K251464 · Jeisys Medical Incorporated · GEX · Oct 3, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251464
Device Name
TRI-BEAM PRO, TRI-BEAM (Nd:YAG Laser unit)
Applicant
Jeisys Medical Incorporated
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 3, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRI-BEAM PRO, TRI-BEAM, Nd: YAG Laser Unit is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 1064 Wavelength in G, TH mode: - Tattoo removal: dark ink (black, blue and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation. - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkle - Treatment of melasma 1064nm in GN (PTP-Off) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris 532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces): -Tattoo removal: light ink (red,tan, purple, orange, sky blue,green) -Removal of Epidermal benign Pigmented Lesions -Removal of Minor Vascular Lesions including but not limited to telangiectasias -Treatment of Lentigines; -Treatment of Cafe-Au-Lait -Treatment of Seborrheic Keratoses -Treatment of Post Inflammatory Hyper-Pigmentation -Treatment of Becker's Nevi, Freckles and Nevi Spilus
Device Story
Nd:YAG laser system for dermatologic applications; utilizes selective photothermolysis to target pigmented/vascular lesions, tattoos, wrinkles, and acne. System includes main unit, color touch screen, handpiece, and foot switch. Operates in multiple modes: PTP_ON (single pulse), PTP_OFF (dual pulse), G (focal energy), TH (flat-top), and GN (flat-top, low irritation). Used in clinical settings by trained professionals. Provider selects parameters via touch screen; laser energy delivered through handpiece to target tissue. Minimizes surrounding tissue damage. Benefits include precise tissue treatment and versatility across various dermatologic conditions.
Clinical Evidence
No clinical performance testing was performed. Evidence is based on bench testing, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993).
Technological Characteristics
Nd:YAG laser; wavelengths 1064nm, 532nm, 585nm, 650nm. Pulse widths 5-10ns (G, TH) and 300μs (GN). Energy delivery modes: PTP (single/dual), G, TH, GN. Connectivity: color touch screen interface. Biocompatibility per ISO 10993. Software validated per IEC 62304 (moderate level of concern).
Indications for Use
Indicated for aesthetic, cosmetic, and surgical soft tissue procedures (incision, excision, ablation, vaporization, coagulation, hemostasis) in dermatology. Specific applications include tattoo removal (dark/light inks), pigmented/vascular lesion removal, hair removal, acne scar/wrinkle treatment, melasma, and inflammatory acne vulgaris.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 3, 2025
Jeisys Medical Incorporated
% Sanghwa Myung
Regulatory Affair Specialist
E&m
D1474, PyeongCheon Arco Tower, 230, Simin-Daro, Dongan-gu
Anyangsi, Gyeonggido 14067
Korea, South
Re: K251464
Trade/Device Name: TRI-BEAM PRO, TRI-BEAM (Nd:YAG Laser unit)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: September 2, 2025
Received: September 2, 2025
Dear Sanghwa Myung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}
K251464 - Sanghwa Myung
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K251464 - Sanghwa Myung
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE
Digitally signed by
TANISHA L. HITHE
Date: 2025.10.03
10:17:39 -04'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
TRI-BEAM PRO, TRI-BEAM
Page 9 of 42
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251464 | ? |
| Please provide the device trade name(s). | | ? |
| TRI-BEAM PRO, TRI-BEAM (Nd:YAG Laser unit) | | |
| Please provide your Indications for Use below. | | ? |
| The TRI-BEAM PRO, TRI-BEAM, Nd: YAG Laser Unit is intended for use in aesthetic, cosmetic and
surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general
dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
1064 Wavelength in G, TH mode:
- Tattoo removal: dark ink (black, blue and brown)
- Removal of Nevus of Ota
- Removal or lightening of unwanted hair with or without adjuvant preparation.
- Treatment of Common Nevi
- Skin resurfacing procedures for the treatment of acne scars and wrinkle
- Treatment of melasma
1064nm in GN (PTP-Off) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne
vulgaris
532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye
handpieces):
-Tattoo removal: light ink (red,tan, purple, orange, sky blue,green)
-Removal of Epidermal benign Pigmented Lesions
-Removal of Minor Vascular Lesions including but not limited to telangiectasias
-Treatment of Lentigines;
-Treatment of Cafe-Au-Lait
-Treatment of Seborrheic Keratoses
-Treatment of Post Inflammatory Hyper-Pigmentation
-Treatment of Becker's Nevi, Freckles and Nevi Spilus | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K251464
# 510(k) Summary
K251464
This summary of 510(k) Safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
Submitter: Jeisys Medical Incorporated
307 Daeryung Techno Town 8th
Gamasan-ro 96, Geumcheon-Gu, Seoul, 08501, South Korea
Telephone: +82-70-7435-4930
Fax: 82-10-2843-3547
E-mail:
Contact Person: E&M
Regulatory Affair/Sang Hwa Myung
Telephone: +80-70-7807-0550
FAX: +82-31-388-9263
Cellphone: +82-10-4952-6638
E-mail: mshenmc@gmail.com
Primary Contact: Sanghwa Myung
Date 510(k) summary Prepared: October 3, 2025
Trade Name: TRI-BEAM PRO, TRI-BEAM
Common Name: Nd:YAG Laser Surgical Unit
Classification: II
Product Code: GEX
Regulation Numbers: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument for Use In General and Plastic Surgery and In Dermatology
Description of Device:
TRI-BEAM PRO, Nd:YAG Laser unit in the wavelength 1064 nm, 532 nm, 585 nm, and 650 nm is intended for treatment for general dermatology. The TRI-BEAM Series, utilizing Nd:YAG laser optical energy, is based on the principle of selective photothermolysis, enabling targeted treatment of pigmented and vascular lesions, tattoos, wrinkles, acne, and other dermatologic conditions while minimizing damage to surrounding tissue. The system comprises the main unit, a color touch screen, a handpiece, and a foot switch.
Handpiece Mode
PTP_ON: Single Pulse
Laser energy is delivered to the target tissue as a single, continuous pulse.
PTP_OFF: Dual Pulse
One pulse is divided into two sub-pulses, delivered sequentially with a short time interval to reduce peak thermal load.
G Mode
Beam profile with energy concentrated at the center and gradually tapering toward the periphery, allowing focal energy delivery.
{5}
K251464
TH Mode
Flat-top energy distribution in which the output intensity at the center and periphery is nearly uniform, ensuring consistent tissue exposure.
GN Mode
Flat-top energy distribution designed to minimize strong irritation to the skin, even when applied at higher energy levels.
Indication for use:
The TRI-BEAM PRO, TRI-BEAM, Nd: YAG Laser Unit is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis
1064 Wavelength in G, TH mode:
- Tattoo removal: dark ink (black, blue and brown)
- Removal of Nevus of Ota
- Removal or lightening of unwanted hair with or without adjuvant preparation.
- Treatment of Common Nevi
- Skin resurfacing procedures for the treatment of acne scars and wrinkle
- Treatment of melasma
1064nm in GN (PTP-Off) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris
532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces):
- Tattoo removal: light ink (red,tan, purple, orange, sky blue,green)
- Removal of Epidermal benign Pigmented Lesions
- Removal of Minor Vascular Lesions including but not limited to telangiectasias
- Treatment of Lentigines;
- Treatment of Cafe-Au-Lait
- Treatment of Seborrheic Keratoses
- Treatment of Post Inflammatory Hyper-Pigmentation
- Treatment of Becker's Nevi, Freckles and Nevi Spilus
Predicate Device:
| Predicate Devices 1 | K241527
Applicant: Wontech Co., Ltd
Trade/Device Name: Pastelle PRO
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in Dermatology
Regulatory Class: Class II
Product Code: GEX |
| --- | --- |
Substantial Equivalence:
Comparison table is as follows.
Table 1: Substantial equivalence comparison
| | Proposed Device | Predicate Device | Comparison Comment |
| --- | --- | --- | --- |
| 510(k) Number | Pending | K241527 | - |
| Manufacturer | Jeisys Medical Inc | Wontech Co., Ltd | - |
{6}
K251464
| Trade/Device Name | TRI-BEAM PRO, TRI-BEAM | Pastelle PRO | - |
| --- | --- | --- | --- |
| Classification Name | Laser surgical instrument for use in general and plastic surgery and in Dermatology | Laser surgical instrument for use in general and plastic surgery and in Dermatology | - |
| Indication for use | The TRI-BEAM PRO, TRI-BEAM, Nd: YAG Laser Unit is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. | The Pastelle Pro system is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. | Same |
| | 1064 nm in G, TH mode - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma | 1064nm in nanosecond mode, including microbeam handpieces: - Tattoo removal: dark ink (black, blue, and brown) - Removal of Nevus of Ota - Removal or lightening of unwanted hair with or without adjuvant preparation - Treatment of Common Nevi - Skin resurfacing procedures for the treatment of acne scars and wrinkles - Treatment of melasma | Same |
| | 1064nm in GN (PTP-Off) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris | 1064nm in Genesis (long-pulse) mode: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris | |
| | 532nm Wavelength (nominal delivered energy of 585 nm and 650 nm with optional dye handpieces): -Tattoo removal: light ink (red,tan, purple, orange, sky blue,green) -Removal of Epidermal benign Pigmented Lesions -Removal of Minor Vascular Lesions | 532nm in nanosecond mode, including microbeam handpieces (nominal delivered energy of 595 nm and 660 nm with optional dye handpieces): - Tattoo removal: light ink (red, tan, purple, orange, sky blue, green) - Removal of Epidermal Pigmented Lesions - Removal of Minor Vascular Lesions including but not limited to telangiectasias - Skin resurfacing procedures for the treatment of acne scars | Same |
{7}
K251464
| | including but not limited to telangiectasias
-Treatment of Lentigines;
-Treatment of Cafe-Au-Lait
-Treatment of Seborrheic Keratoses
-Treatment of Post Inflammatory Hyper-Pigmentation
-Treatment of Becker's Nevi, Freckles and Nevi Spilus | and wrinkles - Treatment of Lentigines - Treatment of Café-au-Lait - Treatment of Seborrheic Keratoses - Treatment of Post Inflammatory Hyperpigmentation - Treatment of Becker's Nevi, Freckles, and Nevi spilus | |
| --- | --- | --- | --- |
| Laser Medium | Nd:YAG laser | Nd:YAG laser | Same |
| Wavelength | 1064nm, 532nm, 585nm, 650nm | 1064nm, 532nm, 585nm, 650nm | Same |
| Duration (Pulse Width) | 1064nm
5-10ns (G, TH mode)
300μs (GN mode)
532nm : 5-10ns (TH mode)
585nm : 5-10ns (TH mode)
650nm: 5-10ns (TH mode) | 532nm/1064nm mode: 5ns~12ns
1064 PTP/Triple/2X PTP: Up to 20ns
Genesis mode: 80~300μs | Substantial Equivalence |
| Pulse Energy (max) | 1064 GN/PTP OFF mode: 3500 mJ
1064 TH/PTP ON mode: 1600 mJ
1064 TH/PTP OFF mode: 1200 mJ
1064 G/PTP OFF mode: 1500 mJ
532 / PTP OFF mode: 500 mJ
585 / PTP OFF mode: 250 mJ
650 / PTP OFF mode: 150 mJ | Genesis mode: 100 ~ 5,000 mJ
1064 nm mode: 50 ~ 1,200 mJ
1064 nm PTP mode: 600 ~ 2,000 mJ
1064 nm Triple mode: 1,000 ~ 1400 mJ
1064 nm 2x PTP mode: 1,000 ~ 1400 mJ
532 nm mode: 30 ~ 500 mJ
1064 nm DOE mode: 60 ~ 900 mJ
595 nm Dye mode: 30 ~ 250 mJ
660 nm Dye mode: 30 ~ 150 mJ | Substantial Equivalence |
| Spot Size(mm) | Zoom: 2~10mm
Collimation: 7mm
1064 nm: 4.5 x 4.5mm²
532 nm: 5 x5 mm²
Dye: 2mm | Zoom: 2mm ~ 10mm
Collimation: 7mm
MLA: 3mm ~ 8mm
DOE: 5mm x 5mm ~ 7mm x 7mm
Dye: 3mm | Different #1 |
| Repetition rate | 1064 nm GN Mode: 1 ~ 10 Hz
1064 nm G mode: 1 ~ 10Hz
1064 nm TH PTP on mode, PTP off mode: 1 ~ | Genesis mode: 1 ~ 10 Hz
1064 nm mode: 1 ~ 10 Hz
1064 nm PTP mode: 1 ~ 10 Hz
1064 nm Tripple mode: 1 ~ 10 Hz | Different #2 |
{8}
K251464
| | 10Hz
532nm: 1 ~ 10Hz
585nm: 1 ~ 2 Hz
650nm: 1 ~ 2 Hz | 1064 nm 2x PTP mode: 1 ~ 10 Hz
532 nm mode: 1 ~ 10 Hz
1064 nm DOE mode: 1 ~ 10 Hz
595 nm Dye mode: 1 ~ 5 Hz
660 nm Dye mode: 1 ~ 2 Hz | |
| --- | --- | --- | --- |
## Discussion
The subject device and predicate device have some difference characteristics such as pulse energy, spot size, repetition rate that do not raise different questions of safety and effectiveness. The Subject device technical characteristics are within the range of the predicate device. The subject device and predicate device have some difference characteristics that do not raise different questions of safety and effectiveness. To support this, we conducted related pre-clinical test.
## Non-clinical Performance Data:
- Basic safety and essential performance of device was evaluated in accordance with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020
- Effect to the device by electromagnetic disturbances is evaluated in accordance with the FDA-recognized consensus standard, IEC 60601-1-2:2014.
- IEC 60601-2-22 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- IEC 60825-1 Ed. 3.0 (2014) Safety of laser products – Part 1: Equipment classification and requirements
- General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability is evaluated in accordance with the FDA-recognized consensus standard, IEC 60601-1-6:2013.
- The software is verified and validated in accordance with its moderate level of concern. Software life cycle processes are evaluated according to the FDA-recognized consensus standard, IEC 62304:2006.
- Application of usability engineering to medical devices is evaluated in accordance with IEC 62366:2008 based on Human Factor Engineering
- Biocompatibility is documented in the reference of ISO 10993-1:2009, ISO 10993-5:2009, and 10993-10:2010.
## Bench testing
Jeisys Medical Inc conducted bench testing to assure that TRI-BEAM PRO, TRI-BEAM operates safely and within the predefined design specifications.
## Risk Management
A risk analysis was conducted based on ISO 14971:2019 Medical devices – Application of risk management to medical devices
## Clinical Data:
No clinical performance testing was performed.
## Conclusion
{9}
K251464
Subject device is substantially equivalent to its predicate devices with same indication for use and technological characteristics. The non-clinical data for proposed device, including biocompatibility, bench testing, hardware, and software documentation shows that the device should perform as intended in the specified use. In addition, the Electromagnetic Compatibility and Electrical Safety testing shows that the device is safe to use and meets require. There are no significant differences between Subject device and the predicate device that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics standards. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification that we conclude that substantially equivalent with predicate device K241527.
{10}
K251464
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.