The T2 Alpha Humerus Nailing System is indicated for the treatment of humerus fractures. Fractures can include, but are not limited to, non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures. The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization. The T2 Humeral Nail is intended to provide temporary stabilization of various types of fractures, malunions, and non-unions of the humerus. The nails are inserted using an open or closed technique and can be static, dynamic, or compression locked. The subject and predicate devices are indicated for use in the humerus. Types of fractures include, but are not limited to, fractures of the humeral shaft, non-unions, malalignments, pathological humeral fractures, and impending pathological fractures.
Device Story
T2 Alpha Humerus Nailing System comprises sterile titanium alloy implants (nails, end caps, compression screws, washers) and non-sterile instrumentation. Used for temporary stabilization of humeral fractures via open or closed surgical techniques. Nails allow static, dynamic, or compression locking. IMN Screws (locking and advanced locking) stabilize the nail-bone construct. Operated by orthopedic surgeons in clinical/OR settings. Provides mechanical fixation to facilitate bone healing. Benefits include stabilization of complex fracture types and pathological conditions. No software or AI components.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including dynamic/static 4-point bending, static torsional stiffness, targeting accuracy, insertion torque, pull-out force, MR compatibility (ASTM F2052, F2213, F2182, F2119), packaging validation (ISO 11607), and biocompatibility (ISO 10993-1).
Technological Characteristics
Materials: Titanium alloy (Ti6Al4V ELI) per ASTM F136. Components: Intramedullary nails, end caps, compression screws, washers, locking screws. Sterilization: Gamma irradiation for implants. Mechanical principle: Intramedullary fixation rod/bone fixation fastener. No software or electronic components.
Indications for Use
Indicated for patients requiring treatment of humerus fractures, including non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures of the humeral shaft.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 20, 2025
Stryker GmbH
Jackie Perri
Staff Specialist, Regulatory Affairs
Bohnackerweg 1
Selzach, 2545
Switzerland
Re: K251400
Trade/Device Name: T2 Alpha Humerus Nailing System; IMN Screws System; T2 Nailing System
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB, HWC
Dated: May 6, 2025
Received: July 21, 2025
Dear Jackie Perri:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251400 - Jackie Perri
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251400 - Jackie Perri
Page 3
Sincerely,
Joseph P. Russell -S
Digitally signed by Joseph P. Russell -S
Date: 2025.08.20 13:17:25 -04'00'
for: Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251400
Device Name
T2 Alpha Humerus Nailing System;
IMN Screws System;
T2 Nailing System
Indications for Use (Describe)
The T2 Alpha Humerus Nailing System is indicated for the treatment of humerus fractures. Fractures can include, but are not limited to, non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures.
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
The T2 Humeral Nail is intended to provide temporary stabilization of various types of fractures, malunions, and non-unions of the humerus. The nails are inserted using an open or closed technique and can be static, dynamic, or compression locked. The subject and predicate devices are indicated for use in the humerus. Types of fractures include, but are not limited to, fractures of the humeral shaft, non-unions, malalignments, pathological humeral fractures, and impending pathological fractures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Stryker GmbH T2 Alpha Humerus & IMN Screws
K251400 - Page 1 of 3
# 510(k) Summary
## Proprietary name:
T2 Alpha Humerus Nailing System
IMN Screws System
T2 Nailing System
## Common name:
Rod, fixation, intramedullary, and accessories
Screw, fixation, bone.
## Regulation Number:
21 CFR 888.3020: Intramedullary fixation rod
21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener
## Product Code(s):
HSB
HWC
## Device Class:
Class II
## Sponsor:
Stryker GmbH
Bohnackerweg 1
2545 Selzach, Switzerland
## Contact Person:
Jackie Perri
Staff Specialist, Regulatory Affairs
Stryker
325 Corporate Drive
Mahwah, NJ 07430
jackie.perri@stryker.com
## Date Prepared:
May 06, 2025
## Predicate Device:
Primary Predicate: T2 Nailing System (K032523)
Additional Predicate: Tornier Humeral Nail and Tornier Long Humeral Nail (K230352)
IMN Screws System (K213328)
## Device Description:
The T2 Alpha Humerus Nailing System is an intramedullary humerus fracture nailing system consisting of sterile implants (Nails, End Caps, Compression Screw, and Washer) and non-sterile indication-specific instrumentation. The Nails, End Caps, Compression Screw, and Washer are made of titanium alloy as per ASTM F136. The T2 Alpha Humerus Nailing System will be used with the existing Locking Screws and Advanced Locking Screws of the IMN Screws System.
The IMN Screws System includes bone screws (Locking Screws and Advanced Locking Screws) that are inserted through the
510(k) Summary
{5}
Stryker GmbH T2 Alpha Humerus & IMN Screws
K251400 - Page 2 of 3
intramedullary nail to stabilize the nail-bone construct. All screws are sterile and made of titanium alloy (Ti6Al4V ELI) per ASTM F136.
The T2 Humeral Nail System is an intramedullary nailing system that allows antegrade and retrograde humeral nailing. The nails, end caps, compression screw, and washer are provided sterile and made of titanium alloy as per ASTM F136.
## Indications for Use:
The T2 Alpha Humerus Nailing System is indicated for the treatment of humerus fractures. Fractures can include, but are not limited to, non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures.
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
The T2 Humeral Nail is intended to provide temporary stabilization of various types of fractures, malunions, and non-unions of the humerus. The nails are inserted using an open or closed technique and can be static, dynamic, or compression locked. The subject and predicate devices are indicated for use in the humerus. Types of fractures include, but are not limited to, fractures of the humeral shaft, non-unions, malalignments, pathological humeral fractures, and impending pathological fractures.
## Comparison to predicate device:
The intended use and indications of the T2 Alpha Humerus devices are comparable to those of the predicate devices. The subject device is introducing new nail diameters (10, 11, and 12 mm). The subject device implants and the predicate device implants are both sterile and manufactured from titanium alloy. There is no change in the fundamental scientific technology shared by both the subject and predicate devices.
The intended use and indications of the IMN Screw devices are identical to those of the predicate device. The advanced locking screw lengths of the subject device are shorter in length (Ø4mm – 20, 22.5, and 25 mm). The subject device implants and the predicate device implants are both sterile and manufactured from titanium alloy. There is no change in the fundamental scientific technology shared by both the subject and predicate devices.
The intended use and indications of the T2 devices are identical to those of the predicate device. The new sterile washers are similar to the current washer previously cleared in this system, made of titanium alloy as per ASTM F136. The subject device implants and the predicate device implants are both sterile and manufactured from titanium alloy. There is no change in the fundamental scientific technology shared by both the subject and predicate devices.
510(k) Summary
{6}
Stryker GmbH T2 Alpha Humerus & IMN Screws
K251400 - Page 3 of 3
## Performance Data:
### Non-Clinical Performance:
The T2 Alpha Humerus Nailing system implants are single-use, sterile devices. These implants are sterilized using gamma irradiation. The T2 Alpha Humerus Nailing system also comprises instruments that are provided non-sterile.
The IMN Screws are single-use sterile devices. These screws are sterilized using gamma irradiation.
Comparative assessment of the predicate and reference systems demonstrated substantial equivalence. The following factors were considered:
- Dynamic 4 Point Bending
- Static 4 Point Bending
- Static Torsional Stiffness
- Targeting Accuracy
- Insertion Torque and Pull-out Force
MR assessment of magnetically induced displacement force per ASTM F2052, magnetically induced torque per ASTM F2213, RF-induced heating per ASTM F2182, and image artifacts per ASTM F2119.
Packaging tests were performed according to ISO 11607-1 and ISO 11607-2.
Biocompatibility evaluation according to ISO 10993-1.
### Clinical Performance:
Clinical data were not needed for the subject devices to demonstrate substantial equivalence to the predicate devices.
## Conclusion:
The subject devices have similar intended use and indications for use as the predicate device. The subject devices use the same operating principle, incorporate the same basic design and labeling, and are manufactured and sterilized using the same materials and processes as the predicate device.
The performance analyses demonstrate that:
- Any differences do not raise different questions of safety and effectiveness; and
- The subject devices are at least as safe and effective as the legally marketed predicate device.
510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.