Mobarn 80:20 Lumbar Interbody Fusion System is intended for spinal fusion procedures to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft in skeletally mature patients with degenerative disc disease (“DDD”) at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had previous non-fusion spinal surgery at the involved spinal level(s) and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Mobarn 80:20 Lumbar Interbody Fusion System is to be combined with supplemental fixation.
Device Story
Family of additively manufactured titanium (Ti6Al4V-ELI) intervertebral body fusion cages; designed for PLIF, TLIF, DLIF, and ALIF surgical approaches. Implants feature open architecture (80% open area) to accommodate bone graft; 20% titanium frame provides structural support. Used by surgeons in clinical settings to stabilize spinal segments and facilitate bony fusion. Implants available in various footprints, heights, and lordotic angles to match patient anatomy. Requires supplemental fixation. Benefits include maximized bone graft contact and structural stabilization for DDD patients.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing performed per FDA guidance for spinal systems and intervertebral body fusion devices, including ASTM F2077 (static/dynamic axial compression, static/dynamic compressive shear) and ASTM F2267 (subsidence), plus expulsion testing.
Technological Characteristics
Material: Ti6Al4V-ELI titanium alloy. Manufacturing: Additive manufacturing. Design: Open architecture (80% open area, 20% frame). Standards: ASTM F2077, ASTM F2267. Sterilization: Not specified. Connectivity: None. Form factor: Interbody cages for PLIF, TLIF, DLIF, ALIF.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with Grade 1 spondylolisthesis or retrolisthesis or previous non-fusion surgery, requiring spinal fusion with autograft/allogenic bone graft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Zyston Strut Open Titanium Spacer System (K171495)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 18, 2025
Mobarn Medical Devices, LLC
% Lisa Ferrara, Ph.D.
Consultant - Regulatory and Technical
Hypoid Ventures, LLC
640 Grand Oaks Drive
Hendersonville, North Carolina 28792
Re: K251479
Trade/Device Name: Mobarn 80:20 Lumbar Interbody Fusion System (Posterior Lumbar Interbody Fusion Cage (PLIF)); Mobarn 80:20 Lumbar Interbody Fusion System (Transforaminal Lumbar Interbody Fusion Cage (TLIF)); Mobarn 80:20 Lumbar Interbody Fusion System (Direct Lateral Lumbar Interbody Fusion Cage (DLIF)); Mobarn 80:20 Lumbar Interbody Fusion System (Anterior Lumbar Interbody Fusion Cage (ALIF))
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: MAX
Dated: August 18, 2025
Received: August 20, 2025
Dear Dr. Ferrara:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251479 - Lisa Ferrara, Ph.D.
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251479 - Lisa Ferrara, Ph.D.
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Mobarn 80:20 Lumbar Interbody Fusion System
Page 9 of 47
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251479 | ? |
| Please provide the device trade name(s). | | ? |
| Mobarn 80:20 Lumbar Interbody Fusion System (Posterior Lumbar Interbody Fusion Cage (PLIF));
Mobarn 80:20 Lumbar Interbody Fusion System (Transforaminal Lumbar Interbody Fusion Cage (TLIF));
Mobarn 80:20 Lumbar Interbody Fusion System (Direct Lateral Lumbar Interbody Fusion Cage (DLIF));
Mobarn 80:20 Lumbar Interbody Fusion System (Anterior Lumbar Interbody Fusion Cage (ALIF)) | | |
| Please provide your Indications for Use below. | | ? |
| Mobarn 80:20 Lumbar Interbody Fusion System is intended for spinal fusion procedures to be used with
autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft in
skeletally mature patients with degenerative disc disease (“DDD”) at one or two contiguous spinal levels
from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and
radiographic studies. These patients should have had six months of non-operative treatment. These DDD
patients may have had previous non-fusion spinal surgery at the involved spinal level(s) and may have up
to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Mobarn 80:20 Lumbar Interbody
Fusion System is to be combined with supplemental fixation. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Traditional 510(k) Premarket Notification
Mobarn Medical Devices, LLC - Lumbar Interbody Fusion System
K251479
Page 1 of 3
# 510(k) Statement – Summary of Safety and Effectiveness
A 510(k) Summary of the safety and effectiveness information upon which this substantial equivalence determination is based, is provided in this section.
| Submitter | Mobarn Medical Devices, LLC |
| --- | --- |
| Address | 900 Camp St Ste 449
New Orleans, LA 70130 |
| Telephone | 504-460-7305 |
| Contact Person | Richard Ehret, Managing Member |
| --- | --- |
| Address | 900 Camp St Ste 449, New Orleans, LA 70130 |
| Telephone | 504-460-7305 |
| Email | richard.b.ehret@gmail.com |
| Date Prepared | April 25, 2025 |
| --- | --- |
| Trade Name | Mobarn 80:20 Lumbar Interbody Fusion System |
| 510k Review Panel | Orthopedic |
| Classification Name | Intervertebral Fusion Device with Bone Graft, Lumbar |
| Regulation Number | 21 CFR 888.3080 |
| Regulatory Class | II |
| Product Code | MAX |
| Predicate Device (Primary) | 510 (k) Number | Manufacturer |
| --- | --- | --- |
| Zyston Strut Open Titanium Spacer System | K171495 | Zimmer Biomet Spine Inc. |
| Reference Device (Additional) | 510 (k) Number | Manufacturer |
| 4Web Medical PLIF STS, TLIF STS, OLIF STS (Spinal Truss System) | K143258 | 4Web Medical Inc. |
{5}
Traditional 510(k) Premarket Notification
Mobarn Medical Devices, LLC - Lumbar Interbody Fusion System
K251479
Page 2 of 3
| Description of Device | The Mobarn 80:20 Lumbar Interbody Fusion System consists of a family of additively manufactured titanium (Ti6Al4V-ELI) lumbar intervertebral body fusion devices (LIBF) designed to facilitate lumbar spinal fusion through various surgical approaches, as listed below.
• Posterior Lumbar Interbody Fusion (PLIF)
• Transforaminal Lumbar Interbody Fusion (TLIF)
• Direct Lateral Lumbar Interbody Fusion (DLIF)
• Anterior Lumbar Interbody Fusion (ALIF)
The Mobarn 80:20 Lumbar Interbody Fusion System is intended for use in lumbar intervertebral fusion procedures. These devices are designed to provide structural support and stabilization to facilitate bony fusion in patients with degenerative disc disease (DDD) and other conditions requiring interbody fusion.
The design of the implants consists of an open architecture surrounded by a titanium frame to accommodate autograft and/or allogenic bone graft, comprised of cancellous and/or cortico-cancellous bone graft. Each implant is engineered with an 80% open area to accommodate autograft and/or allogenic bone graft comprised of cancellous and/or cortico-cancellous bone graft, maximizing bone graft contact and fusion potential, while maintaining mechanical integrity through a titanium frame that makes up the remaining 20% and offers structural support. The implants are available in a variety of footprints, heights, and lordotic angles to approximate anatomical variation in different vertebral levels and/or patient anatomy. |
| --- | --- |
| Indications and Intended Use | Mobarn 80:20 Lumbar Interbody Fusion System is intended for spinal fusion procedures to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft in skeletally mature patients with degenerative disc disease (“DDD”) at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had previous non-fusion spinal surgery at the involved spinal level(s) and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Mobarn 80:20 Lumbar Interbody Fusion System is to be combined with supplemental fixation. |
{6}
Traditional 510(k) Premarket Notification
Mobarn Medical Devices, LLC - Lumbar Interbody Fusion System
K251479
Page 3 of 3
| **Technological Characteristics**
**Substantial Equivalence** | **Summary of Technological Characteristics**
The Mobarn 80:20 Lumbar Interbody Fusion System (Subject device) has the following similarities listed below to the primary and reference predicates which have received 510(k) clearance;
• same indicated use and indications for use,
• same operating principle,
• incorporates similar open architecture geometrical designs, and sizes,
• incorporates similar materials – Ti6Al4V
• additively manufactured devices which have a similar manufacturing process as the predicate
• comparable instruments of similar material to the predicate
• similar mechanical performance (per mechanical testing via ASTM F2077 & ASTM F2267)
• similar Biocompatibility Properties
Based on the design, intended use, and performance characteristics, the Mobarn 80:20 Lumbar Interbody Fusion System is expected to perform as safely and effectively as its predicate devices. |
| --- | --- |
| **Performance** | Mechanical testing was conducted per FDA’s Guidance for Industry and FDA Staff – Spinal System 510(k)s dated May 3, 2004, and Class II Special Controls Guidance Document: Intervertebral Body Fusion Device dated June 12, 2007.
Performance testing was performed to demonstrate substantial equivalence to the predicate device(s).
Performance testing included;
• ASTM F2077 (static and dynamic axial compression, static and dynamic compressive shear)
• ASTM F2267 (subsidence).
• Expulsion |
| **Magnetic Resonance environment** | The Mobarn Medical Devices Mobarn 80:20 Lumbar Interbody Fusion System has not been evaluated for safety and compatibility in the Magnetic Resonance environment. They have not been tested for heating or migration in the Magnetic Resonance environment. |
| **Conclusion** | Based on the indications for use, technological characteristics, design, materials, and comparison to predicate devices, the Mobarn 80:20 Lumbar Interbody Fusion System is substantially equivalent to legally marketed predicate devices and is safe and effective for its intended use. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.