Laboratorios Biogalenic Sterile Water for Inhalation, USP
K251419 · Laboratorios Biogalenic S.A. DE C.V. · BTT · Jan 14, 2026 · Anesthesiology
Device Facts
Record ID
K251419
Device Name
Laboratorios Biogalenic Sterile Water for Inhalation, USP
Applicant
Laboratorios Biogalenic S.A. DE C.V.
Product Code
BTT · Anesthesiology
Decision Date
Jan 14, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Sterile Water for Inhalation, USP is intended to provide a supply of sterile water to unfilled respiratory humidifier chambers. It is intended to be used in institutional and non-institutional care settings. The intended patient population is Adults, Pediatrics, Infants and Neonates.
Device Story
Sterile, single-dose water supply for respiratory humidifier chambers; provided in flexible polyolefin bags with polypropylene twist-off port. Device suspended via integrated hanger; water gravity-fed to humidifier chamber via tubular feed sets. Used in institutional and non-institutional settings. Heated in humidifier to produce water vapor for medical gas/air humidification; prevents airway drying, maintains mucous membrane moisture, and reduces respiratory complications. Not for parenteral administration or irrigation; single-use only.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including physicochemical analysis (TOC, conductivity, bioburden, sterility, endotoxin), packaging integrity (leakage, hermeticity, pressure, rupture, tensile force, drop/compression tests), and biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, pyrogenicity, extractables/leachables). Sterilization validated to SAL 10^-6 using moist heat.
Technological Characteristics
Flexible polyolefin bag; polypropylene twist-off port; integrated hanger. Sterilized via steam-air (moist heat) at 121-122°C for 97 minutes. Complies with USP standards for sterile water. Packaging tested per ASTM D3078-02, ASTM F1929-23, ASTM F1140/F1140M, ASTM D642, and ISO 15747:2018.
Indications for Use
Indicated for adults, pediatrics, infants, and neonates requiring respiratory humidification via unfilled humidifier chambers.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
Predicate Devices
Sterile Water for Inhalation Flex Bag USP (K090915)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 14, 2026
Laboratorios Biogalenic, S.a. De C.v.
% Jennifer Prussman
Quality Assurance and Regulatory Affairs Consultant
Herschel J. Gaddy & Associates
1805 Oak Ridge Circle, Suite 101
St. Joseph, Missouri 64506
Re: K251419
Trade/Device Name: Laboratorios Biogalenic Sterile Water for Inhalation, USP
Regulation Number: 21 CFR 868.5450
Regulation Name: Respiratory Gas Humidifier
Regulatory Class: Class II
Product Code: BTT
Dated: December 12, 2025
Received: December 12, 2025
Dear Jennifer Prussman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K251419 - Jennifer Prussman
Page
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
3K251419 - Jennifer Prussman
Page
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Laboratorios Biogalenic Sterile Water for Inhalation, USP
Page 11 of 52
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251419 | ? |
| Please provide the device trade name(s). | | ? |
| Laboratorios Biogalenic Sterile Water for Inhalation, USP | | |
| Please provide your Indications for Use below. | | ? |
| Sterile Water for Inhalation, USP is intended to provide a supply of sterile water to unfilled respiratory humidifier chambers. It is intended to be used in institutional and non-institutional care settings. The intended patient population is Adults, Pediatrics, Infants and Neonates. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
# APPLICANT
| Company Name | : | Laboratorios Biogalenc, S.A. DE C.V. |
| --- | --- | --- |
| Company Address | : | Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| Company Phone | : | +503-2227-4133 EXT 106 |
| Company Fax | : | +503-2227-4466 |
| Company Email | : | r.q@biogalenc.com.sv, biogal@biogalenc.com.sv |
| Contact Person | : | Roberto Quiñonez |
| Designation | : | Plant Director |
| Phone | : | +503-2227-4133 EXT 106 |
| Email | : | r.q@biogalenc.com.sv |
| Date Summary Prepared | : | September 18, 2025 |
# DEVICE IDENTIFICATION
| Common Name of the device | Sterile Water for Inhalation, USP |
| --- | --- |
| Device Trade Name | Laboratorios Biogalenc Sterile Water for Inhalation, USP |
| Device Classification Name | Respiratory gas humidifier |
| Device Regulation Number | 21 CFR 868.5450 |
| Device Classification | Class II |
| Product code | BTT |
| Classification Panel | Anesthesiology |
{5}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
# PREDICATE DEVICE
Laboratorios Biogalenc S.A. de C.V. identified the following legally marketed device as substantially equivalent:
| Predicate Device | Applicant | 510(k) Number |
| --- | --- | --- |
| Sterile Water for Inhalation Flex Bag USP | Cardinal Health | K090915 |
# DEVICE DESCRIPTION
Sterile Water for Inhalation, USP is a sterile, single-dose medical device provided in flexible polyolefin bags, designed for safe and easy handling. The flexible polyolefin bags are prefilled with sterile water and feature a single polypropylene twist-off port used to transfer the sterile water to humidifier chambers using tubular feed sets. The humidifier chamber is then filled with sterile water from the bags via gravity. The device has a hanger to suspend the product during use.
Sterile Water for Inhalation, USP is not for parenteral administration or irrigation. The product is For Inhalation Therapy Only. It is supplied sterile and intended for single use only. Discard any unused portion.
# INDICATIONS FOR USE
The Sterile Water for Inhalation, USP is intended to provide a supply of sterile water to unfilled respiratory humidifier chambers. It is intended to be used in institutional and non-institutional care settings. The intended patient population is adults, pediatrics, infants and neonates.
# NON-CLINICAL PERFORMANCE TESTING
# PHYSICOCHEMICAL AND BIOLOGICAL TESTING:
The tests listed below for Sterile Water for Inhalation, USP were performed using USP methods and internally validated procedures. The following tests were conducted on the subject device.
- Total Organic Carbon (TOC)
- Water Conductivity
- Process Bioburden
- Volume in Container
- Sterility
- Bacterial Endotoxin
- Appearance
- Particulate Matter
{6}
LABORATORIOS BIOGALENIC S.A. DE C.V.
LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV
| **510 (k) SUMMARY** |
| --- |
| **STERILE WATER FOR INHALATION, USP** |
Based on these tests, Laboratorios Biogalenic S.A. de C.V. has demonstrated that the device functions as intended, supporting substantial equivalence.
## PACKAGING TESTS:
Specific tests conducted on the device demonstrate that it performs as intended and supports substantial equivalence. The results from the tests indicate compliance with the criteria for each test and provide valuable data on the expected performance of the packaging, ensuring it functions as required. The observed results are:
- Resistance to Leakage (ASTM D3078-02): No leaking or deformation of bags.
- Hermeticity (ASTM F1929-23): Compliance with hermeticity requirements.
- Resistance to Pressure (ASTM F1140/F1140M): The device withstands the required pressure limits.
- Resistance to Rupture (ASTM D642): The device resists rupture under specified conditions.
- Tensile Force (ISO 15747:2018): The device hanger remains intact under specified conditions.
- Resistance to Damage from Dropping Tests and Compression Tests (ISO 15747:2018): The device remains intact and functional after dropping and compression tests.
These results validate the device packaging integrity and durability, confirming its suitability for the intended use.
## BIOCOMPATIBILITY
Biocompatibility testing performed in accordance with ISO 10993-1:2018: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process. The following tests were performed to demonstrate substantial equivalence:
- In Vitro Cytotoxicity (L929 Cell Line) conducted per ISO 10993-5:2009, Third Edition: Biological Evaluation of Medical Devices: Tests for in vitro cytotoxicity
- Sensitization (Guinea Pig Maximization Test) conducted per ISO 10993-10:2021 Fourth Edition: Biological Evaluation of Medical Devices: Tests for Skin Sensitization
- Intracutaneous Reactivity conducted per ISO 10993-12:2021, First Edition: Biological Evaluation of Medical Devices: Tests for Irritation
{7}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
- Material Mediated Pyrogenicity conducted per ISO 10993-11:2017 Third Edition: Biological Evaluation of Medical Devices: Tests for Systemic Toxicity
- Extractable/leachable testing per ISO 10993-18:2020/Amd1:2022: Biological Evaluation of Medical Devices: Chemical Characterization of Medical Devices within a Risk Management Process
- Toxicological risk assessment conducted per ISO 10993-17:2023 Biological Evaluation of Medical Devices: Establishment of Allowable Limits for Leachable Substances on the Detected Chemicals within the Extract
## STERILIZATION VALIDATION
The sterilization process was performed in accordance with PDA Technical Report No. 1 (Revised 2007), FDA Guidance for Industry for the Submission Documentation for Sterilization Process Validation in Applications for Human and Veterinary Drug Products, ISO 17665 Sterilization of Health care Products-Moist Heat, which outlines the validation of moist heat sterilization processes, including cycle design, development, qualification, and ongoing control. This process was qualified and is considered robust for controlling hold times during sterilization. The Sterile Water for Inhalation, USP, was validated for sterilization using steam-air to achieve a Sterility Assurance Level (SAL) of $10^{-6}$. Geobacillus stearothermophilus, known for its resistance to moist heat, was used as a biological indicator, and no growth was observed after being subjected to the sterilization cycle and incubated for 48 hours at $55^{\circ}\mathrm{C} \pm 2^{\circ}\mathrm{C}$, indicating that the sterilization conditions were achieved. This process ensures that the Sterile Water for Inhalation, USP is sterile at the point of use. The product was sterilized at $121^{\circ}\mathrm{C}$ to $122^{\circ}\mathrm{C}$ for a period of 97 minutes.
## SUBSTANTIAL EQUIVALENCE
The comparison between the subject device and the predicate device shows that both devices have the same indications for use, specifically, providing a supply of sterile water for inhalation to humidifier chambers. Both devices share the same BTT classification and are manufactured using the same sterile water as the primary material, packaged in flexible single-use bags. They are similar in design and technological characteristics. Various physicochemical, performance, and biological testing have demonstrated that differences in primary packaging do not pose any significant issues. The subject device functions as intended.
The following Substantial Equivalence Comparison Table summarizes the key features of the subject device compared to the predicate device.
{8}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
| Section | Proposed Device
(Sterile Water for Inhalation, USP) | Predicate Device
(Sterile Water for Inhalation Flex Bag USP) | Result |
| --- | --- | --- | --- |
| 510(k) Number | TBD | K090915 | - |
| Manufacturer | Laboratorios Biogalenc S.A. de C.V. | Cardinal Health | - |
| Device Trade Name | Laboratorios Biogalenc Sterile Water for Inhalation, USP | Sterile Water for Inhalation Flex Bag USP | - |
| Generic/ Common Name | Sterile Water for Inhalation, USP | Sterile Water for Inhalation, USP | - |
| Regulation Number | 21 CFR 868.5450 | 21 CFR 868.5450 | Substantially equivalent |
| Device Classification Name | Humidifier, Respiratory Gas | Humidifier, Respiratory Gas | Substantially equivalent |
| Classification Panel | Anesthesiology | Anesthesiology | Substantially equivalent |
| Device Class | Class II | Class II | Substantially equivalent |
| Product Code | BTT | BTT | Substantially equivalent |
{9}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
| Section | Proposed Device
(Sterile Water for Inhalation, USP) | Predicate Device
(Sterile Water for Inhalation Flex Bag USP) | Result |
| --- | --- | --- | --- |
| Indications for Use | Sterile Water for Inhalation, USP is intended to provide a supply of sterile water to unfilled respiratory humidifier chambers. It is intended to be used in institutional and non-institutional care settings. | Sterile Water for Inhalation Flex Bag USP is intended to provide a supply of sterile water to unfilled respiratory humidifier chambers. It is intended to be used in institutional and non- institutional care settings. | The proposed device and the predicate device have the same indications for use. |
| Intended Patient Population | Adults, Pediatrics, Infants and Neonates | unknown | - |
| Prescription/Over the Counter Use | Prescription Use | Prescription Use | Substantially equivalent |
| Single-Use | Yes | Yes | Substantially equivalent |
| Design | The device is a flexible polyolefin bag prefilled with sterile water. It features a single polypropylene twist-off port used to transfer the sterile water to humidifier chambers using tubular feed sets. The humidifier chamber is then filled with sterile water from the sterile water bags via gravity. The device has a hanger to suspend the product during use. | The device is a flexible plastic bag with a single port at the base and is pre-filled with sterile water. Tubular feed sets connect the sterile water bags to unfilled humidifier chambers. The humidifier chamber is then filled with sterile water from the sterile water bags via gravity. The device has a hanger to suspend the product during use. | Similar (SE#1) |
{10}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
| Section | Proposed Device
(Sterile Water for Inhalation, USP) | Predicate Device
(Sterile Water for Inhalation Flex Bag USP) | Result |
| --- | --- | --- | --- |
| Sizes | 1000mL
2000mL | 1000mL
2000mL | Substantially equivalent |
| Spike Access | Twist-Off Port | Twist-Off Port | Substantially equivalent |
| Hanger | Standard Integrated Hanger | Standard Integrated Hanger | Substantially equivalent |
| Primary Packaging Material | Polyolefin | Polyvinyl Chloride (PVC) | Similar |
| Container Clarity | Clear | Clear | Substantially equivalent |
| Mechanism of Action | Sterile Water for Inhalation, USP, is used to humidify air in respiratory therapy. It is transferred to a humidifier chamber where it is heated to produce water vapor. This vapor mixes with medical gases or air, ensuring the delivered air is moist. This prevents drying of the patient's airways, maintains mucous membrane moisture, enhances patient comfort and reduces the risk of respiratory complications. | Sterile Water for Inhalation, USP, is used to humidify air in respiratory therapy. It is transferred to a humidifier chamber where it is heated to produce water vapor. This vapor mixes with medical gases or air, ensuring the delivered air is moist. This prevents drying of the patient's airways, maintains mucous membrane moisture, enhances patient comfort and reduces the risk of respiratory complications. | Substantially equivalent |
| Biocompatibility | Meets the requirements of ISO 10993 | Unknown | - |
{11}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
| Section | Proposed Device
(Sterile Water for Inhalation, USP) | Predicate Device
(Sterile Water for Inhalation Flex Bag USP) | Result |
| --- | --- | --- | --- |
| Chemical Composition | Sterile Water, USP | Sterile Water | Substantially equivalent |
| Sterilization Method | Steam Air | Moist Heat | Substantially equivalent |
| Compounding | Water for the compounding is produced from drinking water that is purified in its final stage by distillation or other equivalent or superior technology that demonstrates the elimination of chemicals, microorganisms and endotoxins and does not contain any added substances. Sterile water is filled in compounding tanks at 70°C and then filled into bags. | Water for the compounding is produced from drinking water that is purified in its final stage by distillation or other equivalent or superior technology that demonstrates the elimination of chemicals, microorganisms and endotoxins and does not contain any added substances. Sterile water is filled in compounding tanks at 70°C and then filled into bags. | Substantially equivalent |
{12}
| | LABORATORIOS BIOGALENIC, S.A. DE C.V.
Calle Claper, Blvd. Del Ejercito Nacional km 5.5, Soyapango, San Salvador, 1639 SV |
| --- | --- |
| 510 (k) SUMMARY | |
| STERILE WATER FOR INHALATION, USP | |
# CONCLUSION
The Laboratorios Biogalenic S.A. de C.V. Sterile Water for Inhalation, USP, has the same technological characteristics and provides the same mechanism of action as the predicate. According to the above analysis, there is no major difference between the proposed device and the predicate device that would impact effectiveness and safety. The proposed Laboratorios Biogalenic S.A. de C.V. Sterile Water for Inhalation, USP, is considered to be substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.