K251325 · Baylis Medical Company, Inc. · DRE · May 29, 2025 · Cardiovascular
Device Facts
Record ID
K251325
Device Name
VersaCross Connect™ Transseptal Dilator
Applicant
Baylis Medical Company, Inc.
Product Code
DRE · Cardiovascular
Decision Date
May 29, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VersaCross Connect Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired.
Device Story
Single-use transseptal dilator; provides torque control and flexibility for catheterization and angiography of heart chambers. Features tapered tip, echogenic shaft, and radiopaque tip for visualization under fluoroscopy and echocardiography. Used with compatible access sheaths (e.g., Agilis NxT) to guide transseptal wires to the atrial septum for puncture; dilates atrial septal defect to enable passage of larger devices. Operated by electrophysiologists or interventional cardiologists in catheter labs. Output is mechanical dilation of tissue; assists clinician in achieving left atrial access for interventional procedures.
Clinical Evidence
Bench testing only. No clinical data presented. Verification activities included torque transmission, shaft outer diameter, snap force, accessory compatibility, guidewire compatibility, shapeability, corrosion, luer tests, torque withstand, flexural rigidity, curve retention, hub-shaft tensile, tip-to-tip cap cantilever/tensile, clamshell tensile/cantilever, and air/liquid leakage. Biocompatibility, sterilization (EtO), pyrogenicity (LAL), and packaging integrity were verified per ISO standards.
Technological Characteristics
Single-use, sterile, vessel dilator. Materials: patient-contacting materials identical to predicate. Features: tapered tip, echogenic shaft, radiopaque tip. Standards: ISO 11070:2014/Amd.1:2018, ISO 80369-7:2021, ISO 80369-20:2015, ISO 10993-1:2018, ISO 11135:2014/A1:2018, ISO 11607-1:2019. Sterilization: Ethylene Oxide (SAL 10^-6). Mechanical operation via manual torque and force transmission.
Indications for Use
Indicated for patients requiring left atrial access via transseptal technique. Used by electrophysiologists, interventional cardiologists, and trained clinicians in catheter labs.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2025
Baylis Medical Company, Inc.
Ezgi Tas
Regulatory Affairs Specialist
5825 Explorer Drive
Mississauga, ON L4W5P6
Canada
Re: K251325
Trade/Device Name: VersaCross Connect™ Transseptal Dilator
Regulation Number: 21 CFR 870.1310
Regulation Name: Vessel dilator for percutaneous catheterization
Regulatory Class: Class II
Product Code: DRE
Dated: April 28, 2025
Received: April 29, 2025
Dear Ezgi Tas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251325 - Ezgi Tas
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251325 - Ezgi Tas
Page 3
Sincerely,
Katherine N. Trivedi -S
Katherine Trivedi
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251325 | |
| Device Name VersaCross Connect Transseptal Dilator | |
| Indications for Use (Describe) The VersaCross Connect Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
VersaCross Connect™ Transseptal Dilator
Baylis Medical Company Inc.
# 13. 510(K) SUMMARY
## Submitter Information
A. Company Name: Baylis Medical Company Inc.
B. Company Address: 5825 Explorer Drive
Mississauga, Ontario L4W 5P6
Canada
C. Company Phone: +1 (905) 602-4875
D. Contact Person: Ezgi Tas
Regulatory Affairs Specialist, III
E. Date Summary Prepared: 28-April-2025
## Device Identification
A. Device Trade Name: VersaCross Connect™ Transseptal Dilator
B. Device Common Name: Dilator
C. Classification Name: CFR 870.1310 – Vessel dilator for percutaneous catheterization
D. Product Code: DRE
E. Device Class: Class II
Special 510(k)
013-Page 1
{5}
VersaCross Connect™ Transseptal Dilator
Baylis Medical Company Inc.
# Identification of Legally Marketed Device
Table 13.1: Predicate Device
| Predicate Device | Manufacturer | 510(k) | Indications for Use |
| --- | --- | --- | --- |
| VersaCross Connect™ Transseptal Dilator | Baylis Medical Company Inc. | K241720 | The VersaCross Connect™ Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. |
# Intended Use/ Indications for Use
The VersaCross Connect™ Transseptal Dilator is indicated for use in procedures where access to the left atrium via the transseptal technique is desired.
# Device Description
The subject VersaCross Connect™ Transseptal Dilator (VXA-AG) is a single-use device that is supplied sterile to the user.
The subject VersaCross Connect™ Transseptal Dilator (VXA-AG) represents modifications to the legally marketed VersaCross Connect™ Transseptal Dilator (K241720).
The VersaCross Connect™ Transseptal Dilator (VXA-AG) is designed for safe and easy catheterization and angiography of specific heart chambers and locations. The dilator provides torque control and is flexible. The dilator features a tapered tip and a shaft that can be reshaped manually. The echogenic shaft and tip and radiopaque tip maximize visualization of the dilator during manipulation in procedures.
The dilator can be used with separately cleared compatible access sheaths such as Agilis™ NxT Steerable Introducer (K061363 or K081645). The dilator provides support and helps guide separately cleared compatible transseptal wires to the atrial septum for puncture. The dilator subsequently dilates the atrial septal defect to enable larger diameter devices to cross the septum. The subject device can be used by electrophysiologists, interventional cardiologists, and other users trained in catheter techniques for any procedure that requires left atrial access via transseptal puncture. Procedures using the devices are performed in fully equipped catheter labs with imaging equipment, including fluoroscopy and echocardiography under sterile technique.
Special 510(k)
013-Page 2
{6}
VersaCross Connect™ Transseptal Dilator
Baylis Medical Company Inc.
# Comparison of Characteristics with Predicate Device
The intended use and indications for use of the proposed VersaCross Connect™ Transseptal Dilator (VXA-AG) remains unchanged from the predicate VersaCross Connect™ Transseptal Dilator (K241720). The subject and predicate devices also share the same fundamental scientific technology, including principles of operation and mechanism of action, environment of use, packaging configuration and sterilization method. The changes between the predicate and subject devices include design, non-patient contacting material, and associated labeling changes.
Differences in technological characteristics between the subject and predicate devices do not raise new or different questions of safety and effectiveness (Table 13.2). The results of verification and validation testing support substantial equivalence of the proposed VersaCross Connect™ Transseptal Dilator with the predicate device.
Table 13.2: Comparison of Subject and Predicate Devices
| Characteristic | Subject Device Compared to Predicate VersaCross Connect™ Transseptal Dilator (K241720) | Comment |
| --- | --- | --- |
| Intended Use/Indications for Use | Identical | Both subject and predicate device are indicated for use in procedures where access to the left atrium via the transseptal technique is desired. |
| Fundamental scientific technology | Identical | Both subject and predicate device rely on transfer of mechanical force to achieve tissue dilation. |
| Operating principles | Identical | The operation principle for both subject and predicate device is the transfer of mechanical force to achieve tissue dilation. |
| Mechanism of action | Identical | Both subject and predicate device dilate tissue to achieve its intended use. |
| Environment of Use | Identical | Both the predicate and subject device are used in facilities equipped with staff to perform diagnostic and interventional procedures. |
| Material | Patient contacting: Identical | Both the predicate and subject device contain the same patient contacting materials. |
| | Non-patient contacting: Similar | Both the predicate and subject device contain similar non-patient contacting materials. |
Special 510(k)
013-Page 3
{7}
VersaCross Connect™ Transseptal Dilator
Baylis Medical Company Inc.
| | | The subject device differs from the predicate device as follows:
• Device hub color change |
| --- | --- | --- |
| Technological characteristics (dimensions, design) | Similar | Both the predicate and subject device share the same fundamental design. The subject device differs from the predicate device as follows:
• Dilator shaft dimensional changes
• Snap fit hub geometry and hub length |
| Labelling | Similar | The subject device labelling has modifications to reflect the associated changes in design:
• Updates to identify the compatible access sheaths, model numbers and device specifications |
| Packaging configuration | Identical | The predicate and subject device contain the same packaging configuration (pouch, die cut card, shelf box and shipper box). |
| Sterilization method | Identical | Both subject and predicate device are Single Use, Ethylene Oxide sterilized |
## Summary of Non-Clinical Performance Verification and Validation
Applicable non-clinical performance verification and validation were completed to demonstrate safety and effectiveness and substantial equivalence of the subject device to the predicate device. All design requirements were met as per applicable standards and test protocols. The following verification and validation activities were completed to demonstrate the safety and effectiveness of the subject device:
## Design Requirements Verification
Applicable design verification was conducted for the subject device to ensure compliance with the applicable requirements of ISO 11070:2014/Amd.1:2018, ISO 80369-7:2021, ISO 80369-20:2015 and Baylis Medical Company Inc. to verify the functional, performance, product interface and compatibility design requirements for the device. The following design requirements have been verified:
- Torque Transmission
- Dilator Protrusion Length
- Aggregate Curve Reach
- Shaft Outer Diameter
- Snap Force
- Accessory Sheath Compatibility
Special 510(k)
013-Page 4
{8}
VersaCross Connect™ Transseptal Dilator
Baylis Medical Company Inc.
- Guidewire Compatibility
- Shapeability
- Corrosion
- Luer Tests
- Torque Withstand
- Flexural Rigidity
- Curve Retention
- Hub-Shaft Tensile
- Tip to Tip Cap Cantilever and Tensile
- Clamshell Tensile
- Clamshell Cantilever
- Air and Liquid Leakage
## Biocompatibility Verification
Biological safety was evaluated for the subject device to verify compliance with the current applicable requirements of ISO 10993-1:2018 and the September 8, 2023 FDA guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Biocompatibility verification of the subject VersaCross Connect™ Transseptal Dilator was conducted through adopted biological data of the predicate device VersaCross Connect™ Transseptal Dilator (K241720).
## Sterilization Verification
Sterilization and residual limit verification were evaluated for the subject device to verify compliance with the current applicable requirements of ISO 11135:2014/A1:2018 and ISO 10993-7:2008/A1:2019. Sterilization was performed with Ethylene Oxide to a Sterility Assurance Level (SAL) of $10^{-6}$.
## Pyrogen Testing
The subject device is supplied non-pyrogenic. Limulus Amoebocyte Lysate (LAL) testing was evaluated using the Kinetic Chromogenic method, as per ANSI/AAMI ST72:2019 and the FDA guidance document, "Guidance for Industry – Pyrogens and Endotoxins Testing: Questions and Answers," to verify the subject device meets current FDA and USP pyrogen limit specifications.
## Packaging Verification
Ship testing was evaluated to verify the integrity of the subject device packaging through the rigors of shipping and handling as well as storage over time. The sterile barrier integrity was also evaluated to verify compliance with the current applicable requirements ISO 11607-1:2019 over the proposed intended shelf life of the subject device.
Special 510(k)
013-Page 5
{9}
VersaCross Connect™ Transseptal Dilator
Baylis Medical Company Inc.
## Benchtop Validation
Customer requirements were validated through benchtop validation activities. Benchtop validation testing was performed to validate the performance of the subject device during normal intended use as per current applicable requirements of ISO 11070:2014/Amd.1:2018 and Baylis requirements.
The proposed VersaCross Connect™ Transseptal Dilator met all test requirements as specified by applicable standards and test protocols. The verification and validation activities for safety and effectiveness, along with the testing completed for the design changes demonstrated the subject device meets its intended use and is as safe, as effective, and performs in a manner that is substantially equivalent to the predicate device.
## Conclusions
The subject and predicate devices share the same indications for use, intended use, and fundamental scientific technology, including principles of operation and mechanism of action. Technological differences between the subject and predicate devices do not raise new or different questions of safety and effectiveness. The results of verification and validation activities support substantial equivalence of the proposed VersaCross Connect™ Transseptal Dilator to the predicate device.
Special 510(k)
013-Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.