Dice's coefficients > 80% for major subcortical structures and > 75% for major cortical structures; mean percentage absolute volume differences ≤ 5%.
—
—
Accuracy and reproducibility study: 72 3D T1 weighted scan cases for accuracy and 72 cases for reproducibility, including data from Philips, Siemens Healthineers, and GE scanners.
Accuracy and reproducibility study: 160 T2 weighted FLAIR scan cases for accuracy and 85 cases for reproducibility, including data from Philips, Siemens Healthineers, and GE scanners.
3 (neuroradiologists and neurologist)
Indications for Use
This device is intended for automatic labelling, visualization and volumetric quantification of segmentable brain structures and lesions from a set of MR images. Volumetric data may be compared to reference percentile data.
Device Story
SwiftSight-Brain is an automated MR image analysis software for clinical assessment of structural MRIs. It accepts 3D T1-weighted and T2-weighted FLAIR MRI scans in DICOM format. The device uses a deep learning-based internal pipeline to perform segmentation, volume calculation, and report generation. It produces segmented images with color overlays and morphometric reports comparing measured volumes to age/gender-matched reference data. Used in routine patient care, the software operates on off-the-shelf hardware (Windows) and integrates with PACS/DICOM workstations. Clinicians review the automated outputs to assist in evaluating brain structures and lesions. Automated quality control functions, including scan protocol verification and tissue contrast checks, ensure data integrity. The device benefits patients by providing standardized, reproducible volumetric measurements to support clinical decision-making regarding brain health and pathology.
Clinical Evidence
Bench testing only. Performance validated using multi-site datasets (US, UK, China, Germany, Brazil, Korea) across Philips, Siemens, and GE scanners (1.5T/3.0T). For 3D T1 images (n=72), Dice coefficients were >80% for subcortical and >75% for cortical structures; mean absolute volume differences were ≤5%. For T2 FLAIR lesion segmentation (n=160), Dice coefficient >0.80; mean absolute lesion volume difference <0.25cc. Ground truth established by two neuroradiologists and one neurologist. Results demonstrate high reproducibility and correlation with expert manual segmentation.
Technological Characteristics
Software-based medical device; operates on off-the-shelf hardware running Windows. Inputs: 3D T1 and T2 FLAIR MRI (DICOM). Processing: Deep learning-based automated segmentation pipeline. Outputs: Segmented images, volumetric measurements, and PDF morphometric reports. Features automated quality control (scan protocol verification, tissue contrast check). Connectivity: DICOM-compliant for PACS integration.
Indications for Use
Indicated for automatic labeling, visualization, and volumetric quantification of brain structures and lesions from MR images in patients ranging from young adults to the elderly (20s to 90s), including healthy subjects and those with mild cognitive impairment, Alzheimer's disease, or small vessel disease.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 23, 2025
AIRS Medical Inc.
HaeRi Lee
RA Manager
13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu
Seoul, 06142
Republic of Korea
Re: K251483
Trade/Device Name: SwiftSight-Brain
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: August 25, 2025
Received: August 25, 2025
Dear HaeRi Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251483 - Lee HaeRi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251483 - Lee HaeRi
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251483 | |
| Device Name SwiftSight-Brain | |
| Indications for Use (Describe) This device is intended for automatic labelling, visualization and volumetric quantification of segmentable brain structures and lesions from a set of MR images. Volumetric data may be compared to reference percentile data. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
AIRS MEDICAL
K251483
AIRS Medical Inc.
13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu, Seoul, 06142, Republic of Korea
www.airsmed.com
Date of the 510(k) summary preparation: September 18th, 2025
# 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
## 1. Submitter
- Name: AIRS Medical Inc.
- Address: 13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu, Seoul, 06142, Republic of Korea
- Phone: +82-10-4820-5200
- FAX: +82-2-6280-3185
## 2. Contact person
- Name: HaeRi Lee
- Position: RA Manager
- Email: lee.haeri@airsmed.com
## 3. Device information
- Device trade name: Swift Sight-Brain
- Model name: SS25-BR-CL
- Primary product specifications
- Regulation No. 21 CFR 892.2050
- Device class: Class II
- Product code: QIH
- Classification name: Automated radiological image processing software
- Associated product specifications
- Regulation No. 21 CFR 892.2050
- Device class: Class II
- Product code: LLZ
- Classification name: System, image processing, radiological
## 4. Predicate device
- Device trade name: Neurophet AQUA (V3.1)
- 510(k) number: K242215
- Product specification
- Regulation No. 21 CFR 892.2050
- Classification: Class II
- Product code: QIH, LLZ
## 5. Device description
SwiftSight-Brain is a fully automated MR image analysis software that provides automatic labeling, visualization, and volumetric quantification of brain structures from a set of MR images and
{5}
A
AIRS MEDICAL
AIRS Medical Inc.
13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu, Seoul, 06142, Republic of Korea
www.airsmed.com
returns segmented images and morphometric reports. The resulting output is provided in morphometric reports that can be displayed on Picture Archive and Communications Systems (PACS). The high throughput capability makes the software suitable for use in routine patient care as a support tool for clinicians in assessment of structural MRIs.
SwiftSight-Brain provides morphometric measurements based on 3D T1 weighted MRI series. The output of the software includes volumes that have been annotated with color overlays, with each color representing a particular segmented region, and morphometric reports that provide comparison of measured volumes to age and gender-matched reference percentile data. In addition, the adjunctive use of the T2 weighted FLAIR MR series allows for improved identification of some brain abnormalities such as lesions, which are often associated with T2 weighted FLAIR hyperintensities.
SwiftSight-Brain processing architecture includes a proprietary automated internal pipeline that performs segmentation, volume calculation and report generation.
The results are displayed in a dedicated graphical user interface, allowing the user to:
- Browse the segmentations and the measures
- Compare the results of segmented brain structures to a reference healthy population
- Read, download and print a PDF report
Additionally, automated safety measures include automated quality control functions, such as scan protocol verification, which validate that the imaging protocols adhere to system requirements.
# 6. Indications for Use
This device is intended for automatic labelling, visualization and volumetric quantification of segmentable brain structures and lesions from a set of MR images. Volumetric data may be compared to reference percentile data.
# 7. Comparison of technological characteristics with the predicate device
The table below compares SwiftSight-Brain (subject device) with Neurophet AQUA V3.1 (predicate device, K242215).
Comparison table
| | Subject device | Predicate device | Substantial Equivalence |
| --- | --- | --- | --- |
| Device name | SwiftSight-Brain | Neurophet AQUA V3.1 | - |
| 510(k) | - | K242215 | - |
| Manufacturer | AIRS Medical Inc. | NEUROPHET, Inc | - |
| Product code | QIH, LLZ | QIH, LLZ | Same |
| Indications for use | This device is intended for automatic labelling, visualization and volumetric quantification of segmentable brain structures and lesions from a set of MR images. Volumetric data may be | Neurophet AQUA is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures and lesions from a set of MR images. Volumetric data may be compared to reference | Same |
{6}
A
AIRS
MEDICAL
AIRS Medical Inc.
13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu, Seoul, 06142, Republic of Korea
www.airsmed.com
| | Subject device | Predicate device | Substantial Equivalence |
| --- | --- | --- | --- |
| | compared to reference percentile data. | percentile data. | |
| Target anatomical sites | Brain | Brain | Same |
| Design and incorporated technology | • Automated measurement of brain tissue volumes, structures, and lesions
• Automatic segmentation and quantification of brain structures using deep learning | • Automated measurement of brain tissue volumes, structures, and lesions
• Automatic segmentation and quantification of brain structures using deep learning | Same |
| Physical characteristics | • Software package
• Operates on off-the shelf hardware (multiple vendors) | • Software package
• Operates on off-the shelf hardware (multiple vendors) | Same |
| Operation system | Windows | Windows | Same |
| Processing architecture | Automated internal pipeline that performs:
• segmentation
• volume calculation
• report generation | Automated internal pipeline that performs:
• segmentation
• volume calculation
• report generation | Same |
| Data source | • MRI scanner: 3D T1 and FLAIR MRI scans acquired with specified protocols
• Supports DICOM format as input | • MRI scanner: 3D T1 and FLAIR MRI scans acquired with specified protocols
• Supports DICOM format as input | Same |
| Output | • Provides volumetric measurements of brain structures and lesions
• Includes segmented color overlays and morphometric reports
• Automatically compares results to reference percentile data and to prior scans when available
• Supports DICOM format as output of results that can be displayed on DICOM workstations and Picture Archive and Communications Systems | • Provides volumetric measurements of brain structures and lesions
• Includes segmented color overlays and morphometric reports
• Automatically compares results to reference percentile data and to prior scans when available
• Supports DICOM format as output of results that can be displayed on DICOM workstations and Picture Archive and Communications Systems | Same |
| Safety | • Automated quality control functions | • Automated quality control functions | Substantially Equivalent |
3
{7}
A
AIRS MEDICAL
AIRS Medical Inc.
13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu, Seoul, 06142, Republic of Korea
www.airsmed.com
| | Subject device | Predicate device | Substantial Equivalence |
| --- | --- | --- | --- |
| | - Tissue contrast check
- Scan protocol verification
- Results must be reviewed by a medical professional | - Image artifact check
- Scan protocol verification
- Results must be reviewed by a trained physician | |
SwiftSight-Brain and the predicate device are identical in most aspects, including intended use, target anatomical site, physical characteristics, processing architecture, data source, and output. While there are differences in terminology related to the automated quality control function and the user who reviews the results, these differences do not affect safety or effectiveness, do not introduce new or increased safety risks, and demonstrate equivalent performance to existing legally marketed devices
# 8. Summary of non-clinical test
The software verification and validation tests were conducted to establish the performance, functionality and reliability characteristics of the subject devices. The device passed all the tests based on pre-determined Pass/Fail criteria.
To demonstrate the 3D T1 weighted and T2 weighted FLAIR analysis performance of the SwiftSight-Brain, the data primarily sourced from U.S. hospitals were utilized. The data presents a diverse mix of clinical, demographic, and technical variables, providing a foundation for both reliable and reproducible testing, as well as comprehensive accuracy assessment. The ground truth was established by the U.S. based two neuroradiologists and one neurologist. The multi-site data collection enhances statistical independence, while the inclusion of varied demographic groups, clinical conditions, and MR field strength addresses potential confounders.
The subjects upon whom the device was trained and tested include healthy subjects, mild cognitive impairment patients, Alzheimer's disease patients, and small vessel disease patients from young adults to elderlies. The multicenter study was adapted to collect scans from various vendors including Philips, Siemens Healthineers, and GE Healthcare and MR scans using general clinical protocols were collected.
For the 3D T1 weighted images, a total of 72 3D T1 weighted scan cases were used for accuracy, and 72 cases were used for reproducibility. The test dataset included MR images from Philips (54 subjects), Siemens Healthineers (53 subjects), and GE (37 subjects) scanners, using both 1.5T and 3.0T field strengths. Subjects were collected from multiple countries, including the United States, the United Kingdom, China, and Germany, with ages ranging from young adults to the elderly (20s to 90s).
As a result, the segmentation accuracy compared to expert manual segmentations of 3D T1 MRI scans was evaluated using Dice's coefficient metric. For major subcortical brain structures Dice's coefficients were above 80% and for major cortical were above 75%.
Brain structural reproducibility of repeated 3D T1 MRI scans for same subjects was evaluated by using the percentage absolute volume differences. The mean percentage absolute volume differences for all major subcortical and cortical structures were 5% or less.
For the T2 weighted FLAIR images, the accuracy test dataset comprised 160 T2 weighted FLAIR scan
{8}
AIRS MEDICAL
AIRS Medical Inc.
13-14F, Keungil Tower, 223, Teheran-ro, Gangnam-gu, Seoul, 06142, Republic of Korea
www.airsmed.com
cases, and the reproducibility test dataset comprised 85 cases. The sample size supports robust performance validation. The test dataset included MR images from Philips (92 subjects), Siemens Healthineers (65 subjects), and GE (88 subjects) scanners, using both 1.5T and 3.0T field strengths. Subjects were collected from multiple countries, including the United States, Brazil and Republic of Korea with ages ranging from young adults to the elderly (20s to 90s).
SwiftSight-Brain performance was then evaluated by comparing segmentation accuracy with expert manual segmentations and by measuring segmentation reproducibility between same subject scans. The system yields reproducible results that are well correlated with expert manual segmentation.
SwiftSight-Brain's lesion segmentation accuracy compared to expert manual segmentations of T2 weighted FLAIR scan was evaluated using Dice's coefficient metric, which exceeds 0.80. The brain lesion segmentation reproducibility was evaluated using repeated T2 weighted FLAIR scan pairs of subjects with brain lesions. The mean absolute lesion volume difference was less than 0.25cc.
# 9. Conclusion
The subject device is substantially equivalent in the areas of technical characteristics, general function, application, and indications for use. The new device does not introduce fundamentally new scientific technology, and the device has been validated through the system level test. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device (K242215, Neurophet AQUA V3.1).
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.