K251159 · Zhenjiang Assure Medical Equipment Co., Ltd. · ITI · Dec 22, 2025 · Physical Medicine
Device Facts
Record ID
K251159
Device Name
Power wheelchair
Applicant
Zhenjiang Assure Medical Equipment Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Power wheelchair (Model EL-001) provides mobility for disabled or elderly users. Device consists of steel/aluminum frame, wheels, seat, armrest, backrest, headrest, battery, motor, and controller. User operates device via joystick to control speed and direction; controller manages brushless DC motors on rear wheels. Features automatic electromagnetic braking. Device is foldable for transport. Used in indoor and outdoor environments. Benefits include increased mobility for individuals with limited walking ability. Healthcare providers do not operate the device; it is for patient self-use.
Clinical Evidence
Bench testing only. Performance verified through compliance with ISO 7176 series (stability, braking, energy consumption, dimensions, speed, strength, climatic, obstacle climbing, power/control systems, EMC, and battery safety) and ISO 10993 series (biocompatibility). Risk analysis conducted per ISO 14971.
Technological Characteristics
Frame: steel and aluminum. Tires: solid PU. Power: 24VDC lead-acid batteries. Drive: dual 350W brushless DC motors. Control: joystick-based electronic controller. Braking: automatic electromagnetic. Foldable design. Standards: ISO 7176 series (performance/safety), ISO 10993 (biocompatibility), ISO 16840-10 (flame retardancy), IEC 60601-1-2/IEC TR 60601-4-2 (EMC).
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
Zhenjiang Assure Medical Equipment Co., Ltd.
% Ariel Xiang
Consultant
Shanghai Sungo Management Consulting Co. Ltd.
14th Floor, 1500# Century Avenue
Shanghai, Shanghai 200122
China
Re: K251159
Trade/Device Name: Power wheelchair
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: November 21, 2025
Received: November 21, 2025
Dear Ariel Xiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251159 - Ariel Xiang
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251159 - Ariel Xiang
Page 3
Sincerely,
Tushar Bansal -S
- Tushar Bansal, PhD
- Acting Assistant Director, Acute Injury Devices Team
- DHT5B: Division of Neuromodulation and Physical Medicine Devices
- OHT5: Office of Neurological and Physical Medicine Devices
- Office of Product Evaluation and Quality
- Center for Devices and Radiological Health
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251159
Device Name
Power wheelchair
Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
ZHENJIANG ASSURE MEDICAL EQUIPMENT CO., LTD. NO.297, CHUQIAO ROAD, ZHENJIANG CITY, JIANGSU PROVINCE, CHINA
# 510(K) Summary
K251159
Document Prepared Date: 2025/04/14
## A. Applicant
Name: ZHENJIANG ASSURE MEDICAL EQUIPMENT CO., LTD.
Address: NO.297, CHUQIAO ROAD, ZHENJIANG CITY, JIANGSU PROVINCE, CHINA
Contact Person: Mei Ying
Email: shelly@assuremedical.com.cn
## B. Device
Trade Name: Power wheelchair
Common Name: Electric wheelchair
Models: EL-001
Regulatory Information
Classification Name: Powered wheelchair Classification: Class II.
Product code: ITI
Regulation Number: 890.3860
Review Panel: Physical Medicine
## C. Predicate device
K244003
Electrically powered wheelchair, HP330
Kunshan Hi-Fortune Health Products Co., Ltd.
## D. Indications for use of the device
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
## E. Device Description
This product consists of frame, wheels, seat, armrest, backrest, headrest, battery, motor and controller with a lightweight and compact design. The whole wheelchair can be folded and it can be easily carried or rolled after folding. The backrest, headrest and armrest can be adjustable. Users can drive the wheelchair by themselves through the control device.
The wheelchair uses lead-acid batteries as its power source. The controller controls the drive
{5}
left/right motor to realize the wheelchair forward, backward and turn functions.
The frame of the device is steel and aluminum. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels. The rear wheels are driving wheels to control the speed and direction. The wheels are Solid PU tires.
When in use, the operator drives the motor of the rear wheel by operating the joystick to achieve the rear wheels movement.
The DC brushless motor and brake system are fixed on the rear wheels.
The max loading of the device is 135KG. Only for one person sit.
## F. Comparison to predicate device
Table 1 General Comparison
| Elements of Comparison | Proposed Device (K251159) | Predicate Device (K244003) | Remark |
| --- | --- | --- | --- |
| Manufacturer | ZHENJIANG ASSURE MEDICAL EQUIPMENT CO., LTD. | Kunshan Hi-Fortune Health Products Co., Ltd | -- |
| Common or Usual name | Power wheelchair | Electrically powered wheelchair | -- |
| Model(s) | EL-001 | HP330 | -- |
| Classification | Class II | Class II | Same |
| Classification regulation | 21 CFR890.3860 | 21 CFR890.3860 | Same |
| Product code | ITI | ITI | Same |
| Indications for use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person | SE. |
{6}
| | seated position. | limited to a seated position. | |
| --- | --- | --- | --- |
| Use condition | indoor and outdoor use | indoor and outdoor use | SE |
| Number of wheels | 6, including two front wheels and two rear wheels, two anti-tip wheels | 6, including two front wheel and two rear wheels, two anti-tip wheels | SE |
| Function of wheels | Front wheels: driven wheels suitable for rotation, acceleration, retrograde
Rear wheels: driving wheels to control the speed and Direction | Front wheels: driven wheels suitable for rotation, acceleration, retrograde
Rear wheels: driving wheels to control the speed and Direction | SE |
| Movement control method | By Joystick control | By Joystick control | SE |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | SE |
| Brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | SE |
| Braking distance | < 1m | ≤1.5m | SE |
| Brake time | < 1s | < 1s | |
{7}
| Armrest | Polyurethane | Polyurethane | SE |
| --- | --- | --- | --- |
| Max loading weight | 135kg | 136kg | SE |
| Maximum safe operational incline degree | 6° | 6° | SE |
| Back cushion | Polyurethane | Polyurethane | SE |
| Seat cushion | Polyurethane | Polyurethane | |
| Main frame material | Steel+Aluminum | Carbon steel material | Similar, Different materials will not affect safety and performance of the Proposed device as all related stability tests are performed according to standard ISO 7176 series. |
| Overall Dimension (length*width*height) | 1132mm*590mm*1086 mm | 950mm*660mm*950 mm | Minor difference on wheelchair dimension will not cause different performance. All safety and |
| Folded Dimension (length*width*height) | 820mm*590mm*808mm | 740mm*600mm*410m m | |
{8}
| | | | performance have been validated with the maximum rated weight dummy. |
| --- | --- | --- | --- |
| Front wheel size/type | 9 " | 6 " | Different sizes and materials will not affect safety and performance of the proposed device as all related stability tests are performed according to standard ISO 7176 series. |
| Rear wheel size/type | 13 " | 10 " | |
| Maximum speed forward | 6 km/h (1.7 m/s) | Up to 5.4km/h (1.5 m/s) | Similar,Both the Proposed device and The predicate device complys with ISO 7176-6:2018 Wheelchairs - Part 6: Determination |
| Maximum Speed backward | Less than 3.24 km/h (0.9m/s) | Less than 2.5km/h (0.7m/s) | |
{9}
| | | | of maximum speed, acceleration and deceleration of electric wheelchairs, so these differences do not affect safety and effectiveness. |
| --- | --- | --- | --- |
| Maximum distance of travel on the fully charged battery | 30.6km | 12km | Different, will not cause new safety and effectiveness concerns. |
| Battery | Lead-acid battery;rechargeable, 12 VDC 50Ah*2 | Lead-acid Battery,;rechargeable; 12V 22AH*2 | Both meet the requirements of the ISO 7176-25 : 2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs..It will not cause new safety |
| Battery charger | Input: 100-240V. 50/60Hz, 2.2-1.1A. Output: 24 V, 5A | Off-board charger Input:100-240VAC,50-60Hz,1.2-0.5A Output:24V,2A | |
{10}
| | | | and effectiveness concerns. |
| --- | --- | --- | --- |
| Motor | Brushless DC motor; 24VDC; 350W; 2pcs | Brushless DC motor; 24VDC; 250W; 2pcs | Similar, Both meet the requirements of the ISO 7176-14. It will not cause new safety and effectiveness concerns. |
| Electronic controller | Dynamic, | Dynamic, | |
| Turning Radius | 900mm | 825mm | Minor difference in turning radius is caused by different size of wheelchair and may |
| Obstacle climbing height(forward) | 40mm | 25mm | Minor difference on obstacle climbing will not cause new safety and effectiveness concerns. |
Table 2 safety
comparison
{11}
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-1, ISO10993-5, ISO10993-10 and ISO10993-23 requirements. | All user directly contacting materials are compliance with ISO10993-1, ISO10993-5, ISO10993-10 and ISO10993-23 requirements. | SE |
| EMC | ISO7176-21/ IEC 60601-1-2/ IEC TR 60601-4-2 | ISO7176-21/ IEC 60601-1-2/ IEC TR 60601-4-2 | SE |
| Performance | ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-22, ISO 7176-25 | ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-22, ISO 7176-25 | SE |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | SE |
## G. Substantial Equivalence Discussion
The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-21, ISO 7176-25, and FDA guidance Submission for Power Wheelchair.
The indications for use for both devices are the same. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3.
The frame materials, turning radius, and maximum obstacle climbing are slightly different while such differences will not impact the safety and effectiveness of the proposed device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-8 and ISO 7176-10.
{12}
The biocompatibility of the predicate device and proposed device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2021 & ISO 10993-23: 2021.
The flame-retardant test of the seat cushion/ backrest of both proposed device and predicate device is carried out according to the ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level. In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the proposed device and predicate devices do not raise new issues of safety or effectiveness.
## H. Product Performance
### Product Material Safety
To ensure the safety of the wheelchair material, the bio-compatibility testing was done to the products. The testing was conducted according to the following standards:
- ISO10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO10993-5 Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity
- ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23 Biological evaluation of medical devices - Part 23: Tests for irritation
From the test result, we can find the material are safety and can meet the requirements.
### Performance of the products
The following performance data were provided to verify that the proposed device met all design specifications and provided support of the substantial equivalence determination.
Risk Analysis developed in accordance with ISO 14971: 2019.
**Software validation**
- ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
- ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
{13}
ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method
ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
ISO 7176-22: 2014 Wheelchairs-Part 22: Set-up procedures
ISO 7176-25: 2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs.
Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 and IEC TR 60601-4-2
I. Conclusion
The proposed device and the predicate products have the same technological characteristics, features, specifications, materials, mode of operation, and intended use of the device. To ensure all the key characteristic of the products can meet the requirements, the testing was done and all the results indicate the positive conclusion.
Based on the analysis above, we confirm these two products are essentially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.