K251370 · Canon Medical Systems Corporation · KPS · Dec 1, 2025 · Radiology
Device Facts
Record ID
K251370
Device Name
Cartesion Prime (PCD-1000A/3) V10.21
Applicant
Canon Medical Systems Corporation
Product Code
KPS · Radiology
Decision Date
Dec 1, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
PET image quality and denoising
Deep learning artificial neural network
—
—
FDG studies of 16 cancer patients (BMI: 19.5-34.5)
—
Bench testing using NEMA NU-2 phantoms (Adult, Pediatric, Small Pool); Clinical evaluation of 10 PET DICOM 18F-FDG whole body cases
3 (physicians)
Respiratory motion-corrected PET imaging
Deep learning artificial neural network
—
—
FDG studies of 27 cancer patients (BMI: 19.82-45.35)
—
Bench testing using 11 clinical raw data patient datasets; Clinical evaluation of 10 patient cases
3 (physicians)
Indications for Use
The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces cross-sectional images of the body by computer reconstruction of X-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images. This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, localization, diagnosis, staging, restaging, follow-up, therapeutic planning and therapeutic outcome assessment of (but not limited to) oncological, cardiovascular, neurological diseases and disorders. Additionally, this device can be operated independently as a whole body multi-slice CT scanner. AiCE-i for PET is intended to improve image quality and reduce image noise for FDG whole body data by employing deep learning artificial neural network methods which can explore the statistical properties of the signal and noise of PET data. The AiCE algorithm can be applied to improve image quality and denoising of PET images. Deviceless PET Respiratory gating system, for use with Cartesian Prime PET-CT system, is intended to automatically generate a gating signal from the list-mode PET data. The generated signal can be used to reconstruct motion corrected PET images affected by respiratory motion. In addition, a single motion corrected volume can automatically be generated. Resulting motion corrected PET images can be used to aid clinicians in detection, localization, evaluation, diagnosis, staging, restaging, follow-up of diseases and disorders, radiotherapy planning, as well as their therapeutic planning, and therapeutic outcome assessment. Images of lesions in the thorax, abdomen and pelvis are mostly affected by respiratory motion. Deviceless PET Respiratory gating system may be used with PET radiopharmaceuticals, in patients of all ages, with a wide range of sizes, body habitus, and extent/type of disease.
Device Story
Cartesion Prime (PCD-1000A/3) V10.21 is a combined PET/CT diagnostic imaging system. It acquires X-ray transmission data for CT reconstruction and PET radiopharmaceutical distribution data. The system uses CT images for attenuation correction and anatomical fusion. Operated by trained healthcare professionals in clinical settings, it provides metabolic and functional information to assist in disease diagnosis, staging, and therapeutic planning. The device features AiCE-i for PET, a deep learning-based algorithm that denoises and improves image quality, and a Deviceless PET Respiratory gating system that uses a neural network to extract respiratory motion signals from list-mode PET data for motion-corrected image reconstruction. Clinicians view the resulting fused or motion-corrected images to evaluate lesions and physiological function. The device benefits patients by providing higher-quality, motion-corrected images, potentially improving diagnostic accuracy and treatment planning.
Clinical Evidence
Bench-only testing. Performance validated via phantom studies (NEMA NU-2) comparing AiCE2 to AiCE1 and DRG2 to DRG1. Metrics included Contrast Recovery Coefficient (CRC), Background Variability (BGV), Contrast to Noise Ratio (CNR), and SUV accuracy. Clinical evaluation involved 10 patients (for AiCE2) and 11 patients (for DRG2) evaluated by three physicians with 20+ years of experience; results confirmed diagnostic quality and improved image sharpness/noise reduction compared to OSEM/Gaussian methods and device-based gating.
Technological Characteristics
Multi-slice helical CT scanner (780 mm aperture, 700 mm FOV) and digital PET detector with SiPM sensors (temporal resolution < 250 ps). Connectivity: DICOM. Software: Deep learning artificial neural networks for image denoising (AiCE-i) and respiratory gating. Standards: IEC 60601-1, IEC 62304, NEMA NU-2, NEMA XR-29.
Indications for Use
Indicated for patients of all ages, sizes, and body habitus requiring PET/CT diagnostic imaging for metabolic/physiologic assessment, including oncological, cardiovascular, and neurological diseases. Used for detection, localization, diagnosis, staging, restaging, follow-up, and therapeutic planning/assessment. Respiratory gating indicated for motion correction in thorax, abdomen, and pelvis.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 1, 2025
Canon Medical Systems Corporation
% Yoshiaki Cook
Sr. Manager, Regulatory Affairs
Canon Medical Systems, USA
2441 Michelle Drive
Tustin, California 92780
Re: K251370
Trade/Device Name: Cartesion Prime (PCD-1000A/3) V10.21
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission Computed Tomography System
Regulatory Class: Class II
Product Code: KPS, JAK
Dated: October 8, 2025
Received: October 8, 2025
Dear Yoshiaki Cook:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251370 - Yoshiaki Cook
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K251370 - Yoshiaki Cook
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Cartesian Prime (PCD-1000A/3) V10.21
# Indications for Use
| Indications for Use | K251370 | ? |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | ? |
| Please provide the device trade name(s). | | ? |
| Cartesion Prime (PCD-1000A/3) V10.21 | | |
| Please provide your Indications for Use below. | | ? |
| The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces cross-sectional images of the body by computer reconstruction of X-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images. | | |
| This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, localization, diagnosis, staging, restaging, follow-up, therapeutic planning and therapeutic outcome assessment of (but not limited to) oncological, cardiovascular, neurological diseases and disorders. Additionally, this device can be operated independently as a whole body multi-slice CT scanner. | | |
| AiCE-i for PET is intended to improve image quality and reduce image noise for FDG whole body data by employing deep learning artificial neural network methods which can explore the statistical properties of the signal and noise of PET data. The AiCE algorithm can be applied to improve image quality and denoising of PET images. | | |
| Deviceless PET Respiratory gating system, for use with Cartesian Prime PET-CT system, is intended to automatically generate a gating signal from the list-mode PET data. The generated signal can be used to reconstruct motion corrected PET images affected by respiratory motion. In addition, a single motion corrected volume can automatically be generated. Resulting motion corrected PET images can be used to aid clinicians in detection, localization, evaluation, diagnosis, staging, restaging, follow-up of diseases and disorders, radiotherapy planning, as well as their therapeutic planning, and therapeutic outcome assessment. Images of lesions in the thorax, abdomen and pelvis are mostly affected by respiratory motion. Deviceless PET Respiratory gating system may be used with PET radiopharmaceuticals, in patients of all ages, with a wide range of sizes, body habitus, and extent/type of disease. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
# 510(k) SUMMARY
K251370
1. SUBMITTER'S NAME:
Junichiro Araoka
Senior Manager, Quality Assurance Department
Canon Medical Systems Corporation
1385 Shimoishigami
Otawara-Shi, Tochigi-ken, Japan 324-8550
2. ESTABLISHMENT REGISTRATION:
9614698
3. OFFICIAL CORRESPONDENT/CONTACT PERSON:
Yoshiaki Cook
Sr. Manager, Regulatory Affairs
Canon Medical Systems USA, Inc
2441 Michelle Drive
Tustin, CA 92780
(657) 270-5595
4. DATE PREPARED:
May 2, 2025
5. TRADE NAME(S):
Cartesion Prime (PCD-1000A/3) V10.21
6. COMMON NAME:
System, Emission Computed Tomography
System, X-ray, Computed Tomography System
7. DEVICE CLASSIFICATION:
Classification Name: Emission Computed Tomography X-ray system
Regulation Number: 21 CFR §892.1200
Regulatory Class: Class II
Classification Name: Computed Tomography X-ray system
Regulation Number: 21 CFR §892.1750
Regulatory Class: Class II
2441 Michelle Drive, Tustin, CA 92780
PHONE: 800-421-1968
https://us.medical.canon
{5}
Canon
CANON MEDICAL SYSTEMS USA, INC.
Made For life
# 8. PRODUCT CODE:
90KPS
90JAK
# 9. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
# 10. PREDICATE DEVICE:
| Product | Marketed by | Regulation Number | Regulation Name | Product Code | 510(k) Number | Clearance Date |
| --- | --- | --- | --- | --- | --- | --- |
| Cartesion Prime (PCD-1000A/3) V10.15 | Canon Medical Systems USA | 21 CFR 892.1200 | Emission Computed Tomography System | KPS | K231748 | September 12, 2023 |
# 11. REASON FOR SUBMISSION:
Modification of a cleared device
# 12. DEVICE DESCRIPTION:
The Cartesian Prime (PCD-1000A/3) V10.21 combines a high-end CT and a high-throughput PET designed to acquire CT, PET and fusion images.
The high-end CT system is a multi-slice helical CT scanner with a gantry aperture of $780~\mathrm{mm}$ and a maximum scan field of view (FOV) of $700~\mathrm{mm}$ . The high-throughput PET system has a digital PET detector utilizing SiPM sensors with temporal resolution of $< 250$ ps (238 ps typical). Cartesian Prime (PCD-1000A/3) V10.21 is intended to acquire PET images of any desired region of the whole body and CT images of the same region (to be used for attenuation correction or image fusion), to detect the location of positron emitting radiopharmaceuticals in the body with the obtained images. This device is used to gather the metabolic and functional information from the distribution of radiopharmaceuticals in the body for the assessment of metabolic and physiologic functions. This information can assist research, detection, localization, evaluation, diagnosis, staging, restaging, follow-up of diseases and disorders, as well as their therapeutic planning, and therapeutic outcome assessment. This device can also function independently as a whole body multi-slice CT scanner.
The subject device incorporates the latest reconstruction technology, AiCE-i for PET (Advanced Intelligent Clear-IQ Engine- integrated), intended to improve image quality and reduce image noise for FDG whole body data by employing deep learning artificial neural network methods which can more fully explore the statistical properties of the signal and noise of PET data. The AiCE algorithm will be able to better differentiate signal from noise and can be applied to improve image quality and denoising of PET images compared to conventional PET imaging reconstruction.
2441 Michelle Drive, Tustin, CA 92780
PHONE: 800-421-1968
https://us.medical.canon
{6}
A Deviceless PET Respiratory gating system has been implemented for use with the subject device. With this subject device, respiration is extracted using a pre-trained neural network. Respiratory-gated reconstruction is performed at a speed equal to or faster than that with "Normal".
# 13. INDICATIONS FOR USE:
The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces cross-sectional images of the body by computer reconstruction of X-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images.
This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, localization, diagnosis, staging, restaging, follow-up, therapeutic planning and therapeutic outcome assessment of (but not limited to) oncological, cardiovascular, neurological diseases and disorders. Additionally, this device can be operated independently as a whole body multislice CT scanner.
AiCE-i for PET is intended to improve image quality and reduce image noise for FDG whole body data by employing deep learning artificial neural network methods which can explore the statistical properties of the signal and noise of PET data. The AiCE algorithm can be applied to improve image quality and denoising of PET images.
Deviceless PET Respiratory gating system, for use with Cartesian Prime PET-CT system, is intended to automatically generate a gating signal from the list-mode PET data. The generated signal can be used to reconstruct motion corrected PET images affected by respiratory motion. In addition, a single motion corrected volume can automatically be generated. Resulting motion corrected PET images can be used to aid clinicians in detection, localization, evaluation, diagnosis, staging, restaging, follow-up of diseases and disorders, radiotherapy planning, as well as their therapeutic planning, and therapeutic outcome assessment. Images of lesions in the thorax, abdomen and pelvis are mostly affected by respiratory motion. Deviceless PET Respiratory gating system may be used with PET radiopharmaceuticals, in patients of all ages, with a wide range of sizes, body habitus, and extent/type of disease.
# 14. SUBSTANTIAL EQUIVALENCE:
The Cartesian Prime (PCD-1000A/3) V10.21 is substantially equivalent to Cartesian Prime (PCD-1000A/3) V10.15, which received premarket clearance under K231748, and is currently marketed by Canon Medical Systems USA.
The subject and predicate devices are the same with the only differences being, expansion of the use of AiCE-i for PET with pediatric imaging and allowing for user-selectable intensity settings as opposed to simply selecting 'on/off', and implementation of a deep-learning artificial neural network based Deviceless PET Respiratory gating system that analyzes respiratory motion in order to reconstruct motion-corrected PET images.
{7}
A comparison of the technological characteristics between the subject and the predicate device is included below.
| | Subject Device | Predicate Device |
| --- | --- | --- |
| Device Name, Model Number | Cartesion Prime (PCD-1000A/3) V10.21 | Cartesion Prime (PCD-1000A/3) V10.15 |
| 510(k) Number | This submission | K231748 |
| Deviceless PET Respiratory gating system | Mode Setting: AI | Mode Setting: Normal |
| AiCE-i for PET | Pediatric Imaging: Available | Pediatric Imaging: Not Available |
| | Intensity Settings: Smooth, Standard, Sharp | Intensity Settings: ON/OFF |
# 15. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC81001-5-1, IEC62366-1, NEMA XR-25, NEMA XR-26, NEMA XR-29 and NEMA NU-2. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
# 16. TESTING
Risk analysis and verification/validation activities conducted through bench testing demonstrate that the established specifications for the device have been met.
# 16.a. AiCE-i for PET (AiCE2)
This feature employs a deep learning artificial neural network method to improve image quality and denoise PET images. This algorithm was developed to explore the statistical properties of signal and noise of input PET images and was trained to automatically adapt to different noise levels to produce consistently high-quality images. The improvement to the algorithm implemented in AiCE2 incorporated a different training methodology relative to the previous version, in order to achieve the development of Sharp, Standard and Smooth intensity settings, although the training data remained identical. From the original training dataset, a subset was assembled using FDG studies of sixteen (16) cancer patients (BMI: 19.5-34.5) over half of which were acquired from the U.S. and selected to cover a wide range of representative anatomy.
# Performance Testing - AiCE for PET Pediatric Use
A series of bench tests was conducted to support marketing claims associated with the use of AiCE for PET for pediatric imaging. This optional pediatric imaging function demonstrates equivalence to the currently cleared imaging methods with regards to: Contrast Recovery Coefficient (CRC), Background Variability (BGV) and Contrast to Noise Ratio (CNR), absence of artifacts, and the quantitativity $(\mathrm{SUV}_{\mathrm{mean}})$ for the intended pediatric use.
{8}
Page 5 of 7
# Performance Testing – AiCE for PET Image Intensity
A series of bench tests was conducted to support marketing claims associated with the tiered imaging intensity function (as opposed to the previously cleared "on/off" function). This method has been demonstrated to be substantially equivalent to the current method of image intensity, and, with respect to quantitative parameters such as accuracy of SUV (max and mean) and tumor volume, is improved over the current image intensity setting.
# Performance Testing - Substantial equivalence between AiCE2 and AiCE1
Phantom testing in accordance with NEMA NU-2, Section 7 was conducted in order to compare the Sharp, Standard and Smooth intensity settings available with AiCE2 to AiCE1, using a 10 mm sphere target to measure SUVmean, background variability (BGV), contrast recovery coefficient (CRC), and signal to noise ratio (SNR (with Std error)) and compare the values between both versions of AiCE-i for PET. The results across all indices demonstrated either equivalence or improvement by AiCE2.
Additional phantom studies were conducted using 1) the Adult and Pediatric NEMA phantoms to confirm that the $10\mathrm{mm}$ CRC, BGV, and CNR of the Sharp, Standard and Smooth intensity settings available with AiCE2 are substantially equivalent to AiCE1, 2) the Small Pool phantom in order to confirm that the quantified SUVmean is equivalent between both versions, and 3) the Pediatric NEMA phantom and Small Pool Phantom to confirm that there are no artifacts in the images reconstructed with either version of AiCE-i for PET. The results of these studies demonstrated equivalent performance between AiCE1 and AiCE2, with respect to the preservation of contrast and improving noise levels relative to conventional imaging methods.
# Performance Testing – Representative Clinical Images
Three (3) physicians having at least 20 years of experience in nuclear medicine evaluated the performance of Sharp, Smooth, and Standard intensity settings of AiCE2 using ten (10) PET DICOM clinical 18F-FDG whole body cases, compared to OSEM and Gaussian filter methods. All three physicians reported that AiCE2 images at all three intensity settings were of diagnostic quality and consistent across all 10 cases, determined to perform better with respect to overall image quality and image sharpness, as well as exhibit lower image noise compared to the predicate methods.
This validation dataset was entirely independent of the dataset used for training the algorithm during its development and was acquired from a total of 10 male and female patients and selected to cover characteristics common to the intended U.S. patient population by including a range of body mass indices, from 20.5 to 38.7, and by the inclusion of cases characterized by small lesions with lower level of FDG uptake.
# 16.b. Deviceless PET Respiratory gating system (DRG2)
This feature employs an algorithm which uses a neural network to extract motion information from acquired PET data and to generate a corresponding gating signal which can be used to reduce the effects of respiratory motion, thereby improving the image quality of reconstructed PET images. This neural network was trained on FDG studies of 27 cancer patients (BMI: 19.82-45.35, 3 instances unknown) acquired entirely from the U.S. and selected to be representative of both sexes as well as a wide range of scan characteristics.
{9}
Page 6 of 7
## Performance Testing – Deviceless PET Respiratory gating System Operational Mode
A series of bench tests was conducted to support marketing claims associated with the AI operating system of the Deviceless PET Respiratory gating system. This system has been demonstrated to be substantially equivalent to the current method of respiratory gating, using an external device, and, with respect to quantitative parameters such as accuracy of SUV (max and mean) and tumor volume, is improved over the current device-based gating method.
## Performance Testing – Substantial equivalence between DRG2 and DRG1
This evaluation used reconstructed clinical raw data acquired from 11 total male and female patients (BMI: 17.7-38.7), each with a representative lesion and covering a range of breathing rates. This dataset was entirely independent of the dataset used for training the algorithm. The primary objective of this evaluation was to demonstrate equivalency between deviceless PET respiratory gating with the new AI mode setting (DRG2) and predicate deviceless PET respiratory gating functionality (DRG1), by comparison of the quantified outputs on high uptake regions (e.g., lesions) of the reconstructed datasets. By satisfying all prespecified criteria, it was demonstrated that DRG2 performs with substantial equivalence to DRG1.
## Performance Testing – Representative Clinical Images
Three (3) physicians having at least 20 years of experience in nuclear medicine evaluated the respiratory gated images of ten (10) patients using Deviceless PET Respiratory gating system, non-gated images and those using device-base gating. All three (3) physicians determined that all images were of diagnostic quality and deviceless gated images demonstrated similar or better performance as device-based gated images and resulted in better motion correction compared to non-gated images.
This validation dataset was entirely independent of the dataset used for training the algorithm during its development and consisted of 10 total male and female patients, selected to cover characteristics common to the intended U.S. patient population, including a range of body mass indices, from 18.9 to 38.7, and cases including representative tumors and lesions.
Software Documentation for a Basic Documentation Level, per the FDA guidance document, “Content of Premarket Submissions for Device Software Functions” issued on June 14, 2023, is included in this submission. This documentation includes justification for the Basic Documentation Level determination as well as testing which demonstrates that the verification and validation requirements have been met.
Cybersecurity documentation, per the FDA guidance document “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”, issued on September 27, 2023, was included in this submission.
## 17. CONCLUSION
The Cartesian Prime (PCD-1000A/3) V10.21 performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this
{10}
submission it is concluded that the subject device has demonstrated substantial equivalence to the predicate device and is as safe and effective for its intended use.
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.