K251303 · Menovex Medical Technology (Shenzhen) Co., Ltd. · GEX · Nov 4, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251303
Device Name
Tamburo Thulium Fiber Laser System
Applicant
Menovex Medical Technology (Shenzhen) Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications:urology, lithotripsy gastroenterological surgery and gynecological surgery.
Device Story
Tamburo Thulium Fiber Laser System is a 1940 nm thulium fiber laser surgical instrument. System components include laser console, footswitch, power cord, interlock connector, ground wire, protective goggles, and optical delivery fibers. Operated by physicians in clinical settings (urology, gastroenterology, gynecology). Laser energy is transmitted via fiber to the treatment site for soft tissue interaction. Device features include air cooling, aiming beam, and LCD touch screen interface. Output is controlled laser energy for surgical procedures; benefits include precise tissue management (incision, ablation, coagulation).
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including optical, mechanical, and hardware performance verification, biocompatibility (ISO 10993 series), and electrical safety/EMC (IEC 60601 series).
Technological Characteristics
Thulium fiber laser; 1940 nm wavelength; 60W max power; 0.025-6J pulse energy; 0.2-50ms pulse duration; up to 2400 Hz frequency. Air cooling. Optical fiber delivery (150-800μm). Sterilization: EO. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, ISO 10993.
Indications for Use
Indicated for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue in urology, lithotripsy, gastroenterological surgery, and gynecological surgery, with or without an endoscope.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 4, 2025
menovex Medical Technology (Shenzhen) Co., Ltd
% Kyra Kang
Official Correspondent
Landlink Healthcare Technology (Shanghai) Co., Ltd.
Room 1308, Baohua International Plaza, West Guangzhong Road
Jingan District
Shanghai,
China
Re: K251303
Trade/Device Name: Tamburo Thulium Fiber Laser System;
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: October 30, 2025
Received: October 30, 2025
Dear Kyra Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251303 - Kyra Kang
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-
{2}
K251303 - Kyra Kang
Page 3
devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2025.11.04
19:43:58 -05'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Tamburo Thulium Fiber Laser System
Page 11 of 55
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251303 | ? |
| Please provide the device trade name(s). | | ? |
| Tamburo Thulium Fiber Laser System | | |
| Please provide your Indications for Use below. | | ? |
| The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications:urology, lithotripsy gastroenterological surgery and gynecological surgery. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K251303
# 510(k) Summary
## I. Submitter
menovex Medical Technology (Shenzhen) Co., Ltd
401, 501, Building 13, Biomedicine Innovation Industrial Park, No. 14, Jinhui Road, Jinsha Community, Kengzi Street, Pingshan District, Shenzhen 518118, China
Contact person: Ms. Xueying Qin
Tel.: +86-0755-89454602
E-mail: qxy100@menovex.com
Preparation date: April 27, 2025
Submission Correspondent
Ms. Kyra Kang
Landlink Healthcare Technology (Shanghai) Co., Ltd.
E-mail: kyra.kang@landlink-health.com
## II. Proposed Device
Device Trade Name: Tamburo Thulium Fiber Laser System
Common name: powered laser surgical instrument
Regulation Number: 21 CFR 878.4810
Regulatory Class: Class II
Product code: GEX
Review Panel: General & Plastic Surgery
## III. Predicate Devices
510(k) Number: K183647
Trade name: SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories).
Common name: powered laser surgical instrument
Classification: Class II
Product Code: GEX
Manufacturer: General & Plastic Surgery
Page 1 of 6
{5}
Page 2 of 6
## IV. Device description
Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications:
urology, lithotripsy gastroenterological surgery and gynecological surgery.
The Tamburo is a Thulium fiber laser device emitting at a wavelength of 1940 nm, which is used by physicians as a tool in surgical procedures.
This device consists of a thulium laser, footswitch, power cord, Interlock connector, ground wire and protective goggles.
The thulium laser includes thulium laser, cooling device, aiming beam device, and LCD touch screen.
Tamburo Thulium Laser Fiber:
Tamburo Thulium Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.
This device consists of a fiber, dust cap, ferrule, outer cap and handle.
The fiber includes core, cladding, coating and outer jacket.
Tamburo Thulium Laser Fibers are compatible for use with Tamburo Thulium Laser Systems.
## V. Indications for use
The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy gastroenterological surgery and gynecological surgery.
## VI. Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate devices are listed in below table
Table 5.1 General Comparison
| Characteristics | Proposed device | Predicate device (K183647) | Discussion |
| --- | --- | --- | --- |
| Trade name | Tamburo Thulium Fiber Laser System | SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium | / |
{6}
| | | SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories). | |
| --- | --- | --- | --- |
| Classification Name | Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology | Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology | Same |
| Product Code | GEX | GEX | Same |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Indications for use | The Tamburo Thulium Fiber Laser System are intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy gastroenterological surgery and gynecological surgery. | The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. | Same |
| Laser Source | Thulium laser | Thulium laser | Same |
| Wavelength | 1938.9nm(1940 ±20 nm) | 1936.5nm(1940 ±20 nm) | Same |
Page 3 of 6
{7}
| Laser mode | Pulsed / CW | Pulsed | Different 1 |
| --- | --- | --- | --- |
| Divergence angle of terminal laser beam | 15.40° (16°±20%) | 12.96° | Different 2 |
| Aiming beam peak wavelength | 522.55nm (520 ±10nm) | 518.50nm (500-550nm) | Similar |
| Frequency | Up to 2400 Hz | Up to 2400 Hz | Same |
| Pulse Energy | 0.025-6J | 0.025-6J | Same |
| Pulse duration | 0.2-50ms | 0.2-50ms | Same |
| Max average power | 60W | 60W | Same |
| Delivery system | Optical fibers | Optical fibers | Same |
| Cooling system | Air cooling system | Air cooling system | Same |
| Fiber size | 150, 200, 272, 365, 550, 800μm | 150, 200, 365, 550, 940 μm | Similar |
| Fiber Sterilization | EO | EO | Same |
# Analysis 1 Laser mode
Both pulsed and CW modes have been rigorously tested and validated to ensure they meet all applicable regulatory requirements and deliver consistent, reliable performance. The difference in laser mode does not affect the safety or effectiveness.
{8}
Analysis 2 Divergence angle of terminal laser beam
By calculation, the distance of the Tamburo laser is 0.84 times that of Olympus. During clinical use, the fiber directly contacts the tissue. The distance difference can be ignored. The device has been rigorously tested and validated to ensure that it meets all applicable safety and performance standards, demonstrating that the divergence angle does not affect its ability to deliver effective and controlled laser energy for its intended use. Therefore, this difference does not pose any risk to the safety or effectiveness of the device.
## VII. Non-Clinical Testing
The device described in this summary, were tested and demonstrated to be in conformance with the following standards:
### Performance testing :
A comprehensive performance test was conducted on the Tamburo Thulium Fiber Laser System and Tamburo Thulium Laser Fiber to verify the performance of the product in optical, mechanical and Hardware aspects, along with comparative testing against equivalent products to demonstrate substantial equivalence.
### Biocompatibility testing
Biocompatibility of the device was evaluated in accordance with the FDA guidance "Use of International Standard ISO 10993-1" The following testing was conducted:
- ISO 10993-1 Fifth edition 2018-08
- ISO 10993-5 Third edition 2009-06-01
- ISO 10993-10 Fourth edition 2021-11
- ISO 10993-11 Third edition 2017-09
- ISO 10993-23 Software First edition 2021-01
Guidance for Industry and Food and Drug Administration Staff-Content of Premarket Submissions for Device Software Functions
### Electrical Safety and EMC
- IEC 60601-1 Edition 3.2 2020-08
- IEC 60601-1-2 Edition 4.1 2020-09
- IEC 60601-4-2 Edition 1.0 2-2024-03
- IEC 60601-1-6 Edition 3.2 2020-07
- IEC 60601-2-22 Edition 4.0 2019-11
Page 5 of 6
{9}
Page 6 of 6
## Shelf -life
The shelf life of the Tamburo Thulium Fiber Laser System is ensured by the stability evaluation performed in both real time and accelerated conditions, by the stability of laser performance. The performance is maintained after transport simulation and after 7 years aging.
The shelf life of the Tamburo Thulium Laser Fiber is set to 3 years. The sterile barrier performance (with the sterile barrier manufactured at validated worst case manufacturing conditions) is maintained after transport simulation and after 3 years aging (corresponding to the shelf-life of the device).
## VIII. Clinical Testing
No clinical study is included in this submission.
## IX. Conclusion
The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.