The VITAL Tubing Sets for Hemodialysis are sterile, single-use arterial and venous blood lines for exclusive use with AK98 (model BLU014E, BLU015E, BLU021E, BLU022E, BLU025E), Fresenius 2008 series (model BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU016E, BLU017E, BLU018E, BLU020E, BLU023E, BLU024E, BLU029E, BLU030E, BLU063E), Fresenius 5008X (model BLU026E, BLU027E) or Bbraun Dialog series Hemodialysis system (model BLU019E, BLU028E). The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the hemodialysis machine which drives the flow of blood through the circuit.
Device Story
VITAL Tubing Sets are sterile, single-use arterial and venous blood lines; designed for extracorporeal blood circulation during hemodialysis. Input: patient blood flow driven by hemodialysis machine pump rotor mechanism. Output: transport of blood through hemodialyzer and return to patient. Used in clinical hemodialysis settings; operated by trained healthcare professionals. Device consists of tubing, drip chambers, connectors, and filters. Benefits: provides necessary extracorporeal circuit for blood purification. Clinical decision-making relies on standard hemodialysis protocols; device facilitates the procedure by maintaining circuit integrity and flow.
Clinical Evidence
No clinical data. Bench testing only, including performance testing, packaging validation, hemodialysis machine compatibility, biocompatibility, and chemical analysis per ISO 10993 series.
Technological Characteristics
Materials: PVC, PP, PE, ABS, PC, Silicone rubber, PCTG, PETG, MABS. Components: Arterial/venous lines, drip chamber, connectors, clamps, filters. Sterilization: Ethylene Oxide (SAL 10⁻⁶). Standards: ISO 8637-2:2024, ISO 80369-7:2021, ISO 11607-1/2, ISO 10993 series. Connectivity: None (mechanical interface).
Indications for Use
Indicated for patients undergoing hemodialysis treatment requiring an extracorporeal blood circuit to transport blood from the patient through a hemodialyzer and back to the patient.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 16, 2026
Vital Healthcare Sdn. Bhd.
Ch'ng Hui Zhen
Regulatory Affair
PT 83718, Jalan Bestari 1A/KU7
Taman Perindustrian Kapar Bestari
Kapar, Selangor 42200
Malaysia
Re: K251341
Trade/Device Name: VITAL Tubing Sets for Hemodialysis – Models Models: BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU014E, BLU015E, BLU016E, BLU017E, BLU018E, BLU019E, BLU020E, BLU021E, BLU022E, BLU023E, BLU024E, BLU025E, BLU026E, BLU027E, BLU028E, BLU029E, BLU030E, BLU063E
Regulation Number: 21 CFR 876.5820
Regulation Name: Hemodialysis system and accessories
Regulatory Class: Class II
Product Code: FJK
Dated: December 19, 2025
Received: December 19, 2025
Dear Ch'ng Hui Zhen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251341 - Ch'ng Hui Zhen
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K251341 - Ch'ng Hui Zhen
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Gema Gonzalez -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251341 | |
| Device Name VITAL Tubing Sets for Hemodialysis | |
| Indications for Use (Describe) The VITAL Tubing Sets for Hemodialysis are sterile, single-use arterial and venous blood lines for exclusive use with AK98 (model BLU014E, BLU015E, BLU021E, BLU022E, BLU025E), Fresenius 2008 series (model BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU016E, BLU017E, BLU018E, BLU020E, BLU023E, BLU024E, BLU029E, BLU030E, BLU063E), Fresenius 5008X (model BLU026E, BLU027E) or Bbraun Dialog series Hemodialysis system (model BLU019E, BLU028E). The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the hemodialysis machine which drives the flow of blood through the circuit. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251341
Page 1 of 8
Page 1 of 8
Tab #29 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K251341
1. Date of Preparation: 16 Jan 2026
2. Sponsor Identification
Vital Healthcare Sdn. Bhd.
PT 83718, Jalan Bestari 1A/KU7, Taman Perindustrian Kapar Bestari, 42200 Kapar, Selangor, Malaysia.
Establishment Registration Number: Not Known
Contact Person: Ch'ng Hui Zhen
Position: Regulatory Affair
Tel: +603 33962329
Fax: +603 33962339
Email: chnghuizhen@vital.com.my
3. Designated Submission Correspondent
Vital Healthcare Sdn. Bhd.
PT 83718, Jalan Bestari 1A/KU7, Taman Perindustrian Kapar Bestari, 42200 Kapar, Selangor, Malaysia.
Ms. Ch'ng Hui Zhen (Primary Contact Person)
Tel: +603 33962329
Fax: +603 33962339
Email: chnghuizhen@vital.com.my
Ms. Gan Pei Hong (Alternative Contact Person)
Tel: +603 33962329
Fax: +603 33962339
Email: ganpeihong@vital.com.my
{5}
K251341
Page 2 of 8
Page 2 of 8
## 4. Identification of Proposed Device
**Trade Name:** VITAL Tubing Sets for Hemodialysis
**Common Name:** Haemodialysis tubing set
**Models:** BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU014E, BLU015E, BLU016E, BLU017E, BLU018E, BLU019E, BLU020E, BLU021E, BLU022E, BLU023E, BLU024E, BLU025E, BLU026E, BLU027E, BLU028E, BLU029E, BLU030E, BLU063E
## Regulatory Information
**Classification Name:** Set, Tubing, Blood, With and Without Anti-Regurgitation Valve
**Classification:** II
**Product Code:** FJK
**Regulation Number:** CFR876.5820
**Review Panel:** Gastroenterology/Urology
## Indications for use:
The VITAL Tubing Sets for Hemodialysis are sterile, single-use arterial and venous blood lines for exclusive use with AK98 (model BLU014E, BLU015E, BLU021E, BLU022E, BLU025E), Fresenius 2008 series (model BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU016E, BLU017E, BLU018E, BLU020E, BLU023E, BLU024E, BLU029E, BLU030E, BLU063E), Fresenius 5008X (model BLU026E, BLU027E) or Bbraun Dialog series Hemodialysis system (model BLU019E, BLU028E). The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the hemodialysis machine which drives the flow of blood through the circuit.
## Device Description
The VITAL Tubing Sets for Hemodialysis consists of an Arterial line and a Venous line. The tubing is soft, transparent, smooth and non-kink to ensure the good liquidity.
The proposed models have the same principles, intended use and manufacturing process, but have some difference in the product configuration and dimension. The typical/representative model is BLU063E among all the models due to it has the most complex structure for Ethylene Oxide sterilization which covers all the components of the other models and the distance from the closed end and ventilated end is the farthest.
The proposed devices are provided in sterile condition, it is subject to ethylene oxide sterilization prior
{6}
K251341
Page 3 of 8
to achieve a Sterility Assurance Level (SAL) of 10⁻⁶.
## 5. Identification of Predicate Device
510(k) Number: K201866
Product Name: NovaLine Tubing Sets for Hemodialysis
Manufacturer: Bain Medical Equipment (Guangzhou) Co., Ltd
## 6. Non-Clinical Test Conclusion
Non clinical tests, such as performance testing, packaging testing, side by side testing, hemodialysis machine compatibility testing, biocompatibility & chemical analysis were conducted to verify that the proposed device meet all design specifications and is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 11607-1:2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ISO 11607-2:2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Device
- ISO 8637-2:2024 Extracorporeal systems for blood purification - Part 2: Extracorporeal blood and fluid circuits for haemodialyzer, haemodiafilters, haemofilter and haemoconcentrators.
- ISO 80369-7:2021 Small bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- ASTM D4169-23 Standard Practice For Performance Testing Of Shipping Containers And Systems.
- ASTM F2096-11 (2019) Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- USP <85> Bacterial Endotoxin Test
- EN ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for in vitro Cytotoxicity
- EN ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation
- EN ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization
- EN ISO 10993-11:2017 Biological Evaluation of Medical Devices - Part 11: Tests for Systemic Toxicity
{7}
K251341
Page 4 of 8
- EN ISO 10993-4:2017 Biological Evaluation of Medical Devices - Part 4: Selection of Tests for Interactions with Blood
- EN ISO 10993-6:2016 Biological Evaluation of Medical Devices - Part 6: Tests for local effects after implantation
- EN ISO 10993-7:2008 Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals
- EN ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process
- EN ISO 10993-17:2023 Biological evaluation of medical devices - Part 17 Toxicological risk assessment of medical device constituents
- EN ISO 10993-18:2020 Biological evaluation of medical devices - Part 18 Chemical characterization of medical device materials within a risk management process
## 7. Clinical Test Conclusion
No clinical study is included in this submission.
## 8. Substantially Equivalent (SE) Comparison
The technological differences between the proposed and predicated devices are stated as below.
| | Proposed Device | Predicated Device | Remark |
| --- | --- | --- | --- |
| Device Name | VITAL Tubing Sets for Hemodialysis | NovaLine Tubing Sets for Hemodialysis | SE |
| Indication for Use | The VITAL Tubing Sets for Hemodialysis are sterile, single-use arterial and venous blood lines for exclusive use with AK98 (model BLU014E, BLU015E, BLU021E, BLU022E, BLU025E), Fresenius 2008 series (model BLU001E, BLU002E, BLU003E, BLU004E, BLU005E, BLU006E, BLU007E, BLU008E, BLU009E, BLU010E, BLU011E, BLU012E, BLU013E, BLU016E, BLU017E, BLU018E, BLU020E, | The NovaLine Tubing Sets for Hemodialysis - Models BL 11 and BL 12 - are sterile, single-use arterial and venous blood lines for exclusive use with the Baxter Healthcare AK98 Hemodialysis System. The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the | Discussion 1 |
{8}
K251341
Page 5 of 8
| | BLU023E, BLU024E, BLU029E, BLU030E, BLU063E), Fresenius 5008X (model BLU026E, BLU027E) or Bbraun Dialog series Hemodialysis system (model BLU019E, BLU028E). The blood lines serve as the extracorporeal blood circuit in patients undergoing hemodialysis treatment, by which blood is transported from the patient through a hemodialyzer and back to the patient. The pump line interfaces with a pump rotor mechanism of the hemodialysis machine which drives the flow of blood through the circuit. | hemodialysis machine which drives the flow of blood through the circuit. | |
| --- | --- | --- | --- |
| Feature | Pre-Pump Post-Pump Color Coded component Sterile Non-pyrogenic Single Use Prescription Device | Pre-Pump Post-Pump Color Coded component Sterile Non-pyrogenic Single Use Prescription Device | SE |
| Main Configuration | Arterial & Venous Line Drip Chamber Main Tubing Branch Tubing Female Luer Lock Clamps Filters | Arterial & Venous Line Drip Chamber Main Tubing Branch Tubing Female Luer Lock Clamps Filters | SE |
| Physical Performance | The priming volume (mL/mm) of the proposed devices are as below: BLU063E is 201±10% BLU001E is 130±10% BLU002E is 128±10% BLU003E is 128±10% BLU004E is 144±10% BLU005E is 142±10% | The priming volume (mL/mm) of BL 11 & BL 12 are 127±10% & 186±10% respectively. | Discussion 2 |
{9}
K251341
Page 6 of 8
| | BLU006E is 156±10% | | |
| --- | --- | --- | --- |
| BLU007E is 155±10% | | |
| BLU008E is 144±10% | | |
| BLU009E is 142±10% | | |
| BLU010E is 161±10% | | |
| BLU011E is 153±10% | | |
| BLU012E is 156±10% | | |
| BLU013E is 153±10% | | |
| BLU014E is 129±10% | | |
| BLU015E is 187±10% | | |
| | BLU016E is 141±10% | | |
| BLU017E is 122±10% | | |
| BLU018E is 132±10% | | |
| BLU019E is 104±10% | | |
| BLU020E is 93±10% | | |
| BLU021E is 127±10% | | |
| BLU022E is 186±10% | | |
| BLU023E is 152±10% | | |
| BLU024E is 124±10% | | |
| BLU025E is 88±10% | | |
| | Positive pressure | Positive pressure Limitation (mmHg) is 500. | |
| Limitation (mmHg) for all proposed devices is 500. | | |
| Negative Pressure | Negative Pressure Limitation (mmHg) is -500. | |
| Limitation (mmHg) for all proposed devices is -500. | | |
| Performance | Conforms to ISO8637-2:2024 ISO80369-7:2021 | Conforms to ISO8638:2010 ISO594-2:1998 | Discussion 3 |
| Materials | It consists of various materials, such as PVC, PP, PE, ABS, PC, Silicone rubber, PCTG, PETG, MABS. | It consists of various materials, such as PP, PVC, Silicone rubber, ABS, PE. | Discussion 4 |
| Biocompatibility | Chemical Analysis Conforms to ISO10993 series standards | Conforms to ISO10993 series standards | Discussion 5 |
| Sterilization | By Ethylene Oxide, SAL (10^{-6}) | By E-Beam Irradiation, SAL (10^{-6}) | Discussion 6 |
{10}
Page 7 of 8
K251341
Page 7 of 8
| Labelling | Direction for Use | Direction for Use | SE |
| --- | --- | --- | --- |
| | Intended Use | Intended Use | |
| | Description | Description | |
| | Warnings and Cautions | Warnings and Cautions | |
## Discussion 1 – Indications for Use
The indication for use of the proposed and predicate device have clearly stated the compatible system to avoid any misuse. The compatibility between the proposed device and the hemodialysis system is included in this submission. Both of them are still used for the same purpose, which is to connect with a hemodialysis system and provide extracorporeal access to the patient's blood during hemodialysis. This difference will not result in any safety and effectiveness issue of the proposed device.
## Discussion 2 – Physical Performance
The priming volume of the proposed and predicate device is different due to their length of main tubes is different. Differences in the tubing length and priming volume can impact the mechanical and performance characteristics of device. The bench tests such as pressure leak test, endurance testing, priming volume test, tensile strength test, tube kinking inspection, blood pump segment validation were conducted to prove that the proposed device meet the requirements of mechanical and performance characteristics. Therefore, the differences in priming volume will not lead to new safety and effectiveness problems.
## Discussion 3 – Performance
The ISO standard of the predicate device, ISO8638-2010 has been withdrawn and replaced with ISO8637-2, whereas ISO594-2 has also been replaced with ISO80369-7. Updated ISO version will not lead to new safety and effectiveness issue in the proposed device.
## Discussion 4 – Material
The material of the proposed device is PVC, PP, PE, ABS, PC, Silicone rubber, PCTG, PETG, MABS. These materials of proposed device have many years of clinical application history and are non-innovative materials. We have conducted chemical analysis and biocompatibility tests according to the requirements of ISO10993 series, and the results were all passed. Therefore, we think the differences on the materials will not cause new problems on the safety and effectiveness.
## Discussion 5 – Biocompatibility
Chemical analysis according to ISO 10993-17 and ISO 10993-18 that supports the biocompatibility of the proposed device is included in this submission. The biocompatibility studies that carried out in the proposed device are also complied with ISO 10993 series standards. Therefore, this item is considered substantially equivalent.
## Discussion 6 – Sterilization Method
The sterilization method of the proposed device is Ethylene Oxide sterilization. The sterilization method of the predicate device is E-Beam sterilization. Both of the sterilization assurance level can reach $10^{-6}$. This difference will not result in any safety and effectiveness issue of the proposed device.
{11}
Page 8 of 8 K251341
Page 8 of 8
Conclusion:
According to the analysis in Substantially Equivalent Discussion, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices. Their differences will not result in new safety and effectiveness issue in the proposed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.