K251398 · Shenzhen Enmind Technology Co., Ltd. · OHT · Oct 2, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251398
Device Name
IPL Hair Removal Device (IPL-001, IPL-002)
Applicant
Shenzhen Enmind Technology Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Oct 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
IPL Hair Removal Device (models IPL-001, IPL-002) is an over-the-counter, home-use device for hair reduction. It utilizes Intense Pulsed Light (IPL) generated by a Xenon Arc Flashlamp to reduce hair growth without cutting or pulling. The device is powered by an external adapter. It features a skin contact sensor that prevents light emission unless the light exit window is in full contact with the skin. Users activate light pulses via a manual finger switch or an auto-emission mode. Model IPL-002 includes an integrated cooling function for user comfort and a sapphire light exit window. The device is operated by the user at home. By delivering controlled light pulses to the skin, the device targets hair follicles to achieve long-term, stable reduction in hair regrowth, providing a non-invasive alternative to traditional hair removal methods.
Clinical Evidence
No clinical data. Bench testing only. Verification included output energy density, pulse duration, and flash count. Compliance with ISO 10993 (biocompatibility), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), IEC 60601-2-83 (light therapy), and IEC 62471 (photobiological safety) was demonstrated. Usability engineering was conducted per FDA guidance.
Technological Characteristics
Intense Pulsed Light (IPL) system using a Xenon Arc Flashlamp. Powered by external 100-240V, 50/60Hz adapter. Wavelengths: 580-1200nm (IPL-001) and 560-1200nm (IPL-002). Energy density: 1.5-5.5 J/cm². Pulse duration: 4.5-15.5ms. Features skin contact sensor, manual/auto emission modes, and cooling function (IPL-002). Sapphire window (IPL-002). Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-83, and IEC 62471.
Indications for Use
Indicated for removal of unwanted body and/or facial hair and permanent reduction in hair regrowth in adult users.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 2, 2025
Shenzhen Enmind Technology Co., Ltd.
% Tangyao Dai
Registration Specialist
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center
No. 3101-90, Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K251398
Trade/Device Name: IPL Hair Removal Device (IPL-001, IPL-002)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: September 2, 2025
Received: September 2, 2025
Dear Tangyao Dai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K251398 - Tangyao Dai
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K251398 - Tangyao Dai
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2025.10.02
16:14:24'04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
IPL Hair Removal Device
Page 9 of 42
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251398 | ? |
| Please provide the device trade name(s). | | ? |
| IPL Hair Removal Device (IPL-001, IPL-002) | | |
| Please provide your Indications for Use below. | | ? |
| IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial
hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term,
stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the
completion of a treatment regime. | | |
| Please select the types of uses (select one or both, as
applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K251398 | 510(k) Summary | Prepared on: 2025-10-02 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Shenzhen Enmind Technology Co., Ltd. | |
| Applicant Address | Room 201, Block A, No.1, Qianhai Road 1, Qianhaishen Port Cooperative District, Shenzhen, 518000 Guangdong, P.R. China | |
| Applicant Contact Telephone | +86 15302788688 | |
| Applicant Contact | Mr. Junfeng Wang | |
| Applicant Contact Email | erik.wang@enmind-tech.com | |
| Correspondent Name | FEIYING Drug & Medical Consulting Technical Service Group | |
| Correspondent Address | Rm 2401 Zhenye International Business Center, No. 3101-90, Qianhai Road Shenzhen Guangdong 518052 China | |
| Correspondent Contact Telephone | +86 18027206248 | |
| Correspondent Contact | Ms. Tracy Che | |
| Correspondent Contact Email | 1273487536@qq.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | IPL Hair Removal Device (IPL-001, IPL-002) | |
| Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology | |
| Classification Name | Light Based Over-The-Counter Hair Removal | |
| Regulation Number | 878.4810 | |
| Product Code(s) | OHT | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K230122 | Hair Removal Device, Model(s): UI04 SD, UI04 DG | OHT |
| K242710 | Hair Removal Device (R3C16-P, R3C16-W, R3C16-G, R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro) | OHT |
| K130315 | iPulse Hair Removal System | OHT |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| IPL Hair Removal Device is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It | | |
{5}
works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
IPL Hair Removal Device includes two models, IPL-001 and IPL-002. The two models adopt identical intended use, similar performance, operation and structure, with main differences in appearance, dimensions and weight, and light output parameters.
The device is only powered by the external power adapter. This device adopts light exit window (the IPL-002 adopts sapphire light exit window) that is suitable for multiple hair removal areas. The device is fitted with a skin sensor to detect appropriate skin contact, if the light exit window of the device is not in full contact with the skin, the device cannot emit light pulses, and the IPL emission activation is by manual finger switch or auto light emission. Model IPL-002 has a cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device and predicate device have the same indications for use, both are used for unwanted hair removal.
## Technological Comparison
21 CFR 807.92(a)(6)
The technical characteristic of IPL Hair Removal Device (IPL-001, IPL-002) are substantially equivalent to the predicate device and reference devices in the following aspects:
1. the same intended use, mode of action;
2. the same source energy and power supply: supplied by external adapter with 100-240V~, 50/60Hz;
3. the same light source: Intense Pulsed Light;
4. the same energy medium: Xenon Arc Flashlamp;
The difference between the subject device and the predicate devices mainly includes the following:
1. Wavelength range: The wavelength range of IPL-001 is 580-1200nm, and IPL-002 is 560-1200nm, which are not exactly the same as the predicate device, however, it's within the minimum range of the predicate device, so this difference will not raise any safety or effectiveness issue.
2. Energy density: The energy density of IPL-001 is 1.5-3.2J/cm² and IPL-002 is 2.7-5.5J/cm² which are different from that of the predicate device, however, they are within the range of the minimum and maximum value of the predicate and reference devices, so this difference will not raise any safety or effectiveness issue.
3. Pulse duration: The pulse duration of IPL-001 is 4.5ms-7ms, and IPL-002 is 11.5ms-15.5ms, which are different from that of the predicate device, however it's within the range of the minimum and maximum value of the predicate and reference devices, so this difference will not raise any safety or effectiveness issue.
Thus the subject device is determined to be substantially equivalent to the predicate device and reference devices.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Non-clinical testing have been conducted to verify that the IPL Hair Removal Device meets all design specifications which supports the conclusion that it's substantially equivalent to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation
IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-1-11 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
IEC 60601-2-83 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
IEC 62471 Photobiological safety of lamps and lamp systems
The device software has been evaluated as per FDA guidance "Content of Premarket Submissions for Device Software Functions".
Usability engineering study has been conducted and demonstrated conformance to FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices Guidance for Industry and Food and Drug Administration Staff February 2016".
{6}
In order to verify and assure the performance, function and quality of the IPL Hair Removal Device, we have conducted the verification of output energy density, pulse duration time, valid times of flashes.
Clinical test is not applicable, thus, there's no clinical data for the device.
Based on the above analysis and tests, it can be concluded that the IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.