8 mm Assist Cannula; 12 mm Assist Cannula

K251490 · Intuitive Surgical, Inc. · GCJ · Jul 2, 2025 · Gastroenterology, Urology

Device Facts

Record IDK251490
Device Name8 mm Assist Cannula; 12 mm Assist Cannula
ApplicantIntuitive Surgical, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 2, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The cannulas, obturators, seals, and related accessories are intended to establish a port of entry for instruments, endoscopes, or accessories. They are intended to be used only in a medical facility by trained medical professionals in accordance with its associated user manual.

Device Story

8 mm and 12 mm Assist Cannulas function as surgical access ports; establish entry for instruments, endoscopes, or accessories during laparoscopic or robotic-assisted procedures. Device consists of stainless steel hollow tube, seal, and obturator; creates air-tight system for insufflated body cavities. Used in hospital operating rooms by trained medical professionals. Output is physical access port for surgical tools; enables instrument manipulation within surgical site. Benefits include maintaining pneumoperitoneum and facilitating instrument exchange. No software or algorithmic components.

Clinical Evidence

No human clinical trials conducted. Evidence consists of bench testing (dimensional, mechanical, functional verification) and simulated use in animal and cadaver models. Human factors evaluation included comparative task analysis, use-related risk analysis, and formative usability testing to confirm design safety.

Technological Characteristics

Stainless steel (17-4 and 304 Full Hard) hollow tube; universal seal with luer fitting for insufflation; reusable; steam sterilized. Dimensions: 6.336" length. Compatible with 5-12 mm instruments. No software or electronic components.

Indications for Use

Indicated for establishing a port of entry for instruments, endoscopes, or accessories in medical facilities by trained professionals. Used in laparoscopic or robotic-assisted surgical procedures in the abdomen and thoracic regions.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 2, 2025 Intuitive Surgical, Inc. Melissa Gonzalez Technical Lead, Regulatory Affairs 1266 Kifer Road Sunnyvale, California 94086 Re: K251490 Trade/Device Name: 8 mm Assist Cannula; 12 mm Assist Cannula Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: May 9, 2025 Received: May 14, 2025 Dear Melissa Gonzalez: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251490 - Melissa Gonzalez Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251490 - Melissa Gonzalez Page 3 Sincerely, James H. Jang -S Digitally signed by James H. Jang -S Date: 2025.07.02 23:37:30 -04'00' James Jang, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} 8 mm Assist Cannula Page 9 of 34 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251490 | ? | | Please provide the device trade name(s). | | ? | | 8 mm Assist Cannula; 12 mm Assist Cannula | | | | Please provide your Indications for Use below. | | ? | | The cannulas, obturators, seals, and related accessories are intended to establish a port of entry for instruments, endoscopes, or accessories. They are intended to be used only in a medical facility by trained medical professionals in accordance with its associated user manual. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K251490 Traditional 510(k) Assist Cannula, 8 mm and 12 mm # 510(k) Summary (21 CFR § 807.92) ## I. Submitter Information 510(k) Owner: Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 Primary Contact: Melissa S. Gonzalez Technical Lead, Regulatory Affairs Phone: 408-523-8684 Email: melissa.gonzalez@intusurg.com Secondary Contact: Taian Chen Senior Regulatory Affairs Specialist Phone: 650-302-8499 Email: taian.chen@intusurg.com Date Summary Prepared: May 9, 2025 ## II. Subject Device Trade Name: 8 mm Assist Cannula and 12 mm Assist Cannula Common Name: Surgical Cannula Classification: Class II Regulation: 21 CFR § 876.1500, Endoscope and Accessories Product Code: GCJ ## III. Predicate Device Information Predicate Device: 12mm & Stapler Cannula (PN 470375, K142683) ## IV. Device Description The subject device 8 mm and 12 mm Assist Cannulas are used as part of a trocar system to establish a port of entry into the surgical site. The 8 mm and 12 mm Assist Cannulas are intended to be used laparoscopically (independently of the da Vinci Systems), but may be used in conjunction with da Vinci systems to assist robotic procedures. INTUITIVE {5} K251490 Traditional 510(k) Assist Cannula, 8 mm and 12 mm Page 2 of 8 ## V. Indications for Use The cannulas, obturators, seals, and related accessories are intended to establish a port of entry for instruments, endoscopes, or accessories. They are intended to be used only in a medical facility by trained medical professionals in accordance with its associated user manual. ## VI. Technological Characteristics The cannula is a stainless steel hollow tube that serves as a port of entry for endoscopic instruments. A seal is attached to a cannula and an obturator is placed through the seal. The seal attaches to the top of the cannula to create an air-tight system that allows an insufflated body cavity to be maintained. The seal has an insufflation port with a universal luer fitting and stopcock which connects to the da Vinci Insufflator Tube Set with Smoke Evacuation. The following table provides a summary comparison of the technological characteristics between the subject device to the predicate devices and reference device. The differences in technological characteristics do not raise new questions of safety and effectiveness. INTUITIVE {6} Traditional 510(k) Assist Cannula, 8 mm and 12 mm Page 3 of 8 K251490 | Attributes | Subject Devices Assist Cannulas (8 mm and 12 mm) | Predicate Device 12 mm & Stapler Cannula (K142683) | Reference Device 8mm Hex Cannula, Standard (K232610) | | --- | --- | --- | --- | | Manufacturer | Intuitive Surgical, Inc. | SAME as subject device Intuitive Surgical, Inc. | SAME as subject device Intuitive Surgical, Inc. | | Product Code | GCJ | SIMILAR to subject device NAY | SIMILAR to subject device NAY | | Regulation Number and Name | 21 CFR 876.1500, Endoscopes and Accessories | SAME as subject device 21 CFR 876.1500, Endoscopes and Accessories | SAME as subject device 21 CFR 876.1500, Endoscopes and Accessories | | Device Classification | Class II | SAME as subject device Class II | SAME as subject device Class II | | Classification Advisory Committee | General and Plastic Surgery | SAME as subject device General and Plastic Surgery | SAME as subject device General and Plastic Surgery | | Intended Use | To establish a port of entry for instruments, endoscopes, or accessories. | SAME as subject device To establish a port of entry for instruments, endoscopes, or accessories. | SAME as subject device To establish a port of entry for instruments, endoscopes, or accessories. | | Anatomical Site | | SAME as subject device | SAME as subject device | INTUITIVE {7} Traditional 510(k) Assist Cannula, 8 mm and 12 mm Page 4 of 8 K251490 | Attributes | Subject Devices Assist Cannulas (8 mm and 12 mm) | Predicate Device 12 mm & Stapler Cannula (K142683) | Reference Device 8mm Hex Cannula, Standard (K232610) | | --- | --- | --- | --- | | | Abdomen and Thoracic | Abdomen and Thoracic | Abdomen and Thoracic | | Where Used (hospital, home, ambulance, etc) | Hospital, Surgical Operating Rooms | SAME as subject device Hospital, Surgical Operating Rooms | SAME as subject device Hospital, Surgical Operating Rooms | | Compatible Trocar Accessories | 8 mm Assist Cannula - Universal Seal - da Vinci Insufflator Tube Set with Smoke Evacuation* - 8 mm Cannula Gage Pin - 8 mm Blunt Obturator (Reusable) - 8 mm Bladeless Optical Obturator (Disposable) * connected through the use of the Universal Seal | SIMILAR to subject device 12 mm & Stapler Cannula - Universal Seal - da Vinci Insufflator Tube Set with Smoke Evacuation* * connected through the use of the Universal Seal | SAME as subject device 8 mm Hex Cannula - Universal Seal - da Vinci Insufflator Tube Set with Smoke Evacuation* - 8 mm Cannula Gage Pin - 8 mm Blunt Obturator (Reusable) - 8 mm Bladeless Optical Obturator (Disposable) * connected through the use of the Universal Seal | INTUITIVE {8} Traditional 510(k) Assist Cannula, 8 mm and 12 mm Page 5 of 8 K251490 | Attributes | Subject Devices Assist Cannulas (8 mm and 12 mm) | Predicate Device 12 mm & Stapler Cannula (K142683) | Reference Device 8mm Hex Cannula, Standard (K232610) | | --- | --- | --- | --- | | | **12 mm Assist Cannula** - Universal Seal - da Vinci Insufflator Tube Set with Smoke Evacuation* - 12 mm & Stapler Blunt Obturator (Reusable) - 12 mm & Stapler Bladeless Obturator (Reusable) *connected through the use of the Universal Seal | **SAME** as subject device | NA | | | | **12 mm & Stapler Cannula** - Universal Seal - da Vinci Insufflator Tube Set with Smoke Evacuation* - 12 mm & Stapler Blunt Obturator (Reusable) - 12 mm & Stapler Bladeless Obturator (Reusable) -*connected through the use of the Universal Seal | | | Compatible Devices | **8 mm Assist Cannula:** Compatible with instruments, imaging devices, and surgical accessories ranging in size from 5 mm to 8 mm. | **SIMILAR** to subject device: Compatible with robotic instruments, imaging devices, and surgical accessories ranging in size from 5 mm to 8 mm. | NA | | | **12 mm Assist Cannula:** Compatible with instruments, imaging devices, and surgical accessories ranging in size from 5 mm to 12 mm. | **SIMILAR** to subject device: Compatible with robotic instruments, imaging devices, and surgical accessories ranging in size from 5 mm to 12 mm. | NA | INTUITIVE {9} Traditional 510(k) Assist Cannula, 8 mm and 12 mm K251490 | Attributes | Subject Devices Assist Cannulas (8 mm and 12 mm) | Predicate Device 12 mm & Stapler Cannula (K142683) | Reference Device 8mm Hex Cannula, Standard (K232610) | | --- | --- | --- | --- | | Inner Diameter | 8 mm Assist Cannula: Funnel Opening: 1.179" Tube: 0.380" Tip: 0.354" | SIMILAR to subject device Funnel Opening: 1.179" Tube: 0.564" Tip: 0.564" | SAME as subject device Funnel Opening: 1.179" Tube: 0.380" Tip: 0.354" | | | 12 mm Assist Cannula: Funnel Opening: 1.179" Tube: 0.564" Tip: 0.564" | SAME as subject device Funnel Opening: 1.179" Tube: 0.564" Tip: 0.564" | NA | | Overall Length | 6.336" | SAME as subject device 6.336" | NA | | Overall Design | 8 mm Assist Cannula: Funnel: Bowl shape Tube: Ribbed | SIMILAR to subject device Funnel: Bowl shape with magnetic fin Tube: Smooth | NA | | | 12 mm Assist Cannula: Funnel: Bowl shape Tube: Ribbed | SIMILAR to subject device Funnel: Bowl shape with magnetic fin Tube: Smooth | NA | | Tip Design | 8 mm Assist Cannula: Hexagonal with chisel tip | SIMILAR to subject device Round with flat tip | SAME as subject device Hexagonal with chisel tip | | | 12 mm Assist Cannula: Round with chisel tip | SIMILAR to subject device Round with flat tip | NA | | Patient Contact Materials | Funnel: 17-4 stainless steel | SAME as subject device Funnel: 17-4 stainless steel | NA | INTUITIVE {10} Traditional 510(k) Assist Cannula, 8 mm and 12 mm Page 7 of 8 K251490 | Attributes | Subject Devices Assist Cannulas (8 mm and 12 mm) | Predicate Device 12 mm & Stapler Cannula (K142683) | Reference Device 8mm Hex Cannula, Standard (K232610) | | --- | --- | --- | --- | | | **Tube:** 304 Full Hard stainless steel (patient contacting material) | **SIMILAR** to subject device **465 Carpenter Steel** (patient contacting material) | NA | | Biocompatibility | All materials have been evaluated for biocompatibility per ISO 10993-1 | **SAME** as subject device All materials have been evaluated for biocompatibility per ISO 10993-1 | NA | | Sterility | Packaged non-sterile; Steam sterilization | **SAME** as subject device Packaged non-sterile; Steam sterilization | NA | | Use/Disposition | Reusable | **SAME** as subject device Reusable | NA | | Packaging | Corrugate insert with retention features, inside a corrugate box | **SAME** as subject device Corrugate insert with retention features, inside a corrugate box | NA | INTUITIVE {11} Traditional 510(k) Assist Cannula, 8 mm and 12 mm Page 8 of 8 K251490 ## VII. Performance Data The following testing was conducted to demonstrate substantial equivalence to the predicate devices. **Performance testing** (bench and animal) demonstrates that the subject devices design output meets the design input requirements and that the devices perform as intended. The testing conducted consisted of dimensional measurements, mechanical and functional verification, and simulated use in animal and cadaver models. **Human Factors** evaluation for the subject devices included the following activities: - Known use-related issues for predicate devices and devices similar to the subject devices were analyzed using post-market data and the MAUDE database. All identified use related issues that are relevant to the use of the subject devices were documented in the risk analysis. - A Comparative Task Analysis (CTA) was conducted to describe all aspects of the user device interaction, through the breakdown of steps into user tasks, and to provide an analysis of comparison to the predicate for each task. - A Use-Related Risk Analysis (URRA) was conducted to identify all use-related risks for each user task identified as New or Modified from the predicate in the CTA. - Formative usability evaluations were conducted during the development process to inform the device user interface design and confirm assessment of use-related risks. ## VIII. Conclusion Based on the intended use, indications for use, technological characteristics, and performance data the subject devices, 8 mm Assist Cannula and 12 mm Assist Cannula are substantially equivalent to the predicate device 12 mm & Stapler Cannula. INTUITIVE
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...