | EFAST Structural Murmur Classification  |   |
| --- | --- | --- |
|  Age Group (Years) | Sensitivity (%) | Specificity (%)  |
|  Pediatric (<18) | 85.9 (95% CI: 81.3 - 90.5) | 76.0 (95% CI: 68.7 - 83.4)  |
|  All Adults (18+) | 81.7 (95% CI: 77.5 - 86.0) | 93.9 (95% CI: 90.6 - 97.1)  |
|  Adults 60+ | 82.2 (95% CI: 77.7 - 86.8) | 90.9 (95% CI: 86.1 - 95.7)  |

Table 3: EFAST Sensitivity and Specificity at detecting Structural Murmurs at different Age Groups

|  Heart Sound Timing | Sensitivity (%) | PPV (%)  |
| --- | --- | --- |
|   |  EFAST | EFAST  |
|  S1 Detection | 98.58 (95% CI:97.21-99.38) | 93.27 (95% CI: 90.94-95.15)  |
|  S2 Detection | 94.81 (95% CI: 92.59-96.53) | 94.29 (95% CI: 91.99-96.09)  |

Table 4: EFAST Heart Sound Timing Performance

The table below shows the observed and adjusted Positive Predictive Values (PPVs) for "High Likelihood" (HL) and "Moderate Likelihood" (ML) predictions for structural murmur. The table shows the probability for Structural Heart Disease given "High Likelihood" and "Moderate

K251494

{11}

Likelihood" predictions for "Structural Murmur" at different age brackets. For the adjusted PPV calculations at the bottom, the assumption of prevalence is 0.25 (p(SHD)=0.25).

|   | EFAST Structural Murmur Likelihood  |   |   |
| --- | --- | --- | --- |
|   |   |  PPVHL | PPVML  |
|  Observed | All Ages | 90.5 (95% CI: 86.7 - 94.3) | 64.7 (95% CI: 55.8 - 73.7)  |
|   |  Age <18 | 83.8% (95% CI: 76.7 - 91.0) | 45.3% (95% CI: 31.3 - 59.3)  |
|   |  Age 18+ | 96.3% (95% CI: 93.1 - 99.5) | 81.3% (95% CI: 71.9 - 90.7)  |
|   |  Age 60+ | 95.9% (95% CI: 92.3 - 99.4) | 80.9% (95% CI: 70.9 - 90.9)  |
|  Adjusted for P(SHD) = 0.25) | All Ages | 76.0 (95% CI: 71.8 - 80.2) | 37.9 (95% CI: 35.5 - 40.3)  |
|   |  Age <18 | 65.4 (95% CI: 58.4 - 72.4) | 23.2 (95% CI: 19.6 - 26.8)  |
|   |  Age 18+ | 88.9 (95% CI: 83.7 - 94.1) | 57.5 (95% CI: 54.3 - 60.7)  |
|   |  Age 60+ | 83.9 (95% CI: 78.4 - 89.5) | 48.8 (95% CI: 45.4 - 52.2)  |

# Algorithm Performance Comparison to Predicate:

A comparative analysis of the Structural Murmur classification algorithm relative to the predicate device (EMAS - K213794) is shown in the Table below.

Table 5: Positive Predictive Values (PPVs) for "High Likelihood" (HL) and "Moderate Likelihood" (ML) predictions for structural murmur

|   | Sensitivity (%) | Specificity (%)  |
| --- | --- | --- |
|  EMAS(K213794) | 85.8(95% CI: 82.9 - 88.7) | 84.5(95% CI: 81.0 - 88.0)  |
|  EFAST | 83.4(95% CI: 80.2 - 86.6) | 86.0(95% CI: 82.2 - 89.8)  |

Table 6: Comparison of sensitivity and specificity between EFAST and EMAS at detecting Structural Murmurs, as measured on the Eko Heart Sound Test Dataset

# ECG Rhythm Classification

A retrospective study of de-identified data was conducted with the EFAST algorithms to evaluate the ECG Rhythm Classification performance. Two sites within the US contributed data for 1,256 ECG recordings from 314 unique subjects. All ECG recordings were collected using the Eko CORE 500 Digital Stethoscope. The median age in the sample was 67.5 (IQR 58-75). 198

K251494

{12}

(63.1%) patients were male, and 116 (36.9%) were female. The majority of patients were White (272 [86.6%]), followed by Black or African American (19 [6.1%]), Other (15 [2.4%]), Asian (3 [1.0%]), Native Hawaiian or Other Pacific Islander (2 [&lt;1%]), and American Indian or Alaska Native (1 [&lt;1%]) and therefore representative of the intended use population. Race information was unavailable for 2 patients.

All ECG recordings were annotated by multiple experts (e.g., electrophysiologists, certified cardiographic technicians) in respect to their quality and the presence of atrial fibrillation. For 97 of the 1,256 ECG recordings, the experts did not come to an agreement, so the recordings were excluded from analysis.

Of the recordings that the experts identified as good signal,  $23.5\%$  had confirmed atrial fibrillation and  $76.5\%$  did not have atrial fibrillation (either sinus rhythm or an unspecified rhythm).

The following table demonstrates the results of the primary performance analysis for ECG Rhythm Classification.

|   | Overall Sensitivity (%) | Overall Specificity (%)  |
| --- | --- | --- |
|   |  EFAST | EFAST  |
|  Atrial Fibrillation Detection | 94.7 (95% CI: 91.5 - 97.8) | 94.1 (95% CI: 92.1 - 96.1)  |

# Algorithm Performance Comparison to Predicate:

A comparative analysis of the ECG Rhythm classification algorithm relative to the predicate device (EAS - K192004) is shown in the Table below.

Table 7: EFAST Atrial Fibrillation Detection Performance

|   | Overall Sensitivity (%) | Overall Specificity (%)  |
| --- | --- | --- |
|  EAS(K192004) | 73.1(95% CI: 66.2 - 80.0) | 97.7(95% CI: 96.5 - 98.9)  |
|  EFAST | 94.7(95% CI: 91.5 - 97.8) | 94.1(95% CI: 92.1 - 96.1)  |

Table 8: Comparison of Overall Sensitivity and Specificity between EFAST and EAS at detecting Atrial Fibrillation, as measured on the Eko ECG Test Dataset

K251494

{13}

K251494
Page 10 of 10

• Heart Rate Calculation
The mean absolute percentage error of the heart-rate algorithm was found to be less than 5%.

Associated Subgroup Analyses

Subgroup assessments of diagnostic performance were conducted to determine if there was heterogeneity in device performance across patient demographics. The results are summarized in Table 9 below.

|  Subgroup | Results of Test Heterogeneity  |
| --- | --- |
|  Age | For structural murmur classification, age was stratified into three groups: Pediatrics (<18), Adults (18+), Adults (60+). While no significant differences in sensitivity were observed among the age groups, the specificity of tests for pediatric patients (76.0% [95%CI: 68.7-83.4]) showed a lower value compared to those performed on adults (18+) (93.9% [95%CI: 90.6-97.2]). This is likely due to confusion of structural murmurs with some innocent murmurs on pediatric patients. Atrial Fibrillation detection is intended for adults only.  |
|  Biological sex | No significant differences were observed across biological sex for both structural murmur classification and Atrial Fibrillation detection.  |
|  Race/ethnicity | No significant differences were observed for both structural murmur classification and Atrial Fibrillation detection.  |

Table 9: Clinical Performance in Pre-Specified Subgroup Analyses

CONCLUSIONS

The indications for use, technological characteristics, and performance testing support that the proposed device, the Eko Foundation Analysis Software with Transformers (EFAST), is substantially equivalent, as safe and effective as the predicate device, and raises no new safety or effectiveness issues.

---

**Source:** [https://fda-staging.innolitics.com/device/K251494](https://fda-staging.innolitics.com/device/K251494)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
