Arm Blood Pressure Monitor (MJ1L, MJ2L, RN3L, RN4L,RN5L)
K251143 · Ningbo Ranor Medical Science & Technology Co., Ltd. · DXN · Dec 19, 2025 · Cardiovascular
Device Facts
Record ID
K251143
Device Name
Arm Blood Pressure Monitor (MJ1L, MJ2L, RN3L, RN4L,RN5L)
Applicant
Ningbo Ranor Medical Science & Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Dec 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Arm Blood Pressure Monitor is used for measuring blood pressure and pulse rate. The monitor can be used in hospitals, families, schools and medical centers. It is suitable for adult, not for neonate or pregnancy.
Device Story
Automatic non-invasive blood pressure monitor; uses inflatable cuff wrapped around upper arm; oscillometric measurement principle; detects brachial artery blood flow movement; converts signals to digital systolic/diastolic blood pressure and pulse rate readings. Device stores up to 2x60 measurement sets; calculates average of last 3 readings; optional voice broadcast function. Used in hospitals, homes, schools, and medical centers by patients or clinicians. No stethoscope required. Automatic shutdown after 1 minute of inactivity. Powered by 3.7V 800mA Li-ion rechargeable battery. Provides digital display of results; assists healthcare providers in monitoring patient cardiovascular status.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018+A1:2020. Blood pressure accuracy evaluated in 150 adult subjects against mercury sphygmomanometer; met requirements within ±5 mmHg. Pulse rate accuracy evaluated in 85 adult subjects (50-100 bpm range) against ECG reference; mean error 0.43 bpm, SD 0.91 bpm, RMSE ~1.0 bpm. Bland-Altman analysis showed good agreement.
Technological Characteristics
Oscillometric measurement; upper arm cuff (brushed fabric, ISO 10993 compliant); LCD digital display; Li-ion rechargeable battery (3.7V 800mA); solenoid valve deflation. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, and IEC 62133-2 (battery safety).
Indications for Use
Indicated for adult patients for the measurement of blood pressure and pulse rate in hospitals, homes, schools, and medical centers. Not indicated for neonates or pregnant individuals.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 19, 2025
Ningbo Ranor Medical Science & Technology Co., Ltd.
% Boyle Wang
General Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lu Jiazui Rd., Pudong
Shanghai, 200120
China
Re: K251143
Trade/Device Name: Arm Blood Pressure Monitor (Model MJ1L, MJ2L, RN3L, RN4L, RN5L)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: December 1, 2025
Received: December 1, 2025
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251143 - Boyle Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251143 - Boyle Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PBC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251143 | |
| Device Name Arm Blood Pressure Monitor(Model MJ1L, MJ2L, RN3L, RN4L,RN5L) | |
| Indications for Use (Describe) The Arm Blood Pressure Monitor is used for measuring blood pressure and pulse rate. The monitor can be used in hospitals, families, schools and medical centers. It is suitable for adult, not for neonate or pregnancy. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
1 / 6
# 510(k) Summary
# K251143
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Ningbo Ranor Medical Science & Technology Co., Ltd.
Address: No. 127 Fenghui Road, Wangchun Industrial Park, Haishu District, Ningbo, China
Tel: 86-574-89258788
Contact: Xu Jia'nan
## Designated Submission Correspondent
Contact: Mr. Boyle Wang
Name: Shanghai Truthful Information Technology Co., Ltd.
Address: Room 1801, No. 161 East Lu Jiazui Rd., Pudong, Shanghai, 200120 China
Tel: +86-21-50313932
Email: Info@truthful.com.cn
Date of Preparation: Dec.19,2025
## 2.0 Device Information
Trade name: Arm Blood Pressure Monitor
Common name: Noninvasive Blood Pressure Measurement System
Classification name: Noninvasive Blood Pressure Measurement System
Model(s): MJ1L, MJ2L, RN3L, RN4L, RN5L
Production code: DXN
Regulation number: 21 CFR 870.1130
Classification: Class II
Panel: Cardiovascular
## 3.0 Predicate Device Information
Manufacturer: Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
Trade/Device name: Electronic Blood Pressure Monitor, YE620B, YE620D, YE660E, YE660F and YE680B
K251143
{5}
510(k) Summary
2 / 6
510(k) number: K200939
## 4.0 Indication for Use Statement
The Arm Blood Pressure Monitor is used for measuring blood pressure and pulse rate. The monitor can be used in hospitals, families, schools and medical centers. It is suitable for adult, not for neonate or pregnancy.
## 5.0 Device Description
The subject device, Arm Blood Pressure Monitor, is an automatic non-invasive blood pressure monitor. It uses an inflatable cuff which is wrapped around the patient's upper arm to measure the systolic and diastolic blood pressure as well as the pulse rate of adult, not for neonate or pregnancy.
The unit uses the oscillometric method of blood pressure measurement. It means the unit detects the movement of your blood through your brachial artery, and converts your blood pressure into a digital reading. The unit is simple to use because a stethoscope is not needed while using an oscillometric monitor.
The unit stores the blood pressure and pulse rate in the memory after completing a measurement each time. 2x60 sets of measurement values can be stored automatically. The unit also calculates an average reading based on the values of the latest 3 times measurement.
This blood pressure monitor has voice broadcast function (optional). During measurement and recall the memory, there will be voice operation tips.
No operation for 1 minute the device will shut down automatically.
## 6.0 Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC 60601-1:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-11:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
K251143
{6}
510(k) Summary
3 / 6
- IEC 80601-2-30:2018, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.
- IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- FDA guidance: Content of Premarket Submissions for Device Software Functions, dated July 20, 2023
## 7.0 Clinical Test Conclusion
### 1) Blood Pressure Clinical Validation
The clinical testing was conducted to demonstrate that the subject device meets the performance specifications for its intended use. The clinical trials for the Arm Blood Pressure Monitor were performed in accordance with ISO 81060-2:2018+A1:2020, Non-Invasive Sphygmomanometers – Part 2: Clinical Validation of Automated Measurement Type.
A total of 150 adult subjects were enrolled to evaluate blood pressure measurement accuracy. Auscultatory measurements obtained using a calibrated mercury sphygmomanometer served as the reference standard. Subjects with conditions that could interfere with accurate non-invasive blood pressure measurement or with study participation, such as significant arrhythmias, implanted cardiac rhythm devices, or inability to undergo standardized cuff-based measurement, were excluded in accordance with the clinical protocol and the applicable standard.
The results demonstrated that the accuracy of the subject device met the requirements of ISO 81060-2:2018 within ±5 mmHg. No adverse effects or complications were observed during the study.
### 2) Pulse Rate Clinical Validation
A clinical validation study was conducted to evaluate the pulse rate (PR) measurement accuracy of the subject device by comparison with manually annotated electrocardiogram (ECG) recordings. A total of 85 adult subjects (age range: 24–73 years) were enrolled and included in the analysis. The PR validation was conducted within the validated pulse-rate range of 50–100 bpm.
Subjects were excluded if they had implanted pacemakers or implantable cardioverter-defibrillators, known severe arrhythmias, or any medical condition or circumstance that, in the judgment of the investigator, could interfere with reliable pulse-rate measurement or ECG interpretation.
For each subject, three consecutive PR measurements were obtained under resting
Page 3 of 6
K251143
{7}
510(k) Summary
4 / 6
conditions, and the mean value was paired with a manually annotated ECG reference. Across all 85 subjects, the mean PR error was 0.43 bpm with a standard deviation of 0.91 bpm, and the overall RMSE was approximately 1.0 bpm. Bland-Altman analysis demonstrated good agreement between the device PR outputs and the ECG reference values, with measurement differences consistently distributed within clinically acceptable limits across the validated range.
These findings confirm that the subject device meets the safety and performance requirements for both blood pressure and pulse-rate measurement within the validated conditions of use.
# 8.0 Technological Characteristic Comparison Table
Table1-General Comparison
| Item | Subject Device K251143 | Predicate Device K200939 | Remark |
| --- | --- | --- | --- |
| Manufacturer | Ningbo Ranor Medical Science & Technology Co., Ltd. | Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. | / |
| Product Name | Arm Blood Pressure Monitor MJ1L, MJ2L, RN3L, RN4L,RN5L | Electronic Blood Pressure Monitor YE620D | / |
| Product Code | DXN | DXN | Same |
| Regulation No. | 21 CFR 870.1130 | 21 CFR 870.1130 | Same |
| Class | II | II | Same |
| Intended Use/Indication for Use | The Arm blood pressure monitor is for home use for measuring blood pressure and pulse rate. It is suitable for adult, not for neonate or pregnancy. | Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia). | Same |
| Application Site | Upper Arm | Upper Arm | Same |
| Arm Circumference | 220mm ~ 420mm | Type A (22cm~32cm, 22-45cm optional)
Type B (22cm~ 32cm, 22-45cm optional) | Same |
| Patients Contacting Materials | Patient contact materials of the cuff: Brushed Fabric
According to ISO-10993 | Patient contact materials of the cuff have been tested in accordance with ISO10993 and FDA guidance. | Same |
| Patient Population | Adult | Adult | Same |
| Measurements Item | SYS,DYS,Pulse | SYS,DYS,Pulse | Same |
| Display | LCD Digital Display | LCD | Same |
| Design Method | Oscillometric Method | Oscillometric Method | Same |
Page 4 of 6
K251143
{8}
510(k) Summary
5 / 6
Table 2 Performance Comparison
| Item | Subject Device K251143 | Predicate Device K200939 | Remark |
| --- | --- | --- | --- |
| Max Cuff pressure | 295 mmHg | 300mmHg | Different (1) |
| BP Range | 0-294 mmHg | 0 ~ 300 mmHg | |
| BP Accuracy | ±3 mmHg | ±3 mmHg | Same |
| PR Range | 50-100 beats/min | 40 ~ 200 beats/min | Different (1) |
| Pulse Accuracy | ±5% of reading value | ±5% of reading value | Same |
| Inflation Method | Automatic inflation by air pump | Automatic Internal Pump | Same |
| Deflation Method | By solenoid valve | Semiconductor Pressure Sensor | Different (2) |
| Memory Size | 2x60 set of data | Up to 2x60sets of data | Same |
| Operation Condition | 10~40℃
Humidity: 15~85%RHz
Atmospheric pressure: 70 - 106kPa | 5~40℃,
Humidity:15~90%RH
Atmospheric: 106kPa~80kPa | Different (3) |
| Storage Condition | -20~55 ℃
15~90%RH (noncondensing) | -20℃~55℃,
15~90%RH (noncondensing) | |
| Power Supply | Li-ion Rechargeable battery, 3.7V 800mA | 4 AA batteries or 6V/600mA AC adapter | Different (4) |
| Performance Standard | Comply with IEC 80601-2-30 | Comply with IEC 80601-2-30 | Same |
Analysis:
1) The Blood Pressure Range of the subject device has minor difference with the blood pressure range of predicate devices, the measuring range of the subject device was validated according to ISO 80601-2-30, the difference does not raise different questions of safety and effectiveness. The pulse rate measurement range of the subject device is within that of the predicate device. In addition, the PR measurement performance of the subject device was clinically validated using manually annotated ECG as the reference method. Therefore, the differences between the subject device and the predicate device do not affect the safety or effectiveness of the subject device.
2) The subject device utilizes a solenoid valve for cuff deflation, whereas the predicate device uses a semiconductor pressure sensor to control the deflation process. This difference represents an implementation difference in the deflation control mechanism and does not change the fundamental oscillometric blood pressure measurement principle.
The subject device was verified through performance testing in accordance with IEC 80601-2-30 and clinical validation conducted per ISO 81060-2. The test results demonstrated that the subject device meets the specified blood pressure accuracy
Page 5 of 6
K251143
{9}
510(k) Summary
6 / 6
and performance requirements. Therefore, the difference in deflation methods does not affect the safety or effectiveness of the subject device.
3) All the differences in operation condition don't affect the safety and effectiveness which is concluded after all the required testing, thus, this difference does not raise different questions of safety and effectiveness.
4) The difference in power source will not affect the safety and effectiveness of the subject device. IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 62133-2 can demonstrate that the subject device can maintain the safety and performance. Thus, this difference does not raise different questions of safety and effectiveness.
Table 3 Safety Comparison
| Item | Subject Device K251143 | Predicate Device K200939 | Remark |
| --- | --- | --- | --- |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| Home Use | Comply with IEC 60601-1-11 | Comply with IEC 60601-1-11 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
| Biocompatibility | Comply with ISO 10993-1, FDA Guidance | Comply with ISO 10993-1, FDA Guidance | Same |
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Page 6 of 6
K251143
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.