XERF

K251327 · Lutronic Corporation · GEI · Aug 11, 2025 · General, Plastic Surgery

Device Facts

Record IDK251327
Device NameXERF
ApplicantLutronic Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateAug 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

XERF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

Device Story

XERF is a monopolar electrosurgical unit for dermatologic and general surgical procedures. It delivers high-frequency electrical currents (2 MHz and 6.78 MHz) non-invasively to skin tissue via one of 12 interchangeable handpiece electrode tips. A neutral return pad is attached to the patient's back to disperse energy. The operator adjusts treatment parameters via an LCD touch screen and activates energy delivery using a foot or finger switch. The device heats tissue to induce coagulation. It is intended for use by clinicians in a clinical setting. The system provides controlled thermal effects for tissue coagulation, supporting clinical goals like tissue repair and dermal remodeling.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and biocompatibility (ISO 10993). Performance was validated via ex-vivo thermal studies on kidney, liver, and skin tissue, demonstrating uniform coagulation. A comparative ex-vivo thermal study against the Thermage FLX predicate confirmed equivalent time-dependent heating and cooling profiles.

Technological Characteristics

Monopolar RF electrosurgical unit; frequencies 2 MHz and 6.78 MHz; max output 400W (6.78 MHz) / 300W (2 MHz). Includes 12 interchangeable electrode tips and neutral return pad. Class I, Type BF applied part. 10.1" LCD touch interface. Operates on 100-120V AC. Biocompatibility per ISO 10993-1, -10, -23. Software lifecycle per IEC 62304.

Indications for Use

Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 11, 2025 Lutronic Corporation Cynthia Aguirre Sr. Regulatory Affairs Specialist Lutronic Center 219, Sowon- Ro Deogyang-Gu, Goyang 410220 Korea, South Re: K251327 Trade/Device Name: Xerf Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: July 11, 2025 Received: July 11, 2025 Dear Cynthia Aguirre: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251327 - Cynthia Aguirre Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251327 - Cynthia Aguirre Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JAMES H. Digitally signed by JAMES H. JANG -S JANG -S Date: 2025.08.11 00:51:07 -04'00' James Jang, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} XERF Page 9 of 53 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251327 | ? | | Please provide the device trade name(s). | | ? | | XERF | | | | Please provide your Indications for Use below. | | ? | | The XERF is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K251327 # 510(k) SUMMARY A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92. | 807.92(a)(1) - Submitter Information | | | --- | --- | | Name | Lutronic Corporation | | Address | Lutronic Center 219, Sowon-Ro Deogyang-Gu, Goyang-Si, KR 410220 | | Phone number | 888-588-7644 | | Fax number | N/A | | Establishment Registration Number | 3004483538 | | Name of contact person | Cynthia Aguirre Senior Regulatory Affairs Specialist 3/5 Carlisle Road Westford, MA, 01886 | | Date prepared | 05 August 2025 | | 807.92(a)(2) - Name of device | | | Trade or proprietary name | XERF | | Common or usual name | Electrosurgical Unit and Accessories | | Classification name | Electrosurgical, Cutting & Coagulation & Accessories | | Classification panel | General and Plastic Surgery | | Regulation | 21 CFR 878.4400 | | Product Code(s) | GEI | | 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | | | | Solta Medical K170758 (09/22/2017) Predicate Cutera Inc K221407 (07/13/2022) Reference | | 807.92(a)(4) - Device description | | | | XERF is an electrosurgical unit which uses high frequency electrical currents for treatment. When high frequency is irradiated on the skin, the treated skin tissue is heated, resulting in coagulation. This equipment generates high frequency of 2 MHz and 6.78 MHz and is delivered non-invasively into skin through the tip of the handpiece. There are a total of 12 types of active electrode tips compatible with the handpiece; the different tips are intended for various applications and treatment zones. The system also includes a return pad corresponding to the neutral electrode which is intended to be attached to the patient's back to disperse the generated energy. For adjustment of treatment parameter, data can be used by adjusting the operator interface through LCD touch screen. | {5} K251327 | | Treatment is conducted using foot switch or finger switch after ready when applied treatment parameter is output to the tissue. | | --- | --- | | 807.92(a)(5) Intended use of the device | | | Indications for use | XERF is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. | 2 {6} K251327 807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate | | Company | Lutronic Corporation | Solta Medical | Cutera Inc. | | --- | --- | --- | --- | --- | | | Model | XERF | Thermage FLX | truSculpt iD | | | K No. | TBD | K170758 | K221407 | | | Type | Monopolar RF | Monopolar RF | Bipolar/ Monopolar RF | | | Frequency [MHz] | 2 MHz, 6.78MHz | 6.78MHz | 1 MHz and 2 MHz | | | Max Output Power [W] | 6.78MHz: 400W 2MHz: 300W | 400W | 300 W | | | Duty Cycle | Intermittent | Intermittent | Unknown | | | Interval | off, 200, 500, 800ms (applies to handpiece mode only) | Unknown | Unknown | | | Energy | -2 MHz: 1 ~ 120 J -6.78 MHz: 1~250 J | 6.78 MHz 15J ~ 600J | Unknown | | | Operational Conditions | Temperature range: 10°C - 30°C Humidity: Up to 90% RH | Temperature range: 10°C - 30°C Humidity: Up to 90% RH | Temperature range: 5°C - 30°C Humidity: Up to 90% RH | | | Tip Pattern Sizes (Treatment Width) | 60 Tip: 27.6 x 17.6mm 40 Tip: 17.6 x 17.6mm 10 Tip: 9 x 9 mm 05 Tip: 9 X 4.5 mm | 0.5 x 0.5cm = 0.25cm² 1.73 x 1.73cm = 3cm² 2.0 x 2.0cm = 4cm² 4.0 x 4.0cm = 16cm² | 40 to 240 cm² 16 and 40 cm² | | | Patient Contact Material | Tip electrode Return pad | Treatment tips Coupling fluid Return pad | Polyvinylidene fluoride (PVDF) for handheld handpieces • Liquid resin adhesive for patient decals (for hands-free handpieces) | | | Contact duration | Less than 24 hours | Less than 24 hours | Less than 24 hours | | | Contact type | Intact skin | Intact skin | Intact skin | | | Single Use/Reusable | Single use accessories | Single use, provided clean | Single use accessories | | Dimension (Including a part holder) | 389mm (W) x 856.4mm (L) x 1604.1mm (H) | 49.5cm (W) x 43cm (L) x 125.7cm (H) | 14"W x 17.5"L x 41.5"H | | | Weight | 47 kg | 45 kg | 39kg | | {7} K251327 | | Type & Degree of Protection Against Shock | Class 1 / type BF applied part | Class 1 / type BF applied part | Class 1 / type BF applied part | | | --- | --- | --- | --- | --- | --- | | | Electrical Ratings | Single phase AC 100-120V, 50/60Hz | AC 100~240V, 50-60Hz, 10A | AC 100-240V. 50/60Hz Power: 2000VA | | | | | Power consumption:1kVA (Fuse: 125V / 10A) | | | | | | User Interface | 10.1” LCD Touch Display | LCD Touch Display | LCD Touch Display | | | | Treatment Activation | Finger switch or Footswitch | Manual or Footswitch | Control Screen Button or Finger switch | | | 807.92(b)(1) NON-CLINICAL TESTS SUBMITTED | | | | | | | Electrical Safety IEC 60601-1:2005+ A 1:2012+ A2:2020 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance, FDA Recognition# 19-49 IEC 60601-2-2:2017 + AMD1: 2013 Medical electrical equipment-Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, FDA Recognition# 6-389 EMC IEC 60601-1-2: 2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests, FDA Recognition# 19-36 IEC TS 60601-4-2:2024 Medical electrical equipment-Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems, FDA Recognition# 19-50 Usability IEC 60601-1-6: 201 0+A1:2013+A2:2020 -Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, FDA Recognition# 5-132 Biocompatibility ISO 10993-1:2018 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity, FDA Recognition# 2-258 ISO 10993-10:2021 Biological evaluation of medical devices -Part 10: Tests for skin sensitization, FDA Recognition# 2-296 ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation, FDA Recognition# 2-291 Software IEC 62304:2006+A1:2015 Medical Device-Software Life Cycle Processes, FDA Recognition# 13-79 Performance An ex-vivo thermal study was performed to assess the effect of the RF power on the tissue, healing of the tissue and potential side effects. Three different types of tissue were tested at a range of treatment settings as recommended by the FDA, ex vivo kidney, liver, and skin. Central sections from the fixed 12 mm biopsy samples were obtained, stained by standard H&E histology method, and analyzed under light microscope. Coagulation was observed when treatment was performed with five stacked pulses with the interval of 200ms (within the same treatment spot without moving the tip). Coagulation surface correlated to the perimeter of the RF tip treatment surface. Uniform coagulation at entire perimeter and depth was observed | | | | | | {8} K251327 | in tested tissues. The correlations noted in the in vivo and ex vivo test data, as well as the uniformly coagulated tissue observed, confirms that treatment by the XERF device ,at the appropriate testing settings, will possess a desirable clinical treatment effect, such as tissue repair responses and establishing long-term dermal remodeling. A second ex-vivo thermal study was conducted comparing temperature-time curves of Thermage FLX (predicate) and XERF (subject device). Ex-vivo dermal temperature profiles were measured using fiber optic temperature sensors inserted at a depth of 3.0 mm of the skin layer under standardized treatment conditions. Temperature- time curves were measured at two levels representing worst-case scenario treatment condition and real-world treatment condition. The ex-vivo dermal temperature testing confirmed that the XERF device’s thermal performance is equivalent to that of the Thermage FLX device in terms of time-dependent heating and cooling behavior. | | --- | | **807.92(b)(2) CLINICAL TESTS SUBMITTED** | | No clinical tests were included as part of this submission. | | **807.92(b)(3) Conclusion** | | Based on the comparison of the technological characteristics, the specifications for the XERF are the same or a subset of the predicate/reference device specifications. The indication is the same as the predicate device, Thermage FLX. | 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...