Root canal repair materials (nRoot BP) is developed for permanent root canal sealer and repair and surgical applications. Repair of Root Perforation Repair of Root Resorption Root End Filling Apexification Pulp Capping
Device Story
nRoot BP is a pre-mixed paste root canal repair material; composed of inorganic powder and organic liquid; packaged in preloaded syringe. Used by dentists/endodontists in clinical settings for permanent root canal sealing and surgical endodontic applications. Material solidifies in bodily fluid environments; provides X-ray opacity; does not stain teeth. Device functions as a restorative/sealing agent for root perforations, resorptions, root end filling, apexification, and pulp capping. Clinical benefit derived from effective sealing of root canal system and promotion of healing in surgical endodontic procedures.
Clinical Evidence
Bench testing only. Performance verified against ISO 6876:2012 standards. Tests included appearance, deliverable volume, setting time, dimensional changes, solubility/disintegration, compressive strength, pH, arsenic/lead content, radio-opacity, and microbiological indicators. Biocompatibility evaluated per ISO 10993-1 and ISO 7405. All results met acceptance criteria and demonstrated similarity to the predicate device.
Technological Characteristics
Pre-mixed paste containing tricalcium silicate, dicalcium silicate, zirconium oxide, calcium hydroxide, calcium phosphate, and thickening agent. Form factor: preloaded syringe. Conforms to ISO 6876:2012 for root canal sealing materials. Biocompatibility evaluated per ISO 10993-1 and ISO 7405.
Indications for Use
Indicated for permanent root canal sealing, repair of root perforation, repair of root resorption, root end filling, apexification, and pulp capping in patients requiring endodontic treatment.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 25, 2025
Enpuno Biotechnology Co., Ltd
Chengfa Huang
RA manager
No. 201, Haiping Park Production Plant3, No.229, Guyuan Road
Changsha High-tech Development Zone Changsha City, Hunan 410205
CHINA
Re: K251465
Trade/Device Name: Root canal repair materials (nRoot BP)
Regulation Number: 21 CFR 872.3820
Regulation Name: Root Canal Filling Resin
Regulatory Class: Class II
Product Code: KIF
Dated: August 25, 2025
Received: August 25, 2025
Dear Chengfa Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251465 - Chengfa Huang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251465 - Chengfa Huang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251465
Device Name
Root canal repair materials (nRoot BP)
Indications for Use (Describe)
Root canal repair materials (nRoot BP) is developed for permanent root canal sealer and repair and surgical applications.
- Repair of Root Perforation
- Repair of Root Resorption
- Root End Filling
- Apexification
- Pulp Capping
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
| 510(k) #: K251465 | 510(k) Summary | Prepared on: 2025-08-25 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Enpuno Biotechnology Co., Ltd | |
| Applicant Address | No. 201, Haiping Park Production Plant3, No.229, Guyuan Road Changsha High-tech Development Zone Changsha City Hunan 410205 China | |
| Applicant Contact Telephone | +86-73184257959 | |
| Applicant Contact | Mr. Chengfa Huang | |
| Applicant Contact Email | jiang13620586569@126.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Root canal repair materials (nRoot BP) | |
| Common Name | Root canal filling resin | |
| Classification Name | Resin, Root Canal Filling | |
| Regulation Number | 872.3820 | |
| Product Code(s) | KIF, 21 CFR 872.3820 | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K092715 | IROOT BP PLUS | KIF |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The root canal repair material is a pre mixed paste material containing an inorganic powder system and an organic mixed liquid. Can solidify in a bodily fluid environment and effectively repair root canals; Has excellent X-ray blocking properties, stable curing time, and does not stain teeth. nRoot BP is packaged in a preloaded syringe. The performance of the Root canal repair materials conforms to ISO 6876:2012 Dentistry - Root canal sealing materials. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| Root canal repair materials (nRoot BP) is developed for permanent root canal sealer and repair and surgical applications.
●Repair of Root Perforation
●Repair of Root Resorption
●Root End Filling
●Apexification
●Pulp Capping | | |
{5}
| Indications for Use Comparison 21 CFR 807.92(a)(5) |
| --- |
| The indications for use are the same between proposed device and predicate device. |
| Technological Comparison 21 CFR 807.92(a)(6) |
| The proposed device is essentially identical to the predicate devices in terms of indication for use, design between our device and the predicate devices. |
| Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) |
| The bench test include tests for
1. Type Test by the third test party:
There are one type test reports, respectively:
Item Test results Report No.
Physical and chemical performance testing ISO 6876 The physical and chemical properties of the samples are complied with ISO 6876:2012, CHTT24110009
CHTT25080250
The test results of the components, Appearance, Deliverable volume, Setting time, Dimensional changes after setting, Solubility and disintegration, Compressive strength, pH, Acid-soluble arsenic and lead content, Radio-opacity and Microbiological indicator etc. of the Device under test are very similar to those of the predicate device.
The Root canal repair materials (nRoot BP) (Tricalcium silicate, dicalcium silicate, zirconium oxide, Calcium Hydroxide, Calcium Phosphate and Thickening Agent.) of the proposed device is basically the same as the predicate device.
2. Performance Test
Factory Performance Test, including: Appearance, Deliverable volume, Setting time, Dimensional changes after setting, Solubility and disintegration, Compressive strength, pH, Acid-soluble arsenic and lead content, Radio-opacity and Microbiological indicator, Packaging Marks, labels and instructions.
Bench testing was performed per ISO 6876:2012 and internal procedures to ensure that the Root canal repair materials met its specifications. All tests were verified to meet acceptance criteria. Test results on Appearance, Deliverable volume, Setting time, Dimensional changes after setting, Solubility and disintegration, Compressive strength, pH, Acid-soluble arsenic and lead content, Radio-opacity and Microbiological indicator etc. of the Device under test are very similar to those of the predicate device. The biocompatibility of the subject device was addressed using FDA’s Biocompatibility guidance document, “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”” and ISO 7405 Dentistry — Evaluation of biocompatibility of medical devices used in dentistry.
The conclusions drawn from the nonclinical tests demonstrate that subject device Root canal repair materials (nRoot BP) is as safe and effective as the predicate device. Root canal repair materials (nRoot BP) is substantial equivalent to the legally marketed predicate device K092715 IROOT BP PLUS. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.