134 patients (111 USA, 23 Israel) comprising 26,362 ultrasound clips
>1 (cardiologists)
Probe guidance
MLM based algorithms
AUC > 0.8
AUC > 0.8
—
—
134 patients (111 USA, 23 Israel) comprising 2.4 million ultrasound frames
>1 (expert sonographers and/or cardiologists)
Indications for Use
The UltraSight Guidance is intended to assist medical professionals (not including expert sonographers) in acquiring cardiac ultrasound images. UltraSight Guidance is an accessory to compatible general-purpose diagnostic ultrasound systems. UltraSight Guidance is indicated for use in two-dimensional transthoracic echocardiography (2D- TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).
Device Story
Software application using machine learning/AI to provide real-time guidance for cardiac ultrasound image acquisition; assists non-expert users in positioning/orienting transducer. Inputs: real-time ultrasound video frames. Processing: AI identifies probe position/orientation relative to target cardiac views; calculates image quality score. Outputs: 3D probe display with graphic cues (rotation, tilt, rock, slide) and image quality bar. Used in clinical settings by non-expert medical professionals. Output helps users manipulate transducer to achieve diagnostic-quality images; facilitates standard view acquisition; improves clinical workflow for non-sonographers.
Clinical Evidence
Bench testing only. Evaluated using 134 patients (26,362 clips for quality bar; 2.4 million frames for probe guidance). Quality bar performance: mean AUC > 0.8, mean FPR < 0.2. Probe guidance performance: mean AUC > 0.8. Subgroup analyses performed across BMI, age, sex, and geography.
Technological Characteristics
Software-based AI guidance system; utilizes machine learning (MLM) algorithms. Provides real-time visual feedback (quality bar, 3D probe orientation cues). Compatible with third-party general-purpose diagnostic ultrasound systems. Operates as an accessory software application.
Indications for Use
Indicated for adult patients undergoing 2D transthoracic echocardiography (2D-TTE) to assist non-expert medical professionals in acquiring standard cardiac views.
Regulatory Classification
Identification
A radiological acquisition and/or optimization guidance system is a device that is intended to aid in the acquisition and/or optimization of images and/or diagnostic signals. The device interfaces with the acquisition system, analyzes its output, and provides guidance and/or feedback to the operator for improving image and/or signal quality.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed, technical device description, including a detailed description of the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, and the associated inputs and outputs.
(ii) A detailed, technical report on the non-clinical performance testing of the subject device in the intended use environments, using relevant consensus standards when applicable.
(iii) A detailed report on the clinical performance testing, obtained from either clinical testing, accepted virtual/physical systems designed to capture clinical variability, comparison to a closely-related device with established clinical performance, or other sources that are justified appropriately. The choice of the method must be justified given the risk of the device and the general acceptance of the test methods. The report must include the following:
(A) A thorough description of the testing protocol(s).
(B) A thorough, quantitative evaluation of the diagnostic utility and quality of images/data acquired, or optimized, using the device.
(C) A thorough, quantitative evaluation of the performance in a representative user population and patient population, under anticipated conditions and environments of use.
(D) A thorough discussion on the generalizability of the clinical performance testing results.
(E) A thorough discussion on use-related risk analysis/human factors data.
(iv) A detailed protocol that describes, in the event of a future change, the level of change in the device technical specifications or indications for use at which the change or changes could significantly affect the safety or effectiveness of the device and the risks posed by these changes. The assessment metrics, acceptance criteria, and analytical methods used for the performance testing of changes that are within the scope of the protocol must be included.
(v) Documentation of an appropriate training program, including instructions on how to acquire and process quality images and video clips, and a report on usability testing demonstrating the effectiveness of that training program on user performance, including acquiring and processing quality images.
(2) The labeling required under § 801.109(c) of this chapter must include:
(i) A detailed description of the device, including information on all required and/or compatible parts.
(ii) A detailed description of the patient population for which the device is indicated for use.
(iii) A detailed description of the intended user population, and the recommended user training.
(iv) Detailed instructions for use, including the information provided in the training program used to meet the requirements of paragraph (b)(1)(iv) of this section.
(v) A warning that the images and data acquired using the device are to be interpreted only by qualified medical professionals.
(vi) A detailed summary of the reports required under paragraphs (b)(1)(ii) and (iii) of this section.
(vii) A statement on upholding the As Low As Reasonably Achievable (ALARA) principle with a discussion on the associated device controls/options.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 17, 2025
UltraSight Ltd.
Noa Avisar
VP Clinical and Regulatory Affairs
8 Pinhas St.
Ness Ziona, 7403631
Israel
Re: K251416
Trade/Device Name: UltraSight Guidance
Regulation Number: 21 CFR 892.2100
Regulation Name: Radiological Acquisition And/Or Optimization Guidance System
Regulatory Class: Class II
Product Code: QJU
Dated: November 19, 2025
Received: November 19, 2025
Dear Noa Avisar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251416 - Noa Avisar
Page 2
established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K251416 - Noa Avisar
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YANNA S. KANG -S
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251416 | |
| Device Name | |
| UltraSight Guidance | |
| Indications for Use (Describe) | |
| The UltraSight Guidance is intended to assist medical professionals (not including expert sonographers) in acquiring cardiac ultrasound images. UltraSight Guidance is an accessory to compatible general-purpose diagnostic ultrasound systems. UltraSight Guidance is indicated for use in two-dimensional transthoracic echocardiography (2D- TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
ULTRASIGHT
510(k) SUMMARY
Page: 1 of 8
510(k) SUMMARY
UltraSight Guidance
K251416
Applicant name: UltraSight Ltd.
8 Pinhas Sapir St.
Ness Ziona 7403631
Israel
Phone: +972-54-8886425
Email: noa@ultrasight.com
Contact Person: Noa Avisar
Date Prepared: 11/11/2025
Name of Device: UltraSight Guidance
Classification Name: Image Acquisition And/Or Optimization Guided By Artificial Intelligence
Classification Code: QJU
Device class: II
Regulation number: 892.2100
Panel: Radiology
Predicate Devices: UltraSight Guidance Software (K223347)
Intended Use / Indications for Use
The UltraSight Guidance is intended to assist medical professionals (not including expert sonographers) in acquiring cardiac ultrasound images. UltraSight Guidance is an accessory to compatible general-purpose diagnostic ultrasound systems. UltraSight Guidance is indicated for use in two-dimensional transthoracic echocardiography (2D- TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).
Device Description
UltraSight Guidance is a software application based on machine learning that uses artificial intelligence (AI) to provide dynamic real-time guidance on the position and orientation of the transducer to help non-expert users acquire diagnostic-quality tomographic views of the heart. The system provides guidance for ten standard cardiac views.
{5}
ULTRASIGHT
510(k) SUMMARY
Page: 2 of 8
Main features:
- Quality Bar: The system displays an image quality bar that is continuously updated while the user scans the subject, while attempting to find the maximal quality. The quality bar is a score for image diagnosability. It represents the classification between high and low quality images, where high quality images are defined as grade 3 or more based on American College of Emergency Physicians (ACEP) guidelines (Rachel B. Liu et al., "Emergency Ultrasound Standard Reporting Guidelines", 2018, American College of Emergency Physicians)
- Probe Guidance: The probe guidance feature provides graphic on-screen instructions for the user to emulate how a sonographer would manipulate the transducer to acquire the target cardiac view. The five possible guidance cues are rotation, tilt, rock, and slides in the lateral-medial and up/down directions (with respect to the subjects' head).
The guidance user interface (UI) is composed of a 3D probe display that shows orientation guidance cues (rotations, tilts and rocks) and a cross that shows slide guidance cues (slides in x and y directions). The users infer the maneuver they should perform from viewing the 3D probe display and the slides cross. Supporting text messages appear on the screen regarding guidance cues.
This device is a modification of a previously marketed device, where the main modification introduced is the addition of a pre-processing step, done prior to feeding images as an input to the device. This modification is aimed to enable future compatibility with potentially additional ultrasound probes that meet pre-requisites (as part of a pre-determined change control plan).
Technological Characteristics
The UltraSight Guidance software application is similar in its technological features to its predicate device. Both systems are intended as an assistive tool to aid medical professionals in the acquisition of cardiac ultrasound images. Specifically, both systems are indicated to aid in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).
The technological characteristics of the UltraSight Guidance software are identical to the technological characteristics of the predicate device. Both systems apply MLM based algorithms to provide real-time guidance on how to position and manipulate the transducer on a patient's body. Both systems are coupled with a third-party, previously cleared ultrasound system and provide to the users real-time guidance during acquisition of 2D echocardiography to assist them in obtaining echocardiography images that represent standard diagnostic views and orientations.
{6}
ULTRASIGHT
510(k) SUMMARY
Page: 3 of 8
A table comparing the key features of the subject and the predicate devices is provided below:
Table 1: Substantial Equivalence
| Parameters | UltraSight AI Guidance (K223347) | Proposed UltraSight Guidance (K251416) | Conclusion |
| --- | --- | --- | --- |
| Classification name | Image Acquisition And/or Optimization Guided by Artificial Intelligence | Same | Identical |
| Product Code | QJU | Same | Identical |
| Intended Use | The UltraSight AI Guidance software is intended to assist medical professionals in the acquisition of cardiac ultrasound images. UltraSight AI Guidance software is an accessory to compatible general purpose diagnostic ultrasound systems | Same | Identical |
| Indications for use | The UltraSight AI Guidance is intended to assist medical professionals (not including expert sonographers) in acquiring cardiac ultrasound images. UltraSight AI Guidance is an accessory to compatible general-purpose diagnostic ultrasound systems. UltraSight AI Guidance is indicated for use in two-dimensional transthoracic echocardiography (2D- TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), | Same | Identical |
{7}
ULTRASIGHT
510(k) SUMMARY
Page: 4 of 8
| Parameters | UltraSight AI Guidance (K223347) | Proposed UltraSight Guidance (K251416) | Conclusion |
| --- | --- | --- | --- |
| | Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC). | | |
| Intended user | Medical professionals (not including expert sonographers) | Same | Identical |
| Compatible Ultrasound system and probe | Philips Lumify with S4-1 probe and compatible tablets defined in the labeling | Same | Identical |
| Clinical Features: | | | |
| Image acquisition guidance | Probe Guidance: this feature, which is the core of the application, provides on-screen instructions to the user to emulate how an expert would manipulate the transducer to acquire the optimal view. | Same | Identical |
| Real time feedback on image quality | Quality bar: real-time feedback on the clip's quality advises the user on the expected diagnostic quality of the current clip. This information can be used by the users to assess how close they are to capturing a diagnostic-quality image. | Same | Identical |
| Deep Learning Based Algorithm | Downloadable application that utilizes artificial intelligence (AI) to provide real-time dynamic guidance of transducer position and orientation to help non-expert users acquire diagnostic-quality tomographic views of the heart. | Same | Identical |
{8}
ULTRASIGHT 510(k) SUMMARY Page: 5 of 8
| Parameters | UltraSight AI Guidance (K223347) | Proposed UltraSight Guidance (K251416) |
| --- | --- | --- |
| Predetermined Change Control Plan (PCCP):
M1 Retraining - to enhance MLM performance, to optimize MLM performance for specific demographic population within the intended use, and to increase stability in consecutive MLM predictions. | Not included | Added to the PCCP |
| M2 Expand compatibility with additional ultrasound systems | Not included | Added to the PCCP |
| M3 Addition of target cardiac views | Not included | Added to the PCCP |
## Substantial Equivalence
The UltraSight Guidance software has the same intended use, indications for use, technological characteristics and principles of operation as the predicate device. As discussed above, while there are minor differences between the subject and predicate devices, the differences do not raise any different questions of safety or effectiveness, and bench performance data is available to demonstrate that the performance of UltraSight Guidance is equivalent to the performance of the predicate device. Therefore, the devices are substantially equivalent.
## Non-Clinical Performance Data
The following performance testing was conducted to evaluate the device:
### Quality Bar:
The quality bar is a score for image diagnosability. It represents the classification between high- and low-quality images, where high quality images are defined as grade 3 or more based on ACEP guidelines.
The test dataset was collected from clinical sites geographically distinct from those used for the development dataset. It included a total of 134 patients - 111 from the USA and 23 from Israel - comprising 26,362 ultrasound clips. The dataset represents a diverse patient population with variations in body mass index (BMI: 18-39.3), age (18 - 91 years), and sex (50.7% female and 49.3% male).
Each clip in the dataset is associated with a target cardiac view and was annotated with a clip diagnosability label (diagnosable / non-diagnosable).
The test assessed the classification performance between "diagnosable" and "non diagnosable" clips of each view. The clips were annotated by two expert cardiologists for its diagnosability. In case of disagreement, a third cardiologist provided additional annotation, and the final label was determined
{9}
JLTRASIGHT
510(k) SUMMARY
Page: 6 of 8
by majority vote among the three experts. The mean area under the receiver operating characteristic curve (AUC) was within the pre-defined acceptance criteria (AUC > 0.8) with 95% CI showing good classification performance, relative to the success criteria. The mean false positive rate (FPR) met the acceptance criteria of FPR < 0.2 with 95% CI showing good classification performance. Subgroup analyses were conducted across BMI categories (<25, ≥25 and <30, ≥30), age groups (<50, ≥50 and <65, ≥65), sex (male and female) and US vs non-US.
## Probe Guidance:
The system provides guidance cues when the heart is visible on the ultrasound image. To provide the correct guidance cues, the system needs to be able to identify the position and orientation of the probe relative to the target cardiac view. There are five possible guidance cues: rotations, tilts, rocks, and slides in the x (lateral-medial) and y (superior-inferior) directions.
The test dataset was collected from clinical sites geographically distinct from those used for the development dataset. It included a total of 134 patients - 111 from the USA and 23 from Israel - comprising 2.4 million ultrasound frames. The dataset encompasses a diverse patient population with variation in body mass index (BMI: 18-39.3), age (18 - 91 years), and sex (50.7% female and 49.3% male).
Ground truth was established by expert sonographers and/or cardiologists using the recorded probe position during ultrasound acquisition. The relative position of the probe at each frame, compared to the position associated with the target view, defined the required rotation, tilt, rock and slides adjustments.
The testing evaluated the frame level accuracy of each guidance cue prediction. For each guidance cue, a series of tests was defined to check if the guidance cue is functioning correctly in a particular direction (for example, clockwise rotations relative to 2C view, tilts up from PLAX). Each test was formulated as a binary classification test. The mean AUC was within the acceptance criteria of AUC>0.8 with 95% CI showing good classification performance. Subgroup analyses were conducted across BMI categories (<25, ≥25 and <30, ≥30), age groups (<50, ≥50 and <65, ≥65), sex (male and female) and US vs non-US.
## Software and Cybersecurity:
The device underwent comprehensive software validation and cybersecurity testing in accordance with the FDA's Guidance "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions". These evaluations ensured the software meets its intended use, functions reliably under expected conditions, and incorporates appropriate risk-based cybersecurity controls. The submission includes documentation of threat modeling, vulnerability testing, and secure design practices to support the safety and effectiveness of the device in today's evolving cybersecurity landscape.
{10}
ULTRASIGHT
510(k) SUMMARY
Page: 7 of 8
# Predetermined Change Control Plan (PCCP)
A Predetermined Change Control Plan (PCCP) is added to describe what modifications will be made to the subject device following the marketing authorization (i.e. 510(k) clearance) and how the modifications will be assessed.
Each modification is provided with an impact assessment and a detailed test plan to ensure the device with a modification maintains safety and effectiveness as the predicate device. The device with the planned modifications has the same intended use, indication for use, technological characteristics and principles of operation as the predicate device, therefore remaining substantially equivalent to the predicate device.
Table 3: PCCP Modifications
| Modification | Rationale for Modification | Testing Methods, Validation activities and performance requirements | Communication to users |
| --- | --- | --- | --- |
| Model re-training | Improvement and optimization of MLM: Enhance MLM performance, Optimize MLM performance for specific demographic population within the intended use, Increase stability in consecutive MLM predictions. | Repeating verification tests and the system level validation tests to ensure the following pre-defined acceptance criteria are met: Quality Bar:
• Mean area under the ROC curve (AUC) > 0.8
• Mean false positive rate (FPR) at the green quality threshold < 0.2
Guidance:
• Lower bound of the 95% CI of the mean AUC > 0.8 | The labeling will be updated in accordance with the authorized PCCP, to provide users with current information regarding updated performance results. |
| Expand Compatibility with additional ultrasound systems | Increase accessibility of UltraSight Guidance | Image parameters from the calibrated candidate ultrasound device are compared to the predefined boundaries of applicability. Compatibility is confirmed when the parameters fall within these boundaries. | The labeling will be updated in accordance with the authorized PCCP, with the newly added compatible ultrasound systems.
In addition, the user manual will include information on the methodology of how a candidate ultrasound device was approved. |
| Expanded list of cardiac views | Enhance the system's clinical use by offering guidance to additional supported cardiac views | Repeating verification tests and the system level validation tests to ensure the following pre-defined acceptance criteria are met for both the complete dataset (including enrichment with | The labeling will be updated in accordance with the authorized PCCP, to provide users with current information regarding updated cardiac views to support the change. |
{11}
ULTRASIGHT
510(k) SUMMARY
Page: 8 of 8
| | | additional supported views) and the original test dataset: | |
| --- | --- | --- | --- |
| | | Quality Bar: | |
| | | • Mean area under the ROC curve (AUC) > 0.8 | |
| | | • Mean false positive rate (FPR) at the green quality threshold < 0.2 | |
| | | Guidance: | |
| | | • Lower bound of the 95% CI of the mean AUC > 0.8 | |
| | | Clinical utility test will evaluate the ability of non-expert users to obtain diagnostic-quality cardiac images using UltraSight software for the newly added cardiac views. Images will be evaluated by experts for a pre-defined clinical outcome related to the view and for their diagnostic-quality. The study will be considered successful when the clinical outcome can be achieved in ≥75% of subjects. | |
# Conclusions
In this Traditional 510(k) premarket notification, UltraSight Ltd. demonstrates substantial equivalence of the proposed UltraSight Guidance software to the currently marketed and predicate device UltraSight AI Guidance (K223347). UltraSight Guidance Software has the same intended use and indications for use, and same technological characteristics, and principles of operation as the predicate device. The modifications implemented in the subject device do not raise different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.