K251177 · Guangdong Maidi Medical Co., Ltd. · FRG · Dec 12, 2025 · General Hospital
Device Facts
Record ID
K251177
Device Name
Self-Seal Sterilization Pouch and Roll
Applicant
Guangdong Maidi Medical Co., Ltd.
Product Code
FRG · General Hospital
Decision Date
Dec 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Self-Seal Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: - Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. - Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature. The Self-Seal Sterilization Pouch and Roll are made with medical grade paper and medical compound film. The Self Sealing Sterilization Pouch and Roll maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color changes from red before sterilization to brown after sterilization. The steam sterilization indicator color changes from blue before sterilization to greenish black after sterilization.
Device Story
Device consists of medical-grade paper and PET/CPP compound film pouches/rolls; used by healthcare workers to enclose medical instruments for terminal sterilization. Input: medical devices placed inside pouch/roll. Operation: pouch is self-sealed or heat-sealed (roll); subjected to Steam (132°C) or EO (695 mg/L) sterilization. Output: sterilized, sealed package. External chemical ink indicators provide visual verification of exposure to sterilization process (color change: blue to greenish black for Steam; red to brown for EO). Benefits: maintains sterility of enclosed devices for up to 6 months post-sterilization. Used in clinical settings to ensure instrument sterility until point of use.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via ISO 11135 (EO) and ISO 17665-1 (Steam) achieving SAL 10⁻⁶. Biocompatibility confirmed per ISO 10993 series. Physical performance (tensile, burst, seal strength, leak, dye penetration) and microbial barrier (ASTM F1608) met all acceptance criteria. Shelf life and sterility maintenance validated via ASTM F1980 and real-time aging.
Technological Characteristics
Materials: medical grade paper (60g/m²) and PET/CPP compound film (52 μm). Form factor: various sizes of pouches and rolls. Sterilization: Steam and EO. Chemical indicators: ink-based color change. Biocompatibility: ISO 10993 compliant. Physical standards: ASTM F2251 (thickness), ISO 1924-2 (tensile), ASTM F88/F88M (seal), ASTM F1140 (burst), ASTM F1929 (dye), ASTM F2096 (leak), ASTM F1608 (microbial barrier).
Indications for Use
Indicated for healthcare workers to enclose medical devices for terminal sterilization via Steam or Ethylene Oxide (EO) gas. Suitable for various medical devices within specified weight limits.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
Guangdong Maidi Medical Co., Ltd.
% Kiwi Xu
Official Correspondent
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Century Ave.
Shanghai,
China
Re: K251177
Trade/Device Name: Self-Seal Sterilization Pouch and Roll
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: FRG, JOJ
Dated: November 13, 2025
Received: November 13, 2025
Dear Kiwi Xu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251177 - Kiwi Xu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251177 - Kiwi Xu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN A. Digitally signed by
ANISKO -S STEPHEN A. ANISKO -S
Date: 2025.12.12
14:55:05 -05'00'
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251177
Device Name
Self-Seal Sterilization Pouch and Roll
Indications for Use (Describe)
The Self-Seal Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature.
The Self-Seal Sterilization Pouch and Roll are made with medical grade paper and medical compound film. The Self Sealing Sterilization Pouch and Roll maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color changes from red before sterilization to brown after sterilization. The steam sterilization indicator color changes from blue before sterilization to greenish black after sterilization.
| Model | Product code | Specification (W*L) | | Maximum load (lbs) | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Inch | mm | Metal | Plastic | Gauze/Linens |
| Sterilization Pouch | WMDS.060-130 | 2 1/4" x 5" | 60mm × 130mm | 0.04 | 0.0057 | 0.005 |
| | WMDS.070-183 | 2 3/4" x 7 1/5" | 70mm × 183mm | 0.07 | 0.03 | 0.005 |
| | WMDS.070-260 | 2 3/4" x 10 1/4" | 70mm × 260mm | 0.09 | 0.08 | 0.07 |
| | WMDS.075-185 | 3" x 7 3/10" | 75mm × 185mm | 0.12 | 0.1 | 0.08 |
| | WMDS.090-170 | 3 1/2" x 6 3/4" | 90mm × 170mm | 0.12 | 0.11 | 0.09 |
| | WMDS.090-230 | 3 1/2" x 9" | 90mm × 230mm | 0.13 | 0.12 | 0.11 |
| | WMDS.090-260 | 3 1/2" x 10 1/4" | 90mm × 260mm | 0.14 | 0.13 | 0.11 |
| | WMDS.120-190 | 4 7/10" x 7 1/2" | 120mm×190mm | 0.21 | 0.2 | 0.12 |
| | WMDS.130-260 | 5 1/10" x 10 1/4" | 130mm×260mm | 0.48 | 0.39 | 0.19 |
| | WMDS.135-290 | 5 1/4" x 11 1/2" | 135m × 290mm | 0.5 | 0.4 | 0.21 |
| | WMDS.150-230 | 6" x 9" | 150mm×230mm | 0.56 | 0.4 | 0.21 |
| | WMDS.180-272 | 7" x 10 7/10" | 180mm×272mm | 0.58 | 0.42 | 0.22 |
| | WMDS.195-360 | 7 3/4" x 14 1/4" | 195mm×360mm | 0.6 | 0.43 | 0.23 |
| | WMDS.240-320 | 9 2/5" x 12 3/5" | 240mm×320mm | 1.94 | 0.46 | 0.24 |
| | WMDS.250-400 | 9 4/5" x 15 3/4" | 250mm×400mm | 3.34 | 0.48 | 0.26 |
| | WMDS.300-400 | 11 4/5" x 15 3/4" | 300mm×400mm | 3.34 | 0.69 | 0.3 |
| | WMDS.390-400 | 15 3/10" x 15 3/4" | 390mm×400mm | 3.34 | 0.8 | 0.32 |
| | WMDS.500-600 | 19 7/10" x 23 3/5" | 500mm×600mm | 3.34 | 1.03 | 0.48 |
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
| Model | Product code | Specification (W*L) | | Maximum load (lbs) | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Inch | mm | Metal | Plastic | Gauze/Linens |
| Sterilization Roll | WMDR.055-200 | 2 1/5" | 55mm×200m | 0.07 | 0.03 | 0.003 |
| | WMDR.075-200 | 3" | 75mm×200m | 0.09 | 0.05 | 0.004 |
| | WMDR.100-200 | 4" | 100mm×200m | 0.14 | 0.13 | 0.11 |
| | WMDR.150-200 | 6" | 150mm×200m | 0.55 | 0.53 | 0.5 |
| | WMDR.200-200 | 8" | 200mm×200m | 0.6 | 0.58 | 0.55 |
| | WMDR.250-200 | 10" | 250mm×200m | 1.06 | 1.05 | 0.62 |
| | WMDR.350-200 | 12" | 300mm×200m | 0.6 | 0.58 | 0.55 |
| | WMDR.300-200 | 14" | 350mm×200m | 1.06 | 1.05 | 0.62 |
| | WMDR.400-600 | 16" | 400mm×200m | 1.08 | 1.06 | 0.63 |
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
{5}
510(k) Summary (K251177)
# 510(k) Summary (K251177)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Guangdong Maidi Medical Co., Ltd.
Address: No.3, Zhihui Rd, Guangzhou (Qingyuan) Industrial Transfer Industrial Park, Shijiao Town, Qingyuan, 511545, Guangdong, P.R. China
Tel: +86-20-82031350
Contact: Shi Xiong
Date Submitted: December 11, 2025.
## Designated Submission Correspondent
Contact: Kiwi Xu
Name: Shanghai SUNGO Management Consulting Co., Ltd.
Address: 14th Floor, Dongfang Building, 1500# Century Ave., Pudong Area, Shanghai, 200120 China
Tel: +86-21-58817802
Email: weijia.xu@sungoglobal.com
## 2.0 Device Information
Trade/Device name: Self-Seal Sterilization Pouch and Roll
Common name: Sterilization Pouch and Roll
Regulation Name:
1) Sterilization Wrap;
2) Sterilization process indicator
Classification Product Code: 1) FRG;
Subsequent Product Code: 2) JOJ
Regulation number:
1) 21 CFR880.6850
2) 21 CFR880.2800
Classification: Class II
Panel: General Hospital
Page 1 of 9
{6}
Page 2 of 9
## 3.0 Predicate Device Information
### Predicate Device:
Manufacturer: Qianjiang Kingphar Medical Packaging & Printing Co., Ltd.
Trade/Device Name: Sterilization Pouch and Roll
510(k) number: K221875
## 4.0 Device Description
The Self-Seal Sterilization Pouch and Roll is composed of medical grade paper (60g/m²) and PET/CPP medical compound film (52 μm), it is intended to be used to contain medical devices to be terminally sterilized by the EtO or Steam sterilization process.
The recommended sterilization cycle parameter is:
- Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature.
The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the EtO or Steam Sterilization Process. The pouch is self-sealed prior to sterilization processing. The roll is not self-sealed; use a plastic sealing machine to seal it, and the temperature of the plastic sealing machine needs to reach 205°C. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.
The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post sterilization.
The Pouch/Roll is printed with a chemical indicator bar that changes from blue to greenish black (Steam) or red to brown (EtO) when exposed to EtO gas or Steam vapor during process. The EtO and Steam Chemical Indicator offers an addition way to verify processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The EtO and Steam Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the EtO or Steam.
## 5.0 Indication for Use Statement
The Self-Seal Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature.
The Self-Seal Sterilization Pouch and Roll are made with medical grade paper and medical compound film. The Self Sealing Sterilization Pouch and Roll maintains the
{7}
sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color changes from red before sterilization to brown after sterilization. The steam sterilization indicator color changes from blue before sterilization to greenish black after sterilization.
Table 1 Model and Dimension
| Model | Product code | Specification (W*L) | | Maximum load(lbs) | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Inch | mm | Metal | Plastic | Gauze/Linens |
| Sterilization Pouch | WMDS.060-130 | 21/4"×5" | 60mm × 130mm | 0.04 | 0.0057 | 0.005 |
| | WMDS.070-183 | 23/4"×71/5" | 70mm × 183mm | 0.07 | 0.03 | 0.005 |
| | WMDS.070-260 | 23/4"×101/4" | 70mm × 260mm | 0.09 | 0.08 | 0.07 |
| | WMDS.075-185 | 3"×73/10" | 75mm × 185mm | 0.12 | 0.10 | 0.08 |
| | WMDS.090-170 | 31/2"×63/4" | 90mm × 170mm | 0.12 | 0.11 | 0.09 |
| | WMDS.090-230 | 31/2"×9" | 90mm × 230mm | 0.13 | 0.12 | 0.11 |
| | WMDS.090-260 | 31/2"×101/4" | 90mm × 260mm | 0.14 | 0.13 | 0.11 |
| | WMDS.120-190 | 47/10"×71/2" | 120mm×190mm | 0.21 | 0.20 | 0.12 |
| | WMDS.130-260 | 51/10"×101/4" | 130mm×260mm | 0.48 | 0.39 | 0.19 |
| | WMDS.135-290 | 51/4"×111/2" | 135m × 290mm | 0.50 | 0.40 | 0.21 |
| | WMDS.150-230 | 6"×9" | 150mm×230mm | 0.56 | 0.40 | 0.21 |
| | WMDS.180-272 | 7"×107/10" | 180mm×272mm | 0.58 | 0.42 | 0.22 |
| | WMDS.195-360 | 73/4"×141/4" | 195mm×360mm | 0.60 | 0.43 | 0.23 |
| | WMDS.240-320 | 92/5"×123/5" | 240mm×320mm | 1.94 | 0.46 | 0.24 |
| | WMDS.250-400 | 94/5"×153/4" | 250mm×400mm | 3.34 | 0.48 | 0.26 |
| | WMDS.300-400 | 114/5"×153/4" | 300mm×400mm | 3.34 | 0.69 | 0.30 |
| | WMDS.390-400 | 153/10"×153/4" | 390mm×400mm | 3.34 | 0.80 | 0.32 |
| | WMDS.500-600 | 197/10"×233/5" | 500mm×600mm | 3.34 | 1.03 | 0.48 |
| Sterilization Roll | WMDR.055-200 | 21/5" | 55mm×200m | 0.07 | 0.03 | 0.003 |
| | WMDR.075-200 | 3" | 75mm×200m | 0.09 | 0.05 | 0.004 |
| | WMDR.100-200 | 4" | 100mm×200m | 0.14 | 0.13 | 0.11 |
| | WMDR.150-200 | 6" | 150mm×200m | 0.55 | 0.53 | 0.50 |
| | WMDR.200-200 | 8" | 200mm×200m | 0.60 | 0.58 | 0.55 |
| | WMDR.250-200 | 10" | 250mm×200m | 1.06 | 1.05 | 0.62 |
| | WMDR.350-200 | 12" | 300mm×200m | 0.60 | 0.58 | 0.55 |
| | WMDR.300-200 | 14" | 350mm×200m | 1.06 | 1.05 | 0.62 |
| | WMDR.400-600 | 16" | 400mm×200m | 1.08 | 1.06 | 0.63 |
{8}
6.0 Technological Characteristic Comparison Table
Table 2 Comparison of Technology Characteristics
| Item | Subject Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| Product Name | Self-Seal Sterilization Pouch and Roll | Sterilization Pouch and Roll | --- |
| 510(k) No. | K251177 | K221875 | --- |
| Product Code | FRG,JOJ | FRG,JOJ | Same |
| Regulation No. | 21 CFR 880.6850
21 CFR 880.2800 | 21 CFR 880.6850
21 CFR 880.2800 | Same |
| Class | II | II | Same |
| Intended Use | The Self-Seal Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
• Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
• Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature. The Self-Seal Sterilization Pouch and Roll are made with medical grade paper and medical compound film. The Self Sealing Sterilization Pouch and Roll maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.
The pouch’s external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam).
Ethylene oxide sterilization indicator color changes from red before sterilization to brown after sterilization. The steam sterilization indicator color changes from blue before sterilization to greenish black after sterilization. | The Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
• Steam Sterilization at 132°C, (270°F) for 4 minutes; Drying time of 20 minutes.
• Ethylene Oxide (EO) with a concentration of 800 mg/L at 55°C (131 °F) and 40% to 90% relative humidity for 6 hour. Aeration time of 7 days at 20°C).
The sterilization pouch and roll are made with medical grade paper and medical compound film. The sterilization pouch and roll maintains the sterility of the enclosed devices for up to 24 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture.
The pouch’s external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.
Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam).
Ethylene oxide sterilization indicator color changes from Pink before sterilization to Yellow after sterilization. The steam sterilization indicator | Similar |
Page 4 of 9
{9}
| | | color change from Blue before sterilization to Dark Grey after sterilization. | |
| --- | --- | --- | --- |
| Material Composition | Composed of medical grade paper and medical compound film, EO and Steam Process Indicator | Composed of medical grade paper and medical compound film, EO and Steam Process Indicator | Same |
| Sterilization Cycles | • Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
• Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature. | • Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
• Ethylene Oxide (EO) with a concentration of 800 mg/L at 55°C (131°F) and 40% to 90% relative humidity for 6 hour. Aeration time of 7 days at 20°C). | Different |
| Configuration/Dimension | Various Size, Heat Sealing | Various Size, Heat Sealing | Same |
| Performance Testing | | | |
| Sterilant Penetration Efficacy | The test meet the requirement of SAL 10^{-6} | The test meet the requirement of SAL 10^{-6} | Same |
| Chemical Indicator (CI) Functionality and Endpoint | The sterilant penetrated through the pouch configuration and affected the CI color change to the endpoint color | The sterilant penetrated through the pouch configuration and affected the CI color change to the endpoint color | Same |
| Device Design of Steam CI | The color of Chemical Indicator changes from Blue to dark greenish black, when exposed to Steam | The color of Chemical Indicator changes from Blue to dark grey, when exposed to Steam | Similar |
| Device Design of EO gas CI | The color of Chemical Indicator changes from red to brown, when exposed to EO gas | The color of Chemical Indicator changes from Pink to Yellow, when exposed to EO gas | Similar |
| Thickness Variations (mm) ASTM F 2251 | Passed | Passed | Same |
| Tensile strength ISO 1924-2 | Passed | Passed | Same |
| Burst Strength (kPa) ASTM F1140 ; ISO 11607-1 | Passed | Passed | Same |
| Leak Test ASTM-F2096 | Passed | Passed | |
| Seal Peel Test (N/15mm) ASTM F88/F88M ; ISO 11607-1 | Passed | Passed | Same |
| Dye penetration Test ASTM F1929 ; ISO11607-1 | No Infiltration Passed | No Infiltration Passed | Same |
| Microbial Barrier | Passed | Passed | Same |
Page 5 of 9
{10}
Page 6 of 9
| Test | | | |
| --- | --- | --- | --- |
| End point stability testing results | The color of chemical indicator for EO sterilization indicator ink is Red, and the color of chemical indicator for steam sterilization indicator ink is Blue after 3 year shelf life before sterilization. | The color of chemical indicator for EO sterilization indicator ink is Pink, and the color of chemical indicator for steam sterilization indicator ink is Blue after 5 year shelf life before sterilization. | Similar |
| | The color of chemical indicator for EO sterilization indicator ink is Brown, and the color of chemical indicator for steam sterilization indicator ink is Blue after EO sterilized and 6 months shelf life. | The color of chemical indicator for EO sterilization indicator ink is Yellow, and the color of chemical indicator for steam sterilization indicator ink is Blue after EO sterilized and 24 months shelf life. | Similar |
| | The color of chemical indicator for EO sterilization indicator ink is Red, and the color of chemical indicator for steam sterilization indicator ink is dark greenish black after steam sterilized and 6 months shelf life. | The color of chemical indicator for EO sterilization indicator ink is Pink, and the color of chemical indicator for steam sterilization indicator ink is Dark Grey after steam sterilized and 24 months shelf life. | Similar |
| Maintenance of Sterility | 6 months | 24 months | Different |
| Shelf Life | 3 years from date of manufacture for EO and Steam Indicators | 5 years from date of manufacture for EO and Steam Indicators | Different |
| Biocompatibility | Conform with ISO10993-1 (ISO10993-4, ISO10993-5, ISO10993-10, ISO10993-11, ISO10993-23) | Conform with ISO10993-1 (ISO10993-4, ISO10993-5, ISO10993-10, ISO10993-11) | Same |
The comparison between the subject and predicate devices is based on the following:
- Same intended use
- Same indications for use
- Similar material types that meet ISO 10993 biocompatibility requirements
- Same sterilization methods (EO and Steam sterilization process)
- Same fundamental technology/principal of operation/user interface
The subject device and the predicate device are identical in most key safety and effectiveness technical characteristics, such as material composition, basic design, sterilization method principles, physical properties (such as sealing strength, burst strength, microbial barrier) and biocompatibility. However, as shown in the comparison table, there are some differences, mainly concentrated in chemical indicator color, sterilization cycle parameters and shelf life. The following briefly discusses these differences:
Color differences of chemical indicators:
Property: The color change of the indicator is a visual indication, and it falls under the category of appearance characteristics.
The color differences do not alter the core function of the indicator, which is to indicate exposure to a specific sterilization process (EO or steam) through clear, distinct, and observable color changes. Both color systems comply with the
{11}
requirements of ISO 11140-1 for chemical indicators and provide equivalent process monitoring capabilities. Users only need to interpret them according to the instructions on the respective equipment labels, without causing confusion or misjudgment.
Differences in sterilization cycle parameters:
The subject device has undergone independent sterilization process verification in accordance with ISO 11135 (EO) and ISO 17665-1 (steam) standards. The results demonstrated that it can achieve a sterilization assurance level of SAL 10⁻⁶ under the specified parameters.
Difference between shelf life:
Properties: The declaration for the aseptic preservation period of the subject device is 6 months (compared to 24 months), and the shelf life (before sterilization) is 3 years (compared to 5 years).
This statement is designed to meet market demands and actual usage patterns and is supported by sufficient test data.
The identified differences (indicator color, specific sterilization parameters, declaration period) are non-core characteristics or process adjustments supported by sufficient verification, nor do they change the basic nature or intended use of the equipment.
## 7.0 Summary of Non-Clinical Testing
Summary of non-clinical and performance testing bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The test results demonstrated the subject device is confirmed to be safe and effective for the intended use. The conducted tests listed as below:
Table 3 Summary of Non-Clinical Testing
| Test | Test Method | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| Sterilization process validation | ISO 11135:2014 | Ethylene oxide: 4 hours at 132 °C; relative humidity between 30%-90%; ethylene oxide concentration is 800 mg/L, 7days aeration time at room temperature. SAL=10⁻⁶ | Pass |
| | ISO 17665-1:2006; ISO TS 17665-2:2009 | Steam; 4 minutes at 132°C; 20 minutes dry time. SAL=10⁻⁶ | Pass |
{12}
| Test | Test Method | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| EO Residuals | ISO 10993-7:2008 | EO ≤ 4 mg/device
ECH ≤ 9 mg/device | Pass |
| Cytotoxicity | ISO 10993-5:2009 | Non-cytotoxic | Pass |
| Sensitization | ISO 10993-10:2021 | Non-sensitizing | Pass |
| Irritation | ISO 10993-23:2021 | Non-irritating | Pass |
| Physical Performance |
| Thickness | ASTM F2251-2013(2023) | Plastic: 52μm±15%
Paper: 90μm±15% | Pass |
| Tensile Strength | ISO 1924-2:2008 | MD: ≥4.4kN/m
CD: ≥2.2kN/m | Pass.
All samples met the requirements for Machine Direction and Cross Direction. |
| Seal Strength | ASTM F88/F88M-2023 | ≥ 2.5N/15mm | Pass.
All tested samples demonstrated seal strength greater than 2.5N/15mm. |
| Burst Strength | ASTM F1140/F1140M-13(R2020) | Burst value > 3 kPa | Pass.
All samples withstood internal pressurization greater than 3 kPa. |
| Dye Penetration | ASTM F1929-2023 | No channel or seal leak | Pass.
No dye penetration was observed in the seal area. |
| Leak Test (Bubble) | ASTM F2096-11:2019 | No leakage | Pass.
No gross leaks were detected. |
| Microbial Barrier | ASTM F1608-2021 | LRV ≥ 1.0 | Pass.
The Log Reduction Value (LRV) was greater than 1.0. |
| Chemical Indicator | ISO 11140-1:2014
Premarket Notification [510(k)] Submissions for Chemical Indicators - Guidance for Industry and FDA Staff | FDA | Color change to endpoint
Steam sterile:The color of Chemical Indicator changes from Blue to dark greenish black, when exposed to Steam;
EO sterile:The color of Chemical Indicator changes from red to brown, when exposed to EO gas.
Remain stable before use based on its shelf life.
Maintain the endpoint stability of the color change after being in the presence of the sterilant. | Pass |
Page 8 of 9
{13}
Page 9 of 9
| Test | Test Method | Acceptance Criteria | Results |
| --- | --- | --- | --- |
| | | Off-set (transference): Migration of the indicator agent through the substrate to the surface opposite the one to which the indicator agent was applied shall not occur before, during or after the sterilization process. In use, the indicator agent shall not offset or bleed, penetrate the substrate to which it is applied or materials in which it is in contact. | |
| Shelf Life & Maintenance | | | |
| Shelf Life | ASTM F1980-21 | Maintain integrity after 3-year aging | Pass. Physical properties (Seal strength, Burst, Dye penetration) met acceptance criteria after aging. |
| Sterility Maintenance | Real-time Aging | Sterile after 6 months post-sterilization storage | Pass. Products remained sterile and package integrity was maintained after 6 months of storage following sterilization. |
## 8.0 Summary of Clinical Testing
No clinical study is included in this submission.
## 9.0 Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the subject device (Self-Seal Sterilization Pouch and Roll) is as safe, as effective, and performs as well as or better than the predicate device K221875.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.